52
Emergency related Pac emaker and ICDs Issues Anne-Marie de Vries, January 2015

Emergency pacemaker and ICD issues

Embed Size (px)

Citation preview

Page 1: Emergency pacemaker and ICD issues

Emergency related Pacemaker and ICDs Issues

Anne-Marie de Vries, January 2015

Page 2: Emergency pacemaker and ICD issues

Contents

1. Clinical case2. Introduction pacemaker / Cardiac Resynchronisation Therapy

(CRT) / Implantable Cardiac Defibrillator (ICD)3. Implant related complications4. Pacing system malfunction5. ED management6. Magnet7. Back to case8. Demo-interrogation (by Medtronic)9. Patient experience (ICD)

Page 3: Emergency pacemaker and ICD issues

Clinical case

Page 4: Emergency pacemaker and ICD issues

ECG

ECG

Page 5: Emergency pacemaker and ICD issues

ECG

Page 6: Emergency pacemaker and ICD issues

Introduction

• Components and lead position

• Pacemakers modes

• Pacemaker/CRT/ICD indications

Page 7: Emergency pacemaker and ICD issues

Pacemaker components and lead position

• Basic components: pulse generator and lead(s)

• Pulse generator; hard and software, battery (5-10 years)

• Leads:conductor, electrode, fixation mechanism, terminal

connector pin, insulator

Page 8: Emergency pacemaker and ICD issues

Lead position

• Atrial leads in the right atrial appendage

• Right atrial septum position to suppress AF

• Right ventricular leads are traditionally placed in the right

ventricular apex, however RVOT or inter-ventriculair position to

avoid dyssynchrony

• Left ventriculair lead via coronary sinus to free left lateral or

posterior left ventriculair wall

• Final position depends on venous cardiac anatomy, avoidance of

phrenic nerve stimulation, location of myocardial scar

Page 9: Emergency pacemaker and ICD issues

Lead fixation

Page 10: Emergency pacemaker and ICD issues

Pacemaker modes

5 Letter code:

I = Chambers paced

II = Chambers sensed

III = Response to sensing

IV = Programmability, rate modulation

V = Multi-site pacing

A= Atria T=Triggered

V=Ventricle I=Inhibited

D=Dual Chamber (A+V) R=rate modulation

O=none

Page 11: Emergency pacemaker and ICD issues

Table 2 NASPE pacemaker code

Page 12: Emergency pacemaker and ICD issues

Pacemaker modes

• ED physicians most likely deal with AAIR, VVIR, DDD

and DDDR

• Rate response modality detect physiological changes;

minute ventilation, QT interval, stroke volume

Page 13: Emergency pacemaker and ICD issues

Indications

• Continuous amendments for pacemaker indications

several categories)

• Most recent guideline ACC/AHA/HRS guideline

pacemaker, ICD and CRT implantation 2008 (update

2012 on CRT)

Page 14: Emergency pacemaker and ICD issues

Pacemaker indication

• Sinus node dysfunction;

Symptomatic SSS

DDDR*

• Acquired AV block;

- III and Mobitz II degree block

- Dual chamber pacing preferrable

• Chronic bifascicualr or trifascicular block

Syncope risk varies based on presence of AV block and

pacemaker indication depends of HV intervals (EP-study)

Page 15: Emergency pacemaker and ICD issues

CRT

• Systolic heart failure due intra- and inter-ventriculair dyssynchrony

• Dyssynchrony occurs independent from QRS duration

• Dyssynchrony determined by Echocardiographic parameters; TDI

and speckle tracking imaging

• In patients with systolic heart failure significant reduction in NYHA

class*(PATH-HF, MUSTIC SR, MIRACLE, COMPANION,CARe-

HF,MIRACLE-ICD,CONTAK-CD,RHYTHM-ICD)

• Non-responders to CRT; 10-40% non LBBB (QRS <130 msec,

RBTB)

Page 16: Emergency pacemaker and ICD issues

TDI (Tissue Dopler Imaging)

