Upload
shashidhar-gs
View
114
Download
4
Embed Size (px)
Citation preview
Cryogenics [email protected]
1 | P a g e
UNIT – 5
ULTRA LOW TEMPERATURE CRYO –
REFRIGERATORS
Magneto Caloric Refrigerator:
An apparatus that utilizes the principle of adiabatic demagnetization to
continuously maintain a constant low temperature is known as a
magnetic refrigerator (see Fig). This is possible through the use of
superconducting thermal "valves." These are strips of metal, such as
lead, which are superconducting below a "transition" temperature in the
absence of a magnetic field and have normal electrical resistance in the
presence of a magnetic field. In the superconductive state, thermal
conductivity is reduced. Thus, a superconductor can be used as a thermal
switch by turning a magnetic field on (switch "closed") or off (switch
"open").
Cryogenics [email protected]
3 | P a g e
Thermodynamic cycle for the magnetic refrigerator
The process that occurs in a magnetic refrigerator is shown in Fig.:
1. With the system at a steady-state temperature (Tl at approximately
1 K), the working salt is brought into thermal contact with the liquid
helium bath by opening the upper thermlll valve and closing the lower
thermal valve.
As the magnetic field is applied for process step 1-2, the entropy of the
salt is decreased isothermally.
2. After applying the magnetic field, both thermal valves are closed and
the magnetic field around the working salt is adiabatically reduced in
process step 2-3 to an intermediate value. This causes the salt to cool to
a temperature of T3 •
Cryogenics [email protected]
4 | P a g e
3. With the upper thermal valve closed and the lower thermal valve
opened, the magnetic field is further reduced in process step 3--4 while
heat isabsorbed isothermally from the space to be cooled by the reservoir
salt and by the working salto
4. In process step 4--1, both thermal valves are closed and the working
salt is adiabatically remagnetized to its starting condition.
The refrigeration effect for this proeess is
= m (
where m is the mass of the working salt. The amount of energy rejected
as heat from the working salt is
= m (
The net work for the cycle is then
= | | | | m (
Therefore, the coefficient of performance for an ideal magnetic
refrigerator is
COP =
=
which is the same as that for a Carnot refrigerator.
Dilution refrigerator:
In 1951, H. London made the suggestion to use a solution of the rarer
isotope in the more common isotope
to obtain low
temperatures. In any dilute solution, the solute molecules can be
considered to behave like a gas whose pressure and volume correspond
to the osmotic pressure and volume of the solution. Dilution of the
solution by adding more solvent causes an "expansion" of the solute
"gas" and cooling should result. A practical dilution cycle was first
developed in 1962.
Cryogenics [email protected]
5 | P a g e
The density of is less than that of
. Therefore, at temperatures
below 0.87 K, solutions of and
exist in two liquid phases with
the rich phase residing above the
phase. The migration of
atoms across the liquid-liquid interface is similar to evaporation in that
the atoms exert no drag force on the
atoms. Since the osmotic
pressure of the dilute phase extrapolates to about 0.0023 MPa at 0 K
instead of 0 MPa, it suggests that, near absolute zero, a stable solution of
these isotopes should contain about 6.4 % 3He. During the phase
transition of into the
solution at constant temperature, the
entropy increases and heat is absorbed by the 3He to increase its
enthalpy. This is the driving force behind the dilution cycle.
In a continuously operating dilution cycle (see Fig.), helium gas
composed of essentially pure is compressed with the aid of a
vacuum pump, cooled, and sent through a heat exchanger followed by
heat exchange in two helium baths, the first at 4.2 K and the second near
Cryogenics [email protected]
6 | P a g e
1 K. A throttling device in the line provides the necessary temperature
adjustment to permit the mixture to supply the necessary energy to
operate the still. After further cooling in another heat exchanger, the
mixture is admitted to the mixing chamber. Two liquid phases appear as
the temperature is further decreased with the less dense -rich mixture
on the top. Since this is a continuous process, diffuses into the
phase in the mixing chamber and must be replenished in order to
maintain equilibrium. The lower part of the mixing chamber contains
in the superfluid form (He II), through which
easily diffuses.
This expansion of the into the more dense
phase provides the
refrigeration effect. The mixture is returned through the heat
exchanger to the still where the is evaporated from the mixture.
This vapor is warmed and returned to the vacuum pump where
recompression of the completes the cycle. This compressed
replaces the originally dissolved in the
phase. Temperatures
down to about 15 mK have been achieved for hundreds of hours using
this method.
