Upload
miza-kamaruzzaman
View
751
Download
9
Embed Size (px)
Citation preview
Chapter 4: Techniques in Biochemical Analysis1
Chapter 4: Techniques in Biochemical Analysis
2
BIO 300BIO 300BIOLOGICAL TECHNIQUES AND SKILLSBIOLOGICAL TECHNIQUES AND SKILLS
SARINI BINTI AHMAD WAKIDSARINI BINTI AHMAD WAKIDFACULTY OF APPLIED SCIENCEFACULTY OF APPLIED SCIENCE
BIO 300BIO 300BIOLOGICAL TECHNIQUES AND SKILLSBIOLOGICAL TECHNIQUES AND SKILLS
SARINI BINTI AHMAD WAKIDSARINI BINTI AHMAD WAKIDFACULTY OF APPLIED SCIENCEFACULTY OF APPLIED SCIENCE
Chapter 4: Techniques in Biochemical Analysis
3
CHAPTER 4CHAPTER 4Techniques in Biochemical Techniques in Biochemical
AnalysisAnalysis
Chapter 4: Techniques in Biochemical Analysis 4
Chromatography is a technique for separating mixtures into their components in order to analyze, identify, purify, and/or quantify the mixture or components.
Separate
• Analyze
• Identify
• Purify
• QuantifyComponentsMixture
What is Chromatography?
Chapter 4: Techniques in Biochemical Analysis 5
Chromatography
Chromatography is a method of separating a mixture of molecules depending on their distribution between a mobile phase and a stationary phase.
The mobile phase (also known as solvent) may be either liquid or gas.
The stationary phase (also known as sorbent) can be either a solid or liquid, a liquid stationary phase is held stationary by a solid.
The solid holding the liquid stationary phase is the support or matrix.
The molecules in the mixture to be separated are the solutes.
Chapter 4: Techniques in Biochemical Analysis
6
Chromatography is used by scientists to:
• Analyze – examine a mixture, its components, and their relations to one another
• Identify – determine the identity of a mixture or components based on known components
• Purify – separate components in order to isolate one of interest for further study
• Quantify – determine the amount of the a mixture and/or the components present in the sample
Uses for Chromatography
Chapter 4: Techniques in Biochemical Analysis 7
Uses for ChromatographyReal-life examples of uses for
chromatography:
• Pharmaceutical Company – determine amount of each chemical found in new product
• Hospital – detect blood or alcohol levels in a patient’s blood stream
• Law Enforcement – to compare a sample found at a crime scene to samples from suspects
• Environmental Agency – determine the level of pollutants in the water supply
• Manufacturing Plant – to purify a chemical needed to make a product
Chapter 4: Techniques in Biochemical Analysis 8
Detailed Definition:Chromatography is a laboratory technique that
separates components within a mixture by using the differential affinities of the components for a mobile medium and for a stationary adsorbing medium through which they pass.
Terminology:• Differential – showing a difference, distinctive
• Affinity – natural attraction or force between things
• Mobile Medium – gas or liquid that carries the components (mobile phase)
• Stationary Medium – the part of the apparatus that does not move with the sample (stationary phase)
Definition of Chromatography
Chapter 4: Techniques in Biochemical Analysis 9
Simplified Definition:Chromatography separates the
components of a mixture by their distinctive attraction to the mobile phase and the stationary phase.
Explanation:• Compound is placed on stationary phase• Mobile phase passes through the stationary
phase• Mobile phase solubilizes the components• Mobile phase carries the individual
components a certain distance through the stationary phase, depending on their attraction to both of the phases
Definition of Chromatography
Chapter 4: Techniques in Biochemical Analysis
10
Illustration of Chromatography
Components
Affinity to Stationary Phase
Affinity to Mobile Phase
Blue ---------------- Insoluble in Mobile Phase
Black
Red
Yellow
Mixture Components
Separation
Stationary Phase
Mobile Phase
Chapter 4: Techniques in Biochemical Analysis
11
Chapter 4: Techniques in Biochemical Analysis 12
• Liquid Chromatography – separates liquid samples with a liquid solvent (mobile phase) and a column composed of solid beads (stationary phase)
• Gas Chromatography – separates vaporized samples with a carrier gas (mobile phase) and a column composed of a
liquid or of solid beads (stationary phase)
• Paper Chromatography – separates dried liquid samples with a liquid solvent (mobile phase) and a paper strip (stationary phase)
• Thin-Layer Chromatography – separates dried liquid samples with a liquid solvent (mobile phase) and a glass plate covered with a thin layer of alumina or silica gel (stationary phase)
Types of Chromatography
Chapter 4: Techniques in Biochemical Analysis
13
Types of chromatography
• Partition chromatography• Adsorption chromatography• Gel filtration• Ion exchange chromatography
Chapter 4: Techniques in Biochemical Analysis
14
(A) uses charge, (B) uses pores, and (C) uses covalent bonds to create the differential affinities among the mixture components for the stationary phase.
Chapter 4: Techniques in Biochemical Analysis
15
Partition chromatography
• The distribution of solutes between two immiscible phases.
• The solute will distribute it self between the two phases according to its solubility in each phase, this is called partitioning.
Chapter 4: Techniques in Biochemical Analysis 16
Examples of partition chromatography
The two most common types of partition chromatography are thin layer chromatography and paper chromatography.
In both cases the stationary phase is a liquid bound to a matrix. In paper chromatography the stationary phase are water molecules
bound to a cellulose matrix. In TLC, the stationary phase is the solvent added to the support to
form the thin layer so the solvent gets bound to the matrix (support).
Partition chromatography is mainly used for separation of molecules of small molecular weight.
Chapter 4: Techniques in Biochemical Analysis
17
Paper chromatography
• The cellulose support contains a large amount of bound water.
• Partitioning occurs between the bound water which is the stationary phase and the solvent which is the mobile phase.
Chapter 4: Techniques in Biochemical Analysis 18
Experimental procedure for paper chromatography A small volume of a solution of a mixture to be separated or identified is
placed at a marked spot (origin) on a sheet or strip of paper and allowed to dry.
The paper is then placed in a closed chamber and one end is immersed in a suitable solvent.
