View
404
Download
1
Embed Size (px)
Citation preview
Chhattisgarh Dental College & Research InstituteRAJNANDGAON, CHHATTISGARH
Ayush And Health Sciences University of Chhattisgarh L i b r a r y D i s s e r t a t i o n
P a g e | 0
Mini Implants
P a g e | 1
Contents
1 INTRODUCTION.................................................................................................8
2 HISTORY OF MINI IMPLANTS.......................................................................12
3 THE DESIGN OF ORTHODONTIC MINI-IMPLANTS...................................22
3.1 Classifications of miniscrew Implants..........................................................22
3.2 Elements of Implant Design:.........................................................................26
4 ANATOMICAL CONSIDERATIONS................................................................37
5 BIOMECHANICAL CONSIDERATIONS.........................................................44
5.1 Maxillary anterior en masse retraction mechanics in extraction cases.........44
5.2 Mandibular anterior en masse retraction mechanics in extraction cases......46
5.3 Anterior Intrusion Mechanics in the Maxillary Arch....................................47
5.4 Anterior Intrusion Mechanics in the Mandibular Arch.................................47
5.5 Anterior en masse Retraction with Anterior Intrusion..................................48
5.6 Molar Intrusion Mechanics for Open bite Cases...........................................48
5.7 Maxillary Anterior Lingual Root Torque Mechanics....................................48
5.8 Molar Distalization Mechanics for Non-Extraction Cases...........................49
5.9 Protraction Mechanics in Extraction Cases...................................................50
5.10 Minor Tooth Movements Using Mini Implant Anchorage........................51
5.11 Buccal Cross bite (Scissors Bite) Correction.............................................52
6 INDICATIONS & CONTRAINDICATIONS.....................................................54
6.1 Indications.....................................................................................................54
6.2 Contraindications..........................................................................................55
Mini Implants
P a g e | 2
6.3 Relative Contraindications............................................................................56
7 CLINICAL APPLICATIONS..............................................................................58
8 BIOLOGY OF ORTHODONTIC IMPLANTS...................................................63
9 COMPLICATIONS WITH MINI IMPLANTS & THEIR MANAGEMENT....69
9.1 Complications During Insertion....................................................................70
9.2 Complications Under Orthodontic Loading..................................................75
9.3 Soft-Tissue Complications............................................................................78
9.4 Complications During Removal....................................................................80
10 ADVANTAGES OF MINI IMPLANTS..........................................................83
11 BIBLIOGRAPHY............................................................................................84
12 APPENDIX I.......................................................................................................86
13 APPENDIX II.....................................................................................................88
Mini Implants
P a g e | 3
INTRODUCTION“Truth is neither objectivity nor the balanced view; truth is a selfless subjectivity.”
-Knut Hamsun
Mini Implants
P a g e | 4
1 INTRODUCTION
The goal of Orthodontic practice is to achieve the desired results with minimum
undesired consequences, which is largely dependent on the availability of anchorage.
One such effect is the loss of anchorage. Anchorage, the degree of resistance to
displacement, is a critical component to successful orthodontic treatment. Anchorage
control is a fundamental problem in the treatment of malocclusions1.
In orthodontics, malpositioned teeth are corrected and proper alignment is achieved
by inducing strategically planned tooth movements through application of force. This
force originates from wires, elastics and other appliances attached to the teeth. To make
this force work in the planned direction, with the planned magnitude and for the
planned time, it is necessary to reduce or completely eliminate unwanted reciprocal
effects by using a reliable anchorage and to respect the principles of orthodontic
biomechanics at the same time.
Often, teeth that are in proper alignment are used to provide the force to move those
that are not, and are referred to as anchorage teeth. In accordance with Newton’s Third
Law of motion, outlined in 1687, which states that an applied force comprises of an
‘Action’ component & an equal & opposite ‘Reaction’ component; There is a reactive
or “equal and opposite force” for every applied force. Unfortunately, these reactive
orthodontic forces often result in undesirable movements of the teeth serving as
anchorage. As a result, orthodontists have historically used a variety of appliances and
strategies to enhance anchorage; particularly when minimal movement of the teeth
providing the anchorage is desired. This allows the movement of malaligned teeth
while leaving teeth that do not need to be moved relatively undisturbed.
However, with dental anchorage dependant on the number and quality of the teeth, if
anchorage is lost during the planned correction of the malocclusion anomaly will not
Mini Implants
P a g e | 5
be achieved. Anchorage enhancing appliances, such as headgear, are highly dependent
upon patient compliance for success. All manner of anchorage control concepts have
come and gone in the ensuing years, including Baker anchorage, headgear, lip
bumpers, Nance holding arches, Tip-back bends, lingual arches, uprighting springs,
sectional mechanics, dual arch mechanics and elaborate methods to set anchorage. But
in reality, absolute anchorage did not practically exist.
In an effort to establish anchorage without significant reliance on patient
cooperation, other forms of anchorage have been investigated2.
Restorative dental implants, despite their stability in bone, have limited use in
orthodontics due to cost, an extensive healing period after surgical placement, and
anatomic placement limitations. Still, the use of these titanium dental implants as a
form of anchorage has provided the potential for absolute, compliance independent,
orthodontic anchorage. The use of prosthetic implants as orthodontic anchorage made a
great advance. In osseointegrated implants, the relationship between the implant and
bone can be defined as a functional ankylosis.
Orthodontic miniscrew implants (read mini-implants) have been designed to
circumvent the limitations posed by restorative dental implants. These smaller bone
screws are significantly less expensive, are easily placed and removed, and can be
placed in almost any intra-oral region, including between the roots of the teeth.
Starting with the use of vitalium screwGainsforth, 1945 #7s3 and progressing to
conventional osseointegrated implants which have been used as orthodontic anchorage,
orthodontic therapy seems to be essentially facilitated.
Mini Implants
P a g e | 6
1.1 List of common terms used for mini-screw implants :
• Microimplants
• Micro-implant anchors
• Mini implants
• Mini implants for orthodontic anchorage
• Mini screws
• Ortho implants
• Ortho TAD
• TAD – Temporary Anchorage Devices
• Titanium screw Anchorage
As described by Ludwig 4
In 2005 by Carano and Melsen5 , it was agreed that the word Mini-Implant should
be applied to all these terms. In 2005 Mah and Bergstrand agreed to this aspect and
they pointed out that mini-implant is more appropriate than micro-implant or screw6,
because the word “micro” is defined as a magnitude of 10-6.
----------------------------------- Ω -----------------------------------
Mini Implants
P a g e | 7
History of Mini Implants“I am the owner of the sphere, the seven stars and of the solar year of Caeser’s
hand, and Plato’s brain, of Lord Christ’s heart, and Shakespeare’s strain, I am
History.”
- Ralph Waldo Emmerson
Mini Implants
P a g e | 8
2 HISTORY OF MINI IMPLANTS
In the 19th century, concepts were designed to anchor teeth to structures other than
teeth. This led to the development of Extra-oral anchorage systems. These appliances,
although mechanically efficient, imposed enough discomfort to the patients to deter
them from continuing their use and created practical compliance issues, affecting the
overall therapy negatively.
The era of skeletal anchorage began later, in 1945, with the failed experiments of
Gainsforth and Higley3 to anchor screws in the jaws of mongrel dogs. Using a 2.4-mm
pilot hole, a 3.4-mm-diameter X 13-mm-long Vitallium screw was placed in the
ascending ramus of 6 dogs. A rubber band delivering between 140 and 200 g of force
was attached from the screw head to a 0.040-inch wire that slid through a tube on the
upper molar band and was soldered to the upper canine band. The system was designed
to distally tip/retract the canine by immediately loading the screws with the rubber
bands. However, all the screws were lost in 16-31 days. The experiment, although
unsuccessful, the authors concluded that “anchorage may be obtained for orthodontic
movements in the future.”
There were no more published reports of attempts to use endosseous implants to
move teeth until the clinical case reports of Leonard Linkow in 19697. He used
mandibular blade-vent implants in a patient to apply Class ll elastics for retraction of
maxillary incisors.
Mini Implants
P a g e | 9
Figure 1 A- Spiral-shaft
implants (Chercheve). B- Various
sizes of early designed vent-plant
implants. C - a tripod implant,
assembled. D- Various early
designed blade implants. 8
In the 1970s, after Branemark and co-workers reported the successful
osseointegration of implants in bone 9, many orthodontists began taking
an interest in using implants for
orthodontic anchorage. Sherman
placed six vitreous carbon dental
implants into the extraction sites of
mandibular third premolars of dogs
and applied orthodontic forces in
1978. Two of the implants were firm
and were considered to be successful 10.
1979, Smith won the Milo Hellman Research Award of the American Association of
Orthodontists by studying the effects of loading bioglass-coated aluminum oxide
implants in monkeys and reported no significant movement of the implants during
force application. He described the interface between the bioglass implants and the
Mini Implants
P a g e | 10
surrounding tissue as fusion or ankylosis, despite the observation that intervening areas
of connective tissue were present 11.
After other similar efforts did not show much success, it was revisited by Creekmore
and Eklund in 1983. They inserted vitallium surgical bone screws below spina nasalis
to anchor upper incisor intrusion 12; the result was 6 mm of intrusion. Later, in similar
experiments, by Turley et al13 , the in-depth clinical and animal experimental
investigations established the base for today’s use of skeletal anchorage techniques.
The most important studies published between 1970 and 2000 were systematically
reviewed and, related to implants for orthodontic anchorage by Favero and Bressan in
their published article Orthodontic anchorage with specific fixtures: related study
analysis. The literature analysis was divided in specifics topics as materials, size and
screws shape, biomechanics, loading and healing time, used strength, surgery and
reasons to evaluate the success. One of very important aspects is that biocompatible
material had to be used.
Mini Implants
P a g e | 11
In Favero et al’s literature review there are different studies by different authors,
done with different materials 14.
EXPERIMENTS WITH TITANIUM
Experiments on animals Experiments on humans
Roberts 1984
Garetto 1985
Turley 1988
Roberts 1989
Linder-Aronson 1990
Wehrbein 1993
Block 1995
Southard 1995
Miotti 1996
Wehrbein 1997
Parr 1997
Akin -Nergiz 1998
Van Roekel 1989
Roberts 1990
Haanaes 1991
Higuchi 1991
Odman 1994
Roberts 1994
Wehrbein 1996
Kanomi 1997
Wehrbein 1998
Umemori 1999
Wehrbein 1999
Mini Implants
P a g e | 12
Experiments With Other Materials
Ceramics Biodegradable
materials
(Polyactide)
Vitallium Ticonium
Sherman 1978
Smith JR 1979
Mendez C 1980
Paige S 1980
Turley PK 1980
Gray 1983
Glatzmaier 1995
Glatzmaier 1996
Turley PK 1983
Gray JB 1983
Creekmore TD 1983
Douglass 1987
It took 60 years to progress from vitallium to the current standard, titanium.
Although, it ranks ninth among the earth’s most abundant elements, titanium was not
discovered until 1791 and was not mass produced before 1948 15. Three times stronger
than stainless steel, it exhibits little response to heat, electricity or magnetic forces. It is
highly biocompatible; and is inert. Type v titanium has the highest tensile strength,
hence most suitable for bone screws.
Mini Implants
P a g e | 13
In 1988, Shapiro and Kokich described the possibility of using dental implants for
anchorage during orthodontic treatment prior to being used for prosthodontic purposes 16. They emphasized the importance of the position of the implants as well as the proper
case selection and implant requirements during diagnosis and treatment planning.
1994, Roberts and co-workers reported on the clinical application of 3.75 mm x 7.0
mm standard Branemark fixture as anchorage in the retromolar area for closing a
mandibular first molar extraction site17. An anchorage wire attached to the implant was
extended to the vertical tube of the premolar bracket. Stabilizing the premolar anterior
to the extraction site allowed mesial movement of the molars without distal movement
of the more anterior teeth.
The osseointegration is analogous to the situation in which ankylosed tooth that can
be orthodontically loaded without moving, working as a stable anchorage unit, keeping
in mind that the absence of the periodontal membrane does not allow the cell
alterations which result in movements18. Nevertheless conventional dental implants can
be placed only in limited areas such as in the retromolar region or in edentulous areas. 19, 20. Another limitation has been the direction of force application such as when a
dental implant is placed on the alveolar ridge and is too large for horizontal orthodontic
traction. Furthermore, dental implants are troublesome for patients because of the
severity of surgery, the discomfort of the initial healing, and the difficulty of oral
hygiene.
