32
© Innova Biosciences Ltd. All rights reserved Quality – Consistency – Expertise

Antibody purification – what you need to know to use antibodies effectively

Embed Size (px)

DESCRIPTION

In this webinar Dr Andy Lane discusses the various methods available for purifying antibodies from different sources, and explains why it is vitally important to understand how your antibodies have been purified to know what you can do with them, either within assays or for further processing such as conjugation to dyes and enzymes.

Citation preview

Page 2: Antibody purification – what you need to know to use antibodies effectively

© Innova Biosciences Ltd. All rights reserved

Welcome to our 15th webinar

ANTIBODY PURIFICATION: What you need to know to use antibodies effectively

Dr Andy Lane

Page 3: Antibody purification – what you need to know to use antibodies effectively

© Innova Biosciences Ltd. All rights reserved

Dr Andy Lane

• Sources of antibodies

• Different methods of purifying antibodies

• Why “purified antibodies” are not all the same

• How to purify antibodies easily without specialist knowledge or equipment

• Considerations for conjugating antibodies to enzymes, dyes and nanoparticles

Page 4: Antibody purification – what you need to know to use antibodies effectively

© Innova Biosciences Ltd. All rights reserved

Sources of antibodies

• Animal serum – polyclonal antibodies

• Ascites – monoclonal antibodies

• Cell culture supernatant – monoclonal antibodies

• Eggs – avian antibodies

• Bacterial expression systems – recombinant antibodies

Each of these sources has contains the specific antibody of interest amongst other proteins, lipids and other components.

Page 5: Antibody purification – what you need to know to use antibodies effectively

© Innova Biosciences Ltd. All rights reserved

Different methods of purifying antibodies

• Fractionation by globulin precipitation

• Ion exchange chromatography

• Size exclusion chromatography

• Protein A affinity chromatography

• Protein G affinity chromatography

• Protein L affinity chromatography

• Antigen affinity chromatography

Page 6: Antibody purification – what you need to know to use antibodies effectively

© Innova Biosciences Ltd. All rights reserved

Globulin precipitation

• First published in 1899 by James Atkinson in J. Exp. Med. – separating albumin from the anti-toxic components of a horse antiserum using magnesium sulphate.

• Ammonium sulphate and sodium sulphate are more commonly used, taking advantage of the principle of “salting-out”. At the appropriate concentration these precipitate globulins, whilst many other proteins, including albumin, remain in solution.

• Whilst this is a simple and gentle procedure, it only provides a partially pure preparation if starting from a complex mixture such as serum.

• This step is often combined with further purification using one of the other methods discussed in this webinar.

Page 7: Antibody purification – what you need to know to use antibodies effectively

© Innova Biosciences Ltd. All rights reserved

Ion exchange chromatography

• Ion exchange chromatography (IEC) uses positively or negatively charged resins to bind proteins based on their net charges in a given buffer system (pH).

• Typically, a complex mixture is added to the column in a certain set of buffer conditions (e.g. low salt), and the buffer conditions are then changed either step-wise or on a gradient basis. Different proteins are released from the column in differing conditions.

• IEC is perhaps more often used to purify polyclonal antibodies than for monoclonals – note that each monoclonal is unique in terms of charge, and will therefore be released under a specific set of conditions – something that is useful for repeated purifications of the same antibody, but also means that optimisation is required for each antibody.

• There are various IEC resins available, with perhaps the most widely used being DEAE-Sepharose.

Page 8: Antibody purification – what you need to know to use antibodies effectively

© Innova Biosciences Ltd. All rights reserved

Starting buffer counter-ions

Substances to be separated

Gradient ions

Ion exchange chromatography

Page 9: Antibody purification – what you need to know to use antibodies effectively

© Innova Biosciences Ltd. All rights reserved

Size exclusion chromatography

• Separation of complex mixtures on the basis of size or molecular shape can be achieved by size exclusion (SEC) or gel filtration chromatography.

• The molecular sieving process takes place as a solute passes through a packed bed stationary phase. Separation depends on the different abilities of the various molecules to enter the pores of the bead-based stationary phase.

• Large molecules, which cannot enter the pores, are excluded and pass through the column quickly. Smaller molecules that can enter the pores are retarded and move through the column more slowly. Very small molecules, such as salt, are able to fully permeate and elute last.

Page 10: Antibody purification – what you need to know to use antibodies effectively

© Innova Biosciences Ltd. All rights reserved

Size exclusion chromatography

• SEC tends not to be used for primary purification of antibodies other than IgMs. However, it may be a valuable “polishing step” following purification by other methods.

• Ranges of chromatography resins are available that separate proteins within different ranges of size (e.g. AcA22, AcA34, AcA44 from Ultrogel®, and S200, S300, S400 in the Sephadex range)

DE-SALTING COLUMNS

• So called de-salting columns are used to undertake a buffer exchange, or to remove small size contaminants. This is simply a special case of SEC, with the most commonly used gel being Sephadex G-25, often in pre-prepared PD10 columns.

