YSI IQ SensorNetYSI IQ SensorNetContinuous Process Monitoring
& Control Instrumentation
September 19th, 2013Presented by Laura St. Pierre
• Quick overview of IQ SensorNet
• New IQ SensorNet Capabilities
2
IQ SensorNet Systems
182 Controller System 2020 XT Controller System182 Controller System
• 1 - 4 parameters
2020 XT Controller System
• Sensor network
• 1 - 20 parameters
3
IQ SensorNet SensorsOptical UV:
- Nitrate
- Nitrite
- COD, BOD
TOC DOCTOC, DOC
- UVT
- SAC
ConductivityDO
- TSSSludge BlanketLevel
pH / ORP Turbidity / Susp. SolidsISE:
- Ammonium
DO Level
4
Ammonium, Nitrate, Chloride, Potassium
IQ Sensors
Detachable, universal sensor cable
5
FDO – Optical-based Dissolved Oxygen
FDO ith l dFDO with angled cap and smart sensor cap checkcheck.
6
UltraClean System
Maintenance FREE lt i lf l iultrasonic self-cleaning
Available on UV UV/VISAvailable on UV, UV/VIS, TSS and Turbidity probesp
Air cleaning available for h bother probes
7
UltraCleanTM – Ultra Sonic Cleaning
without cleaning system (30 days) with cleaning system (30 days)
A clean sensor ensures accurate measurements!A clean sensor ensures accurate measurements!A clean sensor ensures accurate measurements! A clean sensor ensures accurate measurements!
MMaintenanceaintenance--free sensor lowers free sensor lowers ownership and ownership and operational costs.operational costs.
8
Individual Electrode Replacement
Individual electrode replacementreplacement• Not often Lifetime > 18 months
• Recognized automaticallyg y
IFL - Sludge blanket level Sensor
Smart Sensor learns your process.
raw data filter, alignment, low passing result
p
Smart signal processing for reliable sludge levelreliable sludge level measurements + advanced user settings.
Smart filter removes readings from:
• Passing debris, layer clouds
f ll f l d l d• Rising & falling of sludge layer due to passing rakes
10
IQ SensorNet Network cable
All-in-one cable for power and
• Fully protected against EMCpower and communication
• 1 year warranty, Extremely robust
• Integrated lightning protection
• Network cable provides power p pand communication
11
Typical Analog Installation
Typical Modbus Installation
IQ SensorNet Solution
IQ SensorNet Network Solution
Easily Add Measuring Locations
Modular System
Provide:•Sensor connections
d l
Stack mounting – max. 3 modules
+ 3rd party analog•Power (AC & DC)•Communication•Communication
CurrentRelaysRelaysModbusProfibus
S i ll i+ more• Backup Controller
Smart installation…Stack mounting – no cables
17
USB-Interface
• Download logged data
• Backup configuration files
• System Update
• Electronic-key-functionElectronic key function
18
Remote Control via USB
USB-RS232 adapter for pmodem
USB-LAN adapter for LAN connection (WLAN)
Full access to network from a distance for data transfer and adjusting settings
19
MC2 Module with Ethernet and USB
Functions as controller so 2020 can be a display and backup controller
LAN
• Can be used inside or outdoors
USB
• Data transfer• Two separate ports allow
you to continue to communicate via Ethernet while updating the system
• Firmware updates
• Electronic key
via USB.