Mini Implants
P a g e | 14
In 1997, Kanomi described a mini-implant
specifically made for orthodontic use21. He
reported that 1.2 mm diameter titanium mini-
implants provided sufficient anchorage for
intruding the lower anterior teeth. After four
months, the mandibular incisors were intruded
6 mm. Neither root resorption nor periodontal
pathology was evident. Kanomi applied an orthodontic force on the mini-implant
Figure 3 several months after implantation, anticipating osseointegration between the
mini-implant and the bone.
Figure 3 Intrusion achieved by mini implants in studies by Kanomi 21
He also mentioned the possibility of mini-implants being used for horizontal
traction, for molar intrusion, and as an anchor for molar distalization and distraction
osteogenesis.
1998 Costa et al presented a screw with a bracket-like head, using 2mm titanium
miniscrews for orthodontic anchorage22. The screws were inserted manually with a
screw driver directly through the mucosa without making a flap and were loaded
immediately. Of the 16 miniscrews used during the clinical trial, two became loose and
Mini Implants
Figure 2 Placement of mini-implant
P a g e | 15
subsequently were lost before treatment was finished. They suggested that miniscrews
could be placed into
Inferior surface of the anterior nasal spine,
Midpalatal suture,
Infrazygomatic crest,
Retromolar area,
Mandibular symphysis area, and
Between the premolar and molar regions.
Majzoub and colleagues, in 1999, investigated the bone response of endosseous
implants to orthodontic loading23. Twenty-four short-threaded titanium implants were
inserted into the calvarial midpalattal suture of 10 rabbits. Two weeks following
insertion, a continuous distalizing force of 150 gm was applied for a period of eight
weeks. All but one test implant remained stable, exhibiting no mobility or displacement
throughout the experimental loading period.
The clinical applications with different micro-implants shapes and the micro-
implants with bracket head advantages were described in a research24. The micro-
implants with bracket head have two fins and a slot as a bracket making it easy for the
wire and ligature placement. There are two different threads for micro-implants:
clockwise and anticlockwise (must be anticlockwise clamped). This new technique
allows a simplified treatment, letting many possibilities of confection without the
necessity of complete appliances. This way, the treatment is faster, without the patient’s
cooperation. Different head types and shapes available in the market were presented.
Mini Implants
P a g e | 16
The self-puncturing titanium orthodontic micro-screws have changed the concepts
as in regard to surgical orthodontic treatments which use the skeletal anchorage
through a simplified and safe surgical approach. The installation of this new device
requires specific knowledge on surgical techniques, clinical application and judgment
for selecting the micro-screws, as well as the orthodontic activation
----------------------------Ω------------------------------------
Mini Implants
P a g e | 17
The design of Orthodontic
mini-implants.“Simple is good”
-Jim Henson
Mini Implants
P a g e | 18
3 THE DESIGN OF ORTHODONTIC MINI-IMPLANTS.
3.1 Classifications of miniscrew Implants.
Labanauskaite et al in 2005 classified Orthodontic implants. 25
I. According to the shape and size
II. According to the implant bone contact.
Mini Implants
Osseointegrated.Nonosseointegrated.
P a g e | 19
III. According to the application
- Used only for orthodontic purposes (orthodontic implants)
- Used for prosthodontic and orthodontic purposes
(prosthodontic implants).
Cope 26 classified Temporary anchorage devices as –
Biocompatible TADs.
Biologic TADs.
Mini Implants
P a g e | 20
Biocompatible TADs are further classified as follows-
Figure 4. Classification of Biocompatible TADs.
Mini Implants
Biocompatible TADsOsseointegrationDental Implant Palatal ImplantRetromolar ImplantPalatal Onplant Mechanical RetentionFixation Screws Fixation Screws with PlatesMini ImplantsFixation Wires
P a g e | 21
Biologic Temporary Anchorage devices are classsified as -
Figure 5 Classification of Biologic TAD.
Currently, there are a number of commercially available miniscrew
implant systems for orthodontic use (Table I);
Mini Implants
Biological TADsOsseointegrationAnkylosed teethMechanicalDilacerated teeth
P a g e | 22
3.2 Elements of Implant Design:
The main differences between the currently available miniscrew implants relate to
their composition, size, and design and include:
1. The alloy or metal used for their fabrication
2. Biocompatibilty
3. The design of the head portion
4. The design and diameter of threaded portion,
5. The length of the implant
3.2.1 Implant alloy or metal used -
The material must be nontoxic and biocompatible, possess excellent mechanical
properties, and provide resistance to stress, strain, and corrosion.
The materials commonly used for implants can be divided into 3 categories 14:
1. Biotolerant (stainless steel, chromium-cobalt alloy),
2. Bioinert (titanium, carbon), and
3. Bioactive (vetroceramic apatite hydroxide, ceramic oxidized
aluminum).
Commercially pure titanium is the material most often used in implantology. It
consists of 99.5% titanium, and the remaining 0.5% is other elements, such as carbon,
oxygen, nitrogen, and hydrogen. Most studies in the literature report on the use of
traditional titanium for prostheses, sometimes modified in the abutments to adapt to
Mini Implants
P a g e | 23
orthodontic requirements. However, pure titanium has less fatigue strength than
titanium alloys. A titanium alloy—titanium + aluminum+ vanadium (Ti6Al4Va)—is
used to overcome this disadvantage. Its mechanical characteristics, moreover, are well
suited to implant requirements: it is very lightweight, and it has excellent resistance to
traction and breaking, enabling it to withstand both masticatory loads and the stresses
of orthodontic forces.
The Orthodontic Mini Implant (OMI) [Leone, Italy] is made from implant steel
1.4441, which is still used in traumatology.
3.2.2 Biocompatibility:
With the exception of the Orthodontic Mini Implant, which is fabricated from
stainless steel, all other aforementioned systems are made of medical type IV or type V
titanium alloy. Because of its particular characteristics, titanium is considered an
excellent material: no correlation has been shown between titanium and the
development of neoplasm, and no allergic or
immunological reactions have been noted that
could be traced to the use of pure titanium.
Bone grows along the titanium oxide surface,
which is formed after contact with air or tissue
fluid. 27
Mini Implants
Figure 6
P a g e | 24
3.2.3 Head design:
Most miniscrew implant systems are available in different designs to accommodate
both direct and indirect anchorage and avoid tissue irritation. The most frequent is the
button-like design with a sphere or a double sphere-like shape or a hexagonal shape.
Figure 7 Various miniscrew implants. A, The Aarhus Anchorage System. B, The AbsoAnchor. C, The
Spider Screw Anchorage System . D, The IMTEC Mini Ortho Implant
Miniscrew implants available with this design include the Aarhus Anchorage
System, the AbsoAnchor System, the Dual-Top Anchor System, the IMTEC Mini
Ortho Implant, the Lin/Liou Orthodontic Mini Anchorage Screw, the Miniscrew
Anchorage System, the Orthoanchor Kl System, and the Spider Screw Anchorage
Mini Implants
P a g e | 25
System. 28With a hole through the head or the neck of the screw, usually 0.8 mm in
diameter, this design is mostly used for direct anchorage.
A bracket like design is also available, which can be used for either direct or indirect
anchorage as provided by the Aarhus Anchorage System, the AbsoAnchor System, the
Dual-Top Anchor System, the Spider Screw Anchorage System, and the Temporary
Mini Orthodontic Anchorage System. 29.
Finally, a further hook design
is used by the LOMAS miniscrew
implant.
Dalstra 28 found that as Bone
density increases, the resistance
created by the stress surrounding
the screw becomes more
important in removal than in
insertion of the screw. At
removal, the stress is con-
centrated in the neck of the screw.
If an Allen wrench is used for
insertion and removal, the hole in
the center of the screw will weaken
the neck, which may lead to fracture. A hollow neck facilitates the insertion of a
ligature, but also weakens the neck. The strength of the screw is optimized by using a
slightly tapered cortical shape and a solid head with a screwdriver slot.
Mini Implants
Figure 8 From left to right: 1.5 mm cylindrical mini-implant
with a slot at the screw head; 1.6 mm tapered mini-implant with
a slot at the screw head; and 2.0 mm cylindrical mini-implant
with threads at the screw head
P a g e | 26
3.2.4 Thread design:
3.2.4.1 Self-drilling versus self-tapping –
The miniscrew implant can have self tapping or self drilling thread design. A
prerequisite for self tapping placement is pilot drilling to prepare a hole for the implant.
Particularly for interradicular microimplants, pilot drilling has potential dangers, such
as damage to tooth roots, drill-bit breakage, overdrilling, and thermal necrosis of bone.
While self drilling screws doesn’t need pilot drilling for placement and can be placed
conveniently in narrow interdental areas.
Yan Chen 30 in 2008, compared the influences of Self-drilling versus self-tapping
orthodontic microimplants on the surrounding tissues biomechanically and
histologically in dogs. Fifty-six titanium alloy microimplants were placed on the buccal
side of the maxillae and the mandibles in 2 dogs were divided into 2 groups of 28; one
group of microimplants was self-drilling, and the other was self-tapping.
Approximately 200 g of continuous and constant forces were applied immediately
between 2 microimplants by stretching closed nickel-titanium coil springs for 9 weeks.
Peak insertion torque and removal torque were recorded immediately after the implants
were placed and when the dogs were killed, respectively.
They found that success rates were higher in the self-drilling group (93%) than in
self-tapping group (86%). Higher peak insertion torque and peak removal torque values
were seen in the self-drilling group in both the maxilla and the mandible. The mean
Peak Insertion Torque (PIT) values of the Self Drilling Implants (SDIs) in the maxilla
and the mandible were 5.6 and 8.7 Ncm, respectively. In the Self Tapping Implants
(STI) group, PIT values were 3.5 Ncm for the maxilla and 7.4 Ncm for the mandible.
A tendency to fracture was found in self-drilling group. The percentage of bone-to-
implant contact values was greater in the self-drilling group. They further found that
microimplants with high PIT might have closer initial contact with bone and might be
good for bone remodeling. SDIs cause less damage to bone during placement;
Mini Implants
P a g e | 27
therefore, osseointegration might happen earlier and be better than with STIs. From an
anchorage perspective, the greater the contact between the surface of the microimplant
threads and the cortical bone, the higher the initial grip of microimplants would be.
They concluded that Self-drilling microimplants can provide better anchorage and
can be recommended for use in the maxilla and in thin cortical bone areas of the
mandible.
It should be noted that even though SDIs have many advantages, if the bone is
dense, they should not to be chosen. STIs should be considered instead. In the maxilla
and areas with thin cortical bone in the mandible, microimplants would penetrate
easily. Failure due to stripping of bone was infrequent, so pilot drilling was not
necessary.
3.2.4.2 Conical Versus Cylindrical Designs –
Wilmes et al in 2008 found that conical mini-implants achieved higher primary
stabilities than cylindrical designs 31. Shinya Yano in 2006 suggested that tapered
orthodontic miniscrews induce bone-screw cohesion following immediate loading
hence advocated use of tapered rather than cylindrical designs. 32
Many other authors advocated the use of conical, rather than cylindrical designs, of
the implant body.
Another concern of using conical design is a
significant increase of MIT was observed mainly
in the taper type miniscrew which leads to higher
bone to implant contact surface along with higher
initial stability of the implants.
Although the conical group required high
removal torque, which means good initial
stability, it also showed high insertion torque
Mini Implants
P a g e | 28
which could affect adjacent tissue healing. The success rates and histomorphometric
analysis showed no significant difference between the cylindrical and conical groups.
The conical shape may need modification of the thread structure and insertion
technique to reduce the excessive insertion torque while maintaining the high
resistance to removal.
3.2.5 Length and diameter of the miniscrew implant -
Miniscrew implants are available in different lengths and diameters to accommodate
placement at different sites in both jaws. Most miniscrew implants have a thread
diameter ranging from 1.2 to 2.0 mm and a length from 4.0 to 12.0 mm, although some
of them are also available at lengths of 14.17 or even 21 mm. The advantage of thin
screw is the ease of insertion between the roots without the risk of root contact.
Seong-A Lim in 2008 suggested that that the maximum insertion torque increased
with increasing diameter and length of the orthodontic miniscrews as well as increasing
cortical bone thickness 33. An increase in screw diameter can efficiently reinforce the
initial stability of miniscrews, but the proximity of the root at the implanted site should
be considered. Miyawaki et al in 2003 reported that a diameter of 1.0 mm or less were
associated with mobility and failure of the screw 34. While Lin et al in 2003 and
Dalstra in 2004 showed that there increased chances of fracture with miniscrews of
diameters less than 1.2 mm 35.
