Upload
klara-valko
View
848
Download
3
Embed Size (px)
Citation preview
www.bio-mimetic-chromatography.com
Lipophilicity in drug discovery
Klara Valko
Bio-Mimetic ChromatographyConsultancy for Successful Drug Discovery
What is lipophilicity?IUPAC definition Lipophilicity represents the affinity of a
molecule or a moiety for a lipophilic environment.
Hydrophobicity measures the association of non-polar groups or molecules in an aqueous environment which arises from the tendency of water to exclude non-polar molecules.
www.bio-mimetic-chromatography.com
Why do we need to measure lipophilicity?
A compound partitioning between aqueous and organic phase can model compound partitioning in vivo
Compound partitions between two immiscible solvent
www.bio-mimetic-chromatography.com
Lipophilicity is measured by partition coefficient Partition coefficient is the quotient of the compound
concentration in an aqueous and a non-miscible solvent under equilibrium condition.
It is expressed by the quotient of the compound concentrations in the two phases under equilibrium condition.
It depends on the nature of the two phases, the compound properties, pH, and the temperature.
www.bio-mimetic-chromatography.com
Measuring lipophilicity by HPLC
Cs. Horváth, W. Melander, I. Molnár, J. Chromatogr. 125 (1976) 129.
Measuring bio-relevant association constants has great impact on drug discovery!
Based on the solvophobic theory the interaction between the solute and the stationary phase is considered as a reversible association of
the solute molecules with the stationary phase moiety (hydrocarboneous, membrane, or protein). Accordingly solute
retention is governed by the dynamic equilibrium constant.
Vs = volume of the stationary phase
Vm = volume of the mobile phase
www.bio-mimetic-chromatography.com
Gradient retention time for the determination of Chromatographic Hydrophobicity Index (CHI)
The isocratic chromatographic hydrophobicity index (φ0) is linearly related to gradient retention times.
Measure log k values of a set of compounds using 3 to 5 different mobile phase composition. Plot the log k values in the function of the organic phase concentration.
Calculate the organic phase concentration that relates to log k = 0. This is the φ0 value.
Reversed phase gradient retention times can be calibrated by isocratically determined φ0 values.
Use the slope and the intercept of the obtained straight line to convert the gradient retention time of any compound obtained under the same conditions as the calibration set of compounds to CHI.
www.bio-mimetic-chromatography.com
www.bio-mimetic-chromatography.com
Chromatographic Hydrophobicity Index (CHI) Approximates the organic phase concentration (%) when the compound elutes from a reversed
phase column using linear gradient. CHI gives a straight line with the gradient retention time.
LunaC18(2) 50 x 3 mm; 1.00 ml/min; Mobile phase A 50 mM ammonium acetate pH 7.4 and B is 100%
acetonitrile. Gradient: 0 - 2.5 min 0 - to 100% B; 2.5 - 2.7 min 100% B.
Calibration of CHI at pH7.4
y = 54.329x - 71.702R 2 = 0.9972
0.00
20.00
40.00
60.00
80.00
100.00
120.00
1.4 1.9 2.4 2.9 3.4
Compound CHI7.4 at pH 7.4
CHI2 at pH 2
CHI10.5 at pH 10.5
Theophylline 18.4 17.9 5.0 Phenyltetrazole 23.6 42.2 16.0 Benzimidazole 34.3 6.3 30.6 Colchicine 43.9 43.9 43.9 Phenyltheophylline 51.7 51.7 51.7 Acetophenone 64.1 64.1 64.1 Indole 72.1 72.1 72.1 Propiophenone 77.4 77.4 77.4 Butyrophenone 87.3 87.3 87.3 Valerophenone 96.4 96.4 96.4
www.bio-mimetic-chromatography.com
Ultrahigh Performance Liquid Chromatography (uPLC) for CHI
CHI TM1
Time0.00 0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 1.80
AU
0.0
5.0e-2
1.0e-1
1.5e-1
2.0e-1
2.5e-1
3.0e-1
CHI TM1_LunapH74 Diode Array 254
Range: 3.969e-11.02
0.86
0.64
0.590.72
0.92
1.27
1.16
1.10
1.37
The CHI test mix is separated in less than 90 secNow we can determine a compound’s lipophilicity in 90 sec
using various starting mobile phase pHCourtesy of Shenaz Bunally at GSK
www.bio-mimetic-chromatography.com
Parallel measurement of a compound’s retention at various pH to reveal acid/base character
Typical 4-way chromatograms of a base
pH2
pH7.4
pH 10.5
IAM 7.4
www.bio-mimetic-chromatography.com
CHIs measured at 3 pHs provide an automatic way of grouping molecules according to acid/base character without structural information.