The solvent is drawn (moved) through the paper by capillary action. As the solvent passes the origin, it dissolves the sample and moves the
components in the direction of flow. After the solvent front has reached a point near the other end of the paper,
the sheet or strip is removed and dried. The spots are then detected and their positions marked. The ratio of the distance moved by a solute to the distance moved by the
solvent = Rf.
The Rf. is always less than one.
Chapter 4: Techniques in Biochemical Analysis 19
Chromatogram Once a sample is applied on TLC or paper, it’s called
chromatogram. Paper chromatogram can be developed either by
ascending or descending solvent flow. Descending chromatography is faster because gravity
helps the solvent flow. Disadvantages : it’s difficult to set the apparatus. Ascending is simple and inexpensive compared with
descending and usually gives more uniform migration with less diffusion of the sample "spots".
Chapter 4: Techniques in Biochemical Analysis 20
Detection of spots
Spots in paper chromatograms can be detected in 4 different ways:
1. By their natural color
2. By their fluorescence
3. By their chemical reactions that take place after the paper has been sprayed with various reagents for example: during paper chromatography of amino acids, the chromatograms are sprayed with ninhydrin.
4. By radioactivity
Chapter 4: Techniques in Biochemical Analysis
21
Identification of spots
• The spots are usually identified by comparing of standards of known Rf values.
Chapter 4: Techniques in Biochemical Analysis 22
Thin layer chromatography Paper chromatography uses paper which can be
prepared from cellulose products only. In TLC, any substance that can be finely divided
and formed into a uniform layer can be used. Both organic and inorganic substances can be
used to form a uniform layer for TLC. Organic substances include: cellulose, polyamide,
polyethylene Inorganic: silica gel, aluminum oxide and
magnesium silicate
Chapter 4: Techniques in Biochemical Analysis
23
TLC
• The stationary phase is the solvent used to form a layer of sorbent spread uniformly over the surface of a glass or plastic plate
Chapter 4: Techniques in Biochemical Analysis
24
Advantages of TLC over paper chromatography• Greater resolving power because
there is less diffusion of spots.• Greater speed of separation• Wide choice of materials as
sorbents
Chapter 4: Techniques in Biochemical Analysis 25
The separation of compounds by chromatography depends on several factors:
Partition of a solute between a moving solvent phase and a stationary aqueous phase. The solute moves in the direction of a solvent flow at a rate determined by the solubility of the solute in the moving phase. Thus a compound with high mobility is more attracted to the moving organic phase than to the stationary phase.
Chapter 4: Techniques in Biochemical Analysis 26
Cont..
Ion exchange effect: any ionized impurities in the support medium will tend to bind or attract oppositely charged ions (solutes) and will therefore reduce the mobility of these solutes.
Temperature: Since temperature can effect the solubility of the solute in a given solvent temperature is also an important factor.
Chapter 4: Techniques in Biochemical Analysis 27
The molecular weight of a solute also affects the solubility and hence chromatographic performance.
Adsorption of compound (solute) onto support medium: Although the support medium (silica gel) is theoretically inert, this isn't always the case. If a solute tends to bind to the support medium this will slow down its mobility in the solvent system.
The composition of the solvent: since some compounds are more soluble in one solvent than in the other, the mixture of solvents used will affect the separation of compounds.
Chapter 4: Techniques in Biochemical Analysis 28
Expression of the results
The term "Rf" (relative flow) is used to express the performance of a solute in a given solvent system /support medium. The term Rf value may be defined as the ratio of the distance the compound migrates to the distance the solvent migrates. Rf value is constant for a particular compound, solvent system and insoluble matrix.
Rf= Distance of migration of solute
Distance moved by solvent
Chapter 4: Techniques in Biochemical Analysis 29
Rf values qualitative results of TLC
expressed as fractions of 1.0 can be expressed from Rf values (eg Rf x 100) no more than two decimal places
due to inaccuracy of physical measurement may not be reproducible
only give an indication of possible nature of unknown complete identification only obtained if spot is eluted and
micro-scale physical measurements done (MS, UV, IR) standard references should always be used on same
plate for comparison most sprays produce differential colours of fluorescence colour test provides extra evidence with distance
migration
Chapter 4: Techniques in Biochemical Analysis 30
Chapter 4: Techniques in Biochemical Analysis
31
Principles of Paper Chromatography
• Capillary Action – the movement of liquid within the spaces of a porous material due to the forces of adhesion, cohesion, and surface tension. The liquid is able to move up the filter paper because its attraction to itself is stronger than the force of gravity.
• Solubility – the degree to which a material (solute) dissolves into a solvent. Solutes dissolve into solvents that have similar properties. (Like dissolves like) This allows different solutes to be separated by different combinations of solvents.
Separation of components depends on both their solubility in the mobile phase and their differential affinity to the mobile phase and the stationary phase.
Chapter 4: Techniques in Biochemical Analysis
32
Paper Chromatography Experiment
What Color is that Sharpie?
Chapter 4: Techniques in Biochemical Analysis
33
Overview of the Experiment
Purpose: To introduce students to the principles and terminology of chromatography and demonstrate separation of the dyes in Sharpie Pens with paper chromatography.