Miyawaki in 2003 34 suggested use of miniscrew implant longer than 5 mm and
diameter more than 1mm. Costa et al in 2005 reported that miniscrew implants of 4 to
6 mm in length and diameter more than 1.2 mm are safe. Deguchi et al 36 after
evaluating the cortical bone thickness with the help of three-dimensional computed
tomographic suggested use of miniscrew implant of length 6 - 8 mm and 1.2- to 1.5-mm
Mini Implants
P a g e | 29
maximum diameter. In addition, for lingual orthodontics, the recommended location is
mesial to the first molar at 30°, and 8 to 10 mm in length.
Overall recent studies unanimously agree using miniscrew implant of length 6 - 8
mm and 1.2- to 1.5-mm in diameter is safe.
3.2.6 Osseointegration:
Because complete osseointegration of screws used in orthodontic applications is a
disadvantage that complicates the removal process, most of these devices are
manufactured with a smooth surface, thereby minimizing the development of bone in
growth and promoting soft tissue attachment at ordinary conditions and in the
absence of special surface treatment regimens. Till now only one study has been done
by Chaddad 37 in 2008 who have tried to modify miniscrew-implant surface
sandblasted, large grit, acid-etched (SLA).
Chaddad evaluated the clinical performance and the survival rate of machined
titanium versus sandblasted, large grit, acid-etched (SLA) mini-implants under
immediate orthodontic loading. Seventeen machined titanium (MT) mini-implants and
15 sandblasted, large grit, acid-etched (SLA) mini-implants were placed in 10 patients.
The mini-implants were immediately loaded and the patients seen at 7, 14, 30, 60, and
150 days. Clinical parameters such as anatomical location, character of the soft tissue
at the screw head emergence, type of mini-implant system, diameter, and length were
analyzed. In addition, the insertion torque recorded at the time of insertion was also
assessed.
Although the survival rate of the SLA mini-implants in this investigation was higher
compared with the MT group (93.5% to 82.5%), the correlation between the implant
surface characteristics and the rate of success was not statistically significant 37. These
findings suggest that altering an implant surface to create more surface area and
increase bone contact may not be the primary consideration when using mini-implants
Mini Implants
P a g e | 30
as orthodontic anchors. The SLA mini-implants presented a higher level of
osseointegration at the time of removal. This clinical observation was based on the
higher torque necessary for removal of SLA mini-implants when compared with
smooth machined titanium implants. Authors clinical experience indicates that surface
treated (SLA) implants could be advantageous in areas of poor bone quality, and
loading should be delayed for 6 to 8 weeks when initial osseointegration has occurred.
In the failure group, all the fixtures had their screw emergence at the oral mucosa
and recorded a torque range of less than 15 Ncm. The insertion torque statistically
influenced the survival rate of the mini-implants (P <.05). Surface treatment,
anatomical location, as well as soft tissue emergence were not statistically significant.
Authors concluded that surface characteristics did not appear to influence survival
rates of immediately loaded mini-implant. A torque value of more than 15 Ncm
recorded at the time of insertion appears to be one of the critical variables for mini-
implant survival under immediate loading.
3.2.7 Insertion torque -
The initial stability of a miniscrew is important because most incidences of
orthodontic miniscrew failure occur at the early stage. Miyawaki in 2003 reported that
the insertion torque is commonly used to evaluate the mechanical stability of implants
including miniscrews.
Studies have shown that a certain level of insertion torque is necessary in order to
achieve the initial anchorage at the screw and the bone interface, and that the insertion
torque of mini screws is an important factor in determining the appropriate initial
stability of a screw. Furthermore, it was suggested that excessive insertion torque, heat
at the border between the screw and bone, and mechanical injury can cause
degeneration of the bone at the implant-tissue interface.
Motoyoshi determined an adequate implant placement torque (IPT) for obtaining a
better success rate of 124 miniscrew-implants that were screwed into the buccal
Mini Implants
P a g e | 31
alveolar bone of the posterior region in 41 orthodontic subjects 38. The peak value of
IPT was measured using a torque screwdriver. The success rate of the mini-implant
anchor for 124 implants was 85.5%. The mean IPT ranged from 7.2 to 13.5Ncm,
depending on the location of the implants. There was a significant difference in the IPT
between maxilla and mandible. The IPT in the mandible was, unexpectedly,
significantly higher in the failure group than in the success group. Therefore, a large
IPT should not be used always. According to author’s calculations of the risk ratio for
failure, to raise the success rate of 1.6-mm diameter miniimplants, the recommended
IPT is within the range from 5 to 10Ncm.
----------------------------Ω-----------------------------
Mini Implants
P a g e | 32
Anatomical Considerations
Mini Implants
P a g e | 33
4 Anatomical Considerations
Miniscrews, mini implants, and miniplates are relatively easy clinical alternatives.
Orthodontic fixation screws can be placed either with or without flap raising. When
screws are placed without a flap, either drilling with a slow-speed handpiece or self-
tapping with a screwdriver (or a combination of them) can be used. Screws pass
through the soft tissue, and therefore the thickness of the soft tissue and cortical bone at
the surgical site are critical factors for success. Therefore, the use of endosseous
implants for absolute orthodontic anchorage has been the focus of many studies and
clinical trials. When the intraoral anchorage is stable, biocompatible, and free from site
specificity, it can be used effectively without patient compliance . Systems that can
satisfy these criteria include miniplates, and mini implants. These implants can be
placed in the inferior ridge of the pyriform aperture, maxillary alveolar bone, the
infrazygomatic crest, palatal alveolar bone, a maxillary tuberosity, the hard palate, and
the midpalatal suture area. To obtain sufficient screw fixation, understandings of the
bone density, screw shape and length as well as soft tissue and cortical bone thickness
are essential. This is probably responsible for the high prevalence of complications
such as hypersensitivity of the root, root fracture, and alveolar bone fracture resulting
from miniscrew insertion.
Here, the anatomical aspect on the maxilla and mandible, with the special reference
on the alveolar bone of the jaws will be introduced and described.
Mini Implants
P a g e | 34
4.1 Possible sites for placement of Miniscrew Implants in Maxilla –
Figure 9 Maxillary mini-implant locations. A. Below nasal spine. B. In the palate. C. Infrazygomatic
crest 39
Maxilla is the largest among the facial bones and forms the main part of the mid-
face region. Anatomically, the maxillary sinus is occupying as an air-filled space within
the maxilla. In this reason, the cortical plate surrounding the maxillary sinus is very
thin compared with the mandibular cortical plate. Furthermore, the facial aspect of the
maxilla above the level of the root apices is mainly composed of the cortical plate with
little spongy bone. The principal advantage of the miniscrewing on the buccal
interdental alveolar region is the ease in access. Even though the cortical plate is very
thin in this region, the initial anchorage can be provided in the adult patients. However,
care should be taken not to injure the dental roots and maxillary sinus when performing
the miniscrewing.
i. Area below the nasal spine the palate median or the paramedian area,
ii. Infrazygomatic crest,
iii. Maxillary tuberosities, and
iv. Alveolar process (both buccally and palatally) between the roots of the teeth.
Mini Implants
P a g e | 35
4.2 Possible sites for placement of Miniscrew Implants in Mandible –
The mandible is the largest, strongest and lowest bone in
the face. It has a horizontally curved body that is convex
forwards, and two broad rami, that ascend posteriorly. The
body of the mandible supports the mandibular teeth within
the alveolar process. The rami bear the coronoid and
condylar processes, and the condyle articulates with the
temporal bones at the temporomandibular
joints.
i. Symphysis or Parasymphysis,
ii. Alveolar process (between the roots of the teeth), and
iii. Retromolar area.
Deguchi et al quantitatively evaluated cortical bone thickness in various locations in
the maxilla and the mandible with the help of three-dimensional computed
tomographic images reconstructed for 10 patients 36. The distances from intercortical
bone surface to root surface, and distances between the roots of premolars and molars
were measured to determine the acceptable length and diameter of the miniscrew for
anchorage during orthodontic treatment.
They found that significantly less cortical bone thickness at the buccal region distal
to the second molar compared with other areas in the maxilla. Significantly more
cortical bone was observed on the lingual side of the second molar compared with the
buccal side. In the mandible, mesial and distal to the second molar, significantly more
cortical bone was observed compared with the maxilla. Furthermore, significantly
more cortical bone was observed at the anterior nasal spine level than at Point A in the
premaxillary region. Cortical bone thickness resulted in approximately 1.5 times as
Mini Implants
Figure 10 Mandibular mini-implant locations. A.
Retromolar area and molar region. B. Alveolar
process. C. Symphysis
P a g e | 36
much at 30° compared with 90°. Significantly more distance from the intercortical
bone surface to the root surface was observed at the lingual region than at the buccal
region mesial to the first molar. At the distal of the
first mandibular molar, significantly more distance
was observed compared to that in the mesial, and also
compared with both distal and mesial in the maxillary
first molar. There was significantly more distance in
root proximity in the mesial area than in distal area at
the first molar, and significantly more distance was
observed at the occlusal level than at the apical level.
They suggested that suggests that the best available location for a miniscrew is
mesial or distal to the first molar, and the best angulation is 30° from the long axis of
the tooth. From findings of the distance from the intercortical bone surface to the root
surface and the root proximity, the safest length is 6 mm with a diameter of 1.3 mm. In
addition, for lingual orthodontics, the recommended location is mesial to the first molar
at 30°, and 8 to 10 mm in length.
Maria Poggio40 provided an anatomical map to assist the clinician in miniscrew
placement in a safe location between dental roots using volumetric tomographic images
of 25 maxillae and 25 mandibles.
They suggest the order of the safer sites available in the interradicular spaces of the
posterior maxilla is as follows:
a. On the palatal side, the interradicular space between the maxillary first
molar and second premolar, from 2-8 mm from the alveolar crest.
b. On the palatal side, the interradicular space between the maxillary second
and first molars, from 2-5 mm from the alveolar crest.
Mini Implants
P a g e | 37
c. Both on buccal or palatal side between the second and first premolar,
between five and 11 mm from the alveolar crest.
d. Both on buccal or palatal side between the first premolar and canine,
between five and 11 mm from the alveolar crest.
e. On the buccal side, in the interradicular space between the first molar and
second premolar, from five to eight mm from the alveolar crest.
f. In the maxilla, the more anterior and the more apical, the safer the location
becomes.
g. The least amount of bone was in the tuberosity
The following is the order of the safer sites available in the interradicular spaces of
the posterior mandible:
a. Interradicular spaces between the second and first molar.
b. Interradicular spaces between the second and first premolar.
c. Interradicular spaces between the first molar and second premolar at 11
mm from alveolar crest.
d. Interradicular spaces between the first premolar and canine at 11 mm from
the alveolar crest.
e. The least amount of bone was between the first premolar and the canine.
They suggested that the features of the ideal titanium miniscrew for orthodontic
skeletal anchorage in the interradicular spaces should be 1.2- to 1.5-mm maximum
diameter, with 6–8 mm cutting thread and a conic shape.
Mini Implants
P a g e | 38
4.3 Safety Distance -
Huang suggested a method to evaluate the possibility of damaging the periodontal
ligament (PDL) is to calculate the safety distance41.
Safety distance: Diameter of the implant + PDL space (normal range 0.25 mm ±
50%) minimal distance between implant and tooth (1.5 mm) Example: Safety distance
(mm) of mini-implants when inserted between roots 1.2+(0.25 + 50%)+(1.5 +1.5) =
4.575. Therefore, the distance between roots needs to be at least 4.6 mm to reduce the
risk.
4.4 Safety Distance Modified
Gautam P and Valiathan A in 2006 42.
Safety distance = Diameter of the implant + 2 × [PDL space (normal range 0.25 mm
± 50%)] minimal distance between implant and tooth (1.5 mm)
----------------------------Ω----------------------------
Mini Implants
P a g e | 39
Biomechanical Considerations
in Mini Implant Use
Mini Implants
P a g e | 40
5 BIOMECHANICAL CONSIDERATIONS
The type of tooth movement that can be produced with Mini Implant anchorage is
determined by the same biomechanical principles and considerations that operate
during conventional orthodontic treatment, e.g. force, moment, center of resistance,
center of rotation. A Mini Implant can be placed in many different areas of the mouth
and at different heights on the gingiva relative to the occlusal plane, creating several
biomechanical orientations, e.g., low, medium and high. Thus, various types of tooth
movement can be produced depending on the position of the Mini Implant, the height
of the elastomer attachment, and the magnitude of the force applied. The following are
various clinical protocols that can be used routinely for effective tooth movement using
Mini Implant anchorage.
5.1 Maxillary anterior en masse retraction mechanics in extraction cases
For maxillary anterior en masse retraction, the line of action and the moment created
will vary depending on the location of the Mini Implant relative to the occlusal plane.
En masse retraction mechanics in extraction cases can be classified into three
categories much like the descriptors used traditionally for headgear traction: low,
medium, and high-pull mechanics.
