Dr. MAMTA Dept of ophthalmology RNT Medical college, Udaipur OPHTHALMOSCOPY

Ophthalmoscopy

Embed Size (px)

Citation preview

Page 1: Ophthalmoscopy

Dr. MAMTA Dept of ophthalmology RNT Medical college, Udaipur

OPHTHALMOSCOPY

Page 2: Ophthalmoscopy

Ophthalmoscopy is a clinical examination of the interior of the eye by means of an ophthalmoscope.

It is primarily done to assess the state of fundus and detect the opacities of ocular media.

The ophthalmoscope was invented by von Helmholtz in 1850.

Introduction

Page 3: Ophthalmoscopy

Structures of the retina

nasal temporal

Page 4: Ophthalmoscopy

Posterior Pole of fundus

Page 5: Ophthalmoscopy

Optic discDISC: LOCATION –nasal to geometric axis

DIAMETER – 1.5mm [1 disc diameter] COLOR – Pale pink SHAPE – Circular EDGES – Regular TERMINATION OF ALL LAYERS EXCEPT NFL CUP: C/D ratio – 0.3 to 0.5

Page 6: Ophthalmoscopy

RETINAL SYSTEM : CENTRAL RETINAL ARTERY AND CENTRAL RETINAL VEIN

Arterioles Venules

Capillaries

CILIARY SYSTEM : POST.CILIARY ARTERIES Choriocapillaries

Vessels

Page 7: Ophthalmoscopy

MACULA Specialised region of retina Diameter – 5.5 mm Location – 2 DD - temporal margin of disc Color – Yellow; deep pigmented 4 zones : Foveola -0.35 mm

Fovea -1.50 mm Parafovea

Perifovea Retinal vessels Cilioretinal artery

Page 8: Ophthalmoscopy

Ophthalmoscopic methods of examination are-

(1)Distant direct ophthalmoscopy (2)Direct ophthalmoscopy (3) Indirect ophthalmoscopy

Page 9: Ophthalmoscopy

The direct ophthalmoscopeParts-

◦ Light source◦ Eyepiece◦ Lens rack and power dial◦ Aperture selector◦ Filter selector◦ On/off and brightness control◦ Power handle

Page 10: Ophthalmoscopy

For a good view of fundus the pupil should be dilated by instilling few drops of short acting mydriatic drug (e.g. combination of tropicamide and phenylephrine )

The subject should be examined in sitting or lying down position.

Examination room should be semidark. keep the eye as still as possible.

METHOD

Page 11: Ophthalmoscopy

For examining right eye of the patient,

Examiner should stand on right side of the patient.

Hold the instrument in his right hand. Use examiner’s right eye.

If examining left eye, stand on left side, hold instrument in left hand use left eye.

Position of the examiner

Page 12: Ophthalmoscopy

Viewing should begin about half meter away from the eye.

First see the “Red reflex” Initially the lens power in the instrument

should be set to zero, and if refractive error present in patient or examiner, e.g. if the patient is myopic then set the (-ve )lens, if the examiner or patient is hypermetropic then set the lens to (+ve) lens. If both patient & examiner have refractive error then sum together their powers.

Page 13: Ophthalmoscopy

Rotate the lens dial until the optic disc is focused clearly.

(Red numbers/negative numbers in myopia or short-sightedness.Green numbers/positive numbers in hyperopia or far-sightedness.)

Examine the optic disc for:

Shape – normally round or slightly ovalClarity of the outline-clear outline or rimColour-pale pink contrast to rich red of rest of fundus.

Page 14: Ophthalmoscopy

Examine the macula last.

To locate the macula, focus on the disc, then move 2 disc diameters temporally. The macula is darker in colour than the surrounding fundus, and devoid of blood vessels.

May also ask the patient to look at the light, this automatically puts the macula into full view.

Page 15: Ophthalmoscopy

Examiner right eye, hand, right patient eye

Page 16: Ophthalmoscopy

It should be performed routinely before the direct ophthalmoscope, as it gives a lot of useful information.

It can be performed with the help of a self-illuminated ophthalmoscope or a simple plain mirror with a hole at the centre.

Procedure- The light is thrown into patients eye sitting in a semi-darkroom, from a distance of 20-25 cm and the features of the red glow in the pupillary area are noted.

(1)Distant direct opthalmoscopy

Page 17: Ophthalmoscopy

1. To diagnose opacities in the refractive media- any opacity in the refractive media is seen as a black shadow in the red glow.

2. To differentiate between a mole and a hole of the iris- a small hole and mole on the iris appears as a black spot on oblique illumination

3. To recognise detached retina or a tumour arising from the fundus is seen as a greyish reflex.

APPLICATIONS

Page 18: Ophthalmoscopy

It is the most commonly practised method for routine fundus examination.

It works on the basic optical principle of glass plate ophthalmoscope introduced by von helmholtz.

A convergent beam of light is reflected into the patients pupil. The emergent rays from any point of the patients fundus reach the observers retina through the viewing hole in the ophthalmoscope.

