40
Field-Checking and Maintenance of HVAC Humidity and CO 2 Sensors Webinar – October 5th, 2016 Lars Stormbom & Maria Uusimaa

Field Checking and Maintenance of HVAC Humidity and CO2 Sensors

Embed Size (px)

Citation preview

Page 1: Field Checking and Maintenance of HVAC Humidity and CO2 Sensors

Field-Checking and Maintenance of HVAC Humidity and CO2 SensorsWebinar – October 5th, 2016Lars Stormbom & Maria Uusimaa

Page 2: Field Checking and Maintenance of HVAC Humidity and CO2 Sensors

Page © Vaisala

Meet the presenters

2016-10-102

Lars Stormbom. Product Manager at Vaisala, with over 30 years of experience of industrial and HVAC measurements.

Maria Uusimaa. Product Manager for Vaisala’s handhelds and industrial transmitters. She has over 10 years of experience in CO2 measurements in various measurement applications.

Page 3: Field Checking and Maintenance of HVAC Humidity and CO2 Sensors

Page © Vaisala

Covered Topics

The importance of field-calibration in HVACWhat causes RH and CO2 sensors to drift?How to check RH and CO2 measurement reliably on the fieldField-checking options

2016-10-103

Page 4: Field Checking and Maintenance of HVAC Humidity and CO2 Sensors

Page © Vaisala

Why calibrate in HVAC?

Why is calibration important? Anything that is worth measuring is worth measuring right. All electronics and sensors are subject to drift over time.

What kind of energy savings can be achieved with calibration? According to the REHVA ETIAQ project DCV can save 20-50 %

energy in public buildingsWith poorly functioning sensors you may loose part or all of that

potential saving

2016-10-104

Page 5: Field Checking and Maintenance of HVAC Humidity and CO2 Sensors

Page © Vaisala 2016-10-105

Field Calibration

Page 6: Field Checking and Maintenance of HVAC Humidity and CO2 Sensors

Page © Vaisala

Calibration or adjustment?

2016-10-106

Calibration = comparing the output of a measurement instrument against a reference instrument and merely reporting the result

Adjustment = changing the output to correspond to the output of a reference instrument

Page 7: Field Checking and Maintenance of HVAC Humidity and CO2 Sensors

Page © Vaisala 2016-10-10

When to calibrate? Depends on the accuracy

need of the measurement and the environment. No single correct answer! Are there stability

specifications? Do you have a back record of

drift values? Start with a shorter inspection

interval – lengthen gradually

Page 8: Field Checking and Maintenance of HVAC Humidity and CO2 Sensors

Page © Vaisala

Field check vs adjustmentA good strategy may be to first perform field checks against a

reference. This should be relatively fast to perform. If the observed difference is larger than a trigger value, for

instance: <2 %RH difference: Do nothing 2 ... 4 %RH difference: Make a one-point field adjustment >4 %RH difference: Send device for service/ Change the

measurement module

2016-10-108

Page 9: Field Checking and Maintenance of HVAC Humidity and CO2 Sensors

Page © Vaisala

One-point or multipoint?One point adjustment:

The measured condition varies only 20% of the measurement range Usually OK for HVAC

Multipoint adjustment: If the whole measurement range is used Usually requires laboratory adjustment If a large correction is indicated in the filed check

2016-10-109

Page 10: Field Checking and Maintenance of HVAC Humidity and CO2 Sensors

Page © Vaisala

You choose the maintenance strategy –we support it

2016-10-1010

Do nothing because the specified stability is good enough

OR

Use convenient field checking and adjustment options

Buy field exchangeable measurement modules – with certificate!

Page 11: Field Checking and Maintenance of HVAC Humidity and CO2 Sensors

Page © Vaisala 2016-10-1011

Relative Humidity

Page 12: Field Checking and Maintenance of HVAC Humidity and CO2 Sensors

Page © Vaisala

How can we measure water vapor?

2016-10-1012

Water is unique among common ambient gases

N2 CO2H2O

It has a positively and a negatively charged end

+--

Page 13: Field Checking and Maintenance of HVAC Humidity and CO2 Sensors

Page © Vaisala 2016-10-1013

Solution: capacitive measurement

Water molecules align with alternating electrical field between two electrodes.