Page 11: Antibody purification – what you need to know to use antibodies effectively

© Innova Biosciences Ltd. All rights reserved

Mixture of different sized proteins

Different sized proteins

Gel

filt

rati

on

co

lum

n

Size exclusion chromatography

Page 12: Antibody purification – what you need to know to use antibodies effectively

© Innova Biosciences Ltd. All rights reserved

Protein A/G/L chromatography

• In their native form these proteins are expressed by bacteria as part of their defense mechanisms against the mammalian immune response.

• All bind to mammalian immunoglobulins via constant regions – Protein A and G to the Fc region, and Protein L to the kappa light chain.

• Recombinant versions are used immobilised to various matrices to purify IgG by a form of affinity chromatography.

Page 13: Antibody purification – what you need to know to use antibodies effectively

© Innova Biosciences Ltd. All rights reserved

Heavy chain

Light chain

Protein G and A

Protein L

Protein A/G/L chromatography

Page 14: Antibody purification – what you need to know to use antibodies effectively

© Innova Biosciences Ltd. All rights reserved

Protein A/G/L chromatography

• Samples are added to the Protein A/G/L matrix in a suitable binding buffer – pH and ionic strength should be considered

• Column is washed, which effectively leaves only IgG bound

• Low pH elution is carried out, which may be with different buffer formulations and may differ depending upon the isotype of the antibody being purified. It may be important to know the basic formulation of the elution buffer for downstream processing

• Low pH elution requires neutralisation to maintain antibody integrity. Often carried out with Tris buffer, but is this sufficient for your downstream needs?

Page 15: Antibody purification – what you need to know to use antibodies effectively

© Innova Biosciences Ltd. All rights reserved

Recombinant Protein A Recombinant Protein G Recombinant Protein A/G Recombinant Protein L Recombinant Protein A/G/L

Native SourceStaphylococcusaureus

Streptococcus N/APeptostrepto-coccus magnus

N/A

Binding Sites for Ig 5 2 6 4 13

Optimal Binding pH 8 - 9 5 5 to 8.2 7.5 7.5

Typical elution pH 3.0 – 7.0 (isotype dependent)

2.5 – 3.0 2.5 – 3.0 2.0 – 3.0 2.5 – 3.0

Ig Binding Target Fc Fc Fc VL-kappa Fc + VL-kappa

Protein A/G/L chromatography

Page 16: Antibody purification – what you need to know to use antibodies effectively

© Innova Biosciences Ltd. All rights reserved

Species Immunoglobulin Binding to Protein A Binding to Protein G Binding to Protein L

Human IgG1 Strong Strong Strong

IgG2 Strong Strong Strong

IgG3 Negligible Strong Strong

IgG4 Strong Strong Strong

Mouse lgG1 Weak Strong Strong

lgG2a Strong Strong Strong

lgG2b Medium Medium Strong

lgG3 Weak/medium Medium Strong

Rat lgG1 Negligible Weak Strong

lgG2a Negligible Strong Strong

lgG2b Negligible Medium Strong

lgG2c Negligible Medium Strong

Goat lgG Weak Medium Negligible

Rabbit lgG Strong Medium Weak

Sheep lgG Weak Medium Negligible

Protein A/G/L chromatography

Page 17: Antibody purification – what you need to know to use antibodies effectively

© Innova Biosciences Ltd. All rights reserved

Antigen affinity chromatography

• In this case a specific chromatography resin is prepared to which the actual antigen that the antibodies bind to is immobilised.

• The sample containing the specific antibodies is then passed over the column, allowing only specific antibodies to bind, with all other contaminants, including other antibodies, being washed away.

• Buffer conditions are then changed to elute the specific antibodies from the column.

• Typical examples of the uses of affinity chromatography are in preparation of species specific secondary antibodies, in purifying antibodies to peptide antigens and for purifying proteins with epitope tags (e.g. Myc, HIS etc).

Page 18: Antibody purification – what you need to know to use antibodies effectively

© Innova Biosciences Ltd. All rights reserved

Matrix Ligand Immobilized Ligand

Sample Complex Impurities

Antigen affinity chromatography

Preparation of gel matrix

Application of sample

Elution of purified antibody

Page 19: Antibody purification – what you need to know to use antibodies effectively

© Innova Biosciences Ltd. All rights reserved

Why “purified antibodies” are not all the same

• Concentration

• Preservatives / stabilisers – have any been added back in to the preparation after purification?

• Buffer components – including contaminants from elution etc.

• Level of purity - e.g. Ig fraction vs. total IgG vs. specific IgG

• Unexpected contaminants – e.g. bovine IgG

• Endotoxin etc.

