4
Fast Handheld Accurate ichroma™ M is the small, battery operated true mobile diagnostic device with new sensitive technology, comparable to accuracy of auto-analyzer. With respiratory tests panel and high sensitivity, ichroma™ M will brings the test laboratory on your palm. ichroma™ M Flu A+B

Accurate Fast Handheld - iscan.es M_brochure4P_20170719_s.pdf · Pneumonia* *available late 2017 Item Test items Dimensions (LxWxH,mm) Weight Power Battery life Data output Operation

  • Upload
    vothuy

  • View
    214

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Accurate Fast Handheld - iscan.es M_brochure4P_20170719_s.pdf · Pneumonia* *available late 2017 Item Test items Dimensions (LxWxH,mm) Weight Power Battery life Data output Operation

Fast

Handheld

Accurate

ichroma™ M is the small, battery operated true mobile

diagnostic device with new sensitive technology,

comparable to accuracy of auto-analyzer. With

respiratory tests panel and high sensitivity, ichroma™ M

will brings the test laboratory on your palm.

ichroma™ M Flu A+B

Page 2: Accurate Fast Handheld - iscan.es M_brochure4P_20170719_s.pdf · Pneumonia* *available late 2017 Item Test items Dimensions (LxWxH,mm) Weight Power Battery life Data output Operation

Room TemperatureNo need for refrigerator

SimpleMix, drop and insert.

Safety of DataAutomatic backup up to 100 results and LIS connectivity

Versatile All respiratory test (Flu, RSV*, Strep A*, and bacterial pneumonia*)

Test Procedure

Multi Mode

Flu A:Positive +

Flu B:Negative -

Single Mode

Page 3: Accurate Fast Handheld - iscan.es M_brochure4P_20170719_s.pdf · Pneumonia* *available late 2017 Item Test items Dimensions (LxWxH,mm) Weight Power Battery life Data output Operation

Specifications

Influenza A+B

Strep A*

PSV*

Pneumonia*

*available late 2017

Test itemsItem

Dimensions (LxWxH,mm)

Weight

Power

Battery life

Data output

Operation Temperature

Humidity

Memory

202.0 x 83.0 x 45

350 g

Battery (4 x AA)

100 tests

LCD / USB mini B

15 ~ 30°C(59~80℉)

20 - 80 %

100 tests

ichroma™ M Influenza A+B Accuracy

Flu A

vs CultureCulture

Total+ -

ichromaTM M ichromaTM M

ichromaTM M ichromaTM M

+ 53 4 57

- 3 119 122

Total 56 123 179

Sensitivity: 53/56 = 94.6%

Specificity: 119/123 = 96.7%

Total agreement: 172/179 = 96.1%

Flu A

vs PCRPCR

Total+ -

+ 74 0 74

- 1 125 126

Total 75 125 200

Sensitivity: 74/75 = 98.7%

Specificity: 125/125 = 100%

Total agreement: 199/200 = 99.5%

Flu B

vs CultureCulture

Total+ -

+ 53 3 56

- 3 120 123

Total 56 123 179

Sensitivity: 53/56 = 94.6%

Specificity: 120/123 = 97.6%

Total agreement: 173/179 = 96.6%

Flu B

vs PCRPCR

Total+ -

+ 72 0 72

- 3 125 128

Total 75 125 200

Sensitivity: 72/75 = 96.0%

Specificity: 125/125 = 100%

Total agreement: 197/200 = 98.5%

Page 4: Accurate Fast Handheld - iscan.es M_brochure4P_20170719_s.pdf · Pneumonia* *available late 2017 Item Test items Dimensions (LxWxH,mm) Weight Power Battery life Data output Operation

Distributed by 43, Geodudanji 1-gil, Dongnae-myeon, Chuncheon-si, Gang-won-do, Korea Tel: +82-33-243-1400 Fax: +82-33-243-9373

Boditech Med Inc.

For more information, please visit:

www.boditech.co.krFor sales related matters, please write to us at

[email protected]

Rev 02_ichroma M