Upload
ronaldo-degazon
View
204
Download
4
Embed Size (px)
Citation preview
Lab #1: Discussion, Error and Precaution, Conclusion
Discussion
The lengths of a glass prism, stool, beaker, copper wire and metal pin were made using 3 measuring
instruments. The accuracy of the instruments (how close the readings are to the 'true' value) could not
be determined as no true values of the quantities being measured were given. The instruments with the
greatest precision (intention to make observations with the greatest possible x aptitude) were the
vernier calipers because they were able to measure quantities to two significant figures. The meter rule
had the least precision as it only made measurements to one significant figure. The micrometer screw
gauge was the most sensitive (had the largest response to unit change in input) as its s cale would
experience the most graduations for a fixed length amongst the 3 instruments. Finally the meter rule
had the greatest range (interval between minimum and maximum values able to be measured) being
100cm but was also the least sensitive instrument.
It was necessary to take repeated readings of the measured quantities in order to eliminate uncertainty
or random error. However, parallax error was hard to avoid when measuring the stool as the point on
the scale was way higher than my line of sight. As seen in all except one of the readings, the
consistencies in the measurements show that these readings were very precise.
In measuring the thickness of the beaker, the difference between the external and internal diameters
was needed to be divided by 2 so that the final result would not represent the thickness of two sides,
but one only.
The results obtained were as expected with the stool, the largest object having the greatest overall
measurement with the strand of copper wire having the least.
Sources of Error
Parallax Error: a parallax error was experienced when taking the length of the stool because of
the stool’s elevated position and the point on the scale being above my line of sight. However, it
was soon realized and corrected.
There was a bit of uncertainty in reading the micrometer screw gauge when measuring the
width of the strand of copper wire.
Precautions
To avoid parallax error, I made sure that all my readings were made at eye level, with one eye
which was used for throughout.
To avoid uncertainty, repeated readings of the quantities were recorded.
Conclusion
The lengths of the instruments that were to be measured were recorded successfully.