Page 17: Emergency pacemaker and ICD issues

Speckle tracking imaging

Page 18: Emergency pacemaker and ICD issues

ICD; indications 2012 ACCF/AHA/HRS guidelines

Level of Evidence: A:

LVEF<35% due to prior MI who are at least 40 days post MI (NYHA II-III)

LV dysfunction due to prior MI, LVEF < 30%, NYHA I

Survivors of Cardiac arrest due to VF or hemodynamically unstable sustained

VT after evaluation to define the cause of event and excluding any reversible

causes

Level B

Non-ischeamic DCM LVEF < 35%, NYHA II-III

Non sustained VT due to prior MI, LVEF < 35% or inducible VF/FT on EPS

Structural heart disease and spontaneous sustained VT

Syncope of undetermined origen with clinically relevant HD sign sustained VT

or VF induced at EPS

LQTS patients who experience syncope and/or VT while on b-Blockers

Page 19: Emergency pacemaker and ICD issues

Implant-related complications

1. Pocket hematoma2. Infection3. Lead dislodgement4. Pneumothorax/hematothorax5. Venous thrombosis6. Pacemaker syndrome7. Triscupid regurgitation8. Twiddler syndrome9. Pericardial effusion, perforation

Page 20: Emergency pacemaker and ICD issues

Pocket hematoma

• Post procedural bleeding and pocket hematoma are relatively

common complications (on oac)

• Hematoma’s are generally treated conservatively unless large or

very painful they need surgical intervention (increased rate of

infection)

Page 21: Emergency pacemaker and ICD issues

Infection

• About 0.2% (retrospective studie) of surgical site infection

(swelling, dehiscencion, erosion)

• However, fever of unknown origin an implanted device should

always be considered as a source of infection

• BC, Echogardiographic Dx

• Long term IVAB (Stahp A), device removal and lead extraction.

Re-insertion of device contralateral

Page 22: Emergency pacemaker and ICD issues

Heamato- and pneumothorax

• About 1.5% pneumothorax (subclavian puncture)

• Hemothorax due to unintentional arterial puncture

• Avoidance with fluoroscopic insertion techniques or axillary

vein cannulation

Page 23: Emergency pacemaker and ICD issues

Thrombosis

• Subclavian or brachio-chephalic venous thrombosis is a very

common event after pacemaker implantation with an

incidence of 2-22% from several days to up to 9 years after

implantation

• Often an asymptomatic occurrence shown on a venogram

during lead revision

• Symptomatic patients show typical manifestations of acute

DVT (swelling) or even vena cava superior syndrome

• PE is an uncommon event

• Therapy ranges from heparinisation/warfarin unto

percutaneous angioplasty or open surgery

Page 24: Emergency pacemaker and ICD issues

Pacemaker syndrome

• Intolerance to ventricular pacing in the absence of

atrioventricular synchrony (“the AV dyssynchrony syndrome”)

• Ventricular pacing leads to suboptimal AV dyssynchrony

regardless of the pacing mode

• Symptoms result from the loss of physiological timing of atrial

and ventricular contractions

• Symptoms are a constellation of features due to decreased

cardiac output, loss of atrial contribution to ventricular filling,

loss of total peripheral resistance response, non physiological

pressure waves

• Patient needs adjustment of generator or lead to better

coordinate timing of atrial and ventricular contraction

Page 25: Emergency pacemaker and ICD issues

Triscupid regurgitation

• Some patients present with R sided cardiac failure

• mechanism; impingement or adherence to leaflets,

perforation, impairment of valve closure due to entanglement,

, scar formation of thrombosis

• Or due to ventricular dyssynchrony

• Dx is made by 3D-TTE

Page 26: Emergency pacemaker and ICD issues

Twiddler Syndrome

• Lead dislodgement due to a patient’s own compulsive

manipulation of generator causing the leads to become retracted

or coiled around the generator

• Dx ismade by radiographic appearance

• Dacron pouches

• Subpectoral pouches

Page 27: Emergency pacemaker and ICD issues

Pacing system malfunction

• Generator malfunction in 4.6/1000 pacemakers and 20.7/1000

ICDs, most often due to battery malfunction

• Lead failure often due to insulation breaks, however this is a late

complication

• Pacemaker malfunction is a rare cause of syncope (8/162

patients in a retrospective study)

