Upload
kamal-abdurrahman
View
78
Download
5
Tags:
Embed Size (px)
Citation preview
Kamal Abdurahman
Group:B
2/5/2015
2015Multi Desaturator Cell
Supervised By : Mr.Ail Kamal Mr.Hiwa Mss.Sana
FACULTY OF ENGINEERING
SCHOOL OF PETROLEUM AND CHEMICAL
RESERVOIR ENGIEERING LAB
Contents
1 -Aim.
2-Inroduction.
3 -Theory.
4 -Apparatus.
5 -Procedure.
6-Calculation.
7 -discussion .
8 -references.
Aim of experiment:
In this test we are determined the curves of capillary pressure
Versus saturation water.
INTRODUCTION
The capillary pressure desaturation cell is dedicated to enable
generation of air-brine capillary curves on core samples.
Capillary curve is the relationship between pressure applied
and pressure stabilized, and water content in the core samples.
The equipment is mainly composed of a console (on left on
first page picture) and a sample extractor (blue vessel on right
on first page picture).
The console controls the air pressure supplied to the vessel. It
is possible to humidify the air used in the process.
In the extractor, core samples are installed on a capillary
(ceramic) pressure plate.
TheoryThe principle involved in the operation of the Capillary pressure
de-saturation cell is that water is removed from sample by
suction wherein a porous ceramic wall serves as a connecting
link and at the same time a means of maintaining a pressure
difference between the liquid phase of the water in the soil and
the water at lower pressure on the opposite side of the wall.
The illustration in Fig. 3 shows a magnified view of soil particles
in contact with the porous ceramic plate inside the Pressure
Extractor during an extraction run.
A wetted porous ceramic plate is backed by a fine mesh screen
which also provides a passage way for the extracted solution,
and is futher sealed by a rubber membrane backing.
The rock samples initially saturated with brine and weighted
individually are installed in the vessel on the ceramic plate
preliminary wetted to ensure good capillary contact.
After bolting the Extractor lid onto the Extractor, air pressure
may be increased to the value of the test (0.1 to 1500 kPa). As
soon as air pressure inside the chamber is raised above
atmospheric pressure, the higher pressure inside the chamber
forces excess water through the microscopic pores in the
ceramic plate. The high pressure air, however, will not flow
through the pores since they are filled with water and the
surface tension of the water at the gas-liquid interface at each
of the pores supports the pressure much the same, When the
air pressure is increased inside the Extractor, the radius of
curvature of this interface decreases (Fig.4). However, the
water films will not break and let air pass throughout the whole
pressure range of the Extractor.
At any given air pressure in the chamber, soil moisture will flow
from around each of the soil particles and out through the
ceramic plate until such time as the effective curvature of the
water films throughout the soil are the same as at the pores in
the membrane. When this occurs, equilibrium is reached and
the flow of moisture ceases. When the air pressure in the
Extractor is increased, flow of soil moisture from the samples
starts again and continues until a new equilibrium is reached.
At equilibrium, there is an exact but opposite relationship
between the air pressure (positive force) in the Extractor and
the soil suction (negative force).
Water content by weight or by volume can be determined for
the sample that was at equilibrium with the pressure in the
Extractor.
According to the pressure value and to the sample
characteristics, it spends from 1 hour to several days. Then, the
operator bleeds off pressure, open the vessel to determine
water production by weight.
Finally, the operator plots the water (or brine) content in
core versus the applied (capillary) pressure. The water
content is usually expressed in % of pore volume of the
sample.
Apparatus of experiment
Procedure
Test Preparation (humidifier filling)
The air in the extractor should be humidified to prevent sample
drying and cracking of the ceramic plate.
1. Unscrew four ¼`` fittings as shown in the nearby picture
2. Unscrew 3xCHC M6 screws with allen key provided
3. Open the top lid of the humidifier tank by hand. Maybe you
need pull off strongly to remove the top lid, due to the mounted
oring.
4. Fill 2/3 of height with water and tight the top lid of the
humidifier tank as was showing before
5. Connect the ¼’’ lines before the experience
6. Dry sample in oven not over 82°C (180 °F) so as not remove
water of hydratation.
7. Leach if necessary to remove salt:
a) Leach sample by flowing fresh or distilled water trough until
all salt is removed. Salt concentration can be detected in
effluent water by resistivity.
b) Do not leach if sample is suspected to contain clay, shale or
anhydrite.
8. Run air permeability measurements, and select sample to
cover desired permeability range.