Ethernet Communication Technology:
• Direct communication with PLC`s via Ethernet/IP protocolDirect communication with PLC s via Ethernet/IP protocol
• Modbus TCP/IP (ODVA certified)
• Remote-Control with PC Terminal Software via cellular-Router
d h h b• View measuring data with your Smart-Phone via Web
New Parameters for UV/VIS probes
•SAC, COD,
(Total andMeasures broad spectrum:
f
(Total and Soluble)
254 wavelengths for each measurement
•BOD, DOC, TOC •
•UVT 254•UVT – 254•TSS
•Nitrate
200 nm200 nm 720 nm720 nm
New UV SensorsNew UV Sensors
NiCaVis 701 IQ NI
NiCaVis 705 IQ NINiCaVis 705 IQ NI
Carbon parameters, UVT-254, Nitrate and Nitrite,
NewNew UV Spectral Sensors UV Spectral Sensors NiCaVisNiCaVis®® 70X IQ NI70X IQ NINiCaVisNiCaVis 70X IQ NI 70X IQ NI
•resolution < 1 nm -Requirement forRequirement for differentiation of NO3
-
and NO2-:
•Measurement of 256 wavelength
•Measuring range 200-390 nm
wavelength [nm]
23
UV-VIS Scans 256 Wave Lengths
256 wave lengths compared to single wave lengthto single wave length measurement.
• Better correlation• Better correlation
• Better turbidity correction
• Better compensation for a• Better compensation for a changing wastewater process
More information, more reliable data
Measuring Principle
Algorithms based on correlations with reference measurements are stored in the sensor.
Calculating concentration:
measurements are stored in the sensor.
1. Recording spectral absorbance.
2. Calculation of concentrations according to2. Calculation of concentrations according to algorithms.
COD, TOC, BOD, COD, TOC, BOD, NONO33--N, TSSN, TSS mg/lmg/lalgorithmsalgorithms
Customer application
• Environmental Operating Solutions, EOSi, sells MicroC asSolutions, EOSi, sells MicroC as a supplemental carbon source for denitrification
• Uses new UV sensor’s nitrite data to optimize denitrificationand phosphorus removaland phosphorus removal
• Poster available on our websitewebsite
26
NEW! Phosphate Analyzer
P 700 IQ Analyzer
New Sensor for the IQ SENSOR NET
Used with:
MIQ/MC2MIQ/MC2MIC/TC 2020 XT
27
Vandomolybdate (yellow) method
P 700 IQ
Orthophosphate (PO4-P)
V d / M l bd M h dVanadate / Molybdate -Method
(yellow-method)
samplei h PO 3
reagent
with PO43-
mixing photometeryellow coloration
420g
=420 nm
NEW Ortho Phosphate
Filter moduleFilter module
Extendable rail
E l iEasy cleaning and exchange of filter
29
P 700 IQl blTwo measurement ranges available in every unit:
A: 0.05 … 15.00 mg/l PO4-P
B: 1 … 50 mg/l PO4-P
Resolution
A: 0.05 mg/l PO4-P
B: 1 mg/l PO4-PB: 1 mg/l PO4 P
Accuracy
A: ± 2 %, ± 0.05 mg/l
B 2 % 1 /lB: ± 2 %, ± 1 mg/l
Integrated pump
Permeate pump integrated in housing
Short tubes reduce risk of freezing
Easy cleaning
Wide temperature range
30
-20°C … +40°C using Outdoor Option
Outdoor option includes Climate Package
P 700 IQ
Low reagent consumption
Mode A: 2 5 l for 8 monthsMode A: 2.5 l for 8 months
Mode B: 2.5 l for 4 months
R d d t f hiReduced cost of ownership
Automatic Calibration
User-defined interval
Reduced maintenance
31
ApplicationsP 700 IQ P 700 IQ P 700 IQ
Primary Effluent
Aeration basins•Regulation of phosphate elimination
Effluent• process control
•process optimization
• process control
Dr. Rob and our Educational Offerings
• Wastewater online instrumentation: how to useinstrumentation: how to use, install and get the best data.
• Obtain CEUs while you learn something!
• Let us know if we can id d ti lprovide an educational
seminar for you!
33
Videos and other resources
34
Thank you for your attention!
Visit us at: WEFTEC, booth 1443 d i1443 and at ysi.com
Register to win an IQ SensorNetSensorNet
Sign up for our wastewater eNewslettereNewsletter
Email us at: [email protected]@ysi.com
Call us at: 800-897-4151,
937 767 7241937-767-7241