Time Required: Prep. time: 10 minutesExperiment time: 45 minutes
Chapter 4: Techniques in Biochemical Analysis
34
• 6 beakers or jars• 6 covers or lids • Distilled H2O• Isopropanol• Graduated cylinder• 6 strips of filter paper• Different colors of
Sharpie pens• Pencil• Ruler• Scissors• Tape
Materials List
Chapter 4: Techniques in Biochemical Analysis
35
Preparing the Isopropanol Solutions
• Prepare 15 ml of the following isopropanol solutions in appropriately labeled beakers:
- 0%, 5%, 10%, 20%, 50%, and 100%
Chapter 4: Techniques in Biochemical Analysis
36
Preparing the Chromatography Strips
• Cut 6 strips of filter paper
• Draw a line 1 cm above the bottom edge of the strip with the pencil
• Label each strip with its corresponding solution
• Place a spot from each pen on your starting line
Chapter 4: Techniques in Biochemical Analysis
37
Developing the Chromatograms
• Place the strips in the beakers
• Make sure the solution does not come above your start line
• Keep the beakers covered• Let strips develop until
the ascending solution front is about 2 cm from the top of the strip
• Remove the strips and let them dry
Chapter 4: Techniques in Biochemical Analysis
38
Developing the Chromatograms
Chapter 4: Techniques in Chapter 4: Techniques in Biochemical AnalysisBiochemical Analysis 3939
Developing the Developing the ChromatogramsChromatograms
Chapter 4: Techniques in Biochemical Analysis
40
Chapter 4: Techniques in Chapter 4: Techniques in Biochemical AnalysisBiochemical Analysis 4141
Observing the Observing the ChromatogramsChromatograms
Concentration of Isopropanol
0% 20% 50% 70% 100%
Chapter 4: Techniques in Biochemical Analysis
42
Black Dye
Concentration of Isopropanol
0% 20% 50% 70% 100%
1. Dyes separated – purple and black
2. Not soluble in low concentrations of isopropanol
3. Partially soluble in concentrations of isopropanol >20%
Chapter 4: Techniques in Chapter 4: Techniques in Biochemical AnalysisBiochemical Analysis 4343
Blue DyeBlue Dye
Concentration of Isopropanol
0% 20% 50% 70% 100%
1. Dye separated – blue2. Not very soluble in low
concentrations of isopropanol
3. Completely soluble in high concentrations of isopropanol
Chapter 4: Techniques in Chapter 4: Techniques in Biochemical AnalysisBiochemical Analysis 4444
Green DyeGreen Dye
Concentration of Isopropanol
0% 20% 50% 70% 100%
1. Dye separated – blue and yellow
2. Blue – Soluble in concentrations of isopropanol >20%
3. Yellow – Soluble in concentrations of isopropanol >0%
Chapter 4: Techniques in Chapter 4: Techniques in Biochemical AnalysisBiochemical Analysis 4545
Red DyeRed Dye1. Dyes separated – red and yellow1. Dyes separated – red and yellow
2. Yellow –soluble in low concentrations of isopropanol and 2. Yellow –soluble in low concentrations of isopropanol and
less soluble in high concentrations of isopropanolless soluble in high concentrations of isopropanol
Concentration of Isopropanol
0% 20% 50% 70% 100%
3. Red – slightly soluble in low concentrations of isopropanol, and more soluble in concentrations of isopropanol >20%
Chapter 4: Techniques in Biochemical Analysis
46
Alternative Experiments
• Test different samples:– Other markers, pens, highlighters– Flower pigments– Food Colors
• Test different solvents:– Other alcohols: methanol, ethanol,
propanol, butanol
• Test different papers:– Coffee filters– Paper towels– Cardstock– Typing paper
Chapter 4: Techniques in Biochemical Analysis
47
Alternative Experiments
Chapter 4: Techniques in Biochemical Analysis
48
Alternative Experiments
Chapter 4: Techniques in Biochemical Analysis
49
Alternative Experiments
Chapter 4: Techniques in Biochemical Analysis
50
Chromatography Chromatography InstrumentsInstruments
Chromatography Chromatography InstrumentsInstruments
Chapter 4: Techniques in Biochemical Analysis
51
• Chromatography techniques• Affinity Chromatography (AC)• Hydrophobic Interaction Chromatography (HIC)• Ion Exchange Chromatography (IEC)• Gel Filtration (GF)• Capillary electrochromatography (CEC)
Chapter 4: Techniques in Biochemical Analysis
52
Affinity Chromatography
Surface bound with
Epoxy, aldehyde or aryl ester groups
Metal Interaction Chromatography
Surface bound with
Iminodiacetic acid + Ni2+/Zn2+/Co2+
Affinity Chromatography
(Christian G. Huber, Biopolymer Chromatography, Encylcopedia in analytical chemistry, 2000)
Chapter 4: Techniques in Biochemical Analysis
53
Metal Interaction Chromatography (AC)
Points to Note:
1. Avoid chelating agents
2. Increasing incubation time
3. Slow gradient elution
(www.qiagen.com)
Chapter 4: Techniques in Biochemical Analysis 54
Affinity Chromatography
Binding Capacity (mg/ml) medium 12mg of histag proteins (MW= 27kDa)
Depends on Molecular weight
Degree of substitution /ml medium~15mol Ni2+
Backpressure ~43psiChange the guard column filter
(Christian G. Huber, Biopolymer Chromatography, Encylcopedia in analytical chemistry, 2000)
Chapter 4: Techniques in Biochemical Analysis 55
Biopolymer (phenyl agarose - Binding Surface)
Driving force for hydrophobic adsorptionWater molecules surround the analyte and the binding surface.
When a hydrophobic region of a biopolymer binds to the surface of a mildly hydrophobic stationary phase, hydrophilic water molecules are effectively released from the surrounding hydrophobic areas causing a thermodynamically favorable change in entropy.