The refrigeration effect for the dilution refrigerator can be determined by
an energy balance around the mixing chamber
= )
where is the molar flow rate of
in the refrigerator, is the
molar enthalpy of the in the more dense phase leaving the mixing
chamber, and is the molar enthalpy of the in the less dense
phase entering the mixing chamber. Below temperatures of 40 mK the
enthalpies in units of J/mol can be approximated by
where and are the temperatures of the streams leaving and
entering the mixing chamber, respectively. Typical refrigeration effects
are on the order of 2-10μW.
The simplest dilution refrigerator may be operated as a single cycle. The
-
mixture is condensed into the mixing chamber, precooled, and
then circulated. Temperatures of 4-5 mK have been obtained with such a
device since heat is not added continuously with the liquid feed.
Cryogenics [email protected]
7 | P a g e
Pomeranchuk Cooling:
Satoh's Pomeranchuk cooling machine
A method for achieving temperatures around 2 mK by using the
properties of solid was originally proposed in 1950 by
Pomeranchuk and subsequently developed in 1969 by Wheatley. Pure
liquid will not solidify at 0 K unless a pressure of about 3.5 MPa is
applied to the liquid. The melting curve has a minimum at 0.3 K for
which the required pressure is close to 3 MPa. Below 0.3 K, the solid
has a high er molar entropy than the liquid. In fact, at 20 mK the molar
entropy of the liquid is only 1/7 that of the solid.
This implies that a substantial cooling effect could be gained by
adiabatic solidification under compression at these low temperatures.
Cryogenics [email protected]
8 | P a g e
However, for this process to work, the compression has to be performed
without frictional heating.
This frictional effect is so significant that at 8 mK, a 1 % conversion of
mechanical work into heat reduces the available refrigerating capacity to
zero.
A schematic of a Pomeranchuk cooling machine proposed by Satoh is
shown in Fig. A major problem in the operation of such cooling devices
on a large scale is the method of precooling . In Satoh's device, two
Leiden dilution refrigerators (LDR) use to precool the
. After
cooling the liquid below 300 mK the liquid is sealed into the
compressor cell and the bellows are used to mechanically and
adiabatically compress the cell.
Once the liquid is compressed up to the solid line, a "Pomeranchuk cell"
will form with liquid isolated from the surroundings by a solid shell. The
cooling effect can be realized by compressing this cell until it becomes
solid.
For the most efficient cooling effect, a limit must be set on the solid
content in the cell to avoid friction caused by crushing. The performance
of this cooling machine, however, is rather poor since a cooling effect is
produced that is about 1 % of the work input to the process.
Nevertheless, Pomeranchuk cooling provides a greater cooling effect
than that obtained from -
dilution systems. Its major
disadvantage is that it operates on a batch basis and requires
considerable time to precool the working fluid to low enough
temperatures.
Measurement Systems for low temperatures:
Accurate measurement of Thermophysical properties is very essential
for controlling the process parameters and efficient performance of the
individual pieces of equipment of the cryogenic process plants.
Measurement of temperature, pressure or vaccum level, liquid level, and
flow rate, is often needed for production, storage and transportation of
cryogenic fluids.
Temperature measurement at low temperatures:
Cryogenics [email protected]
9 | P a g e
Temperature is one property, which cannot be measured directly by
compairing it with a ‘standard’ as in the case of length, mass and time. It
is determined indirectly by measuring another temperature-dependent
property, such as
1. Length of thread of a liquid (e.g. mercury or alcohol) in a glass
capillary,
2. Electrical resistance of a metallic wire.
3. Pressure of a gas of known volume.
4. Pressure of a boiling liquid.
5. E.m.f generated between two dissimilar metals.
6. Speed of sound in gas.
7. Refractive index.
8. Magnetic susceptibility of a paramagnetic salt.
9. Thermal expansion of two different metals, etc.
For each such type, the values of the property at an interval of two fixed
temperatures, such as freezing point and boiling point of water must be
known in order to define the temperature scale. The smallest unit in the
scale is defined by dividing the difference in these two fixed points by a
fixed number, say 100, to obtain the value of a degree. However,
different devices may not show agreement at intermediate temperatures,
though they may show satisfactory agreement at the fixed temperatures,
as most properties do not vary linearly with temperature. The selection
of temperature measurement devices or thermometers often depends on
the range of operation of the device.
In some cases, there is no alternative but to select only one device
because of nonexistence of any choice. When more choices are
available, then the selection of the device is made based on the accuracy,
reproducibility, stability, simplicity, heat conduction, heat capacity,
sensitivity, cost and convenience. Sensitivity of measurement is the most
important criterion, which includes the other primary factors for
selection and needs a systematic analysis. The optimum choice,
however, depends on the requirement of the specific application and
available instrumentation.
