Upload
eric-f-shaver-phd
View
57
Download
0
Embed Size (px)
Citation preview
Copyright©2017HumanFactorsMD.Allrightsreserved.Confidential.
BestPracticesofSafetyInformationPresentationinIFUsEricF. Shaver,PhDDirectorofHumanFactors
6March2017
Copyright©2017HumanFactorsMD.Allrightsreserved.Confidential.
Overview
1. Challengesofdesigningsafetyinformation.
2. IssuescommonlyseeninIFUs.
3. Bestpracticesfordesigningsafetyinformation.
4. Designresourcesforfuturereference.
2
Copyright©2017HumanFactorsMD.Allrightsreserved.Confidential.
DevelopmentChallenges
Numerousdifferentpotentialconsumersofthesafetyinformation.
Oftenhavetodealwithlanguagebarriers.
Devicelabeling- includingsafetyinformation- isoftenlefttothelastminute.
Resourcesareoften limited.
Manypossibledesignfeaturescanbeimplemented.
5
Copyright©2017HumanFactorsMD.Allrightsreserved.Confidential.
DevelopmentChallenges,cont.
Havetoconsiderdevicecomplexity.
6
Copyright©2017HumanFactorsMD.Allrightsreserved.Confidential.
DevelopmentChallenges,cont.
7
DevicelabelingismorethanjusttheIFU.
Copyright©2017HumanFactorsMD.Allrightsreserved.Confidential.
DevelopmentChallenges,cont.
8
Moretoconsiderthansafetyinformation.
Copyright©2017HumanFactorsMD.Allrightsreserved.Confidential.
DevelopmentChallenges,cont.
9
Lotsofresearch;littledesignguidance.
Copyright©2017HumanFactorsMD.Allrightsreserved.Confidential.
BestPractice#1
14
Reference:ANSIZ535.6(2011)
Copyright©2017HumanFactorsMD.Allrightsreserved.Confidential.
BestPractice#1,cont.
15
Reference:ANSIZ535.6(2011)
Copyright©2017HumanFactorsMD.Allrightsreserved.Confidential.
BestPractice#1,cont.
Useariskassessmentprocesstoidentifyhazardsanddeterminewhensafetyinformationiswarranted.
Ensuresafetyinformationisidentifiedwiththecorrectsignalword.
Donotdownplayhazardsbyselectingasignalwordthatconnotesadecreasedlevelofhazardseriousness.
Followdesignconventionsforpresentationofsignalwords.
16
Copyright©2017HumanFactorsMD.Allrightsreserved.Confidential.
BestPractice#2
Beconcisewhenpresentingsafetyinformation.
Minimizethecombiningofinstructionswithsafetyinformation.
Testcomplexsafetyinformationearlywithintendedusers.
19
Copyright©2017HumanFactorsMD.Allrightsreserved.Confidential.
Issue#3,cont.
Seo,etal.(2016)had86healthcareprovidersevaluate38symbolsfromAAMI/ANSI/ISO15223-1.
Only6of38metmodifiedacceptancecriteriafromANSIZ535.3.
Fiveof6containedtextinthevisual.
22
Copyright©2017HumanFactorsMD.Allrightsreserved.Confidential.
BestPractice#3
Provideanexplanationofthesafetysymbolsinaglossary.
Testsymbolsearlywithintendedusers.
Iflowcomprehensionsymbolsarerequired,providecorrespondingtextualexplanation.
23
Copyright©2017HumanFactorsMD.Allrightsreserved.Confidential.
Issue#4
Directsuserstoneworuniquesafetyinformation.
Directsuserstosafetyinformationwithinadocument.
Informsusersthatsafety-relatedinformationislocatedwithinthedocument.
Decreasestheneedtorepeatgenericconsequenceinformationregardingfailuretoreadthedocument.
25
Reference:ANSIZ535.6(2011)
Copyright©2017HumanFactorsMD.Allrightsreserved.Confidential.
BestPractice#5
31
To Clean and Disinfect the System
1. Press the power button to turn the system off.
2. Unplug the power cord from the electrical outlet. WARNING! Failure to unplug the power cord before cleaning and disinfecting could result in serious injury or death.
3. Disconnect the collapsible arm from the system.
4. Detach the end piece from the collapsible arm.
5. Inspect and remove debris from the end piece.
Copyright©2017HumanFactorsMD.Allrightsreserved.Confidential.
BestPractice#5,cont.
32
To Clean and Disinfect the System
1. Press the power button to turn the system off.
2. Unplug the power cord from the electrical outlet. Failure to unplug the power cord before
cleaning and disinfecting could result in serious injury or death.
3. Disconnect the collapsible arm from the system.
4. Detach the end piece from the collapsible arm.
5. Inspect and remove debris from the end piece.
Copyright©2017HumanFactorsMD.Allrightsreserved.Confidential.
BestPractice#5,cont.
33
To Clean and Disinfect the System
1. Press the power button to turn the system off.
2. Unplug the power cord from the electrical outlet. Failure to unplug the power cord before
cleaning and disinfecting could result in serious injury or death.
3. Disconnect the collapsible arm from the system.
4. Detach the end piece from the collapsible arm.
5. Inspect and remove debris from the end piece.
Copyright©2017HumanFactorsMD.Allrightsreserved.Confidential.
AdditionalBestPractices
1. Knowyouraudience.
2. Makeuseofaprocess.
3. Asktherightquestions.
4. IFUsarepartofanecosystem.
5. Usealternativedesignformats.
6. Maintainconsistencywithin- andacross- products.
7. Testearlyandoften.
35
Copyright©2017HumanFactorsMD.Allrightsreserved.Confidential.
References
ANSIZ535.6(2011).ProductsSafetyInformationinProductManuals,Instructions,andOtherCollateralMaterials.NewYork:AmericanNationalStandardsInstitute.
Seo,D.C.,Landoni,M.,Brunk,E.,Becker,M.W.,&Bix,L.(2016).Dohealthcareprofessionalscomprehendstandardizedsymbolspresentonmedicaldevicepackaging?:Animportantfactorinthefightoverlabelspace.PackagingTechnologyandScience.Advanceonlinepublication.10.1002/pts.2199.
40
Copyright©2017HumanFactorsMD.Allrightsreserved.Confidential.
ContactInformation
EricF.Shaver,PhDDirectorofHumanFactorsHumanFactorsMD
Email:[email protected]:www.humanfactorsmd.comLinkedIn:https://www.linkedin.com/company/human-factors-md/Phone:1.800.639.7941
41