(2)Direct ophthalmoscopy

Page 19: Ophthalmoscopy

The emergent rays from the patients eye are parallel and brought to focus on the retina of the emmetropic observer when accommodation is relaxed.

In hypermetropic patient,the emergent rays will be divergent and with the help of convex lens can be brought to focus on the observers retina.

In myopic patient,the rays will be convergent and use concave lens for focus on the observers retina.

Page 20: Ophthalmoscopy

Optics of ophthalmoscopy

Divergentlight, if subjecthypermetropic

Reflectedlight

source

Convergentlight, if subject

myopic

Divergentlight

Convergentlight

Parallellight

Focussedon retina

Corrective lens is placed along pathway

Page 21: Ophthalmoscopy

The image is erect, virtual and about 15 times magnified in emmetropes.

Technique- should be performed in a semi-darkroom with the patient seated and looking straight ahead , while the observer standing or seated slightly over to the side of the eye to be examined by the observer with his or her right eye and left with the left.

Page 22: Ophthalmoscopy

The observer should reflect beam of light from the ophthalmoscope into patients pupil. Once the red reflex is seen the observer should move as close to the patients eye as possible.

Once the retina is focused the details should be examined systematically starting from disc,blood vessels, the four quadrant of the general background and the macula.

Page 23: Ophthalmoscopy

Monocular Indirect OphthalmoscopyIt consists of- Illumination rheostat at its base, Focusing lever for image refinement, Filter dial with red free and yellow filters, Forehead rest for proper observer head

positioning, and Iris diaphragm lever to adjust the

illumination beam diameter.

Page 24: Ophthalmoscopy

Optics- an internal relay lens system re-inverts the initially inverted image to a real erect one, which is then magnified. This image is focusable using the focusing lever.

Indications are- Need for an increased field of view Small pupils Uncooperative children Patients intolerance of bright light of

binocular indirect ophthalmoscope.

Page 25: Ophthalmoscopy

Extent of view- although vitreous base views are possible with monocular IO, its greatest effectiveness extends anteriorly to the peripheral equatorial region.

Advantage- field of view similar to IO and, Erect real image similar to DO. Disadvantage –are lack of stereopsis, Limited illumination and, Fixed magnification.

Page 26: Ophthalmoscopy

It is very popular method for examination of posterior segment introduced by Nagel in 1864.

PRINCIPLE- To make the eye highly myopic by placing a strong convex lens in front of patients eye so that the emergent rays from an area of the fundus are brought to the focus as a real,inverted image between the lens and the observers eye.

(3) Indirect ophthalmoscopy

Page 27: Ophthalmoscopy

Indirect ophthalmoscope

Page 28: Ophthalmoscopy

An inverted reverse real image Magnification = 2 to 4 X Field of view = 40 to 50 degrees Optimal working distance = 40 to 50 cms Good illumination & stereopsis Ease of use with scleral indentor Lenses from 14 to 30 D range

Page 29: Ophthalmoscopy

CHARACTERISTICS- Magnification of image depends upon the dioptric power of convex lens, position of the lens in relation of the eyeball and refractive state of eyeball.

About 5 times magnification is obtained with +13 D lens.

With a stronger lens,image will be smaller, but brighter and field of vision will be more.

Page 30: Ophthalmoscopy

(1) Dark room (2) source of light and concave mirror or self illuminated indirect ophthalmoscope(3) Convex lens(4) pupils of the patient should be dilated.

Prerequisties

Page 31: Ophthalmoscopy

The patient is made to lie in the supine position, with one pillow on a bed or couch and instructed to keep both eyes open.

The examiner throws the light into patients eye from an arms distance.

In practise, Binocular ophthalmoscope with head band or that mounted on the spectacle frame is employed most frequently.

Technique

Page 32: Ophthalmoscopy

Keeping his or her eyes on the reflex, the examiner then interposes the condensing lens in the path of beam of light, close to the patient eye, and then slowly moves the lens away from eye until the image of retina is clearly seen.

Page 33: Ophthalmoscopy

The examiner moves around the head of the patient to examine different quadrants of the fundus.

He or she has to stand opposite to clock hour position to be examined.

By asking the patient to look in extreme gaze, and using of scleral indenter, the whole peripheral retina up to ora serrata can be examined.

Page 34: Ophthalmoscopy

Scleral indentation- helps in making prominent the barely perceptible lesions, done with the depressor placed on patients lid.

Examiner should move the scleral depressor in a direction opposite to that in which he or she wishes the depression to appear.

Scleral depressor should be rolled gently and tangentially over the eye surface.

The temporal part of upper lid is sufficiently lax so depressor can be placed inferiorly in the horizontal meridian.

