Accurate Fast Handheld - iscan.es M_brochure4P_20170719_s.pdf · Pneumonia* *available late 2017...

Preview:

Citation preview

Fast

Handheld

Accurate

ichroma™ M is the small, battery operated true mobile

diagnostic device with new sensitive technology,

comparable to accuracy of auto-analyzer. With

respiratory tests panel and high sensitivity, ichroma™ M

will brings the test laboratory on your palm.

ichroma™ M Flu A+B

Room TemperatureNo need for refrigerator

SimpleMix, drop and insert.

Safety of DataAutomatic backup up to 100 results and LIS connectivity

Versatile All respiratory test (Flu, RSV*, Strep A*, and bacterial pneumonia*)

Test Procedure

Multi Mode

Flu A:Positive +

Flu B:Negative -

Single Mode

Specifications

Influenza A+B

Strep A*

PSV*

Pneumonia*

*available late 2017

Test itemsItem

Dimensions (LxWxH,mm)

Weight

Power

Battery life

Data output

Operation Temperature

Humidity

Memory

202.0 x 83.0 x 45

350 g

Battery (4 x AA)

100 tests

LCD / USB mini B

15 ~ 30°C(59~80℉)

20 - 80 %

100 tests

ichroma™ M Influenza A+B Accuracy

Flu A

vs CultureCulture

Total+ -

ichromaTM M ichromaTM M

ichromaTM M ichromaTM M

+ 53 4 57

- 3 119 122

Total 56 123 179

Sensitivity: 53/56 = 94.6%

Specificity: 119/123 = 96.7%

Total agreement: 172/179 = 96.1%

Flu A

vs PCRPCR

Total+ -

+ 74 0 74

- 1 125 126

Total 75 125 200

Sensitivity: 74/75 = 98.7%

Specificity: 125/125 = 100%

Total agreement: 199/200 = 99.5%

Flu B

vs CultureCulture

Total+ -

+ 53 3 56

- 3 120 123

Total 56 123 179

Sensitivity: 53/56 = 94.6%

Specificity: 120/123 = 97.6%

Total agreement: 173/179 = 96.6%

Flu B

vs PCRPCR

Total+ -

+ 72 0 72

- 3 125 128

Total 75 125 200

Sensitivity: 72/75 = 96.0%

Specificity: 125/125 = 100%

Total agreement: 197/200 = 98.5%

Distributed by 43, Geodudanji 1-gil, Dongnae-myeon, Chuncheon-si, Gang-won-do, Korea Tel: +82-33-243-1400 Fax: +82-33-243-9373

Boditech Med Inc.

For more information, please visit:

www.boditech.co.krFor sales related matters, please write to us at

sales@boditech.co.kr

Rev 02_ichroma M