Temperature plays a strong role
Ammonium sulfate, by virtue of its good salting-out properties and high solubility in water is used as an eluting buffer
Hydrophobic Interaction Chromatography
Hydrophobic region
(Christian G. Huber, Biopolymer Chromatography, Encylcopedia in analytical chemistry, 2000)
Chapter 4: Techniques in Biochemical Analysis
56
Fractogel matrix is a methacrylate resin upon which polyelectrolyte Chains (or tentacles) have been grafted. (Novagen)
Ion Exchange Chromatography
Globular Protein
Deformation due to interaction with conventional ion exchanger
Maintenance of conformation while interacting with tentacle ion exchanger
(www.novagen.com)
Chapter 4: Techniques in Biochemical Analysis
57
Gel Filtration
(http://lsvl.la.asu.edu/resources/mamajis/chromatography/chromatography.html)
Chapter 4: Techniques in Biochemical Analysis 58
Capillary Electrochromatography
• CEC is an electrokinetic separation technique
• Fused-silica capillaries packed with stationary phase
• Separation based on electroosmotically driven flow
• Higher selectivity due to the combination of chromatography and
electrophoresis
Fused silica tube filled with porous methacrylamide-stearyl methacrylate-
dimethyldiallyl ammonium chloride monolithic polymers, 80 x 0.5mm i.d.,
5.5kV. High Plate count ~ 400,000
Height Equivalent to a Theoretical Plate /Plate Count (HETP) H = L/Nnumber of plates N = 16(t/W)2 where L = column length, t = retention time, and W = peak width at baseline
(http://www.capital-hplc.co.uk)
Chapter 4: Techniques in Biochemical Analysis
59
CEC columns AC, IEC columns
CEC column NP, RP columns
Chapter 4: Techniques in Biochemical Analysis 60
Schematic of a Multi-dimensional Separation System
Chapter 4: Techniques in Biochemical Analysis 61
Fast Protein Liquid Chromatograph (FPLC)
1
2
3
5
4
• No air bubbles (Priming)• Use degassed buffers
Injector Module
Column Inlet
DetectorFractionCollector
(www.pharmacia.com)
Chapter 4: Techniques in Biochemical Analysis 62
Chromatography systems
ÄKTAprime:
simple automated purification
ÄKTAFPLC: high performance
purification of proteins & other biomolecules
ÄKTApurifier: high performance
purification and characterizationÄKTAexplorer: for fast method
development and scale-upÄKTApilot: rapid process
development and pilot-scale
ÄKTAxpress: for high throughput tagged
protein purification
Chapter 4: Techniques in Biochemical Analysis 63
High Performance Liquid Chromatography (HPLC)
What is HPLC? Types of Separations Columns and Stationary Phases Mobile Phases and Their Role in Separations Injection in HPLC Detection in HPLC
Variations on Traditional HPLC Ion Chromatography Size Exclusion Chromatography
Chapter 4: Techniques in Biochemical Analysis 64
What is HPLC? High Performance Liquid Chromatography
High Pressure Liquid Chromatography (usually true]
Hewlett Packard Liquid Chromatography (a joke)
High Priced Liquid Chromatography (no joke)
HPLC is really the automation of traditional liquid chromatography under conditions which provide for enhanced separations during shorter periods of time!
Probably the most widely practiced form of quantitative, analytical chromatography practiced today due to the wide range of molecule types and sizes which can be separated using HPLC or variants of HPLC!!
Chapter 4: Techniques in Biochemical Analysis 65
Chapter 4: Techniques in Biochemical Analysis 66
Chapter 4: Techniques in Biochemical Analysis 67
Types of HPLC Separations (partial list) Normal Phase: Separation of polar analytes by partitioning onto a polar,
bonded stationary phase.
Reversed Phase: Separation of non-polar analytes by partitioning onto a non-polar, bonded stationary phase.
Adsorption: In Between Normal and Reversed. Separation of moderately polar analytes using adsorption onto a pure stationary phase (e.g. alumina or silica)
Ion Chromatography: Separation of organic and inorganic ions by their partitioning onto ionic stationary phases bonded to a solid support.
Size Exclusion Chromatography: Separation of large molecules based in the paths they take through a “maze” of tunnels in the stationary phase.
Chapter 4: Techniques in Biochemical Analysis 68
Chapter 4: Techniques in Biochemical Analysis 69
Chapter 4: Techniques in Biochemical Analysis 70
Chapter 4: Techniques in Biochemical Analysis 71
What does the analyst do? Select the correct type of separation for the analyte(s) of interest, based on the sample type (among other factors).
Select an appropriate column (stationary phase) and mobile phase
Select an appropriate detector based on whether universal or compound-specific detection is required or available
Optimize the separation using standard mixtures
Analyze the standards and sample
Chapter 4: Techniques in Biochemical Analysis 72
Chapter 4: Techniques in Biochemical Analysis 73
Columns and Stationary Phases. HPLC is largely the domain of packed columns
some research into microbore/capillary columns is going on. Molecules move too slowly to be able to reach and therefore
“spend time in” the stationary phase of an open tubular column in HPLC. In solution, not the gas phase Larger molecules in HPLC vs. GC (generally)
Stationary phases are particles which are usually about 1 to 20 m in average diameter (often irregularly shaped) In Adsorption chromatography, there is no additional phase
on the stationary phase particles (silica, alumina, Fluorosil). In Partition chromatography, the stationary phase is coated
on to (often bonded) a solid support (silica, alumina, divinylbenzene resin)
Chapter 4: Techniques in Biochemical Analysis 74
Chapter 4: Techniques in Biochemical Analysis 75
Stationary Phases Polar (“Normal” Phase):
Silica, alumina Cyano, amino or diol terminations on the bonded phase
Non-Polar (“Reversed Phase”) C18 to about C8 terminations on the bonded phase Phenyl and cyano terminations on the bonded phase
Mixtures of functional groups can be used!!
Packed particles in a column require: Frits at the ends of the column to keep the particles in Filtering of samples to prevent clogging with debris High pressure pumps and check-valves Often a “Guard Column” to protect the analytical column
Chapter 4: Techniques in Biochemical Analysis 76
Optimization of Separations in HPLC Correct choice of column so the above equilibrium has some
meaningful (non-infinity, non-zero) equilibrium constants. Correct choice of mobile phase Decision on the type of mobile phase composition
constant composition = isocratic varying composition = gradient elution
Determination if flow rate should be constant usually it is
Decision on heating the column heating HPLC columns can influence the above equilibrium….
Chapter 4: Techniques in Biochemical Analysis 77
Chapter 4: Techniques in Biochemical Analysis 78
The Mobile Phase in HPLC... Must do the following:
solvate the analyte molecules and the solvent they are in be suitable for the analyte to transfer “back and forth”
between during the separation process
Must be: compatible with the instrument (pumps, seals, fittings,
detector, etc) compatible with the stationary phase readily available (often use liters/day) of adequate purity
spectroscopic and trace-composition usually! Not too compressible (causes pump/flow problems)
Free of gases (which cause compressability problems)
Chapter 4: Techniques in Biochemical Analysis 79
Typical HPLC Pump (runs to 4,000+ psi)
Chapter 4: Techniques in Biochemical Analysis 80
Chapter 4: Techniques in Biochemical Analysis 81
Polarity Index for Mobile Phases….. The polarity index is a measure of the relative polarity of a solvent.
It is used for identifying suitable mobile phase solvents. The more polar your solvent is, the higher the index. You want to try to choose a polarity index for your solvent (or
solvent mixture) that optimizes the separation of analytes usually the index is a starting point the polarity of any mixture of solvents to make a mobile phase can be
modeled to give a theoretical chromatogram Usually, optimization of solvent composition is experimental
A similar number is the Eluent Strength (Eo] Increasing eluent strength or polarity index values mean increasing
solvent polarity. Remember, the analyte(s) and samples must be mobile phase and
stationary phase compatible!