Page 35: Ophthalmoscopy

Scleral Indentation-Thimble scleral depressor

-Pencil type depressor-Cotton tipped applicator

To examine periphery between equator and ora serrata by creating a mound to view. Start superonasal superior ,superotemporal,Inferotemporal, inferior, inferonasal

Page 36: Ophthalmoscopy

FUNDUS DRAWING- The image seen with IO is vertically inverted

and laterally reversed; top of the retinal chart is placed towards the foot end of the patient [i.e.upside down]

Fundus drawing is made on a special Amsler chart,which has 12 clock hours marked and has three concentric circles made on it-

Innermost circle,middle circle and outermost circle represents to the equator,the ora serrata and,the midpoint of pars plana respectively.

Page 37: Ophthalmoscopy

Fundus diagram

Page 38: Ophthalmoscopy
Page 39: Ophthalmoscopy

-Technique is difficult and can be mastered by hours of practice.

-Magnification is less compare to DO.-It is impossible with very small pupil.-Patient is usually more uncomfortable with

intense light of IO and with scleral indentation.

-Reflex sneezing can occur on exposure to bright light.

Disadvantages

Page 40: Ophthalmoscopy

Advantages of Indirect system• Image not affected by the patients refractive power

• In eyes with nystagmus

• Delivery of LASER

• Binocular examination of fundus up-to the periphery

• Large field of view allow for the panoramic view

Page 41: Ophthalmoscopy

• Better Resolution

• Use in operating room for cryo/scleral buckling

• Better view in presence of media opacities

• Increased illumination• Reduced distortion

Page 42: Ophthalmoscopy
Page 43: Ophthalmoscopy

Dropout in glaucoma,myelination at disc margin.

Crossings: nipping insystemic hypertension

Cupping and notchingin glaucoma

Swelling and blurredmargins in papilloedemaand optic neuritis

Vessel walls: sheathing in systemic hypertension, leakage and neovascularisation in diabetes.

Bifurcations: embolisms,branch occlusions

Pigmentation changesin retinitis pigmentosa

Haemorrhagesin diabetes, vessel occlusion,hypertension

Myopic crescentin myopia

ARMD, drusen,macular holes

Retinal tears,detachments

Anterior ischaemic opticneuropathy in diabetes

Vessels

Optic disc

Fovea

Background/periphery

Retinal nerve fibre layer

VitreousAsteroid hyalosis, floaters,haemorrhages

Page 44: Ophthalmoscopy

(1) Hruby lens biomicroscopy (2) Contact lens biomicroscopy (3) Indirect fundus biomicroscopy

Slit lamp biomicroscopic examination of the fundus

Page 45: Ophthalmoscopy

Hruby lens is a planoconcave lens with diopteric power 58.6D, which neutralizes the optical power of normal eye (60D) and forms a virtual,erect image of the fundus.

This lens provides a small field with low magnification and cannot visualize the fundus beyond equator.

HRUBY LENS BIOMICROSCOPY

Page 46: Ophthalmoscopy
Page 47: Ophthalmoscopy

It combines stereopsis, high illumination and high magnification with the advantages of slit beam.

Modified Koeppe lens, i.e. posterior fundus contact lens can be used to examine the posterior segment, provides a virtual and erect image.

Goldmanns three-mirror contact lens, consist of a central contact lens and three mirror placed in the cone,each with different angles of inclination.

Contact lens biomicroscopy

Page 48: Ophthalmoscopy

With this, the central as well as peripheral parts of the fundus can be visualized.

It also provides a virtual and erect image. By flat central portion can see posterior

vitreous and pole.

Angled mirror - 73 deg - Peripheral fundus 67 deg - Pars plana 59 deg - Gonioscopy

Eliminates total internal reflection by replacement with cornea – goldmann contact lens interface.

Page 49: Ophthalmoscopy
Page 50: Ophthalmoscopy
Page 51: Ophthalmoscopy

Also known as non-contact fundus biomicroscopy, most commonly used lenses are +78D and +90D, which are mostly double aspheric lenses, so it does not matter which side is held towards the patient.

Optics similar to IO, thus a real, inverted image is formed between the condensing lens and objective lens of slit-lamp.

Magnification provided by fundus non-contact lenses is calculated by dividing power of the eye by the power of lens, e.g. +90D lens provides a magnification of 60/90=0.75X,i.e. a minification of the image.

Indirect Fundus Biomicroscopy

Page 52: Ophthalmoscopy

However, the Magnified image seen because of the magnification provided by the slit-lamp.

Magnifying knob to 10X or 16X. Field of view high powered lens provides

larger field of view but lesser magnification. Drawing the slit lamp biomicroscopic view: Done on an inverted fundus chart and paper

is turned as patients gaze direction changes in respective clock hour meridian.

Page 53: Ophthalmoscopy

Performing indirect slit lamp biomicroscopy

Evaluating fundus :

Page 54: Ophthalmoscopy

Image formation in indirect fundus biomicroscopy

Page 55: Ophthalmoscopy

PanfundoscopyProvides wide field 130 degree and high power lens

A real inverted magnified image is formed

It is used in both posterior fundus examination and also Laser pan retinal photo Coagulation.

Page 56: Ophthalmoscopy