• An assesment of pacemaker malfunction should consist of ECG,

radiology, patients acid-base status (VBG), medication use and

interrogation (lead problems can result in very high impedance)

Page 28: Emergency pacemaker and ICD issues

Lead dislodgement

• Usually after recent implantation, however up to 3 months after

implant is possible

• Undersensing, failure to capture or change in DFTs

• Telemetry, device interrogation

• Obtain a CXR in 2 directions (compare with previous CXR)

• Malposition near phrenic nerve or diaphragm may cause

hiccougs or muscle twitching

• Seldom leads may cause perforation and tamponade (TTE)

Page 29: Emergency pacemaker and ICD issues

Pacing system malfunction

1. Failure to capture

2. Failure to sense

3. Failure to pace

4. Electromagnetic interference

Page 30: Emergency pacemaker and ICD issues

Failure to capture

• Failure to deliver a pacing stimulus without subsequent

depolarisation

• Functional failure to capture when myocardial tissue is in

refractionary state

• Pathological failure to capture due to numerous conditions;

myocardial disease, electrolyte disturbances (Potassium !), anti-

arrhythmic drugs (Class Ic and QTc lengthening)

• Latency; might be confused with failure to capture; obtain a

multichannel recording recording (interrogation)

• Lead dislodgement, perforation and fracture, increasing DFT

may all lead to failure to capture

Page 31: Emergency pacemaker and ICD issues

Failure to sense

• On myocardial depolarisation a signal is send via the lead wire

and filtered by the generator

• Undersensing could be caused by a change in morphology or

vector of the depolarisation front

• PVC, BBB or VTs may need exceed the sensing threshold

(depending on the programming)

• Break in wire insulation or Battery depletion may cause under

sensing

• Undersensing immediate after implantation may be caused by

dislodgement or perforation

• Again; Class Ic or HYPERKALEMIA

• Long blanking and refractory periods of a pacemaker can lead to

relatively or functional undersensing

Page 32: Emergency pacemaker and ICD issues

Failure to pace

• Failure to deliver a stimulus to the heart

• Absence of pacing artefacts on intrinsic rates lower than the lower rate

of the device

• Causes; oversensing, partial lead fracture or insulation defects,

crosstalk

• Oversensing could be caused by retrograde P waves, t waves,

ventricular EADs or DADs skeletal myo-potentials (pectoralis,

diaphragm, rectus abdomens)

• External electromagnetic interference

• Blunt trauma to pulse generator, lead fracture, dislodgement or loose

connection

• Pacing artefacts may not be visible on 12 lead ECG and give a false

impression of failure to pace

• PACs or PVCs on a single telemetry lead can give a false impression of

failure to pace

Page 33: Emergency pacemaker and ICD issues

Pacemaker tachycardia

• Also known as pacemaker re-entry tachycardia or endless-loop

tachycardia (like other re-entry dysrhythmias, pacemaker is now a part

of the re-entry circuit)

• only in patients with dual chamber devices

• An intrinsic premature complex is sensed by an atrial lead of the

pacemaker which responds by generating a ventricular impulse. The

ventricular impuls is retrogradely conducted via the AV node to the atria,

now the atrial lead senses the atrial impulse and generates a ventriculair

pulse (completed loop)