Chapter 4: Techniques in Biochemical Analysis 82
Chapter 4: Techniques in Biochemical Analysis 83
Chapter 4: Techniques in Biochemical Analysis 84
Optimization of Mobile Phase Polarity…
Changing the mobile phase composition alters the separation.
Chapter 4: Techniques in Biochemical Analysis 85
Isocratic versus Gradient Elution Isocratic elution has a constant mobile phase composition
Can often use one pump! Mix solvents together ahead of time! Simpler, no mixing chamber required Limited flexibility, not used much in research
mostly process chemistry or routine analysis.
Gradient elution has a varying mobile phase composition Uses multiple pumps whose output is mixed together
often 2-4 pumps (binary to quarternary systems) Changing mobile phase components changes the polarity index
can be used to subsequently elute compounds that were previously (intentionally) “stuck” on the column
Some additional wear on the stationary phase Column has to re-equiluibrate to original conditions after each run
(takes additional time).
Chapter 4: Techniques in Biochemical Analysis 86
Chapter 4: Techniques in Biochemical Analysis 87
Chapter 4: Techniques in Biochemical Analysis 88
Chapter 4: Techniques in Biochemical Analysis 89
Injection in HPLC Usually 5 to 1000 L volumes, all directly onto the column
not much worry about capacity since the columns have a large volume (packed).
Injector is the last component before the column(s) A source of poor precision in HPLC
errors of 2-3 %RSD are due just to injection other errors are added to this due to capillary action and the small dimensions/cavities inside
the injector 6-PORT Rotary Valve is the standard manual injector Automatic injectors are available Two positions, load and inject in the typical injector Injection loop internal volume determines injection volume.
Chapter 4: Techniques in Biochemical Analysis 90
LOAD (the sample loop)
Inject (move the sampleloop into the mobilephase flow)
Chapter 4: Techniques in Biochemical Analysis 91
Chapter 4: Techniques in Biochemical Analysis 92
Detection in HPLC Numerous Types (some obscure) Original HPLC Detectors were common laboratory
instruments such as spectrophotometers, etc. Must be solvent -compatible, stable, etc. Universal
respond to all analytes Analyte Specific
respond to specific properties of analytes Non-destructive
most Destructive
ELSD, MS and a few others.
Chapter 4: Techniques in Biochemical Analysis 93
Chapter 4: Techniques in Biochemical Analysis 94
Standard Absorbance Detector…. Single Beam UV-VIS instrument with a flow-through cell (cuvette) Can use any UV-VIS with a special flow cell
Extra connections lead to band-broadening if UV-VIS is far from HPLC column exit.
Usually utilize typical UV-VIS lamps and 254 nm default wavelenth Can be set to other wavelengths (most) Simple filter detectors no longer widely used
adjustable wavelength units are cost-effective Non-destructive, not-universal
not all compounds absorb light can pass sample through several cells at several different wavelenghts
Usually zeroed at the start of each run using an electronic software command. You can have real-time zeroing with a reference cell.
Chapter 4: Techniques in Biochemical Analysis 95
Chapter 4: Techniques in Biochemical Analysis
96
SPECTROSCOPYSPECTROSCOPY SPECTROSCOPYSPECTROSCOPY
Chapter 4: Techniques in Biochemical Analysis
97
Definition• Spectroscopy - The study of the
interaction of electromagnetic
radiation with matter
Chapter 4: Techniques in Biochemical Analysis
98
Introduction• Spectroscopy is an analytical technique
which helps determine structure.• It destroys little or no sample.• The amount of radiation absorbed by
the sample is measured as wavelength is varied.
Chapter 4: Techniques in Biochemical Analysis 99
Major Types of Spectroscopy Infrared (IR) spectroscopy measures the bond vibration
frequencies in a molecule and is used to determine the functional group.
Mass spectrometry (MS) fragments the molecule and measures the masses.
Nuclear magnetic resonance (NMR) spectroscopy detects signals from hydrogen atoms and can be used to distinguish isomers.
Ultraviolet (UV) spectroscopy uses electron transitions to determine bonding patterns.
Chapter 4: Techniques in Biochemical Analysis 100
Introduction of Spectrometric Analyses
The study how the chemical compound interacts with different wavelenghts in a given region of electromagnetic radiation is called spectroscopy or spectrochemical analysis.
The collection of measurements signals (absorbance) of the compound as a function of electromagnetic radiation is called a spectrum.
Chapter 4: Techniques in Biochemical Analysis
101
Energy AbsorptionThe mechanism of absorption energy is different in the Ultraviolet, Infrared, and Nuclear magnetic resonance regions. However, the fundamental process is the absorption of certain amount of energy.
The energy required for the transition from a state of lower energy to a state of higher energy is directly related to the frequency of electromagnetic radiation that causes the transition.
Chapter 4: Techniques in Chapter 4: Techniques in Biochemical AnalysisBiochemical Analysis 102102
Wave Number (cycles/cm)
X-Ray UV Visible IR Microwave
200nm 400nm 800nm
Wavelength (nm)
Spectral Distribution of Radiant Energy
Chapter 4: Techniques in Chapter 4: Techniques in Biochemical AnalysisBiochemical Analysis 103103
Electromagnetic SpectrumElectromagnetic Spectrum
Chapter 4: Techniques in Chapter 4: Techniques in Biochemical AnalysisBiochemical Analysis 104104
Electromagnetic SpectrumElectromagnetic Spectrum
Chapter 4: Techniques in Chapter 4: Techniques in Biochemical AnalysisBiochemical Analysis 105105
V = Wave Number (cm-1)
Wave Length
C = Velocity of Radiation (constant) = 3 x 1010 cm/sec.