Figure 11
Mini Implants
P a g e | 41
5.1.1 Medium—Pull Mechanics for the Maxillary Arch
Maxillary Mini Implants usually can be placed buccally between the second
premolar and first molar roots for anterior en masse retraction. When a maxillary Mini
Implant is placed about 8 to 10 mm above the main archwire, the term medium-pull en
masse retraction mechanics is used. If force is applied from a medium-pull Mini
Implant to a hook located between the lateral incisor and canine that extends 6 to 7 mm
vertically, the maxillary occlusal plane ordinarily can be maintained. Thus, medium-
pull mechanics are useful in treating patients who have normal overbite relationships.
5.1.2 Low-Pull Mechanics for the Maxillary Arch
When a Mini Implant is placed buccally between the roots of the maxillary second
premolar and first molar and is less than 8 mm away from the main archwire, the term
low-pull en masse retraction mechanics is used. If force is applied from a low-pull
Mini Implant to an anterior hook extending 6 to 7mm above the main archwire, the
maxillary Occlusal plane usually can be rotated in a clockwise direction. Therefore,
low-pull mechanics are useful in treating patients who have an open bite or an open-
bite tendency.
5.1.3 High-Pull Mechanics for the Maxillary Arch
When a Mini Implant is placed buccally between the maxillary second premolar and
the first molar roots, and is more than l0 mm away from the main archwire, the term
high-pull en masse retraction mechanics is used. If force is applied from a high-pull
Mini Implant to an anterior hook extending 6 to 7 mm above the main archwire, the
maxillary occlusal plane usually will rotate in a counterclockwise direction. Thus,
high-pull mechanics are useful in treating patients who have a deep bite or deep bite
tendency.
Mini Implants
P a g e | 42
5.2 Mandibular anterior en masse retraction mechanics in extraction cases
5.2.1 Medium-Pull Mechanics for the Mandibular Arch
Mandibular Mini Implants usually are placed buccally between the second premolar
and first molar roots for anterior en masse retraction. When a mandibular Mini Implant
is placed 6 to 8 mm away from the main archwire, the term medium-pull en masse
retraction mechanics is used. If force is directed from a medium-pull Mini Implant to a
hook located between the lateral incisor and canine that extends 4 to 6 mm below the
main archwire, the mandibular occlusal plane usually can be maintained. Therefore,
medium-pull mechanics are useful in treating patients who have normal overbite
relationships.
5.2.2 Low-Pull Mechanics for the Mandibular Arch
When a Mini Implant is placed buccally between the roots of the mandibular second
premolar and first molar and is less than 6 mm away from the main archwire, the term
low-pull en masse retraction mechanics is used. If force is applied from a mini implant
in a low-pull location to an anterior hook extending 4 to 6 mm below the main
archwire, a counterclockwise rotation of the mandibular occlusal plane typically can be
achieved. Low-pull mechanics are useful in treating patients who have an open bite or
open bite tendency.
5.2.3 High-Pull Mechanics for the Mandibular Arch
High-pull en masse retraction mechanics result when a Mini Implant is placed
buccally between the mandibular second premolar and first molar roots and more than
8 mm away from the main archwire. If force is applied from a high-pull Mini Implant
Mini Implants
P a g e | 43
to an anterior hook extending 4 to 6 mm below the main archwire, the mandibular
occlusal plane usually can be rotated in a clockwise direction. Therefore, high-pull
mechanics are useful in treating patients with a deep bite or deep bite tendency.
5.3 Anterior Intrusion Mechanics in the Maxillary Arch
For intrusion of the maxillary anterior teeth, Mini Implants can be placed between
the roots of the upper incisors. Force can be applied from the mini implant directly to
the main archwire. Usually a force originating from a single Mini Implant placed
between the maxillary central incisor roots is adequate to intrude the anterior dentition.
However, if there is a transverse cant to the occlusal plane. Two mini implants can be
placed bilaterally between the central and lateral incisor roots. Forces of differing
magnitudes then can be applied on each side for improvement of the canted occlusal
plane during intrusion.
5.4 Anterior Intrusion Mechanics in the Mandibular Arch
For intrusion of the mandibular anterior teeth, Mini Implants can be placed between
the roots of the lower incisors. Again, force can be applied from the Mini Implant
directly to the main archwire. One mini implant placed between the lower central
incisor roots usually is sufficient to allow for intrusion of the entire mandibular anterior
segment. However, if the occlusal plane is canted transversely, two mini implants can
be inserted between the central and lateral incisor roots bilaterally. Differential forces
then can be applied for improvement of the canted occlusal plane during intrusion.
Mini Implants
P a g e | 44
5.5 Anterior en masse Retraction with Anterior Intrusion
In deep bite extraction cases, high-pull mechanics are recommended in the maxillary
arch for intrusion of the anterior teeth during en masse retraction. In reality, it is
difficult to place Mini Implants higher in the buccal vestibule. In addition, high-pull
mechanics do not produce much of a horizontal force compared to low-pull or
medium-pull mechanics. Thus, we recommend using two posterior Mini-Implants in
low or medium-pull orientation combined with one or two anterior Mini Implants.
Posterior Mini Implants usually will be more effective in retracting the anterior teeth,
whereas anterior Mini Implants will be more effective in their intrusion. Furthermore,
anterior intrusion Mini Implants will counteract the tendency for incisors to tip
lingually during their retraction
5.6 Molar Intrusion Mechanics for Open bite Cases
‘In that it is possible to intrude molar teeth using Mini Implants, open bites can be
corrected relatively easily, especially skeletal open bites. If 1mm of absolute molar
intrusion is achieved posteriorly, an anterior open bite of 2 to 3 mm will be closed
anteriorly. A Mini Implant can be placed between the roots of the maxillary second
premolar and first molar and/or the first molar and second molar buccally and/or
palatally, for the intrusion of maxillary molar teeth. A transpalatal arch (TPA) is used
for palatal support in the absence of palatally-placed Mini Implants. In the mandibular
arch, however, it is not advisable to insert mini implants lingual to the molar roots: a
lingual holding arch can be used for support instead.
5.7 Maxillary Anterior Lingual Root Torque Mechanics
After anterior en masse retraction in an extraction case, severe lingual tipping of the
maxillary anterior teeth sometimes is observed. Whenever lingual root torque then is
applied, labial crown tipping usually is observed instead. To prevent this kind of labial
crown tipping, Class II elastics are required. Moreover, to prevent the side effects of
Class II elastics, up-and-down vertical elastics and high-pull headgear are used. Mini
Mini Implants
P a g e | 45
Implants in the maxillary buccal area, however, also can prevent labial crown tipping
during lingual root torque application; ligature wires are connected from the Mini
Implants to the anterior portion of the main archwire.
5.8 Molar Distalization Mechanics for Non-Extraction Cases
To correct Class II or Class III molar relationships, sometimes it is necessary to
distalize molar.
Mini Implants can be placed between the roots of the second premolar and first
molar, and nickel titanium coil springs can be used. After molar distalization, the
anterior teeth will need to be retracted. The first Mini Implant can be removed if it
interferes with this
retraction, & a second mini
implant is placed just distal
to the first one or between
the first molar & second molar roots
5.8.1 Retraction of the Entire Maxillary or Mandibular Dentition
Two buccally placed mini implants can provide
sufficient anchorage to move the entire maxillary or
mandibular dentition posteriorly. Usually mini implants
are inserted between the roots of second premolar and
first molar.
Retraction of the entire dentition is more positive in
patients who have mesially tipped posterior teeth. Thus, Mini Implants function nicely
in combination with the multiloop edgewise archwire (MEAW) technique of Kim
(Kim, 1999a,b, 141_.; Chang and Moon, 1999) for retraction of the entire dentition.
Mini Implants
Figure 12
P a g e | 46
5.8.2 Midpalatal Mini Implant Placement For Molar Distalization
The midpalatal area also is a good site for mini implant placement because the
palate is covered with relatively thin keratinized mucosa and has adequate bone
volume. Additionally, there is no concern about a Mini Implant touching the roots of
adjacent teeth during implantation. However, if the patient has a broad unossified
suture, sufficient mechanical stability is not possible. In this situation it is better to
place the Mini Implants parasagittally. For molar distalization, orthodontic forces can
be applied from a Mini Implant to the center point of a transpalatal arch. Furthermore,
a mid-palatal mini implant can be used to anchor orthodontic force, which is applied
from a level high above the center of rotation of the molars. Thus, distalization traction
or distal tipping of roots can be achieved rather easily with midpalatal implants than
with buccally placed Mini implants.
If a bracket head type of Mini Implant is placed in the midpalatal area, a transpalatal
arch can be inserted directly into the bracket slots. Force can be applied directly to the
teeth from the transpalatal arch, much in the same manner as with a pendulum
appliance (Hilgem 1992;)_ The incorporation of a midpalatal mini implant into a
transpalatal arch can be technically challenging, however, because of the level of
precision required. In addition the application of elastomers for tooth movement can be
difficult especially in patients who have high and narrow palates
5.9 Protraction Mechanics in Extraction Cases
Sometimes molar protraction is needed in minimum or moderate anchorage cases or
unusual extraction cases. However, molar protraction is one of the most difficult tooth
movement to accomplish, especially in patients with a low mandibular plane, Class 2 &
a deep bite. If mini implants are incorporated into the treatment protocol, molars can be
moved forward more effectively and without disturbing the anterior teeth. Mini
implants, for this purpose can be placed between the roots of the mandibular canine
and first premolar or first premolar and 2nd premolar.
Mini Implants
P a g e | 47
5.10 Minor Tooth Movements Using Mini Implant Anchorage
5.10.1 Retromolar Mini Implants for Single Molar Uprighting -
A single retromolar Mini Implant is useful in uprighting a
mesially tipped molar. Elastic chain or ligature wire can be
connected from a retromolar Mini Implant to an attachment on
the tipped molar (Fig.5-31). These mechanics produce an
intrusive force during molar uprighting and prevent the
occlusal trauma that normally would occur with
conventional uprighting techniques. However, this type of
simple retromolar Mini Implant mechanics cannot control
the movement of a tooth precisely.
5.10.2 Molar Distalizatien Using a Single Mini Implant in an Edentulous Area
For distalization of a molar tooth (or teeth) that is (are)
adjacent to an edentulous area, an open coil spring can be
used on the main archwire in the edentulous region. To
prevent tipping, the anterior teeth labially, a Mini Implant
can be placed in the edentulous area and connected by a
ligature wire to a sliding hook on the main archwire as
shown in Figure 14. The open coil NiTi spring is
compressed by tying the ligature to the sliding hook and
moving the hook posteriorly along the archwire.
5.10.3 Molar Uprighting and/or Protraction or
Distalization Mechanics Using Two Mini Implants in an Edentulous Area
Mini Implants
Figure 14
Figure 13
P a g e | 48
A single Mini implant cannot resist rotational or torquing forces. To resist rotational
and torquing forces, two Mini Implants can be placed
side-by-side in an edentulous area and then joined
together using light-cured resin. Subsequently, a
bracket can be bonded to the resin of the Mini Implant-
supported structure. A rectangular wire inserted into
this bracket will facilitate three-dimensional
movement of the involved tooth (Figure 12)
5.11 Buccal Cross bite (Scissors Bite) Correction
Correcting a scissors bite with conventional orthodontic mechanics requires the use
of through-the bite elastics (Graber, I972; Moyers, 1988;) However, if these elastics are
used, undesirable extrusion of the posterior teeth may occur. To correct a scissors bite
without causing molar extrusion, intra-arch mechanics, rather than interarch
mechanics, must be used. Transpalatal and lingual arches can be used to reinforce
anchorage in conventional intra-arch methods. If mini implants are used, the same type
of uprighting and intrusion is observed during buccal
cross bite correction. With only one Mini Implant,
however, it is difficult to apply orthodontic force in
the proper direction; if a bracket head type of Mini
Implant is selected, a wire can be extended from the
slot of the bracket to allow the force to be applied
more effectively. Different handed screws, i.e., right and left-handed, are inserted
depending upon the moment and force to be applied
Mini Implants
Figure 15
Figure 16
P a g e | 49
Indications & Contraindications
“That, which does not kill us, makes us stronger.”
– Freidrich Nietzsche
Mini Implants
P a g e | 50
6 Indications & Contraindications
Indications, Contraindication & Relative Contraindications
Absolute anchorage represents new orthodontic paradigm and it is maybe the most
important advancement in recent times, because it offers the orthodontics of " action
without reaction", practically eliminating the third principle of Newton. Indications for
miniimplants can be for achieving the absolute anchorage or forces directions which
would be very difficult to achieve using the traditional mechanics.