• This pacemaker re-entry tachycardia will not exceed the programmed

upper rate, however can be significant enough to cause symptoms

• Adenosine may not be effective where as magnet therapy will be

• NB ! pacemakers atrial sensing tresholds need to be adjusted

• NB ! Mode switch can cause the pacemaker to discontinue atrial

tracking when the atrial upper rate is exceeded

Page 34: Emergency pacemaker and ICD issues

Pax

Page 35: Emergency pacemaker and ICD issues

ED management

1. History

2. Physical examination

3. 12 lead ECG

4. VBG !!!

5. Radiology (CXR PA and lat)

6. Device interrogation/magnet function

Page 36: Emergency pacemaker and ICD issues

ED Management

1. Hyperkalemia

2. Myocardial infarction (Sgarbossa)

Page 37: Emergency pacemaker and ICD issues

Hyperkalemia

• Widened QRS complex (widened paced complexes)

• Severe hyperkalemia causes a sine wave following pacing

artefacts

• Elevation of pacing threshold; increased latency, intermittent

capture, or continuous loss of capture, loss of sensing

• Latency may be temporarily overcome by increasing the maximum

voltage of output HOWEVER treatment of hyperkaliema is a

priority

Page 38: Emergency pacemaker and ICD issues

Acute myocardial infarction

• The depolarisation and depolarisation from ventricular paced

rhythms may complicate the diagnosis of acute MI (also in LBBB)

• ST and T waves shift in a discordant direction which can mask or

mimic AMI, “appropiate discordance”

• SERIAL ECGs show dynamic ST-segment changes

• 3 criteria used to Dx infarction in patients with LBBB are

• Concordant ST-elevation> 1 mm (score 5)

• Concordant ST-depression>1 mm in V1-V3 (score 3)

• Excessively discordant ST elevation > 5 mm (score 2) in leads

with a negative QRS complex

Page 39: Emergency pacemaker and ICD issues

New or old LBBB ?

• “NEW developed LBBB” reflects AMI is NOT true ! Recognition goes

back to 1917 !

• New LBBB should be either due to a LARGE anterior or

anteroseptal infarction resulting in a large injury OR very focal lesion

just distal from the His bundel

• New or old, not always clearly visible in acute settings

• Patients with previous LBBB generally have underlying myocardial

or conduction system abnormalities predisposing to a larger risk

(older, hypertension, CHF)

• Historical studies do not make a fair differentiation in patients with

(L)BBB and ACS

• New studies show that in only 29% of new LBBB; ACS is present*

*Chang et al, Am J Emerg Med, 2009,; 27:916-21

Page 40: Emergency pacemaker and ICD issues

The Ventriculair conduction system

Page 41: Emergency pacemaker and ICD issues

Sgarbossa criteria

Criteria A Sgarbossa: (90% specifity): 5 point

Concordant ST-elevation> ANY 1mm in lead with a positive QRS

Criteria B Sgarbossa: (90% specifity): 3 point

Concordant ST- depression > 1 mm in ANY lead from V1-V3

Criteria C Sgarbossa: Excessively ST-elevation > 5mm score 2

(high sensivity, lower specifity, ratio of R vs > 20%)

> 3 points ; 90% specifity for ACS

Criteria from Sgarbossa based on a retrospective review from 17

patients from GUSTO-1

Page 42: Emergency pacemaker and ICD issues

ECG ACS in LBTB

Page 43: Emergency pacemaker and ICD issues

ICD trouble shooting

• Device and procedural complications (as with pacemakers)

• Arrhythmia related complications; inappropriate shocks,

appropriate shocks; ventricular storm and incessant tachycardia,

phantom shocks

• pro-arrhythmic effects

• Increased DFT (fibrosis, amiodarone),

• Heart-failure

• Phantom shocks

• Emotional and quality of life issues: anxiety, depression, fear for

Icd discharge, negative life style changes; unable to return to

work or to drive, decrease in physical activity

Page 44: Emergency pacemaker and ICD issues

Inappropriate shocks in ICD patients

• Most often caused by SVTs or non sustained VTs

• Other causes; electromagnetic interference, inappropriate

sensing, ICD malfunction (lead fracture)

• Most inappropriate shocks are avoided with modern and

adequate ICD programming (high cut-off rates, arrhythmia

discrimination detection)