= Frequency of Radiation (cycles/sec)
The energy of photon:
h (Planck's constant) = 6.62 x 10-27 (Ergsec)
V =C
E = h = hC
C
= C =
Electromagnetic Radiation
Chapter 4: Techniques in Biochemical Analysis
106
Equation Definitions• E = energy (Joules, ergs)• c = speed of light (constant) = wavelength• h = Planck’s constant = “nu” = frequency (Hz)• nm = 10-9 m• Å = angstrom = 10-10 m
Chapter 4: Techniques in Biochemical Analysis 107
Visible
Ultra violet
Radio
Gamma ray
Hz
cmcm-1Kcal/mol eV
Type
Quantum Transition
Type
spectroscopy
Type
Radiation
Frequency
υ
Wavelength
λ
Wave
Number VEnergy
9.4 x 107 4.9 x 106 3.3 x 1010 3 x 10-11 1021
9.4 x 103 4.9 x 102 3.3 x 106 3 x 10-7 1017
9.4 x 101 4.9 x 100 3.3 x 104 3 x 10-5 1015
9.4 x 10-1 4.9 x 10-2 3.3 x 102 3 x 10-3 1013
9.4 x 10-3 4.9 x 10-4 3.3 x 100 3 x 10-1 1011
9.4 x 10-7 4.9 x 10-8 3.3 x 10-4 3 x 103 107
X-ray
Infrared
Micro-wave
Gamma ray emission
X-ray absorption, emission
UV absorption
IR absorption
Microwave absorption
Nuclear magnetic resonance
Nuclear
Electronic (inner shell)
Molecular vibration
Electronic (outer shell)
Molecular rotation
Magnetically induced spin states
Spectral Properties, Application and Interactions of Electromagnetic Radiation
Chapter 4: Techniques in Biochemical Analysis 108
Chapter 4: Techniques in Biochemical Analysis
109
Spectrum of Radiation
Chapter 4: Techniques in Biochemical Analysis
110
Visible Light
Chapter 4: Techniques in Biochemical Analysis
111
Visible Light
Red
Orange
Yellow
Green
Blue
Indigo
Violet
R
O
Y
G
B
I
V
700 nm
650 nm
600 nm
550 nm
500 nm
450 nm
400 nm
Chapter 4: Techniques in Chapter 4: Techniques in Biochemical AnalysisBiochemical Analysis 112112
Dispersion of Polymagnetic Light with a Prism
Polychromatic Ray
Infrared
RedOrange
Yellow
Green
Blue
Violet
Ultraviolet
monochromatic Ray
SLIT
PRISM
Polychromatic Ray Monochromatic Ray
Prism - Spray out the spectrum and choose the certain wavelength( that you want by slit.
Chapter 4: Techniques in Biochemical Analysis
113
Ultra Violet Spectrometry
The absorption of ultraviolet radiation by molecules is dependent upon the electronic structure of the molecule.
So the ultraviolet spectrum is called electronic spectrum.
Chapter 4: Techniques in Biochemical Analysis
114
INTRODUCTION TO INTRODUCTION TO SPECTROPHOTOMETRYSPECTROPHOTOMETRY
Chapter 4: Techniques in Biochemical Analysis
115
SpectrophotometrySpectrophotometry
• Spectrophotometry: An analytical Spectrophotometry: An analytical method using several spectra (lights). method using several spectra (lights). (State each spectrum used in (State each spectrum used in spectrophotometry.)spectrophotometry.)
• Spectrophotometer: An instrument for Spectrophotometer: An instrument for measuring absorbance that uses a measuring absorbance that uses a monochromatormonochromator to select the to select the wavelength.wavelength.
Chapter 4: Techniques in Biochemical Analysis
116
SpectrophotometrySpectrophotometry
-Advantages of spectrophotometers-Advantages of spectrophotometers
i.i. relatively inexpensive relatively inexpensive
ii.ii. inexpensiveinexpensive
iii.iii. easy to maintaineasy to maintain
iv.iv. portableportable
Chapter 4: Techniques in Biochemical Analysis
117
BACKGROUNDBACKGROUNDwhite light is observed, what is actually seen is a
mixture of all the colors of light
Why do some substances appear colored?
When this light passes through a substance, certain energies (or colors) of the light are absorbed while other color(s) are allowed to pass
through or are reflected by the substance.
If the substance does not absorb any light, it appears white (all light is reflected) or colorless (all light is transmitted). A solution appears a
certain color due to the absorbance and transmittance of visible light. For example, a blue solution appears blue because it is absorbing all of
the colors except blue.
Chapter 4: Techniques in Biochemical Analysis
118
BACKGROUNDBACKGROUND
Chapter 4: Techniques in Biochemical Analysis
119
BACKGROUNDBACKGROUND• The amount of light absorbed by a solution is The amount of light absorbed by a solution is
dependent on the ability of the compound to dependent on the ability of the compound to absorb light (absorb light (molar absorptivitymolar absorptivity), the distance ), the distance through which the light must pass through the through which the light must pass through the sample sample (path length(path length) and the ) and the molar molar concentrationconcentration of the compound in the solution. of the compound in the solution.
• If the same compound is being used and the If the same compound is being used and the path length is kept constant, then the path length is kept constant, then the absorbance is directly proportional to the absorbance is directly proportional to the concentration of the sample.concentration of the sample.
Chapter 4: Techniques in Biochemical Analysis
120
SpectrophotometerSpectrophotometer • A spectrophotometer is used to provide a A spectrophotometer is used to provide a
source of light of certain energy source of light of certain energy (wavelength) and to measure the quantity (wavelength) and to measure the quantity of the light that is absorbed by the sample. of the light that is absorbed by the sample.
Light Bulb Prism Filter Slit Sample Detector
Chapter 4: Techniques in Biochemical Analysis
121
Spectrophotometer Spectrophotometer • The basic operation of the spectrophotometer includes a white light The basic operation of the spectrophotometer includes a white light
radiation source that passes through a radiation source that passes through a monochromatormonochromator. The . The monochromator is either a prism or a monochromator is either a prism or a diffraction grating diffraction grating that separates that separates the white light into all colors of the visible spectrum. After the light is the white light into all colors of the visible spectrum. After the light is separated, it passes through a separated, it passes through a filter filter (to block out unwanted light, (to block out unwanted light, sometimes light of a different color) and sometimes light of a different color) and a a slitslit (to narrow the beam of (to narrow the beam of light--making it form a rectangle). Next the beam of light passes through light--making it form a rectangle). Next the beam of light passes through the the samplesample that is in the sample holder. The light passes through the that is in the sample holder. The light passes through the sample and the unabsorbed portion strikes a sample and the unabsorbed portion strikes a photodetectorphotodetector that that produces an electrical signal which is proportional to the intensity of the produces an electrical signal which is proportional to the intensity of the light. The signal is then converted to a light. The signal is then converted to a readable output readable output that is used in that is used in the analysis of the sample. the analysis of the sample.
