01.11.2012
A manual Multileaf Collimator
for use in Cobalt-60 teletherapy
Gernot Echner, DKFZ Heidelberg
Stefan Ueltzhoeffer, Precisis AG
AMPICON 2012Mangalore, India
Page 204/13/2023 |
G. EchnerDept. of Med. Phys. in RT
Introduction
• Recent reports: Need for affordable Radiotherapy in LMI1
• Quest for MLC for Co-60 machines
• Industrial partner for development: Precisis AG, Germany
IAEA, Radiation Oncology Physics Handbook, 2005
Rosenblatt et.al (IAEA), Radiotherapy&Oncology 104, (2012) 269-271
Page 304/13/2023 |
G. EchnerDept. of Med. Phys. in RT
Background / Development constraints
• Adaptable to Co-60
• Up-to-date Dosimetry: <2% Leakage, Single Focus, Rounded Leafends (small
Penumbra)
• Field size 20x30 cm2
• Power source independent / Mechanical MLC
• Low Maintenance: Stable mechanism, Standard parts
• Low cost
Image: © Ideenkoch - Fotolia.com
Page 404/13/2023 |
G. EchnerDept. of Med. Phys. in RT
Technical Solution (1)
Drive: Closed-loop Pneumatics
• Optimal “Spring” properties• High force• Low wearing
Leaf Material: Brass
Density = 8.47 g/cm3
μCuZn39Pb3 = 0.4311 cm-1 HVL = 1.6079 cm 6*HVL = 9.65 cm ≙ 1.563 %
Page 504/13/2023 |
G. EchnerDept. of Med. Phys. in RT
Technical Solution (2)
• Drive integrated into Leaves (Compactness)
• Workflow: Mechanical field adjustment
Page 604/13/2023 |
G. EchnerDept. of Med. Phys. in RT
Dosimetry: Material & Methods
Material• PTW Ion chamber• Theratron Elite80 (Ø1.5cm) & Theratronics Equinox (Ø2.0cm)
• GafChromic EBT film• PTW Phantom/Virtual Water
Methods• 80cm/85cm/100cm SSD• Varying field size • pdd & skin dose• Interleaf & End2End Leakage• Penumbra: 80% / 20%
Images: © Precisis AG
Page 704/13/2023 |
G. EchnerDept. of Med. Phys. in RT
Dosimetry: Results Penumbra
Penumbra with depth Penumbra with fieldsize
Source Ø 2.0 cm:Reference Penumbra: 0.85 cm
Source Ø 1.5 cm:Reference Penumbra: 0.55 cm
Source Ø 2.0 cm, Depth 10 cmPenumbra: 0.69 cm – 1.39 cm
Pen
um
bra
/ cm P
enu
mb
ra /
cm
Page 804/13/2023 |
G. EchnerDept. of Med. Phys. in RT
Dosimetry: Results Leakage
Interleaf Leakage End2End Gap
Average Leakage ~1.5%Max < 3%
End2End Gap ~5%Improved planned
Page 904/13/2023 |
G. EchnerDept. of Med. Phys. in RT
Summary
Features:• Drive: Closed-loop Pneumatics, power independent• Field size: 20cm x 30cm• Leaf width: 10mm• Focus: Single focus + rounded leaf ends• Penumbra: < 10mm (Reference Field)• Leakage: < 2%• Compact design (65cm x 35cm x 12,5cm; 45kg)
Marketed by Precisis AG as „Cob(alt)Ra(diation)Leaf“
CobRaLeaf®
Page 1004/13/2023 |
G. EchnerDept. of Med. Phys. in RT
Outlook
• mark -> patient treatment
• Automatic Interlocking system
• Improve ergonomics (easier movement)
• Adaptation also to non-Theratronics machines
Image: © Anatoly Maslennikov - Fotolia.com
Page 1104/13/2023 |
G. EchnerDept. of Med. Phys. in RT
Acknowledgments
Martin Baumann1, Gernot Echner1, Armin Runz 1, Marco Langhans1, Roland Schwandtner2,
Dr. Jozef Kurzeja2, Prof. Irenaeus Admietz2, Mr. Stefan Ueltzhoeffer3, Andrei Ciresianu4, Mark Xu4,
Shawn Luimes4, Prof. Wolfgang Schlegel1
1Deutsches Krebsforschungszentrum, Heidelberg2Klinik f.Strahlentherapie, Evangelisches Krankenhaus Witten 3Precisis AG, Heidelberg4Best Theratronics, Ottawa, Canada
Thank you for your attention
Page 1204/13/2023 |
G. EchnerDept. of Med. Phys. in RT
Backup Information
Skin dose for 85 cm SSD
Field size Skin dose, 85 cm SSD Skin dose, 100 cm SSDEquinox Jaw size (cm)
5x5 54,70% 54,70% 6x6 10x10 59,90% 58,70% 12x12 20x20 79,50% 74,50% 25x25 20x30 79,40%
Properties Tungsten (Densimet®, Plansee)Density: 18.0 g/cm3 Minimum Attenuation w/o coherent scattering (cm2/g) for E=1.332 MeV: 0.0521 μW= 0.9378 cm-1
HVL = ln2/μ=0.739 cm 6*HVL ( Residual:1.563 %) = 4.43 cm (target) Residual: 2.0 % -> 5.644 HVL = 4.17 cm (minimum)
Maximum edge length
Tumor/PTV
10cm 20cm 30cm 40cm0.00%
20.00%
40.00%
60.00%
80.00%
100.00%Co-60
Linac_EN
Linac_DE
Gesamt
Maximum Edge length is >=
Cas
es
Fieldsize Study
Max. Edge length
=>Approx.80% of casestreatable in 1 field