There are two basic forms of absolute anchorage: -
Direct anchorage:
When active segment is pulled directly from min implant
.
Indirect anchorage:
When active segment is pulled from the reactive segment and this segment is
fixed to miniimplant to increase anchorage.
6.1 INDICATIONS
The most frequent indications are:
1. Molar intrusion.
2. Molar uprighting by crown distalizing or by root mesializing.
3. Anterior open bite treatment with molar intrusion (with or without
extractions).
Mini Implants
P a g e | 51
4. Anterior deep bite treatment with incisal intrusion (with or without
extractions) .
5. Leveling of transverse tipping of occlusal plane.
6. Extraction cases.
7. Distalizing or anchorage after distal movement with other kinds of
appliances, such as Pendulum.
8. Forced eruption of included or non-included teeth.
9. Asymmetric expansion.
10. Bodily movement of teeth or a group of teeth.
11. As surgical fixation with lingual brackets.
12. Absolute anchorage in lingual orthodontics.
13. They can be used in a growing patient.
14. Edentulous spaces closure.
6.2 CONTRAINDICATIONS
1. Systemic diseases such as diabetes, osteoporosis, osteomyelitis, blood
dyscrasias, metabolism disorders,etc.
2. Patient undergoing the radiotherapy in arches.
3. Psychological disorders.
Mini Implants
P a g e | 52
4. Presence of active oral infections.
5. Uncontrolled periodontal disease.
6. Presence of pathological formations in the zone, such as tumors or cysts.
7. Insufficient space for insertion of miniimplant.
8. Thin cortical bone and insufficient retention.
9. Deficient quality of the bone.
10. Soft tissue lesions, such as lichen planus, leucoplakia, etc.
11. Patient who does not accept miniimplant treatment.
6.3 RELATIVE CONTRAINDICATIONS
1. Tobacco, alcohol and drugs abuse.
2. Mouth breather.
3. Absence of 'ability to maintain the correct oral hygiene’.
----------------------------Ω----------------------------
Mini Implants
P a g e | 53
Clinical Applications
“That, which does not kill us, makes us stronger.”
– Freidrich Nietzsche
Mini Implants
P a g e | 54
7 CLINICAL APPLICATIONS
Guire reported potential uses of TADs for Orthodontic Purposes 43:
Three Dimensions and Three Tissue Considerations (3D/3T).
3D/3T Anterior-Posterior Vertical Transverse
Skeletal Possibly as anchors to
prevent unwanted dental
movement during
conventional orthopedic
corrections such as Herbst.
Possibly eliminate
compensatory
eruption of teeth occurring as
a result of natural growth to
yield a more anterior rather
than vertical growth.
Intrusion of upper
and/or lower facial
height in cases of excess
vertical growth through
counterclockwise
rotation of the mandible.
Possibly intrusion of
entire upper and/or
lower dental arch to
eliminate excess alveolar
display ("gummy
smile")
in cases of maxillary
alveolar hyperplasia.
Possibly true
orthopedic
maxillary
expansion, without
the undesirable
tipping of posterior
teeth that occurs in
traditional
expansion.
Mini Implants
P a g e | 55
Soft tissue Close spaces completely
from the posterior to maintain
the incisor position for
optimal lip support.
Close spaces completely
from the anterior to reduce
excessive lip protrusion.
In selected cases,
eliminate lip
incompetency through
decreasing lower-face
height.
Dental Close spaces completely
from anterior or posterior
congenitally missing teeth,
thus eliminating the need for
bridge or implant.
Close spaces in cases of
previously extracted or lost
teeth, thus possibly
eliminating a bridge or
implant.
Retract maxillary and/or
mandibular anterior segments
completely without unwanted
anterior molar movement,
i.e., loss of posterior
anchorage.
Uprighting tipped molars
without extruding.
Intrusion of over-
erupted Upper and/or
lower Anterior teeth in
cases of Deep bite.
Intrusion of
overerupted single or
multiple posterior teeth.
Extrusion of impacted
teeth without unwanted
reciprocal effects on
anchor teeth
True unilateral
movement of
buccal segments to
eliminate true
unilateral cross-
bites
Mini Implants
P a g e | 56
Maintain torque control of
incisors during retraction by
directing forces through the
center of resistance.
En masse movement of
arches mesially or distally to
correct Class II or III molar
and canine relations.
Asymmetric correction of
Class II or III dental relation.
Correction of a canted
occlusal plane.
3D/3T Anterior-Posterior Vertical Transverse
Other factors:
1. Oral health May decrease orthodontic treatment time, thereby
minimizing deleterious effect of orthodontic appliances
on oral hygiene procedures.
2. Perimeter In cases requiring the extraction of permanent teeth,
the orthodontist has the ability to choose a malformed,
previously restored, or otherwise compromised tooth
Mini Implants
P a g e | 57
rather than a virgin tooth that might normally be needed
in planning for anchorage requirements.
In Class I cases having a non-extraction lower arch
and congenitally missing one or two maxillary lateral
incisors, there is the possibility of unilateral or bilateral
space closure from the posterior, thus giving patient the
option of substituting a canine for the lateral incisor
instead of a bridge or implant.
Ability to distalize molars through translation (i.e.,
bodily movement) rather than tipping
3. Interactions Eliminate the need for patient compliance in headgear
or elastic wear thus reducing treatment time.
----------------------------Ω-------------------------------
Mini Implants
P a g e | 58
Biological aspects of
Orthodontic Mini-implants
Mini Implants
P a g e | 59
8 Biology of orthodontic implants
Biologic aspects of orthodontic implantation, the healing process, and the formation
of the tissue-implant interface.
8.1 Bone-to-Implant interface:
Anabolic modeling on bone surfaces is the first osseous healing reaction following
implantation of a biocompatible device into cortical bone. Similar to fracture healing, a
bridging callus forms at the periosteal and endosteal surfaces. Under optimal
conditions (minimal trauma and vascular compromise) the callus originates within a
few millimeters from the margin of the implantation site. In rabbits, the lattice of
woven bone reaches the implant surface in about 2 weeks and is sufficiently compacted
and remodeled by 6 weeks to provide adequate resistance to loading. There is no
quantitative data for the early healing process in humans. Extrapolating from relative
durations of the remodeling cycles (6 weeks vs. 4 months), timing for the primary
callus (woven bone) may be similar to rabbits but the remodeling-dependent
maturation process probably requires 3 times longer (up to 18 weeks).
If periosteum is stripped, the callus must originate in the nearest untraumatized
osteogenic tissue. Since healing reactions are self limiting, extensive loss of the
osteogenic (inner) layer may preclude periosteal bridging altogether. Reapproximating
retracted periosteum when the wound is closed positions the nonosteogenic fibrous
(outer) layer near the implant. A compromised osteogenic reaction, associated with a
defect in the periosteal margin of bone, may favor invasion of fibrous connective
tissue. Extensive stripping of periosteum substantially inhibits the initial healing re-
sponse. Even though stimulating cytokines and growth factors are released from the
blood clot at the surgical site, essentially no competent osteoprogenitor cells survive
periosteal stripping. These cells must be reintroduced by ingrowth of new vascular
Mini Implants
P a g e | 60
tissue. Therefore the surgeon should minimize periosteal trauma consistent with
adequate access and appropriate soft tissue management.
Efficient reduction of an osseous defect by a bridging callus requires relative
stability of the approximating segments. An unloaded healing phase (two-stage implan-
tation procedure) is widely used to prevent extensive functional movement during
healing. However, there are other important biomechanical considerations:
Mechanical retention of the implant within the wound
Approximation of the periosteal margin of the cortex to the implant surface and
Functional flexure of the implanted bone.
Healing implants in functioning bones are never really "unloaded." The initial callus
reaction near the implant is primarily driven by local cytokines and growth factors;
however, the overall size and extent of the periosteal callus is mechanically dependent.
A surgical defect weakens the bone and, as a result, may increase peak strains at a
distance from the surgical site. Focal areas of new bone formation (additional regions
of woven or lamellar bone) are often noted around the bone. Remodeling of the callus
begins early in the healing period. According to the principle of adequate strength with
minimal mass, the callus reduces in size and reorients as internal maturation and
strength are attained. 18
Interface remodeling is essential in establishing a viable interface between the
implant and original bone. About a millimeter of compacta adjacent to the osseous
wound dies postoperatively despite optimal surgical technique. This is probably
because of inflammation and the relatively poor collateral circulation within cortical
bone. Dead bone is not useless tissue; it provides important structural support during
the initial healing phase. However, it must be replaced with vital bone (via remodeling)
to strengthen the interface and provide adaptable tissue for long-term maintenance.
Mini Implants
P a g e | 61
Remodeling of the nonvital interface is achieved by cutting/filling cones emanating
from the endosteal surface. The mechanism is similar to typical cortical remodeling
except that many of the cutting/filling cones are oriented perpendicular to the usual
pathway (long axis of the bone). In longitudinal sections, cutting/filling cones
occasionally deviate from the plane of the interface, turn 90 degrees, and form a
secondary osteon perpendicular to the interface. At the/same time the interface is
remodeled, the adjacent nonvital cortex (viewed in cross-section) is penetrated by
typical cutting/filling cones.
Maturation of the interface and supporting bone has been suggested to require an
elapsed time after implant placement of about 3 sigma (12 months) 44. The first 4
months (1 sigma) is the initial "unloaded" healing process. During maturation the
callus volume is decreased and interface remodeling continues. The bone maturation
phase requires an additional 2 sigma (8 months).
It was previously believed that maturation involved two physiologic transients:
The regional acceleratory phenomenon (RAP) and
Secondary mineralization of newly formed bone.
Extensive remodeling (RAP) in cortical bone is a well-known healing reaction to
surgical wounds and is certainly evident in bone surrounding implants. In general, the
remodeling sites decrease with increasing distance from the wound 27. It is also well
accepted that stiffness and strength of lamellar bone are directly related to mineral
content. Because of this, it was previously suggested that full strength of bone
supporting an implant would not be achieved until about 12 months (54 weeks) after
the bone is formed, that is, after completion of the secondary mineralization process.
However, recent evidence has shown that elevated remodeling is an ongoing response
of bone adjacent (<1 mm) to an implant 45. Because of rapid remodeling at the
interface, it is likely that the secondary mineralization process is not completed and the
Mini Implants
P a g e | 62
mineral content of bone supporting an implant remains lower than that of the
surrounding bone.
Long-term maintenance of the rigid osseous fixation involves continuous
remodeling of the interface and supporting bone. Bone, like other relatively rigid
materials, is subject to fatigue. Repetitive loading results in microscopic cracks
(microdamage). If allowed to accumulate, these small defects can lead to structural
failure. Because osteoclasts preferentially resorb more highly mineralized tissue,
cutting cones tend to remodel the oldest and presumably most weakened bone. This
physiologic mechanism helps to maintain structural integrity indefinitely. Human
cortical bone from long-bone midshaft diaphysis and rib remodels at a rate of about 2%
to 10% per year 27. No human data is available for mandibular cortical bone. However,
based on recent studies in dogs (which have diaphyseal and rib remodeling rates
similar to humans), cortical bone supporting the teeth may have a substantially higher
remodeling rate (30% to 40% per year). The interaction of mechanical and metabolic
factors in controlling adult bone remodeling is not well understood. Cortical bone
around an endosseous implant continues to remodel. Long-term maintenance of rigid
osseous fixation involves a remodeling rate of as much as approximately 500% per
year in bone immediately adjacent (within 1 mm) to an implant 27. Although the precise
reason for the sustained remodeling response is unclear, its physiologic control appears
strongly related to the mechanical stress distribution concentrated in bone adjacent to
the interface. Data from four animal species (including humans) indicate that the
elevated remodeling response is a universal mechanism necessary for the long-term
retention of rigidly integrated implants.
8.2 Support and load transfer /mechanisms:
The three basic means of retention of an endosteal dental implant in function are
Mini Implants
P a g e | 63
i. Fibroosseous retention,
ii. Biointegration and
iii. Osseointegration.
According to the American Academy of Implant Dentistry (AAID) Glossary of
Terms (1986) 46
Fibrosseous retention is defined as tissue-to-implant contact; interposition of
healthy, dense collagenous tissue between the implant and bone.
Osseointegration is defined as contact established between normal and remolded
bone and an implant surface without the interposition of non-bone or connective tissue.
Bioactive retention is achieved with bioactive materials such as hydroxyapatite
(HA), which bond directly to bone, similar to ankylosis of natural teeth. Bone matrix is
deposited on the HA layer as a result of some type of physiochemical interaction
between the collagen of bone and HA crystals of implant.