Light Bulb Prism Filter Slit Sample Detector
Chapter 4: Techniques in Chapter 4: Techniques in Biochemical AnalysisBiochemical Analysis 122122
Spectrophotometer
An instrument which can measure the absorbance of a sample at any wavelength.
Light Lens Slit Monochromator
Sample Detector Quantitative Analysis
Slits
Chapter 4: Techniques in Biochemical Analysis
123PGCC CHM 103 SinexPGCC CHM 103 Sinex
Io I
Cell withPathlength, b,
containing solution
lightsource detector
blank where Io = I
concentration 2concentration 1
b
with sample I < Io
The process of light being absorbed by a solution
As concentration increased, less light was transmitted (more light absorbed).
Chapter 4: Techniques in Chapter 4: Techniques in Biochemical AnalysisBiochemical Analysis 124124
Beer – Lambert Law
Glass cell filled with concentration of solution (C)
IILight
0
As the cell thickness increases, the transmitted intensity of light of I decreases.
Chapter 4: Techniques in Biochemical Analysis 125
R- Transmittance
R = I0 - Original light intensity
I- Transmitted light intensity
% Transmittance = 100 x
Absorbance (A) = Log
= Log = 2 - Log%T
Log is proportional to C (concentration of solution) and is
also proportional to L (length of light path through the solution).
I
I0
I
I0
I0
I
1
T
I
I0
Chapter 4: Techniques in Biochemical Analysis 126
A CL = ECL by definition and it is called the Beer - Lambert Law.
A = ECL
A = ECL
E = Molar Extinction Coefficient ---- Extinction Coefficient of a solution containing 1g molecule of solute per 1 liter of solution
Chapter 4: Techniques in Chapter 4: Techniques in Biochemical AnalysisBiochemical Analysis 127127
E =Absorbance x Liter
Moles x cm
UNITS
A = ECL
A = No unit (numerical number only)
E = Liter
Cm x Mole
L = Cm C = Moles/Liter
A = ECL = (Liter
Cm x Mole) x
Mole
Literx Cm
Chapter 4: Techniques in Biochemical Analysis 128
PGCC CHM 103 SinexPGCC CHM 103 Sinex
The BLANKThe BLANK
The blank contains The blank contains allall substances substances expect the analyte.expect the analyte.
Is used to set the absorbance to zero:Is used to set the absorbance to zero:
AAblankblank = 0 = 0 This removes any absorption of light This removes any absorption of light
due to these substances and the cell.due to these substances and the cell. All measured absorbance is due to All measured absorbance is due to
analyte.analyte.
Chapter 4: Techniques in Biochemical Analysis
129PGCC CHM 103 SinexPGCC CHM 103 Sinex
Beer’s LawBeer’s Law
A = abcA = abc
where a – molar absorptivity, b – where a – molar absorptivity, b – pathlength, and c – molar concentrationpathlength, and c – molar concentration
See the See the Beer’s Law Simulator
Chapter 4: Techniques in Biochemical Analysis
130
SpectrophotometerSpectrophotometer
The spectrophotometer displays this quantity in one of two ways:
(1)Absorbance -- a number between 0 and 2
(2) Transmittance -- a number between 0 and 100%.
The sample for a spectral analysis is prepared by pouring it into a cuvette which looks similar to a small test tube. A cuvette is made using a special optical quality glass that will itself absorb a minimal amount of the light. It is also marked with an indexing line so that it can be positioned in the light beam the same way each time to avoid variation due to the differences in the composition of the glass
Chapter 4: Techniques in Biochemical Analysis
131
Fundamentals of Spectrophotometry Absorption of Light
Beer’s Law The relative amount of a certain wavelength of light absorbed (A) that passes
through a sample is dependent on:- distance the light must pass through the sample (cell path length - b)- amount of absorbing chemicals in the sample (analyte concentration – c)- ability of the sample to absorb light (molar absorptivity - )
Increasing [Fe2+]
Absorbance is directly proportional to concentration of Fe+2
Chapter 4: Techniques in Biochemical Analysis
132
Fundamentals of Spectrophotometry Absorption of Light
3.) Beer’s Law Absorbance is useful since it is directly related to the analyte concentration,
cell pathlength and molar absorptivity. This relationship is known as Beer’s Law
where: A = absorbance (no units)= molar absorptivity (L/mole-cm)b = cell pathlength (cm)c = concentration of analyte (mol/L)
Beer’s Law allows compounds to be quantified by their ability to absorb light, Relates directly to concentration (c)
A = abc
Chapter 4: Techniques in Biochemical Analysis
133
Fundamentals of Spectrophotometry Absorption of Light
4.) Absorption Spectrum By choosing different wavelengths of light (A vs. B) different compounds
can be measured
A B
Chapter 4: Techniques in Biochemical Analysis
134
Fundamentals of Spectrophotometry Spectrophotometer
1.) Basic Design An instrument used to make absorbance or transmittance measurements is
known as a spectrophotometer
Chapter 4: Techniques in Biochemical Analysis
135
Single Beam Spectrophotometer
Chapter 4: Techniques in Biochemical Analysis
136
Dual Beam Spectrophotometer
Chapter 4: Techniques in Biochemical Analysis
137
Fundamentals of Spectrophotometry Spectrophotometer
1.) Basic Design Light Source: provides the light to be passed through the sample
- Tungsten Lamp: visible light (320-2500 nm)
- Deuterium Lamp: ultraviolet Light (160-375 nm)
In presence of arc, some of the electrical energy is absorbed by D2 (or H2) which results in the disassociation of the gas and release of light
D2 + Eelect D*2 D’ + D’’ + h(light produced)
Excited state
Low pressure (vacuum)
Tungsten Filament
- based on black body radiation:heat solid filament to glowing, light emitted will be characteristic of temperature more than nature of solid filament
Chapter 4: Techniques in Biochemical Analysis
138
Fundamentals of Spectrophotometry Spectrophotometer
1.) Basic Design Wavelength Selector (monochromator): used to select a given wavelength
of light from the light source- Prism:
- Filter:
Chapter 4: Techniques in Biochemical Analysis
139
Fundamentals of Spectrophotometry Spectrophotometer
1.) Basic Design Wavelength Selector (monochromator): used to select a given wavelength
of light from the light source- Reflection or Diffraction Grating:
Chapter 4: Techniques in Biochemical Analysis
140
Fundamentals of Spectrophotometry Spectrophotometer
1.) Basic Design Sample Cell: sample container of fixed length (b).