9. Obtain dry weight of sample.
10. Place sample in saturator and evacuate for at least four
hours, longer evacuation will be necessary for tight samples.
11. Pressurize sample in saturator with degassed evacuated
brine at 2,000 psi. Unless otherwise specified, use brine of
91,000 ppm NaCl. As a general rule, allow sample to remain at
2000 psi in brine for a period of 8 to 16 hours, depending on the
permeability of the sample.
12. Remove sample from depressurized saturator, wipe excess
brine from sample with hand only, taking care to rub off any of
the sample grains. Determine saturated weight.
Operation
It is assumed that:
- One type of test is selected:
a. maximum desaturation and
b. desaturation step by step
- Flush the ceramic with brine to be used before placing sample
in desaturator.
- The core sample (s) is loaded in the single / multi desaturator
cell and prepared for the test.
- The humidifier is filled with 2/3 of water
- Air connection is done
- All fittings are tight
These detailed procedure and calculation instruction are
based on Dr R. MONICARD work.
a. For a maximum desaturation:
1. Close the low pressure gas regulator valve
2) Open all other valves (HV01, HV02 and HV03)
3) Extract fresh sample by centrifuge or Dean Stark method.
4) Set the air supply. Close low pressure gas regulator and
open the high pressure gas regulator to perform maximum
desaturation experience.
5) Turn clockwise progressively the regulator knob until the cell
pressure display monitors the required operating pressure.
6) Start with a capillary pressure step of 1 psi for a period of at
least 48 hours. Open gently the valve HV04 to pressurize the
desaturator cell.
7) After this time length in desaturator cell, isolate the pressure
regulator by setting HV02 in mid position.
8) Open gently the fitting on desaturator cell to depressurize the
vessel.
9) Open the desaturator, remove sample.
b. Step by step:
Repeat the previous steps, using different pressures and using
the same 48 hours interval (remember you have a maximum
ceramic pressure plate of 250psi). This time interval is the
minimum required for searching equilibrium. Tighter samples
may require a longer time for reaching equilibrium.ecord
weight.
6.5 Shutoff procedure:
After the last step:
1. Isolate the air supply.
2. Isolate the pressure regulators by closing HV02 (in the mid
position).
3. Turn anticlockwise both regulators knob.
4. Open gently the valve HV03 to depressurize the vessel.
5. Open HV05 to release gas
6. Open desaturator cell.
7. Place the sample(s) in a tare container(s) and dry in an oven
temperature of 71 - 82°C (160 - 180 °F) for a period of at least
2 4 hours, and again longer for tighter samples. The use of a
tare container in the drying step will eliminate the grain loss.
The tare will retain for final weight measurement any particle
which is displaced from the sample during drying.
8. Remove sample and tare from oven and measure final dry
weight of sample after it reaches room temperature in
desiccator.
9. Run gas expansion porosity. Unless the samples are
leached, the salt from the water displaced during the capillary
test will be left in the sample, and must be considered as part of
the porosity in the calculation.
5. Drying the Cell after the Run
When a Pressure Cell is to be dried for storage after a run, it is
very important to keep evaporation deposits on the surface
to a minimum.
To do this, cover the surface of the ceramic plate with a thin
layer of fine dry soil and allow it to set for several days until dry.
After it is dry, remove the soil and store the cell.
This procedure forms the evaporation deposits on the soil
particles rather than on the surface of the cell.
After a period of time, if the flow rate of a Cell drops due to
deposits, they should be replaced.
Discussion
-In this test the main effect is time because there are many
cores samples need to a many weeks to saturated.
-The ceramics plate resistant to the 3bar pressure we are must
be takes less than 3bar pressure if higher than 3bar this plate
cracked or broken.
-In this test the De-saturation of the core decrease when you
have a capillary pressure, then the saturation decrease when
you takes another capillary pressure .
Pc
Sw
-If the core has low permeability to saturated this core need
usually one week, If the core has high permeability to saturated
this core need to two or three or more days.
Reference
http://www.malvern.com/en/products/measurement-type/
desaturation/default.aspx
Jiao, D. and M.M. Sharma, “desaturation,” Journal of Colloidal
and Interfacial Science, 1994. 162:p. 454-462.
http://www.glossary. desaturation .slb.com/en/Terms/m/
mudcake.aspx
Fisk, J.V., and Jamison, D.E., SPE Reservoir Engineering,
December 1989, pp. 341-46.