Basics of - HPLC
- Kaumil Bhavsar Date -21/08/07
Index
What is HPLC ? Applications Types of HPLC HPLC Instrumentation Solvent system Pumping system Sample injector system Column Detectors P9 RP-HPLC salient features
What is HPLC ? Originally referred to as
High Pressure Liquid Chromatography Now more commonly called
High Performance Liquid Chromatography
HPLC is an analytical technique used to separate component of mixture by using variety of chemical interaction between the analyte and the chromatography column.
Why HPLC? Automated version of LC Provides enhanced separations in short time duration. It’s a most practiced quantitative, chromatographic technique which
can separates wide range of molecule types and sizes.
Applications of HPLC Pharmaceutical / Biopharmaceutical Pharmaceutical quality control. Shelf-life determinations of pharmaceutical products. Identification of counterfeit drug products. Complex molecules separation.
Environmental Biomonitoring of pollutents. Water monitoring - Phenol content and toxic componants checking.
Clinical Analysis of antibiotics and blood substances. Detection of endogenous neuropeptides in brain extracellular fluids.
Food and Flavor Sugar analysis in fruit juices. Ensuring soft drink consistency and quality.
Types of HPLC Normal Phase: Separation of polar analytes by partitioning onto a
polar, bonded stationary phase using non polar mobile phase.
Reversed Phase: Separation of non-polar analytes by partitioning onto a non-polar, bonded stationary phase using polar mobile phase.
Adsorption: In Between Normal and Reversed. Separation of moderately polar analytes using adsorption onto a pure stationary phase (e.g. alumina or silica)
Ion Exchange Chromatography: Separation of organic and inorganic ions by their partitioning onto ionic stationary phases bonded to a solid support.
Size Exclusion Chromatography: Also called as Gel filtration chromatography. Separation of molecules based on their size.
Selection of Chromatography
Separation based on selection of column material
Separation based on selection of column material
HPLC instrumentation
HPLC components
1) Solvent system
2) Pumping system
3) Sample injection system
4) Column
5) Detector
DetectorMobile phase Pump Injection Valve Separation column=> => => =>
Solvent system Mobile phase reservoirs: Several reservoirs – Use to carry mobile
phase Degassing: Remove dissolved gas - interfering detection Protect band spreading Sparging: fine bubble of gas (He) Vacuum pumping. Dust removal: Dust interference with detection, column clogging,
damage pumping system Millipore filter under vacuum Isocratic elution: constant composition Gradient elution: different solvent systems during elution, continuous
change or step wise, solvent proportion valve
Pumping system
Requirements for HPLC pump.
High pressure (up to 6000 psi)
• pulse free, prevents remixing of solutes
• control flow rate from 0.1 to 10 mL/min
Types of HPLC pumps
1) Reciprocating pumps: Most commonly used.
Disadvantage: pulses from single piston.
2) Syringe pumps: Generates pulse free output.
Disadvantage: Limited mobile phase capasity.
Reciprocating pump : Single piston pump
Reciprocating pump : Double piston pump
Syringe pump
Sample injection system Sample Injection system: Limit of precision of HPLC Sample size: 0.5 ~ 500 μL No interference with the
pressure Sample loop, 1 ~ 100 μL, Auto sampler: inject
continuously Volume uptake: 1 μL – 1 mL Controlled temperature
environment
Sample loading & Injection.
Column Made of Stainless steel tubing for high pressure. Analytical column: straight, L(5 ~ 25 cm), d (3 ~ 5
mm), dp(3~5 μm). N (40 k ~ 70 k plates/m) Microcolumn: L (3 ~ 7.5 cm), d (1 ~ 5 mm), dp: (3 ~
5 μm) , N: ~100k plates/m, high speed and minimum solvent consumption
Guard column: protects analytical column, similar packing, remove particulate matter and contamination.
Temperature control: 0.1 °C - 150 °C. Column packing: silica, alumina, a polystyrene-
divinylbenzene, synthetic or an ion-exchange resin Pellicular particle: original, spherical,nonporous
beads,proteins and large biomolecules separation (dp: 5 μm)
Porous particle: commonly used, dp: 3 ~ 10 μm. Narrow size distribution, porous microparticle coated with thin organic films.
Detectors Ideal Detector Properties: High Sensitivity Universality or predictable specificity Large linear response range Low dead volume Non-Destructive Insensitive to temperature & mobile phase Continuous operation Reliable and easy to use No single detector fits all these criteria
Types of HPLC detectors
UV-Vis Works w/all molecules Non-specific; complex samples; absorption wavelength
DAD Works for all wavelengths High LOD
Fluorescence Very specific; low LOD Not everything fluoresces
IR Works w/all molecules Many solvents IR active
Refractive Index Works w/nearly all molecules
Temperature sensitive; high LOD
Scattering Uniform response; 5ng/25L LOD
Non-specific; interference from solvent
Electrochemical Commercially available Non-specific; high LOD
Mass Spec Low LOD; analyte identification
Ability to ionize analyte
Name Advantage Disadvantage
UV-VIS detector. Single Beam UV-VIS instrument with a flow-
through flow cell (cuvette) Usually typical UV-VIS lamps used having
254 nm default wavelenth Can be set to other wavelengths (most) Simple filter detectors no longer widely
used Non-destructive, not-universal not all compounds absorb light can pass sample through several cells at
several different wavelenghts Usually zeroed at the start of each run
using an electronic software command. You can have real-time zeroing with a reference cell.
Diode array detector (DAD)
Refractive Index Detector.
ELSD (Evaporative Light Scattering Detector)
P9 RP- HPLC Salient features. Reverse phase HPLC. 15 min method, RT- Between 4 to 6 min. Mobile phase : A - Water + TFA
B - ACN + TFA Column: Brand Name - Discovery Type of Column:Reverse phase C18 column Specification of ColumnDiameter: 4.6mm Length: 50mm Porosity:
300•A, Particle size: 5 Micron Guard column Brand Name - Zorbex 300 SB (Agilent) Type of Column:Reverse phase Zorbax 300 SB-C18 Specification of ColumnDiameter: 4.6mm, Length: 35 mmPorosity:
300•A, Particle size: 5 Micron
Typical P9 RP-HPLC peak (Agilent-1200).
min0 2 4 6 8 10 12 14
mAU
-100
0
100
200
300
400
500
*VWD1 A, Wavelength=214 nm (10APRIL07\10HP250407 P9STD119\10HP250407B.D - 10APRIL07\10HP250407 P9STD119\B250407A.D)
4.65
4