Reference quality upper-air measurements: GRUAN data

Preview:

Citation preview

Reference quality upper-air measurements: GRUAN data processing for the Vaisala RS92

radiosonde

Ruud Dirksen, Michael Sommer, Franz Immler, Dale Hurst, Rigel Kivi, Holger Vömel

GRUAN RS92 data product CIMO TECO 27-30 September 2016 www.gruan.org

Introduction

• Globally: 800 radiosoundings/day

High-res profiles of important ECVs (PTU, wind)

• Climate research requires reference quality data

Requirements for reference data (Keynote presentation G. Bodeker, Tuesday)

GRUAN RS92 data product CIMO TECO 27-30 September 2016 www.gruan.org

GCOS Reference Upper Air Network (2008)

• Traceability to an SI unit

• Uncertainty analysis

• Raw + meta-data

• Documentation

• Validation

GRUAN RS92 data product CIMO TECO 27-30 September 2016 www.gruan.org

Establishing reference quality

Best estimate+

Uncertainty

Uncertainty of input data

Traceable sensor

calibration

Transparent processing algorithm

Disregarded systematic effects

Black box software

Proprietary methods

GRUAN data product for RS92

GRUAN RS92 data product CIMO TECO 27-30 September 2016 www.gruan.org

Good performance, employed at 50% of GRUAN sitesVaisala calibration of PTU sensors traceable to SI standards

Proprietary algorithms

• Characterization & correction of instrument errors/biases• Vertically resolved error estimates• Well-documented

Error sources for: Temperature HumidityRadiation

Sensor time-lag

Calibration

Vaisala RS92 radiosonde

• PTU, wind

• 0-35km

• Interval: 1 s

• Ascend rate 5 m/s

• 2 RH sensors, alternate heating

• Capacitive sensors (T, U)

• Reflective, hydrophobic coating

GRUAN RS92 data product CIMO TECO 27-30 September 2016 www.gruan.org

T-sensor

GPS

RH-sensors

P-sensor

Radiation error: Laboratory experiments

Shadow RS92 records background temperature & ambient pressure

Simultaneous testing of 3 radiosondes

p=[3 hPa , ambient]

Difference illuminated – background radiosonde

GRUAN RS92 data product CIMO TECO 27-30 September 2016 www.gruan.org

Parameterisation of sensor response

GRUAN RS92 data product CIMO TECO 27-30 September 2016 www.gruan.org

Radiation error correction model

GRUAN RS92 data product CIMO TECO 27-30 September 2016 www.gruan.org

cloud

cloud

Actinic flux on sensor

• LUT actinic flux

• Streamer model, I+D

• Clear sky & cloudy

• Mid-lat atmosphere

T-correction profile

GRUAN RS92 data product CIMO TECO 27-30 September 2016 www.gruan.org

In processing: average of Vaisalaand GRUAN correction• Update in version 3

Temperature: uncertainties

GRUAN RS92 data product CIMO TECO 27-30 September 2016 www.gruan.org

Dominant uncertainty:Sensor orientation wrt Sun

RS92 dual soundingA

ltitu

de [k

m]

Night (N=17) Day (N=29)

Estimated uncertainty vs observed differences

GRUAN RS92 data product CIMO TECO 27-30 September 2016 www.gruan.org

RH: Dry biases

Heating of humidity sensor

• ∆T: radiation correction of T-sensor

• f: enhancement factor [6, 13] (laboratory experiments)

Uncertainties: ∆T, f

Temperature-dependent dry bias (-30, -70°C)

Based on RS92 - CFH comparison (empirical)

Max at 7% at -60°C (similar to Voemel, 2007)

Uncertainty: comparable to correction

GRUAN RS92 data product CIMO TECO 27-30 September 2016 www.gruan.org

RH: sensor time-lag

Low-pass filter

Flattens features in humidity profile

Relevant below -40°C (tropopause)

Correction: numeric inversion + a-posteriori filtering

GRUAN RS92 data product CIMO TECO 27-30 September 2016 www.gruan.org

RH: corrections & uncertainties

Dominant uncertainties:•Calibration•Cal. correction•Dry bias

IPW: ~4%

RS92: reliable RH profiles up to UTLS

Yangjiang 20 July 2010

GRUAN RS92 data product CIMO TECO 27-30 September 2016 www.gruan.org

Manufacturer’s ground check

RS92 recalibration at 0%RH.

Quality of desiccant

Introduces bias

GRUAN RS92 data product CIMO TECO 27-30 September 2016 www.gruan.org

Lindenberg

Ground check in SHC

• Traceability• 4% change over ~9

years• SHC readings enter

uncertainty budget• Future: use SHC to

scale profile

Lindenberg

GRUAN RS92 data product CIMO TECO 27-30 September 2016 www.gruan.org

Outlook

GRUAN RS92 data product CIMO TECO 27-30 September 2016 www.gruan.org

• Improve radiation correction

• Redesign architecture processing software (modular, versatile)

• Develop GRUAN data products for other radiosondes

• Change management RS92-RS41 transition P1(11)