2
Portable Syngas Monitoring System Portable Syngas Analyser The Portable Syngas Monitoring System (SMS-P) is an anaylser that utilises non- dispersive IR and other techniques to provide the measurement of CO, CO 2 , CH 4 , H 2 & O 2 The SMS-P provides a measurement of all the major constituents of the syngas and calculates the live CV value, essential to determine your qualification for renewable obligation certificates (ROCS). In modern gasification where syngas composition is paramount for determining the calorific value, the a1-cbiss SMS- P provides an accurate, cost-effective solution. The SMS-P is powered by Li-on battery to provide on the spot measurement of syngas monitoring in pyrolysis and gasification processes Small, Robust & Lightweight Designed to be used in conjunction with the a1-cbiss portable sampling probe, which features a selection of sampling tubes for varied process conditions. Internal Data Logger for Real-Time Data Readings Equipped with an internal data logger, the SMS can easily connect to a laptop (via a RS232 cable) to provide real time data and data storage on a laptop KEYBENEFITS Uses an internal Li-on battery to work without an AC power supply Temperature regulated enclosure for NDIR and TCD detectors Equipped with an internal sampling pump Lightweight, moulded casing and shoulder strap for increased portability & ease of transportation Internal data logger connects to PC / laptop to display real time data The SMS-P quickly measures syngas constituents and aids in achieving ROCs accreditation GASESANALYSED CO H 2 CO 2 C n H m CH 4

Sms portable syngas analyser datasheet 1

Embed Size (px)

Citation preview

Portable Syngas Monitoring SystemPortable Syngas Analyser The Portable Syngas Monitoring System (SMS-P) is an anaylser that utilises non-dispersive IR and other techniques to provide the measurement of CO, CO2, CH4, H2 & O2

The SMS-P provides a measurement of all the major constituents of the syngas

and calculates the live CV value, essential to determine your qualification for

renewable obligation certificates (ROCS). In modern gasification where syngas

composition is paramount for determining the calorific value, the a1-cbiss

SMS- P provides an accurate, cost-effective solution.

The SMS-P is powered by Li-on battery to provide on the spot

measurement of syngas monitoring in pyrolysis and gasification

processes

Small, Robust & Lightweight Designed to be used in conjunction with the a1-cbiss portable

sampling probe, which features a selection of sampling tubes for

varied process conditions.

Internal Data Logger for Real-Time Data Readings Equipped with an internal data logger, the SMS can easily connect to a

laptop (via a RS232 cable) to provide real time data and data

storage on a laptop

KEYBENEFITS• Uses an internal Li-on battery to work

without an AC power supply

• Temperature regulated enclosure for NDIR

and TCD detectors

• Equipped with an internal sampling

pump

• Lightweight, moulded casing and

shoulder strap for increased portability &

ease of transportation

• Internal data logger connects to PC /

laptop to display real time data

The SMS-P quickly measures syngas constituents and aids in achieving ROCs accreditation

“ “

GASESANALYSED

CO H2CO2 CnHmCH4

Cert No. 996ISO 9001

11 Ark Royal Way, Lairdside Technology Park, Tranmere, Wirral, CH41 9HTT: +44(0)151 666 8300 F: +44(0)151 666 8329 E: [email protected] W: www.a1-cbiss.com Buy Online: shop.a1-cbiss.com

TECHNICALSPECIFICATION

GENERAL• Set up time <15 minutes

• Highly visible yellow casing

MEASUREMENT TECHNOLOGY • CO, CO2, CH4, CnHm: proprietary dual beam NDIR detectors

• O2: industrial electrochemical cell

• H2: proprietary thermal conductivity detector

MEASUREMENT RANGES • CO: 0-100%

• CO2: 0-50%

• CH4 : 0-10%

• CnHm : 0-10%

• O2: 0-25%

• H2: 0-50%

• H2: 0-50%

• Additional ranges available at no extra cost

CALCULATION • BTU index (gas calorific value)

• High Heating Value or Low Heating Value (in MJ/m3 or

kcal/m3)

• Optional N2

OPERATING TEMP & PRESSURE • Tamb: 0-50°C

• Pamb: 86 to 108 kPa

• RH: 0-95% non condensing

PERFORMANCE • Resolution: 0,01%

• Accuracy: ≤ ± 2% FS

• Repeatability: ≤ 2%

• Zero: Auto-zeroing function via keyboard interface

• Response time: (T90) ≤ 15 sec

• Warm-up time: 800 seconds (auto-zeroing cycle during

the last 120 seconds)

DATA LOGGING • Up to 1500 sets of data

• Logging rate adjustable from 3 to 99 sec

• Identify 10 different sites and up to 100 measuring points

• Simultaneous indication of the 7 measures and units

POWER • Power supply: External 220 VAC-50Hz

• Internal with battery and charger; Autonomy of > 4h with

pump in operation

• Electrical connection: Cold appliance plug with 1.5 m of

cable

PHYSICAL • Dimensions: L 380 × D140 × H255 mm

• Weight: Approx. ± 5 kg (7 kg max with packaging)

• Casing: Robust aluminium casing

STANDARD CONFIGURATIONS • HV = Heating Value

GAS 3160 HV: CO, CO2, CH4, H2, O2, CnHm, HV

GAS 3150 HV: CO, CO2, CH4, H2, O2, HV

GAS 3140 HV: CO, CO2, CH4, H2, HV