Presentation on Dosimetry with calorimeter By Emmanuel W. Fiagbedzi Master of medical physics student ICTP-University of trieste Italy 03/14/22 1

Dosimetry with calorimeter

Embed Size (px)

Citation preview

Presentation on Dosimetry with calorimeter

By

Emmanuel W. Fiagbedzi

Master of medical physics student

ICTP-University of trieste

Italy

04/15/23 1

INTRODUCTIONCalorimetry dosimetry is the most precise of all

absolute dosimetry techique.It is a way of measuring energy imparted to

matter by radiation through temperature change This means that the dose absorbed in the

sensitive volume is proportional to change in temperature.

Only relatively small corrections for thermal leakage and for chemical reactions are necessary

04/15/23 2

In principle any kind of thermometer can be applied in a calorimeter if the temperature change is large enough to measure with sufficient accuracy and precision

In practice only thermocouples and thermistors are sufficiently sensitive and small; thermistors are usually preferable because of their greater sensitivity

04/15/23 3

The temperature increase per unit of absorbed dose to the material in the calorimeter’s sensitive volume depends on its thermal capacity, which is usually expressed in cal/g °C or J/kg °C

The exact value of the calorie (i.e, the energy required to raise 1 g of water 1 °C) depends upon the temperature of the water to which it refers

Usually thermal-capacity (or specific-heat) tables assume the value of the calorie for water at 15 °C; hence 1 cal = 4.185 J, and 1 cal/g °C = 4185 J/kg °C

04/15/23 4

For a sensitive volume containing a material of thermal capacity h (J/kg C), mass m (kg), and thermal defect , and that absorbs E joules of energy, the temperature increase is given by

where ̅D is the average absorbed dose (Gy)

in the sensitive volume

(1 ) (1 ) ( C)

E DT

hm h

04/15/23 5

Thus a measurement of ̅D does not require explicit knowledge of m if h is known

The thermal defect is the fraction of E that does not appear as heat, due to competing chemical reactions, if any

is negative for exothermic reactionsA few typical values of h are given in the

following table

04/15/23 6

Thermal capacity of several calorimetric materials

04/15/23 7

For example, in Al a dose of 1 Gy causes a temperature increase of 1.12 10-3 C

To measure this temperature rise with 1 % precision would require a thermometer capable of detecting temperature changes of the order of 10 C

04/15/23 8

Thermocouples typically have temperature coefficients of 40 – 70 V/C

A temperature change of 10 C would then give a potential change of (4 – 7) 10-10 V

This is too small to detect with available instruments, such as a nanovoltmeter

Increasing the dose to 100 Gy would cause a temperature rise of 0.112 C, requiring detection of (4 – 7) 10-8 V for 1 % precision, which can be accomplished with a nanovoltmeter

04/15/23 9

Thermocouples are generally found to be most useful in calorimeters where large doses (> 10 Gy) are given, usually in a short enough time period for thermal leakage to be negligible (i.e., under adiabatic conditions)

Thermocouple sensitivity can be multiplied by constructing a thermopile, consisting of a number of thermocouples in series, but this is usually not practical for calorimetric dosimetry because of the increase in perturbation of the medium and the number of thermal leakage paths

04/15/23 10

Thermistors can be obtained in sized comparable to thermocouples

They are semiconductors made of metallic oxides and other constituents that are usually not specified by the manufacturer

They exhibit negative temperature coefficients of the order of several percent per °C at room temperatures, increasing in negative coefficient with decreasing temperature.

04/15/23 11

The resistance of a thermistor at room temperature is typically 103 – 105 , which can be conveniently measured with great precision and accuracy by a Wheatstone bridge as shown in the following diagram

The bridge null detector must be sensitive enough so that the power dissipated in the thermistor is negligible compared to the radiation heating

04/15/23 12

A Wheatstone bridge circuit for measuring the resistance of a thermistor in the sensitive volume (or “core”) of a calorimetric dosimeter. When Rx is set to produce a null current reading,

Rc/Rx = R1/RJ, from which Rc can be determined.

04/15/23 13

There are three general types of radiometric calorimeter designs, depending on whether absorbed dose in a reference medium, energy fluence in a radiation beam, or power output of a radioactive source is to be measured.

They are 1.Energy fluence calorimeter 2.Absorbed dose calorimeter 3. calorimeters for measuring the power ouput of a radioactive source.

We shall discuss only one which is the Energy fluence calorimeter.

04/15/23 14

An energy-fluence calorimeter contains a core,usually consisting of a cylindrical piece of dense material such as lead or gold,large enough to stop an incident beam of radiation.

The core is suspended by nylon strings in an insulated vacuum chamber,sometimes adjacent to a twin core that serves as a control to determine thermal leakage.

More than one thermistors may be necessary to sample the temperature adequately and also the heater should be designed to distribute the heat uniformly.

04/15/23 15

04/15/23 16

The energy fluence of the radiation beam passing through the aperture of the area A is calculated by

h in the equation can be determined through electrical calibration

04/15/23 17

04/15/23 18

At high dose rate where other dosimeters shows saturation effects,calorimeters are at their best

Attix, Introduction to radiological physics and Radiation Dosimetry. pages 426-436

Padovani, Dosimetry 6 lecture notes slides

GRAZIE!!!