- -+

- -+

Energy stored and increased capacitance

Electrodes

Page 14: Field Checking and Maintenance of HVAC Humidity and CO2 Sensors

Page © Vaisala 2016-10-1014

With a polymer between the electrodes we can multiply the number of water molecules and increase the sensitivity

air Polymer

Use a polymer film to increase sensitivity

Page 15: Field Checking and Maintenance of HVAC Humidity and CO2 Sensors

Page © Vaisala

What can cause drift of RH measurement?

2016-10-1015

• Cleaning• Detergents, floor

wax

• New materials• Chipboards• Paint

• Stored materials• Packaging materials• Truck exhausts

A humidity sensor must breath the ambient air – it can´t be hermetic like a temperature sensor

Page 16: Field Checking and Maintenance of HVAC Humidity and CO2 Sensors

Page © Vaisala

What does humidity sensor drift look like?

2016-10-1016

Page 17: Field Checking and Maintenance of HVAC Humidity and CO2 Sensors

Page © Vaisala 2016-10-1017

Vaisala HUMICAP®

Page 18: Field Checking and Maintenance of HVAC Humidity and CO2 Sensors

Page © Vaisala

The HUMICAP® Advantage

2016-10-1018

Page 19: Field Checking and Maintenance of HVAC Humidity and CO2 Sensors

Page © Vaisala

Vaisala HUMICAP® Technology

We use in-house synthesized polymers for best stabilityThe HUMICAP® structure with a porous upper electrode

makes the sensor insensitive to chemicals as well as dust and dirt

2016-10-1019

Page 20: Field Checking and Maintenance of HVAC Humidity and CO2 Sensors

Page © Vaisala

Humidity calibration options

Page 21: Field Checking and Maintenance of HVAC Humidity and CO2 Sensors

Page © Vaisala

Calibration Options for RH

Reference instrumentUsing salt solution calibrator - HMK15Exchange of module and probesSending the product to Vaisala service

2016-10-1021

Page 22: Field Checking and Maintenance of HVAC Humidity and CO2 Sensors

Page © Vaisala

Using a handheld device Handheld devices are especially good for quick pass/fail checks, but can

with a bit of care be used also for one-point adjustments in the field. Allow enough time for the temperature to stabilize. Remember that an 1°C

temperature difference can mean a 6 %RH difference. Graphical displays are good tools to check stabilization. Avoid heating the handheld sensor with your hand. Avoid breathing on the sensor and the device you are checking. Some wall sensors give off a lot of heat – keep the handheld below a wall

sensor. Get close enough to the checked device – there are significant temperature

gradients in most rooms! For duct sensors it is good to plan for field checking in advance: Make a

sealed hole for the handheld probe when installing the duct sensor.

2016-10-1022

Page 23: Field Checking and Maintenance of HVAC Humidity and CO2 Sensors

Page © Vaisala

Using salt calibrators Can be used for 2 ... 3 point calibrations at room temperature for applications

where accuracy is needed over a larger humidity range. A fixed set of humidity values are available (e.g. 11%, 33%, 54%, 75%, 97%) Can be used with an accredited probe as reference if needed Suitable only for probe type sensors Relatively straightforward to use When properly used gives highly repetable and reliable results Sensors are usually removed from the site a calibrated in a lab or other suitable

space with relatively stable temperature conditions – not really field calibration? Requires ~30 min per humidity point depending on the calibrated sensor Bringing the sensor into the same space as the calibrator well in advance for

temperature stabilization speeds up the process. Again: it is good practice to monitor stabilization of the readings!

2016-10-1023

Page 24: Field Checking and Maintenance of HVAC Humidity and CO2 Sensors

Page © Vaisala

Exchaging probes and modules

2016-10-1024

Exchangeable measurement modules or probes can be an easy way to maintain measurement accuracy and traceability, especially if you have only a few units Make sure you get a calibration certificate with the new probe/module One drawback is that there is no information on the performance of the

old measurement. Has there been drift? How much? Is the exchange interval suitable? May be combined with spot-checking with a handheld to determine if

some units need changing

Page 25: Field Checking and Maintenance of HVAC Humidity and CO2 Sensors

Page © Vaisala 2016-10-1025

Carbon Dioxide

Page 26: Field Checking and Maintenance of HVAC Humidity and CO2 Sensors

Page © Vaisala

CO2 Technology in HVACOptical (infra-red) measurement technologyLight source, filter and detector CO2 measurement at 4.26 µmAttenuation in intensity gives the CO2 concentration (e.g. 1000

ppmCO2)Changes in sensor structure cause drift: Light source intensity change Dirt in optical cuvette Any mechanical changes!