- Usually round or square cuvet- Made of material that does not absorb light in the wavelength range of
interest
1. Glass – visible region
2. Quartz – ultraviolet
3. NaCl, KBr – Infrared region
Chapter 4: Techniques in Biochemical Analysis
141
Cuvettes (sample holder)
• Polystyrene– 340-800 nm
• Methacrylate– 280-800 nm
• Glass– 350-1000 nm
• Suprasil Quartz– 160-2500 nm
Chapter 4: Techniques in Biochemical Analysis
142
Spectrophotometer
1.) Basic Design Light Detector: measures the amount of light passing through the
sample.- Usually works by converting light signal into electrical signal
Fundamentals of Spectrophotometry
Photomultiplier tubeProcess: a) light hits photoemissive cathode and e- is emitted. b) an emitted e- is attracted to electrode #1 (dynode 1), which is 90V more positive. Causes several more e- to be emitted. c) these e- are attracted to dynode 2, which is 90V more positive then dynode 1, emitting more e-. d) process continues until e- are collected at anode after amplification at 9 dynodes. e) overall voltage between anode and cathode is 900V. f) one photon produces 106 – 107 electrons. g) current is amplified and measured
Chapter 4: Techniques in Biochemical Analysis 143
Applications of Spectrophotometry
Quantitative Applications
• Usually using UV-Vis• IR can be used
- Environmental applications; analysis waters & waste waters
- Clinical applications: analysis of glucose
- Industrial analysis; analysis of iron content in food
- Forensic applications: Determination of blood alcohol
Chapter 4: Techniques in Biochemical Analysis 144
Advantage of Advantage of spectrophotometer over spectrophotometer over colorimetercolorimeter
can be used to profile printers & scanners, measure colors "in can be used to profile printers & scanners, measure colors "in the wild", measure your illuminationthe wild", measure your illumination
colorimeter measures only 3 points on the specturm (RGB), colorimeter measures only 3 points on the specturm (RGB), while a spectrophotometer measures while a spectrophotometer measures manymany points across the points across the entire spectrumentire spectrum
colorimeters use a single type of light (such as incandescent or colorimeters use a single type of light (such as incandescent or pulsed xenon) Spectrophotometers can compensate for this shift, pulsed xenon) Spectrophotometers can compensate for this shift, making spectrophotometers a superior choice for accurate, making spectrophotometers a superior choice for accurate, repeatable color measurement.repeatable color measurement.
Chapter 4: Techniques in Biochemical Analysis
145
Sample Cells
UV Spectrophotometer
Quartz (crystalline silica)
Visible Spectrophotometer
Glass
Chapter 4: Techniques in Biochemical Analysis
146
Light Sources
UV Spectrophotometer
1. Hydrogen Gas Lamp
2. Mercury Lamp
Visible Spectrophotometer
1. Tungsten Lamp
Chapter 4: Techniques in Biochemical Analysis
147
Chemical Structure & UV Absorption
Chromophoric Group ---- The groupings of the molecules which contain the electronic system which is giving rise to absorption in the ultra-violet region.
Chapter 4: Techniques in Biochemical Analysis
148
UV Spectrometer Application
Protein
Amino Acids (aromatic)
Pantothenic Acid
Glucose Determination
Enzyme Activity (Hexokinase)
Chapter 4: Techniques in Biochemical Analysis
149
Flurometric Application
Thiamin (365 nm, 435 nm)
Riboflavin
Vitamin A
Vitamin C
Chapter 4: Techniques in Biochemical Analysis
150
Visible Spectrometer Application
Niacin
Pyridoxine
Vitamin B12
Metal Determination (Fe)
Fat-quality Determination (TBA)
Enzyme Activity (glucose oxidase)
Chapter 4: Techniques in Biochemical Analysis 151
Major Types of Light Spectroscopy
Absorption spectroscopy Measures amount of light absorbed Most common, non-destructive Concentration, pH measures, purity, ID
Atomic emission spectroscopy Measures light emitted from burned sample Elemental analysis
Fluorescence spectroscopy Samples fluoresce when they emit at higher than what they absorb Measures solvent interactions, distances, molecular shape, and motion
Circular Dichroism spectroscopy Absorption of circular polarized light Chiral compound identification
Transmission spect. (colorimetry)
Chapter 4: Techniques in Biochemical Analysis 152
Introduction
Atomic absorption is the absorption of light by free atoms. An atomic absorption spectrophotometer is an instrument that uses this principle to analyze the concentration of metals in solution. The substances in a solution are suctioned into an excited phase where they undergo vaporization, and are broken down into small fragmented atoms by discharge, flame or plasma.
Chapter 4: Techniques in Biochemical Analysis 153
Atomic Emission Spectroscopy
By exposing these atoms to such temperatures they are able to “jump” to high energy levels and in return, emit light. The versatility of atomic absorption an analytical technique (Instrumental technique) has led to the development of commercial instruments. In all, a total of 68 metals can be analyzed.
Chapter 4: Techniques in Biochemical Analysis 154
Advantages of AA
Determination of 68 metals Ability to make ppb determinations on major components of a
sample Precision of measurements by flame are better than 1% rsd.
There are few other instrumental methods that offer this precision so easily.
AA analysis is subject to little interference. Most interference that occurs have been well studied and
documented. Sample preparation is simple (often involving only dissolution in
an acid) Instrument easy to tune and operate
Chapter 4: Techniques in Biochemical Analysis 155
Atomic emission spectrometer
Chapter 4: Techniques in Biochemical Analysis
156
NEXT CLASS:Chapter 5
DNA Technology
THANK YOU