Page 45: Emergency pacemaker and ICD issues

Appropriate shocks in ICD patients

• Increasing arrhythmia duration and ATP have diminished

delivered shocks (ATP terminates at least 90% of VTs)

• Electrical storm (definition in ICD carriers is 2 or more adequately

delivered shocks < 24 hours);

• Amiodarone loading, b Blockers (reduction of effects of sympatic

nerve system) and sometimes proceeding towards sedation. both

electrical storm and incessant VT indicate a worse prognosis due

to the underlying mechanism

Page 46: Emergency pacemaker and ICD issues

Pro-arrhythmic effect of ICD

• Arrhythmogenic effect of implanted lead; irritation of local

myocardial tissue, fibrosis.

• Inserted in a prior scar site may induce reentry

• RV pacing causes interventricular and intraventricular

dyssynchrony

• “normal” bradycardia back up pacing may result in short-long

pacing sequence leading towards induction of VT

• Biventriculair pacing reduces SCD however may induce Vt due to

an abnormal sequence of depolarisation and depolarisation fronts

Page 47: Emergency pacemaker and ICD issues

Magnet function

• Know where the magnet is located in your ED

• Know the effects of applying a magnet to a device

• Know the difference between a magnet application and

an electrical reset

• Most pacemakers and ICDs turn off the sensing mode

when a magnet is applied over the generator

• The pacemaker starts to pace in an asynchronous

mode in a fixed rate (Medtronic 65/min reflects EOL)

• Rarely a device may be programmed in a “magnet off”

setting

Page 48: Emergency pacemaker and ICD issues

When to apply a magnet ?

1. To terminate a pacemaker mediated tachycardia

2. Inappropriate sensing (due to electromagnetic interference;

diathermy, MRI)

3. Over sensing due to lead problems

4. Stop ICD ATP or defibrillation shock therapy (backup

bradycardia pacing will still be provided).NB an alert sound will

be initiated when applying a magnet to a medtronic ICD (also on

lead and impedance problems or EOL)

Page 49: Emergency pacemaker and ICD issues

ACLS in an ICD patient

• ICD and CRT-D patients should have same approach

• SHOCKS delivered are of NO risk for the MET-team

• If arrhythmia persists after a device defibrillation, attach patient to

anterior-posterior placed leads preferably 8 cm from generator

• Consider magnet placing to avoid unintentional device

defibrillation's

• Consider; increased DFTs due to ipsilateral pneumothorax

• `consider femoral access instead of subclavian due to often

thromboses veins (or in appropiate response of a pacemaker due

to wire interrogation)

Page 50: Emergency pacemaker and ICD issues

More ?

go to: Medtronic Academy (for Physicians)

LIFTL Fellowship exam SAQ and SMACC 2014

Amal Mattu

ECG pedia

Page 51: Emergency pacemaker and ICD issues

Back to clinical case

81 male

4/7 post dual chamber pacemaker insertion (pAF)

Re-currence of pre implantation dizziness, palpitations and

SOB

CXR: correct lead position

Interrogation: A not sensing, P wave in bipolar, re-programmed

to bipolar, treshold and impedence normal.

Loaded on amiodarone

Page 52: Emergency pacemaker and ICD issues

Literature

1. Up to Date; Electrical storm and incessant ventricular tachycardia, Overview of

pacing in heart failure, implantable cardioverter defibrillators;

2. EMP: Managing Pacemaker related Complications and malfunctions in the ED

3. Cardiac Pacing and ICDs; 4th Edition, K.A Ellenbogen

4. A practical guide to cardiac pacing, 5th edition, H.Weston Moss

5. Electrofysiological Testing, 3d edition, R.N Fogoros

6. AHHF/AHA/HRS 2012 Guidelines for device based therapy

7. Medtronic Academy

8. LIFTL

9. Neeland et al, J am Coll Cardiol, 2012; 10;60 (2):96-105. “Evolving

considerations in the management of patients with LBTB and suspected AMI