Many antibodies are sold by companies as “purified” – but what does that mean?

Page 20: Antibody purification – what you need to know to use antibodies effectively

© Innova Biosciences Ltd. All rights reserved

Your final buffer……..

• The final buffer that your antibody is in may have a significant effect on what you can use your antibody for. For instance, the presence of glycine in a buffer can inhibit conjugation reactions, or sodium azide may kill live cells. Some buffers may be suitable for conjugating to proteins, but not to nanoparticles.

• If you undertake a dialysis step post-purification to transfer an antibody into a particular buffer, you need to be sure that the dialysis has been effective.

• e.g. To remove (i.e. to 1nM level) 100mM glycine from 10ml of antibody you need to dialyse against 1 million litres of PBS. Alternatively, changing the dialysis 4 times will achieve the same effect with only 4 litres of PBS.

• However, a single dialysis into 1 litre leaves glycine present at 1mM.

Why “purified antibodies” are not all the same

Page 21: Antibody purification – what you need to know to use antibodies effectively

© Innova Biosciences Ltd. All rights reserved

Total IgG vs. specific IgG

• The serum produced when any polyclonal antibody is raised will contain only approximately 10% of antibodies that are specific to the antigen of interest

• Therefore if you purify all of the IgG present, only 10% of that IgG will be specific for the antigen.

• Only antigen affinity chromatography effectively provides IgG that is 100% specific for antigen – Protein A/G/L chromatography provides total IgG, albeit apparently 100% pure when analysed by SDS-PAGE.

• These differences lead to significant differences in activity – a working dilution of an antigen affinity purified antibody is likely to be 10-fold higher than that of a Protein A/G purified antibody.

Why “purified antibodies” are not all the same

Page 22: Antibody purification – what you need to know to use antibodies effectively

© Innova Biosciences Ltd. All rights reserved

Unexpected contaminants – e.g. bovine IgG

• By definition, your purified monoclonal antibody must be 100% pure. Is that right?

• Be aware that if your cell culture media contains foetal/newbornbovine serum there may be significant levels of bovine IgG present – which may co-purify with your monoclonal antibody when using media such as Protein A or G.

• Avoid these either by using a mouse IgG specific purification system or converting all of your hybridoma culture to serum free media.

• If you use ascites you avoid bovine IgG, but you will have normal mouse IgG present which also co-purifies.

Why “purified antibodies” are not all the same

Page 23: Antibody purification – what you need to know to use antibodies effectively

© Innova Biosciences Ltd. All rights reserved

http://en.wikipedia.org/wiki/Fast_protein_liquid_chromatography

Purification doesn’t have to bethis complicated…..

Page 24: Antibody purification – what you need to know to use antibodies effectively

© Innova Biosciences Ltd. All rights reserved

BSA removal kit (820-0010):remove BSA and concentrate your antibody in one easy to use kit!

The AbSelect purification range

Page 25: Antibody purification – what you need to know to use antibodies effectively

© Innova Biosciences Ltd. All rights reserved

AbSelect Rat purification kit

Page 26: Antibody purification – what you need to know to use antibodies effectively

© Innova Biosciences Ltd. All rights reserved

BSA removal kit

Page 27: Antibody purification – what you need to know to use antibodies effectively

© Innova Biosciences Ltd. All rights reserved

What is Lightning-Link® technology?

The worlds fastest, easiest to use and most efficient conjugation technology!

• Only 30 seconds hands-on time! • Over 50 labels available including:

Enzymes, fluorescent proteins / dyes, tandems, biotin & streptavidin

Lightning-Link®

Antibodies – Proteins – Peptides

Fast – Easy-to-use – Reliable

Page 29: Antibody purification – what you need to know to use antibodies effectively

© Innova Biosciences Ltd. All rights reserved

Conjugation considerations

You need to know some things about your reagent.Lightning-Link®, Thunder-Link® and InnovaCoat® conjugations are really simplebut you need protein in the right format to work effectively.

Concentration – 1mg/ml or higher is preferred

Purity – ensure other proteins have been removed, and also make sure they haven’t been put back again afterwards!

Buffer formulation – most common formulations are suitable,but ensure that amines such as glycine are truly absent,as well as thiols such as DTT or mercaptoethanol. Tris is OK up to 20mM

Preferred buffer formulations for the antibodies to be labelled with the various kits differ (especially for nanoparticles), and purification methodsare important to understand

Page 32: Antibody purification – what you need to know to use antibodies effectively

© Innova Biosciences Ltd. All rights reserved

Innova Biosciences Ltd.

Babraham Research Campus,

Cambridge, UK,

CB22 3AT

www.innovabiosciences.com

Lightning-Link® is a registered trademark of Innova BiosciencesDyLight® is a registered trademark of Thermo Fisher Scientific Inc. and its subsidiaries