To a certain extent, it appears to be within the hands of the clinician to control which
method of load bearing will he established by the host. Relative movement during
healing has been shown to be the determining factor 47. If the implant is loaded so that
the implant moves relative to its surrounding bone during the healing phase, then a
tendon or ligament will occur around a plate or subperiosteal implant. If the implant is
allowed to heal without relative movement, then ankylosis can be expected to occur.
Mini Implants
P a g e | 64
Complications with Mini
Implants & their
management
Mini Implants
P a g e | 65
9 Complications with use of Mini Implants & their management
Risks, Complications of Orthodontic miniscrews and Measures to prevent it.
Complications can arise during miniscrew placement and after orthodontic loading
that affect stability and patient safety 48 .
COMPLICATIONS DURING INSERTION
i. Trauma to the periodontal ligament or the dental root.
ii. Miniscrew slippage.
iii. Nerve involvement.
iv. Air subcutaneous emphysema.
v. Nasal and maxillary sinus perforation.
vi. Miniscrew bending, fracture, and torsional stress.
COMPLICATIONS UNDER ORTHODONTIC LOADING
i. Stationary anchorage failure.
ii. Miniscrew migration.
SOFT-TISSUE COMPLICATIONS
i. Aphthous ulceration.
ii. Soft-tissue coverage of the miniscrew head and auxiliary.
iii. Soft tissue inflammation, infection, and peri-implantitis.
Mini Implants
P a g e | 66
COMPLICATIONS DURING REMOVAL
i. Miniscrew fracture.
ii. Partial osseointegration.
9.1 COMPLICATIONS DURING INSERTION
9.1.1 Trauma to the periodontal ligament or the dental root.
Interradicular placement of orthodontic miniscrews risks trauma to the periodontal
ligament or the dental root. Potential complications of root injury include loss of tooth
vitality, osteosclerosis, and dentoalveolar ankylosis. Trauma to the outer dental root
without pulpal involvement will most likely not influence the tooth's prognosis 49.
Dental roots damaged by orthodontic miniscrews have demonstrated complete repair of
tooth and periodontium in 12 to 18 weeks after removal of the miniscrew. 50
Interradicular placement requires proper radiographic planning, including surgical
guide with panoramic and periapical radiographs to determine the safest site for
miniscrew placement.
During inter-radicular placement in the posterior region, there is a tendency for the
clinician to change the angle of insertion by inadvertently pulling the hand-driver
toward their body, increasing the risk of root contact. 48 To avoid this, the clinician may
consider using a finger-wrench or work the hand-driver slightly away from their body
with each turn. If the miniscrew begins to approximate the periodontal ligament, the
patient will experience increased sensation under topical anesthesia. If root contact
occurs, the miniscrew may either stop or begin to require greater insertion strength 51. If
trauma is suspected, the clinician should unscrew the miniscrew 2 or 3 turns and
evaluate it radiographically.
Mini Implants
P a g e | 67
9.1.2 Miniscrew slippage
The clinician might fail to fully engage cortical bone during placement and
inadvertently slide the miniscrew under the mucosal tissue along the periosteum. High-
risk regions for miniscrew slippage include sloped bony planes in alveolar mucosa
such as the zygomatic buttress, the retromolar pad, the buccal cortical shelf, and the
maxillary buccal exostosis if present. Slippage in the retromolar pad can lead to the
greatest risk of iatrogenic harm if the miniscrew moves lingually in the submandibular
or lateral pharyngeal space near the lingual and inferior alveolar branch nerves. In the
retromolar region, serious consideration should be given to flap exposure for direct
visualization and a predrilled pilot hole, even for self-drilling miniscrews. If the al-
veolar tissue is thin and taut, some clinicians advocate placing the pilot hole with a
transmucosal method, using a slow-speed bur to perforate both tissue and cortical bone
without making a flap 52.
Miniscrew slippage can occur in dentoalveolar regions of attached gingiva if the
angle of insertion is too steep. Placement of miniscrews less than 30° from the occlusal
plane, typically to avoid root contact in the maxilla or to gain cortical anchorage in the
mandible, can increase the risk of slippage. To avoid this, the clinician can initially
engage bone with the miniscrew at a more obtuse angle before reducing the angle of
insertion after the second or third turn. Miniscrews should engage cortical bone after 1
or 2 turns with the hand-driver. Only minimal force should be used with the hand-
driver, regardless of bone density. Greater forces increase the risk of miniscrew
slippage.
9.1.3 Nerve involvement:
Nerve injury can occur during placement of miniscrews in the maxillary palatal
slope, the mandibular buccal dentoalveolus, and the retromolar region. Most minor
nerve injuries not involving complete tears are transient, with full correction in 6
Mini Implants
P a g e | 68
months. Long-standing sensory aberrations might require pharmacotherapy
(corticosteroids), microneurosurgery, grafting, or laser therapy 48.
Placement of miniscrews in the maxillary palatal slope risks injury to the greater
palatine nerve exiting the greater palatine foramen. The greater palatine foramen is
located laterally to the third molar or between the second and third molars. Location,
size, and shape of the foramen can vary with ethnicity. 53 The greater palatine nerve
exits the foramen and runs anteriorly, 5 to 15 mm from the gingival border, to the
incisive foramen. Miniscrews inserted in the palatal slope should be placed medial to
the nerve and mesial to the second molar. Placement of the miniscrew above the nerve
could increase the risk of palatal root contact and reduce biomechanical control.
Placement of the miniscrews in the mandibular buccal dentoalveolus risks injury to
the inferior alveolar nerve in the mandibular canal. The mandibular canal travels
forward in an S-shaped curve moving from buccal to lingual to buccal. The inferior
alveolar nerve occupies its most buccal position within the body of the mandible at the
distal root of the second molar and the apex of the second premolar, before exiting
from the mental foramen. Miniscrews inserted near the mandibular second molar and
the second premolars are at greatest risk for accidental damage to the inferior alveolar
nerve. 54 The soft-tissue appearance of the dentoalveolus can be deceptive, and a
panoramic radiograph should be taken to determine the vertical position of mandibular
canal and the location of the mental foramina. Greater caution is needed in adult
patients who might have a more occlusal position of the mandibular canal due to
resorption of the alveolar ridge.
Placement of miniscrews in the retromolar pad risks injury to the long buccal nerve
and the lingual nerve. The long buccal nerve branches off the mandibular nerve trunk
and crosses high on the retromolar pad supplying the mucosa of the cheek. The lingual
nerve runs immediately under the floor of the mouth and supplies general sensory
innervation to the anterior two thirds of the tongue. To avoid nerve involvement and
Mini Implants
P a g e | 69
slippage, the retromolar miniscrews should be no longer than 8 mm and placed in the
buccal retromolar region below the anterior ramus is been recommended.
9.1.4 Air subcutaneous emphysema:
Air subcutaneous emphysema is the condition in which air penetrates the skin or
submucosa, resulting in soft-tissue distention. Subcutaneous emphysema can occur
during routine operative dental procedures if air from the high-speed or air-water
syringe travels under the gingival tissues. The main symptom of air subcutaneous
emphysema is immediate mucosal swelling with or without crepitus (crackling).
Additional sequelae include cervicofacial swelling, orbital swelling, otalgia, hearing
loss, mild discomfort, airway obstruction, and possibly interseptal and interproximal
alveolar necrosis. Clinically visible swelling of the skin and mucosa occurs within
seconds to minutes after air has penetrated the submucosal space and typically spreads
to the neck (in 95% of cases) or the orbital area (in 45% of cases).
The clinician should be alert for subcutaneous emphysema during miniscrew
placement through the loose alveolar tissue of the retromolar, mandibular posterior
buccal, and the maxillary zygomatic regions. If a purchase point or pilot hole is to be
drilled through the mucosa, the clinician should use slow speed under low rotary
pressure. If either a pilot hole or a mucosal punch is placed, an air-water syringe should
never be used. Air from the syringe can enter the submucosal space through the small
tissue opening, even in attached tissue. Bleeding and saliva should be controlled with
suction, cotton, and gauze, rather than an air-water syringe. 48 In case of subcutaneous
emphysema, the clinician should immediately discontinue the procedure and take
periapical and panoramic radiographs to determine the extent of the condition. The
patient should not be dismissed until the swelling begins to regress and an infection can
be ruled out. Upon dismissal, the patient should be instructed to apply light pressure
with an ice pack for the first 24 hours. The clinician could prescribe a mild analgesic,
an antibacterial rinse, such as chlorhexidine, and an antibiotic prophylaxis for a week.
Mini Implants
P a g e | 70
In most cases of subcutaneous emphysema, careful observation for further problems or
infection is adequate, and swelling and symptoms generally subside in 3 to 10 days.
9.1.5 Nasal and maxillary sinus perforation:
Perforation of the nasal sinus and the maxillary sinuses can occur during miniscrew
placement in the maxillary incisal, maxillary posterior dentoalveolar, and zygomatic
regions. A posterior atrophic maxilla is a major risk factor for sinus perforation. The
sinus floor is deepest in the first molar region and can extend to fill a large part of the
alveolar process in posterior edentulous spaces. Penetration of the Schneiderian
membrane is a well-documented phenomenon that often occurs when the thin, lateral
wall of the sinus is infractured from the buccal side. Small (<2 mm) perforations of the
maxillary sinus heal by themselves without complications. 55 Ardekian et al and
Branemark et al 56 reported that immediately loaded dental implants that perforated the
nasal and maxillary sinuses showed no differences in implant stability.
If the maxillary sinus has been perforated, the small diameter of the miniscrew does
not warrant its immediate removal. Orthodontic therapy should continue, and the
patient should be monitored for potential development of sinusitis and mucocele. For
miniscrews placed in pneumatized, edentulous regions of the maxilla, or placed higher
in the posterior maxilla when intrusive forces are desired, the clinician should consider
angulating the miniscrew perpendicular to the alveolar ridge to avoid damage to the
sinus.
9.1.6 Miniscrew bending, fracture, and torsional stress:
Increased torsional stress during placement can lead to implant bending or fracture,
or produce small cracks in the peri-implant bone, that affect miniscrew stability. 39 Self-
drilling miniscrews should be inserted slowly, with minimal pressure, to assure
maximum miniscrew-bone contact. A purchase point or a pilot hole is recommended in
regions of dense cortical bone, even for self-drilling miniscrews. During miniscrew
placement in dense cortical bone, the clinician should consider periodically derotating
Mini Implants
P a g e | 71
the miniscrew 1 or 2 turns to reduce the stresses on the miniscrew and the bone. The
clinician should stop inserting the miniscrew as soon as the smooth neck of its shaft has
reached the periosteum. Over insertion can add torsional stress to the miniscrew neck,
leading to screw loosening and soft-tissue overgrowth. Once the miniscrew has been
inserted, torsional stress from wiggling the hand-driver off the miniscrew head can
weaken stability. When removing the hand-driver from the miniscrew head, the
clinician should gently separate the hand-driver handle from its shaft and then gently
remove the shaft from the miniscrew head.
9.2 COMPLICATIONS UNDER ORTHODONTIC LOADING
9.2.1 Stationary anchorage failure:
According to the literature, the rates of stationary anchorage failure of miniscrews
under orthodontic loading vary between 11% and 30%. If a miniscrew loosens, it will
not regain stability and will probably need to be removed and replaced. 57 Stability of
the orthodontic miniscrew throughout treatment depends on bone density, peri-implant
soft tissues, miniscrew design, surgical technique, and force load. 58
The key determinant for stationary anchorage is bone density. 59 Stationary
anchorage failure is often a result of low bone density due to inadequate cortical
thickness 49. Bone density is classified into 4 groups (Dl, D2, D3, and D4) based on
Hounsfield units (HU)—an x-ray attenuation unit used in computed tomography scan
interpretation to characterize the density of a substance.132
a) D1 (>1250 HU) is dense cortical bone primarily found in the anterior mandible
and the maxillary midpalatal area.
b) D2 (850-1250 HU) is thick (2 mm), porous cortical bone with coarse trabeculae
primarily found in the anterior maxilla and the posterior mandible.
Mini Implants
P a g e | 72
c) D3 (350-850 HU) is thin (1 mm), porous cortical bone with fine trabeculae
primarily found in the posterior maxilla with some in the posterior mandible.
d) D4 (150-350 HU) is fine trabecular bone primarily found in the posterior
maxilla and the tuberosity region.
Figure 17 Areas according to Bone density
Sevimay et al 60 reported that osseointegrated dental implants placed in Dl and D2
bone showed lower stresses at the implant-bone interface. D1-D3 bone is optimal for
self-drilling miniscrews. Placement of miniscrews in Dl and D2 bone might provide
greater stationary anchorage under orthodontic loading. Placement of miniscrews in D4
bone is not recommended due to the reported high failure rate. In general, stationary
Mini Implants
P a g e | 73
anchorage failure is greater in the maxilla, with the exception of the midpalatal region,
due to the greater trabeculae and lower bone density. 61 Loss of mid-palatal miniscrews
is likely a result of tongue pressure.