Cryogenics [email protected]
10 | P a g e
The fundamental thermodynamic temperature scale is defined in terms
of the properties of the ideal gas, i.e., the working substance of a Carnot
engine and the two temperatures between which it is operating,
according to the second law of thermodynamics. This concept relates to
the efficiency of a Carnot engine, which depends on these two
temperatures only and is independent of the working substance.
Accordingly, the fundamental thermodynamic scale represents the
absolute scale or Kelvin scale. The numerical value of the unknown
temperature in this scale is conceptually obtained in terms of one fixed-
point temperature, say, the triple point of water, i.e., 273.16K, and the
ratio of the heat rejected to the heat absorbed by a Carnot engine
operating between these two temperatures. However, this scale is not
practically possible.
A practical and reproducible temperature scale is, however, accepted to
be one that closely approaches the absolute temperature scale, like the
one first introduced in 1927 as the International Temperature Scale (ITS-
27). This included six fixed points out of which only one, namely, the
normal boiling boiling point of oxygen is in the range of cryogenic
temperatures. In 1968, the International Practical Temperature Scale
(ITPS-68) extended the range of temperature scale to the triple point of
hydrogen (13.81K) and the values of the primary fixed points were
revised as those listed in table below.
Fixed Point Condition Temperature (K)
Gold point Normal melting point 1337.58
Silver point Normal melting point 1235.08
Zinc point Normal melting point 692.73
Steam point (water) Normal boiling point 373.15
Triple point (water) Standard triple point 273.16
Oxygen point Normal boiling point 90.188
Triple point (oxygen) Standard triple point 54.361
Neon point Normal boiling point 27.102
Hydrogen Normal boiling point 20.28
Hydrogen (at 25 torr) LP boiling point 17.042
Cryogenics [email protected]
11 | P a g e
Triplepoint(hydrogen) Standard triple point 13.81
The standard measuring instrument selected for more accurate
measurements between the triple point of hydrogen and the freezing
point of antimony (903.89 K) was a platinum resistance thermometer,
though a platinum resistance thermometer is more sensitive for
measurement of cryogenic temperatures.
Resistance Thermometers:
Metallic Resistance Thermometers:
Since the resistivity of pure metals varies with change in temperature,
metals have been used as simple and reliable temperature-measuring
devices. Many elements or compounds, however, are not suitable for use
in low-temperature resistance thermometry because they lack one or
more of the desirable properties of an ideal resistance thermometer.
These properties include the following:
1. A resistivity that varies linearly with temperature to simplify
interpolation.
2. High sensitivity.
3. High stability of resistance so that its calibration is retaiined over long
periods of time and is not affected by thermal cycling.
4. Capability of being mechanically worked.
Although a number of metals such as lead, nickel, copper or indium are
more or less suitable for resistance thermometry, platinum has come to
occupy a predominant position, partly because of its excellent
characteristics, such as chemical inertness and ease of fabrication, and
partly because of custom. Certain desirable features such as ready
availability in high purity and the existence of a large body of
knowledge about its behavior have come into being as its use grew and
have tended to perpetuate that use. Its sensitivity down to 20K and its
stability are excellent. Its principal disadvantages are (1) low resistivity,
(2) insensitive low about 10 K, and (3) a variation in the form of the
resistance– temperature relation from specimen to specimen below about
Cryogenics [email protected]
12 | P a g e
30 K. for platinum, the dependence of resistance on temperature is
expressed by the following relationship:
Where is the resistance at T°C, is the resistance at the ice point
273.16 K, and A, B, C are constants determined by calibration of the
device for three out of the remaining fixed points. Typical values of the
constants for a 25-ohm platinum resistance thermometer are: A = 3.946×
° ; B = -1.108 × ° ; and C = 3.33 × ° .
The sensitivity of a resistance thermometer is defined as (1/R)(dR/dT). It
is desirable that the temperature coefficient (dR/dT) of a metal be
significantly high to make it sensitive as an element to be used for a
resistance thermometer over the entire range of temperature. Also, the
purer is the metal, lower is its resistance, R, and so more is its sensitivity
with temperature. For example, a 25-ohm (at room temperature)
platinum resistance thermometer will have a resistance of less than 0.2
ohm at 20K, whereas its temperature coefficient at 20K is one-fifth of
that at room temperature. Accordingly, platinum resistance thermometer
has a higher sensitivity at 20K than at room temperature. However, the
instrument should be able to measure the very low resistance at the low
temperature with precision. Similarly, indium is also found to have good
sensitivity over the entire range from 4K to room temperature. Lead has
high temperature coefficient, but it becomes superconducting at 7.2K,
beyond which it cannot be used as a resistance thermometer.