0
10
20
30
40
50
60
70
80
90
100
Neutral(Zwitterionic)
Strong acid Weak acid Strong base Weak base Amphoteric
pH2
pH7.4pH10.5
The change of CHI values by changing the pHCHI
CHI values at pH 2, pH 7.4 and pH 10.5 reveal acid/base character of compounds
www.bio-mimetic-chromatography.com
We can use biological macromolecules as stationary phase to obtain bio-mimetic properties Reversed phase retention reveals how compounds
interact/partition/bind to hydrophobic environment such as aliphatic hydrocarbons. This environment is common in the inner part of phospholipid bi-layers.
Applying protein stationary phases can be used to reveal compounds’ interaction/affinity to biologically important proteins, such as plasma proteins
Immobilized artificial membrane (IAM stationary phases were developed to emulate the phospholipid membrane on solid surfaces.
www.bio-mimetic-chromatography.com
Bio-mimetic HPLC measurement of Human Serum Albumin (HSA), α-1acidglycoprotein (AGP) and Immobilized Artificial Membrane (IAM) partition
HSA AGP IAM
pH 7.4 aqueous mobile phases
IPA IPA ACN
Bio-Mimetic Stationary phases
www.bio-mimetic-chromatography.com
Measuring bio-mimetic properties The retention times obtained on bio-mimetic
stationary phase should be calibrated
Measure a set of compounds retention for which the bio-mimetic binding properties are available using other methods, such as equilibrium dialysis or ultra filtration
www.bio-mimetic-chromatography.com
Biomimetic lipophilicity measurements (Membrane partition) using Immobilised Artificial Membrane stationary phase
Stationary phaseIAM calibration y = 26.647x - 37.653
R2 = 0.9966
0
10
20
30
40
50
60
1 1.5 2 2.5 3 3.5gtR
CHI
Typical calibration
Compound gtR CHI IAMOctanophenone 3.269 49.4Heptanophenone 3.145 45.7Hexanophenone 3.001 41.8Valerophenone 2.822 37.3Butyrophenone 2.601 32Propiophenone 2.341 25.9Acetophenone 2.013 17.2Acetanilide 1.83 11.5Paracetamol 1.591 2.9
Column: IAM PC2 (CH2)12 150 x 4.6
Mobile Phase flow rate: 2 ml/min
Gradient: 0 to 3 min 0 to 80% acetonitrile
3 to 3.5 min 80% acetonitrile
3.5 to 3.7 min 0% acetonitrile
Cycle time: 5 min
K. Valko et al. J. Pharm. Sci.89 (2000) 1085-1096
www.bio-mimetic-chromatography.com
Calibration ploty = 2.177x + 0.1304
R2 = 0.9612
-0.50
0.00
0.50
1.00
1.50
2.00
2.50
-0.40 -0.20 0.00 0.20 0.40 0.60 0.80 1.00
logtR
logK
lite
ratu
re
Calculate %Binding
logK = slope * log(tR) + intK = %B / (101-%B)
y = 0.9309x - 0.3329R2 = 0.879
0
20
40
60
80
100
0 20 40 60 80 100
HSA Column
Lite
ratu
re %
bin
ding n=71
K. Valko et al. J. Pharm. Sci. 92 (2003) 2236
Column: HSA 50 x 3 mm (Chrom Tech, Chiral Technologies)Flow rate: 1.8 ml/min at 300CMobile phase: 50 mM ammonium acetate pH7.4Gradient: 0 - 3 min 0 to 30% 2-propanol;
3 to 10 min 30% 2-propanol; 10 to 10.5 min 0% 2-propanolCycle time: 15 min
Serum albumin binding measurement using chemically bonded serum albumin stationary phases
www.bio-mimetic-chromatography.com
Retention time of compounds can be converted to % binding or K association constant
)(logexplog HSAKHSAkHSAke
2 min
www.bio-mimetic-chromatography.com
Accuracy and reproducibility The reproducibility of the retention time
measurements are very good. Variations can be expected only in the third
decimal places in the minutes. The accuracy increase as the binding
increases. We have still several minutes left after the
peak of 99% bound compound, that we can accurately measure.