2016-10-1026

One IR source, one filter CO2 absorption wavelength

Page 27: Field Checking and Maintenance of HVAC Humidity and CO2 Sensors

Page © Vaisala

Ways to Overcome Drift: Adding Components

Two light sources

OR

Two filters and detectors

2016-10-1027

Two IR sources, one filter reference source

One IR source, two filters CO2 and reference wavelengths

Page 28: Field Checking and Maintenance of HVAC Humidity and CO2 Sensors

Page © Vaisala

Ways to Overcome Drift: Using Software Algorithms

ABC (Automatic background calibration) Zeroing against an assumed

background CO2 concentration (~400 ppmCO2)

2016-10-1028

One IR source, one filter CO2 absorption wavelength

Page 29: Field Checking and Maintenance of HVAC Humidity and CO2 Sensors

Page © Vaisala

Vaisala CARBOCAP® - It´s unique

2016-10-1029

Vaisala CARBOCAP® sensorOne IR source, one tunable filter CO2 and reference wavelengths = true referencing!

Page 30: Field Checking and Maintenance of HVAC Humidity and CO2 Sensors

Field calibration options for CO2

Page 31: Field Checking and Maintenance of HVAC Humidity and CO2 Sensors

Page © Vaisala

Field calibration options for CO2

Reference gasReference instrument – a handheld

Exchange of modules or probesSending the product to Vaisala service center

2016-10-1031

Page 32: Field Checking and Maintenance of HVAC Humidity and CO2 Sensors

Page © Vaisala

Making a good CO2 measurement in the fieldChoose a reference – either reference

gas or a reference instrumentAllow time for calibrationCalibration is best done in a stable

environmentNote temperature and pressure settings,

compensate if possible!

2016-10-1032

Page 33: Field Checking and Maintenance of HVAC Humidity and CO2 Sensors

Page © Vaisala

CO2 field adjustment using calibration gas

CO2 adjustment against high accuracy bottled gasAllows multipoint calibration!

With gas bottle: 0.3 ... 0.5 l/min gas flow to gas port approximately 2 minutes of stabilization time

Vaisala does not sell gas bottles

2016-10-1033

Page 34: Field Checking and Maintenance of HVAC Humidity and CO2 Sensors

Page © Vaisala

CO2 field adjustment

2016-10-1034

One good solution is using Linde gas Ecocyl

Integrated flow adjustment with preset flow rates: 0.25, 0.3, 0.5, 1 l/min On/OFF valve Pressure meter Light-weight aluminiun cylinder contains 150 l gas

with ±0.5% accuracy specs. Good for about 75 adjustments with 2 min stabilization time at 0.5 l/min

Page 35: Field Checking and Maintenance of HVAC Humidity and CO2 Sensors

Page © Vaisala

Reference instrument – a handheldCalibration against a freshly calibrated

hand-held deviceEasy to use!Vaisala CARBOCAP® Carbon Dioxide

Meter GM70 is a good choice for CO2 field checks Add HM70 probe for simultaneous humidity

and temperature check

2016-10-1035

Page 36: Field Checking and Maintenance of HVAC Humidity and CO2 Sensors

Page © Vaisala

Exchange of modules

2016-10-1036

Page 37: Field Checking and Maintenance of HVAC Humidity and CO2 Sensors

Page © Vaisala

Calibration orders via:http://store.vaisala.com (US & Europe)or www.vaisala.com/calibration

2016-10-1037

We are happy to calibrate Vaisala instruments for you. Information at:www.vaisala.com/calibration

Page 38: Field Checking and Maintenance of HVAC Humidity and CO2 Sensors

Page © Vaisala

Vaisala Service Agreements

2016-10-1038

Page 39: Field Checking and Maintenance of HVAC Humidity and CO2 Sensors

Page © Vaisala

Summary

Define the approach for field-calibration based on your measurement needs!There are plenty of easy-to-use options for Vaisala HVAC

products

2016-10-1039

Page 40: Field Checking and Maintenance of HVAC Humidity and CO2 Sensors

Thank You

Download the HVAC eBook, see videos or get other useful information at:www.vaisala.com/hvac