The metal element for the resistance thermometer needs to be free from
any mechanical strain, as it increases the resistance. It is thus necessary
to mount the metallic wire on a support ensuring that mechanical and
thermal strains are eliminated. One method of mounting designed,
involves winding the metallic wire in the form of a helix and then
winding the helix around a notched mica support, as shown in fig. The
sensing element is annealed in order to remove any mechanical strains
and is then inserted in a platinum tube. The tube is filled with helium gas
at 30 – 40 torr at room temperature prior to sealing. Several variations in
Cryogenics [email protected]
13 | P a g e
encapsulation and shielding are available for commercial applications
with time constants in the range of 1.3 to 1.6 seconds.
Platinum resistance thermometers. ( Capsule, strain–free)
Thermocouples:
Thermocouples are temperature measuring devices commonly used for
measurement of cryogenic as well as for above-ambient temperatures.
Thermocouples are simple, easy to install, inexpensive and temperature-
sensitive having low heat capacity. It is used as a secondary standard at
cryogenic temperature regions till as low as 12K. A thermocouple
consists of two dissimilar wires connected to each other at one end
where the temperature needs to be measured. The temperature difference
between the joined and unjoined ends (at reference temperature)
generates a difference in electrical potential (e.m.f) which may be quite
small and requires a sensitive instrument such as a galvanometer,
potentiometer, standard cell, low temperature bath for reference point.
For example, for measurement of low temperature, copper-constantan
thermocouple is very common and it generates 40Mv per degree at room
Cryogenics [email protected]
14 | P a g e
temperature whereas 17μV at 90K (liquid oxygen temperature), 5μV at
20 K (liquid hydrogen temperature) and almost zero at absolute zero.
Only limitation in measuring a very low temperature using a
thermocouple is that the reference point should be at the boiling point of
liquid nitrogen or liquid hydrogen at a known pressure. Otherwise, an
e.m.f., of the order of 5mV, should be measured with an accuracy of
1μV in order to get an accuracy of 0.1K, in case an ice bath is used to
provide a reference temperature.
Thermocouple types as defined by the ASTM are discussed below.
Type E Thermocouples. The ASTM designation type E indicates a
thermocouple
pair consisting of a Ni-Cr alloy and a Cu-Ni alloy. This type of
thermocouple has the highest Seebeck coefficient, S, of the three ASTM
standard thermocouple types commonly used at low temperatures, types
E, K, and T. Also, both elements of this thermocouple have low thermal
conductivity, reasonable homogeneity, and corrosion resistance in moist
atmospheres. This type is the best thermocouple to use for temperatures
down to
about 40 K.
Type K Thermocouples. The ASTM designation type K indicates a
thermocouple pair consisting of a Ni-Cr alloy and a Ni-Al alloy. The
sensitivity is only about half that of the type E combination at 20 K
(4.1μV /K compared with 8.5μV/K). The negative element is a bit more
homogeneous than the EN element. Both materials have low thermal
conductivity and are corrosion resistant in moist atmospheres. Type K
thermocouples are recommended by the ASTM for continuous use at
temperatures within the range 3-1533 K in inert atmospheres.
Type T Thermocouples. The ASTM designation for type T indicates a
thermocouple pair consisting of Cu and Cu-Ni alloy. Type T is one of
the older, more popular combinations, and is the only one of the
standardized types for which limits of error below 273.15 K have been
Cryogenics [email protected]
15 | P a g e
established. Type T thermocouples are recommended by the ASTM for
use in the temperature range from 89 to 644 K in vacuum or in
oxidizing, reducing, or inert atmospheres.
Gold-Iron Alloy Thermocouples. The increasing use of liquid
hydrogen and liquid helium has created a demand for specialized
thermometry below 25 K. Ordinary thermocouple combinations are only
marginally acceptable because of their low sensitivity in this range.
Dilute alloys of noble metals and transition metals, however, have
relatively high temperature sensitivity below 25 K; Au-2.1 at. % Co is
perhaps the best known of this type. Unfortunately, this alloy is a
supersaturated solid solution. Powell et al. found that the thermoelectric
power decreases with time when stored at room temperature. This
gradual change is probably due to diffusion of cobalt atoms to grain
boundaries. Dilute alloys of iron in gold are metallurgically stable and
have extremely useful thermoelectric properties at very low
temperatures. A differential thermocouple made with Au---0.02, 0.03, or
0.07 at. % Fe as the negative element and copper, "normal" silver (Ag-
O.37 at. % Au), or KP as the positive element provides a usable
sensitivity, even below 4 K.