Peri-implant soft-tissue type, health, and thickness can affect stationary anchorage of
the miniscrew. Miniscrews placed in nonkeratinized alveolar tissues have greater
failure rates than those in attached tissues. 62 The movable, nonkeratinized alveolar
mucosa is easily irritated; soft-tissue inflammation around the miniscrew is directly
associated with increased mobility 34. Additionally, miniscrews placed in regions of
thick keratinized tissue, such as the palatal slope, are less likely to obtain adequate
bony stability. Thin, keratinized tissue, seen in the dentoalveolar or midpalatal region,
is ideal for miniscrew placement.
Miniscrew geometry and surgical technique directly influence the stress distribution
of peri-implant bone. Most miniscrew losses occur as a result of excessive stress at the
screw-bone interface. Self-drilling miniscrews can have greater screw-bone contacts
(mechanical grip) and holding strengths compared with self-tapping screws. 33
Heidemann et al 63 reported greater residual bone between screw threads of self-drilling
miniscrews compared with self-tapping miniscrews. Self-tapping miniscrews, like self-
drilling screws, can be placed without a predrilled pilot hole in the dentoalveolar region
if the cortical bone is thin. If a pilot hole is to be used, for either self-drill ing or self-
tapping miniscrews, the pilot hole size should be no greater than 85% of the diameter
of the miniscrew shaft for optimal stability.
It is still not clear the maximum force-load a miniscrew can withstand in regard to
stationary anchorage. Dalstra et al 35reported that miniscrews inserted into thin cortical
bone and fine trabeculae should be limited to 50 g of immediate loaded force. Buchter
et al 58 reported that miniscrews placed in dense mandibular bone remained clinically
stable with up to 900 g of force. Many articles reported miniscrew stability with
loading forces of 300 g or less. In regions of poor bone density, simply placing a longer
miniscrew under smaller orthodontic force does not ensure stationary anchorage. 64
Mini Implants
P a g e | 74
9.2.2 Miniscrew migration:
Orthodontic miniscrews can remain clinically stable but not absolutely stationary
under orthodontic loading.64 Unlike an endosseous dental implant, that osseointegrates,
orthodontic miniscrews achieve stability primarily through mechanical retention, and
can be displaced within the bone. Liou et al reported that orthodontic miniscrews
loaded with 400 g of force for 9 months extruded and tipped -1.0 to 1.5 mm in 7 of 16
patients. To account for potential migration, the clinician should allow a 2-mm safety
clearance between the miniscrew and any anatomical structures.
9.3 SOFT-TISSUE COMPLICATIONS
9.3.1 Aphthous ulceration:
Minor aphthous ulcerations, or canker sores, can develop around the miniscrew shaft
or on the adjacent buccal mucosa in contact with the miniscrew head. Aphthi are
characterized as mildly painful ulcers affecting nonkeratinized mucosa. Minor
aphthous ulcerations are typically caused by soft-tissue trauma but might occur as a
result of genetic predisposition, bacterial infection, allergy, hormonal imbalance,
vitamin imbalance, and immunologic and psychologic factors. Minor aphthous
ulcerations are self-limiting and resolve within 7 to 10 days without scarring. 65
Placement of a healing abutment, a wax pellet, or a large elastic separator over the
miniscrew head, with daily use of chlorhexidine (0.12%, 10 mL), typically prevents
ulceration and improves patient comfort. The occurrence of aphthous ulceration does
not appear to be a direct risk factor for miniscrew stability, but its presence might
forewarn of greater soft-tissue inflammation.
9.3.2 Soft-tissue coverage of the miniscrew head and auxiliary:
Miniscrews placed in alveolar mucosa, particularly in the mandible, might become
covered by soft tissue. The bunching and rubbing of loose alveolar tissue can lead to
coverage of both the miniscrew head and its attachments (ie, coil spring, elastic chain)
Mini Implants
P a g e | 75
within a day after placement. Soft-tissue coverage might be a risk factor for miniscrew
stability, as well as a clinical concern for the patient, who might think that the
miniscrew has fallen out. Miniscrew attachments (elastic chain, coil spring) that rest on
tissues will likely become covered by tissue. The soft-tissue overlaying the miniscrew
is relatively thin and can be exposed with light finger pressure, typically without an
incision or local anesthetic. Soft-tissue overgrowth can be minimized by placement of a
healing abutment cap, a wax pellet, or an elastic separator. In addition to its
antibacterial properties that minimize tissue inflammation, chlorhexidine slows down
epithelialization and might reduce the likelihood of soft-tissue overgrowth. The authors
suggest partial insertion with a longer miniscrew (10 mm) in regions of loose alveolar
mucosa, leaving 2 or 3 threads of the shaft exposed to minimize the possibility of soft-
tissue coverage.
9.3.3 Soft tissue inflammation, infection, and peri-implantitis:
Healthy peri-implant tissue plays an important role as a biologic barrier to bacteria.
Tissue inflammation, minor infection, and peri-implantitis can occur after miniscrew
placement. 66 Inflammation of the peri-implant soft tissue has been associated with a
30% increase in failure rate. Peri-implantitis is inflammation of the surrounding
implant mucosa with clinically and radiographically evident loss of bony support,
bleeding on probing, suppuration, epithelia infiltrations, and progressive mobility. The
clinician should be forewarned of soft-tissue irritation if the soft tissues begin twisting
around the miniscrew shaft during placement. Some clinicians advocate a 2-week soft-
tissue healing period for miniscrews placed in the alveolar mucosa before orthodontic
loading.135
Mini Implants
P a g e | 76
9.4 COMPLICATIONS DURING REMOVAL
9.4.1 Miniscrew fracture:
The miniscrew head could fracture from the neck of the shaft during removal. The
authors recommend a minimum diameter of 1.6 mm for self-drilling miniscrews that
are 8 mm or longer placed in dense cortical bone. The proper placement technique can
minimize the risk of miniscrew fracture during its removal. If the miniscrew fractures
flush with the bone, the shaft might need to be removed with a trephine. 49
9.4.2 Partial osseointegration:
Although orthodontic miniscrews achieve stationary anchorage primarily through
mechanical retention, they can achieve partial osseointegration after 3 weeks,
increasing the difficulty of their removal 19. The miniscrew typically can be removed
without complications a few days after the first attempt of removal.111
9.4.3 Ingestion of the mini-implant :
Byung-Ho Choi et al evaluated 10 dogs that each ingested 1 screw and 1 reamer. All
screws were passed spontaneously within 1 day, suggesting that spontaneous
evacuation usually occurs for orthodontic anchorage screw ingestion. Patients who
swallow these screws should be carefully observed before resorting to early surgical
removal of the screws 67. Eight of the 10 reamers ingested passed by the fourth day, but
2 became lodged.
Patients who swallow reamers should be carefully observed to determine whether
surgical removal is necessary. 68
Mini Implants
P a g e | 77
Advantages of Mini-implant
use
“That, which does not kill us, makes us stronger.”
– Freidrich Nietzsche
Mini Implants
P a g e | 78
10 Advantages of Mini Implants
• Absolute anchorage.
• New directions of forces.
• Major effectiveness of "en masse" dental movements.
• Reduction of treatment time.
• Less necessity for patient cooperation, especially if we compare microimplants
with intermaxillary elastics or headgear.
• Even though the failure index of microimplants is still high, it shouldn't be the
reason for worries if we take into account the following:
- Headgear and intermaxillary elastics also have a high percentage of failure due to
the absence of patient cooperation.
- Anchorage with auxiliary appliances such as Nance appliance also have a high
index of failure: Anchorage loss of approximately 2mm, lesions due to pressure
against palatine mucosa, infections (for example, candida albicans), debondings
and fractures of appliances, etc.
- Consequences of failures of a microimplant are nothing else but mobility or loss of
microimplant, which can be reinserted in the same site (another microimplant
with the larger diameter if the space permits it and if there is no inflammation in
the zone), or in different zone.
Mini Implants
P a g e | 79
11 BIBLIOGRAPHY
1. Nanda, R., Biomechanics in clinical orthodontics. 1997: WB Saunders Co.
2. Feldmann, I. and L. Bondemark, Orthodontic anchorage: a systematic review. The Angle orthodontist, 2006. 76(3): p. 493-501.
3. Gainsforth, B.L. and L.B. Higley, A study of orthodontic anchorage possibilities in basal bone. American journal of orthodontics and oral surgery, 1945. 31(8): p. 406-417.
4. Ludwig, B., et al., Mini-Implantate in der Kieferorthopädie: Innovative Verankerungskonzepte. 2008: Quintessence Publishing Company.
5. Carano, A. and B. Melsen, Implants in orthodontics. Progress in Orthodontics, 2005. 6(1): p. 62-69.
6. Mah, J. and F. Bergstrand, Temporary anchorage devices: a status report. Journal of clinical orthodontics: JCO, 2005. 39(3): p. 132.
7. Linkow, L., The endosseous blade implant and its use in orthodontics. International journal of orthodontics, 1969. 7(4): p. 149.
8. Linkow, L.I. and R. Cherchève, Theories and techniques of oral implantology. Vol. 1. 1970: CV Mosby Co.
9. Brånemark, P.-I., et al., Intra-osseous anchorage of dental prostheses: I. experimental studies. Scandinavian Journal of Plastic and Reconstructive Surgery and Hand Surgery, 1969. 3(2): p. 81-100.
10. Sherman, A.J., Bone reaction to orthodontic forces on vitreous carbon dental implants. American journal of orthodontics, 1978. 74(1): p. 79.
11. Smith, J.R., Bone dynamics associated with the controlled loading of bioglass-coated aluminum oxide endosteal implants. American journal of orthodontics, 1979. 76(6): p. 618-636.
12. Creekmore, T.D. and M.K. Eklund, The possibility of skeletal anchorage. Journal of clinical orthodontics: JCO, 1983. 17(4): p. 266.
13. Turley, P., P. Shapiro, and B. Moffett, The loading of bioglass-coated aluminium oxide implants to produce sutural expansion of the maxillary complex in the pigtail monkey (< i> Macaca nemestrina</i>). Archives of Oral Biology, 1980. 25(7): p. 459-469.
14. Favero, L., P. Brollo, and E. Bressan, Orthodontic anchorage with specific fixtures: related study analysis. American journal of orthodontics and dentofacial orthopedics, 2002. 122(1): p. 84-94.
15. Zardiackas, L.D., Titanium and Titanium Alloys. Wiley Encyclopedia of Biomedical Engineering.
16. Shapiro, P. and V. Kokich, Uses of implants in orthodontics. Dental Clinics of North America, 1988. 32(3): p. 539.
17. Roberts, W.E., C.L. Nelson, and C.J. Goodacre, Rigid implant anchorage to close a mandibular first molar extraction site. Journal of clinical orthodontics: JCO, 1994. 28(12): p. 693.
Mini Implants
P a g e | 80
18. Roberts, W.E., et al., Osseous adaptation to continuous loading of rigid endosseous implants. American journal of orthodontics, 1984. 86(2): p. 95-111.
19. Ohmae, M., et al., A clinical and histological evaluation of titanium mini-implants as anchors for orthodontic intrusion in the beagle dog. American journal of orthodontics and dentofacial orthopedics, 2001. 119(5): p. 489-497.
20. Park, H.S., H. Kyung, and J. Sung, A simple method of molar uprighting with micro-implant anchorage. Journal of Clinical Orthodontics, 2002. 36(10): p. 592-596.
21. Kanomi, R., Mini-implant for orthodontic anchorage. Journal of clinical orthodontics: JCO, 1997. 31(11): p. 763.
22. Costa, A., M. Raffainl, and B. Melsen, Miniscrews as orthodontic anchorage: a preliminary report. The International journal of adult orthodontics and orthognathic surgery, 1998. 13(3): p. 201.
23. Majzoub, Z., et al., Bone response to orthodontic loading of endosseous implants in the rabbit calvaria: early continuous distalizing forces. The European Journal of Orthodontics, 1999. 21(3): p. 223-230.
24. Hee-Moon Kyung, D., Solutions With Microimplants.
25. Labanauskaite, B., et al., Implants for orthodontic anchorage. Meta-analysis. Stomatologija, 2005. 7(4): p. 128-132.
26. Cope, J.B. Temporary anchorage devices in orthodontics: a paradigm shift. in Seminars in Orthodontics. 2005: Elsevier.
