DIALYZER Dialyzer shell
(polyurethane) is a box or tube with four port (2 blood and 2 dialysis)
Dialyzer – blood and dialysis solution
Movement of molecule across semipermeable membrane
DIALYZER
An ideal dialyzer should have High clearance of small and medium MW toxins Adequate UF Negligible losses of protiens and amino acids Non toxic composition Minimal activation of cell or thrombotic pathways Minimal blood volume Reliability Reusability Low cost
DIALYZER TECHNICAL SPECIFICATION Priming volume very from 40-150 ml. Surface area vary from 0.5-2.2m2
KUF from 2.5 to 85ml/h/mmHg KoA urea varies from 200 to 1200 (<300 represents a
low efficiency dialyzer and >600 a high efficiency dialyzer)
Specification sheets also provide clearance for urea and vitamin B12.
Dialyzer are sterilzed by gamma irradiation, ethylene oxide or by steam
Must be pre rinsed with >2L rinsing solution prior to connecting to patient to prevent release of fragments from the dialysis circuit
DIALYZER MEMBRANECellulose: Obtained form processed cotton (regenerated
cellulose, cuprammonium cellulose, cuprammonium rayon and saponified cellulose)
Substituted cellulose: Cellulose polymer has a large number of free hydroxyl
group at its surface Free hydroxyl group are responsible for blood cell
activation causing bio-incompatibiltiy of the dialyzer (cellulose acetate, cellulose diacetate, triacetate)
DIALYZER MEMBRANECellulosynthetic: A synthetic material ( a tertiary amino compound) is
added to liquefied cellulose during formation of the membrane as result the surface of the membrane is altered and biocompatibility is increased (cellosyn or hemphan)
Synthetics: Not cellulose based but are synthetic plastics and
materials used include polyacrylonitrile (PAN) polysulfone, polycarbonate, polyamide and polymethyl-methacrylate (PMMA)
DIALYZER TYPESCoil dialyzer: Flattened cellulose tubing wrapped as a coil and
through which patients blood flow during dialysis. The blood channels was long to obtain the needed
surface area, and resistance was high UF was unpredictable and blood leak were frequent.
PARALLEL PLATE DIALYZER Sheets of membrane are placed between supporting plates. The plates have ridges and grooves to support the membrane and
allow flow of dialysate along it Resistance to blood flow is low. The surface area vary from 0.25 to 1.5
msq.Advantage: Blood volume is about 50-100 ml at 100 mmHg increase with high TMP Heparin requirement usually low, minimal clotting in the blood
compartment Ultrafiltration is reasonably predictable and controllable
Disadvantage: Formation of local thrombi around inlet and outlet ports and corners May lead to bacterial growth and endotoxin formation therefore plates
are not often reused
HOLLOW FIBER DIALYZER Numerous hollow fiber Hollow fiber are tiny its diameter 150-250um Number of fiber20000 or more, depending upon length, kind of
membrane and surface area of dialyzer
Advantage : Low blood volume 60-90ml Resistance to blood flow is low Ultrafiltration can be precisely controlled Well adapted to reuse
Disadvantage: Deaeration of fiber predialysis is necessary to prevent air lock of the
fibers More heparin is required for most of the patients Adverse patient reaction
REACTION TO MEMBRANESType A reaction Within minutes of
starting dialysis with Dyspnoea, wheeze A feeling of warmth,
urticaria Cough, hypotension Collapse or cardiac
arrest
Type B reaction More common but much
milder often occur 20-40 min after starting dialysis
Usually cause back and chest pain