Thermistors: Thermistors (TM) are essentially thermally sensitive resistors made up
of metal oxides. Frequently used materials are nickel, manganese, and
cobalt oxides. The temperature-resistance relationship for this type of
resistor has a negative slope much like the carbon or germanium
resistors. These resistors are becoming increasingly popular in
measurement and control circuits because they are small with short
response times, typically are of high resistance, which reduces the
overall effect of lead resistances, and have temperature-resistance
characteristics that are dependent on materials and procedures that allow
thermometers to be developed that are particularly sensitive over limited
ranges of temperature. Reproducibilities have been investigated
experimentally by Sachse and determined to be about ± 30 mK after
Cryogenics [email protected]
16 | P a g e
cycling between room temperature and liquid oxygen. After 1000
cyclings, the error was on the order of tenths of degrees.
Gas Thermometry: If the equation of state, i.e., the function relating pressure, volume, and
temperature, of a gas is known, the gas can be used to measure
temperature. If one of the variables is held constant and a second is
measured, the third can be calculated from the equation of state. In
thermometry, it has been found most advantageous to hold the volume
of a fixed amount of gas constant and measure the pressure. The
temperature of the gas is then determined. When properly corrected,
such a constant-volume gas temperature very closely approximates an
ideal gas thermometer, and is used to fix points on the absolute
temperature scale.
One of the simplest forms of the constant-volume gas thermometer
involves a bulb, which is at the low temperature to be measured. This
bulb is connected by a fine capillary to a Bourdon-type pressure gauge
maintained at room temperature. If the volume of the capillary is
assumed to be negligible, and if the gas can be assumed to be ideal over
the temperature range to be measured, then the temperature in the bulb is
given by the relation
T =
[ ] (1)
where is the volume of the gas-thermometer bulb, the volume of
the Bourdon pressure gauge, the temperature at ambient conditions,
and p the pressure measured in the gauge in consistent units. This simple
device can be quite accurate at low temperatures since most of the gas
will be in the gas-thermometer bulb. At higher temperatures, however, a
substantial amount of the gas will be in the Bourdon gauge, making the
thermometer rather insensitive to T. If the ratio of / is made
sufficiently large, such a device can have an accuracy of ±0.05 K at
temperatures below 30 K. In the same vein, Scott has noted that, if room
temperature is taken at 300 K and the ratio of / is equal to 20, a
Cryogenics [email protected]
17 | P a g e
straightline p-T relation through 20 and 77 K will deviate from Eq. (1)
by less than 0.2 K maximum over the 20-90 K temperature range.
Equation (1) may be considered the first step in obtaining an accurate
temperature with a gas thermometer, in that it corrects for the "nuisance
volume" or the volume of gas that is not at the temperature being
measured.
However, in order to obtain accurate results in gas thermometry, it is
also necessary to correct for the imperfection of the gas, the change in
volume of the bulk with changing temperature, variations in the amount
of gas adsorbed on the walls of the gas-thermometer bulb, and the
thermomolecular pressure gradient encountered at extremely low
pressures.
One engineering application of gas thermometers is achieved by
deliberately increasing the volume of gas maintained at ambient
temperature. This makes the gas thermometer nonlinear, increasing the
sensitivity at low temperatures, and compressing the scale at high
temperatures. At low temperatures, most of the gas in this arrangement
is at the measuring temperature and contributes to the measurement. At
higher temperatures, most of the gas is in the room temperature
reservoir, and does not contribute to the measurement. By using a
Bourdon gauge to show the pressure in this device, a simple gas
thermometer is achieved, which is useful for such purposes as
monitoring the cooldown of cryogenic apparatus. Accordingly, for
engineering applications, gas thermometry is recommended for
indicating the approximate temperature, or temperature trend. For
precision temperature measurements, the corrections are usually too
demanding for common use in the field.
Liquid level sensors:
Measurement of liquid level is important to know the amount of the
liquid remaining within the cryogenic storage vessel or dewar flask at
any instant of time. Liquid level measurement may pose problem, as it
has to be carried out under totally closed and insulated conditions. It can
be performed by a variety of methods, the simplest being the float and a
Cryogenics [email protected]
18 | P a g e
long stem with a pointer. The float may be used along with a mechanical
–electrical transducer for remote indication.