www.bio-mimetic-chromatography.com
AGP binding measurement by HPLC
Same principle as HSA binding measurements:AGP column2-propanol gradientpH 7.4 ammonium acetateCalibration with AGP binding data derived from published % AGP bound values
Typical AGP column calibration plot y = 2.7976x - 0.5289R2 = 0.9744
-0.400
-0.200
0.000
0.200
0.400
0.600
0.800
1.000
1.200
1.400
0.00 0.10 0.20 0.30 0.40 0.50 0.60 0.70
log tR
logK
AG
P
Calibration set of compounds: Nizatidine, Bromazepam, Warfarin, Propranolol, Imipramin, Nicardipine, Chlorpromazine
www.bio-mimetic-chromatography.com
How can we use the bio-mimetic chromatographic data?
We can build models for in vivo distribution of drug molecules
We can predict Volume of distribution Unbound volume of distribution Drug efficiency max Minimum efficacious dose when the potency data are available
Ask for help to set up these measurements! You need only an HPLC, several HPLC columns and calibration set of compounds.
In three days you can be up and running your bio-mimetic chromatography!
ReferencesC. Hansch, A. Leo, ρ-σ-π analysis. A method for correlation of biological activity and chemical structure. J. Amer. Chem.Soc.
86(1964) 1616-1624 Horvath, Cs., Melander, W., Molnar, I. Solvophobic interactions in liquid chromatography with non-polar stationary phases,
Journal of Chromatography, 125 (1976) 129-156Valko, K.; Snyder, L.R.; Glajch, G.L. Retention in reversed-phase liquid chromatography as a function of mobile-phase
composition. J. Chromatogr. A 656, (1993) 501–520.Giaginis, A. Tsantili-Kakoulidou, Current state of the art in HPLC Methodology for lipophilicity assessment of basic drugs
(Review) Journal of Liquid Chromatography & Related Technologies, 31: (2008) 79–96.Lombardo, F., Shalaeva, M. Y., Tupper, K. A., Gao, F., ElogDoct: A tool for Lipophilicity Determination in Drug Discovery. 2.
Basic and neutral compounds. (2001) 2490-2497Gocan, S., Cimpan, G., Comer, J., Lipophilicity measurements by liquid chromatography in Advances in Chromatography,
Eds: E. Grushka, N. Grinberg, 44 (2005) 79-176, Taylor & Francis Group, 1574447343Valko, K., Bevan, C., Reynolds, D., Chromatographic hydrophobicity index by fast-gradient RP-HPLC: A high throughput
alternative to log P/log D. Analytical Chemistry 69 (1997) 2022-2029 Valko, K. Measurements of lipophilicity and acid/base character using HPLC methods. In “Pharmaceutical profiling in drug
discovery for lead selection” Eds. Borchardt, R., Kerns, E., AAPS (2004)Arlington, VA 127-182
K. Valko, C.M. Du, C.D. Bevan, D.P. Reynolds, M.H. Abraham, Rapid-gradient HPLC method for measuring drug interactions with immobilized artificial membrane: Comparison with other lipophilicity measures, J. Pharm. Sci. 89 (2000)
1085– 1096. K. Valko, C.M. Du, C. Bevan, D.P. Reynolds, M.H. Abraham, Rapid Method for the Estimation of Octanol / Water Partition Coefficient ( Log P oct ) from Gradient RP-HPLC Retention and a Hydrogen Bond Acidity Term, (2001) 1137–1146.K. Valko, S. Nunhuck, C. Bevan, M.H.M.H. Abraham, D.P.D.P. Reynolds, Fast gradient HPLC method to determine compounds binding to human serum albumin. Relationships with octanol/water and immobilized artificial membrane lipophilicity., J. Pharm. Sci. 92 (2003) 2236–48.
www.bio-mimetic-chromatography.com