27. Misch, C.E., Contemporary implant dentistry. 2007: Mosby.
28. Cattaneo, P.M., M. Dalstra, and B. Melsen, Analysis of stress and strain around orthodontically loaded implants: an animal study. The International journal of oral & maxillofacial implants, 2007. 22(2): p. 213.
29. Leung, M.T., A.B.M. Rabie, and R.W. Wong, Stability of connected mini-implants and miniplates for skeletal anchorage in orthodontics. The European Journal of Orthodontics, 2008. 30(5): p. 483-489.
30. Chen, Y., H.-I. Shin, and H.-M. Kyung, Biomechanical and histological comparison of self-drilling and self-tapping orthodontic microimplants in dogs. American journal of orthodontics and dentofacial orthopedics, 2008. 133(1): p. 44-50.
31. Wilmes, B. and D. Drescher, Impact of insertion depth and predrilling diameter on primary stability of orthodontic mini-implants. The Angle orthodontist, 2009. 79(4): p. 609-614.
32. Yano, S., et al., Tapered orthodontic miniscrews induce bone–screw cohesion following immediate loading. The European Journal of Orthodontics, 2006. 28(6): p. 541-546.
33. Song, Y.-Y., J.-Y. Cha, and C.-J. Hwang, Mechanical characteristics of various orthodontic mini-screws in relation to artificial cortical bone thickness. The Angle orthodontist, 2007. 77(6): p. 979-985.
34. Miyawaki, S., et al., Factors associated with the stability of titanium screws placed in the posterior region for orthodontic anchorage. American journal of orthodontics and dentofacial orthopedics, 2003. 124(4): p. 373-378.
Mini Implants
P a g e | 81
35. Dalstra, M., P. Cattaneo, and B. Melsen, Load transfer of miniscrews for orthodontic anchorage. Orthodontics, 2004. 1: p. 53-62.
36. Deguchi, T., et al., Quantitative evaluation of cortical bone thickness with computed tomographic scanning for orthodontic implants. American journal of orthodontics and dentofacial orthopedics, 2006. 129(6): p. 721.e7-721.e12.
37. Chaddad, K., et al., Influence of surface characteristics on survival rates of mini-implants. The Angle orthodontist, 2008. 78(1): p. 107-113.
38. Motoyoshi, M., et al., Recommended placement torque when tightening an orthodontic mini‐implant. Clinical oral implants research, 2006. 17(1): p. 109-114.
39. Melsen, B., Mini-implants: Where are we? Journal of Clinical Orthodontics, 2005. 39(9): p. 539.
40. Poggio, P.M., et al., “Safe zones”: a guide for miniscrew positioning in the maxillary and mandibular arch. The Angle orthodontist, 2006. 76(2): p. 191-197.
41. Huang, L.-H., J.L. Shotwell, and H.-L. Wang, Dental implants for orthodontic anchorage. American journal of orthodontics and dentofacial orthopedics, 2005. 127(6): p. 713-722.
42. Gautam, D.P. and P. Valiathan, Implants for anchorage. American journal of orthodontics and dentofacial orthopedics, 2006. 129(2): p. 174-174.
43. McGuire, M.K., E.T. Scheyer, and R.L. Gallerano, Temporary anchorage devices for tooth movement: a review and case reports. Journal of periodontology, 2006. 77(10): p. 1613-1624.
44. Roberts, W.E., Bone tissue interface. Journal of Dental Education, 1988. 52(12): p. 804-809.
45. Garetto, L.P., et al., Remodeling dynamics of bone supporting rigidly fixed titanium implants: a histomorphometric comparison in four species including humans. Implant dentistry, 1995. 4(4): p. 235-243.
46. American Academy of Implant Dentistry. Glossary of implant terms. J Oral Implantol, 1986. 12(2): p. 284-94.
47. Brunski, J.B., et al., The influence of functional use of endosseous dental implants on the tissue-implant interface. I. Histological aspects. Journal of dental research, 1979. 58(10): p. 1953-1969.
48. Kravitz, N.D. and B. Kusnoto, Risks and complications of orthodontic miniscrews. American journal of orthodontics and dentofacial orthopedics, 2007. 131(4): p. S43-S51.
49. Melsen, B. and C. Verna. Miniscrew implants: the Aarhus anchorage system. in Seminars in Orthodontics. 2005: Elsevier.
50. Asscherickx, K., et al., Root repair after injury from mini screw.‐ Clinical oral implants research, 2005. 16(5): p. 575-578.
51. Kyung, H.-M., et al., Development of orthodontic micro-implants for intraoral anchorage. Journal of clinical orthodontics: JCO, 2003. 37(6): p. 321.
52. Giancotti, A., C. Arcuri, and A. Barlattani, Treatment of ectopic mandibular second molar with titanium miniscrews. American journal of orthodontics and dentofacial orthopedics, 2004. 126(1): p. 113-117.
Mini Implants
P a g e | 82
53. Sujatha, N., K. Manjunath, and V. Balasubramanyam, Variations of the location of the greater palatine foramina in dry human skulls. Indian journal of dental research: official publication of Indian Society for Dental Research, 2005. 16(3): p. 99.
54. Denio, D., M. Torabinejad, and L.K. Bakland, Anatomical relationship of the mandibular canal to its surrounding structures in mature mandibles. Journal of Endodontics, 1992. 18(4): p. 161-165.
55. Ardekian, L., et al., The clinical significance of sinus membrane perforation during augmentation of the maxillary sinus. Journal of oral and maxillofacial surgery, 2006. 64(2): p. 277-282.
56. Brånemark, P., et al., An experimental and clinical study of osseointegrated implants penetrating the nasal cavity and maxillary sinus. Journal of oral and maxillofacial surgery, 1984. 42(8): p. 497-505.
57. Carano, A., et al., Clinical applications of the miniscrew anchorage system. J Clin Orthod, 2005. 39(1): p. 9-24.
58. Büchter, A., et al., Load related implant reaction of mini implants used for orthodontic‐ ‐ anchorage. Clinical oral implants research, 2005. 16(4): p. 473-479.
59. Misch, C.E., Density of bone: effect on treatment plans, surgical approach, healing, and progressive boen loading. The International journal of oral implantology: implantologist, 1990. 6(2): p. 23.
60. Sevimay, M., et al., Three-dimensional finite element analysis of the effect of different bone quality on stress distribution in an implant-supported crown. The Journal of prosthetic dentistry, 2005. 93(3): p. 227-234.
61. Kinner, F. and K.D. Schlegel, The anatomic basis for palatal implants in orthodontics. Int J Adult Orthod Orthognath Surg, 2002. 17: p. 133-139.
62. Cheng, S.-J., et al., A prospective study of the risk factors associated with failure of mini-implants used for orthodontic anchorage. The International journal of oral & maxillofacial implants, 2004. 19(1): p. 100.
63. Heidemann, W., H. Terheyden, and K. Louis Gerlach, Analysis of the osseous/metal interface of drill free screws and self-tapping screws. Journal of cranio-maxillofacial surgery, 2001. 29(2): p. 69-74.
64. Liou, E.J., B.C. Pai, and J.C. Lin, Do miniscrews remain stationary under orthodontic forces? American journal of orthodontics and dentofacial orthopedics, 2004. 126(1): p. 42-47.
65. Murray, B., et al., A comparative study of the efficacy of AphthealTM in the management of recurrent minor aphthous ulceration. Journal of oral pathology & medicine, 2005. 34(7): p. 413-419.
66. Costa, A., G. Pasta, and G. Bergamaschi. Intraoral hard and soft tissue depths for temporary anchorage devices. in Seminars in Orthodontics. 2005: Elsevier.
67. Choi, B.-H., et al., Ingestion of orthodontic anchorage screws: An experimental study in dogs. American journal of orthodontics and dentofacial orthopedics, 2007. 131(6): p. 767-768.
Mini Implants
P a g e | 83
68. Pablo Echarri, Tae-Weon Kim, and L. Favero, eds. Orthodontics and Microimplants. 1st ed. 2007: Madrid.
Mini Implants
P a g e | 84
APPENDIX - IProduct Company Postal address Web site
Aarhus Anchorage System
MEDICON eG Gänsäcker 15, D-78532, Tuttlingen, Germany
www.medicon.de
ScanOrto A/S Hans Edvard Teglers Vej 2, 2920-Charlottenlund, Denmark
www.aarhus-mini-implant.com
AbsoAnchor System Dentos 258 BunJi, Dong-In Dong, Jung-Gu, Taegu, Korea
www.dentos.co.kr
C-Implant Dentium Inc. 6F Dahn World B/D, 154-11 Samsung-dong, Kangnum-gu,
135-897 Seoul, Korea
www.implantium.com
Cizeta Titanium Miniscrew
Cizeta Surgical San Lazzaro di Savena, Bologna,
Italy
www.cizetasurgical.it
Dual-Top Anchor System
Jeil Medical Corporation
775-3 Daesung B/D, Daelim 3
Dong Youngdeungpoku, Seoul, Korea
www.jeilmed.co.kr
Distributed by RMO Inc.
P.O. Box 17085, Denver, CO
80217
www.rmortho.com
IMTEC Mini Ortho Implant
IMTEC Corporation
2401 N. Commerce, Ardmore, OK 73401
www.imtec.com
Mini Implants
P a g e | 85
Lin/Liou Orthodontic Mini
Anchorage Screw (LOMAS)
Mondeal Medical Systems GmbH
Moltkestrasse 39, D-78532 Tuttlingen, Germany.
www.mondeal.de
Distributed by Mondeal North
America, Inc.
6895 Lake Bluff Drive, Comstock Park, MI 49321
Miniscrew Anchorage System (MAS)
Micerium S.p.a. Via Marconi
Via Marconi 83, 16030 vegno, Italy
www.micerium.it
Product Company Postal address Web site Orthoanchor K1 System
Dentsply Sankin Corporation
Tokyo, Japan www.dentsply-sankin.com
Orthodontic Mini Implant (OMI)
Leone S.p.A. Via P. a Quaracchi 50, 50019
Sesto Fiorentino, Firenze, Italy
www.leone.it
Distributed by Leone America
501 W. Van Buren, Suite S, Avondale, AZ 85323
Spider Screw Anchorage System
HDC Via dell’Industria 19, 36030 Sarcedo, Italy
www.hdc-italy.com
Mini Implants
P a g e | 86
APPENDIX – II
FIGURE 1 A- SPIRAL-SHAFT IMPLANTS (CHERCHEVE). B- VARIOUS SIZES OF EARLY DESIGNED VENT-PLANT IMPLANTS. C - A
TRIPOD IMPLANT, ASSEMBLED. D- VARIOUS EARLY DESIGNED BLADE IMPLANTS. 8.................................13
FIGURE 3 INTRUSION ACHIEVED BY MINI IMPLANTS IN STUDIES BY KANOMI 21..................................................18
FIGURE 2 PLACEMENT OF MINI-IMPLANT..................................................................................................18
FIGURE 4. CLASSIFICATION OF BIOCOMPATIBLE TADS..................................................................................24
FIGURE 5 CLASSIFICATION OF BIOLOGIC TAD............................................................................................25
FIGURE 6............................................................................................................................................27
FIGURE 7 VARIOUS MINISCREW IMPLANTS. A, THE AARHUS ANCHORAGE SYSTEM. B, THE ABSOANCHOR. C, THE SPIDER
SCREW ANCHORAGE SYSTEM . D, THE IMTEC MINI ORTHO IMPLANT..............................................28
FIGURE 8 FROM LEFT TO RIGHT: 1.5 MM CYLINDRICAL MINI-IMPLANT WITH A SLOT AT THE SCREW HEAD; 1.6 MM
TAPERED MINI-IMPLANT WITH A SLOT AT THE SCREW HEAD; AND 2.0 MM CYLINDRICAL MINI-IMPLANT WITH
THREADS AT THE SCREW HEAD....................................................................................................29
FIGURE 9 MAXILLARY MINI-IMPLANT LOCATIONS. A. BELOW NASAL SPINE. B. IN THE PALATE. C. INFRAZYGOMATIC CREST 39.........................................................................................................................................38
FIGURE 10 MANDIBULAR MINI-IMPLANT LOCATIONS. A. RETROMOLAR AREA AND MOLAR REGION. B. ALVEOLAR
PROCESS. C. SYMPHYSIS............................................................................................................39
FIGURE 11..........................................................................................................................................44
FIGURE 12..........................................................................................................................................49
FIGURE 13..........................................................................................................................................51
FIGURE 14..........................................................................................................................................51
FIGURE 15..........................................................................................................................................52
FIGURE 16..........................................................................................................................................52
FIGURE 17 AREAS ACCORDING TO BONE DENSITY.....................................................................................76
Mini Implants