SERVO-s Ventilator SystemProduct End-of-Life Disassembly Instructions
SERVO-s Ventilator System Product End-of-Life Disassembly Instructions
2 MTS-07-0010
Important• The SERVO-s Ventilator System is a medical
device used in hospital environments. The unitmust be properly cleaned before disassembling.
Proper cleaning of the unit before Product End-of-Life disassembly must be performed by the formerowner (hospital).
• Special waste. This product containselectronic and electrical components.Discard disposable, replaced and left-over parts in accordance withappropriate industrial and environmentalstandards.
• Recycling. Worn-out batteries must berecycled or disposed of properly inaccordance with appropriate industrialand environmental standards.
Environmental declarationComponents with special environmental concern
Components listed below shall be disposed of inaccordance with appropriate industrial andenvironmental standards.
PC boards (Printed circuit boards)
• PC 1771 Control, including Lithium battery
• PC 1772 Monitoring, including Lithium battery
• PC 1777 Panel including Backlight Inverter
• PC 1781 Pressure transducer, 2 each
• PC 1784 Expiratory channel
• PC 1785 Expiratory channel connector
• PC 1786 Expiratory channel cassette
• PC 1860 Main back-plane
• PC 1861 Pneumatic back-plane
• PC 1862 DC/DC & Standard connectors
• PC 1863 Power control
Other electronics
• TFT assembly including backlight
• Touch screen (glass)
• O2 cell, containing caustic lime and lead (Pb)
• O2 Sensor, containing PC boards
• Gas module Air, containing multiple PC boards
• Gas module O2, containing multiple PC boards
• AC/DC Converter, containing PC boards
• Expiratory cassette
• Expiratory valve coil
• Safety valve pull magnet
• Battery modules Nickel-Metal Hydride
Construction materials
The construction materials used in SERVO-s in % ofthe total weight.
Metal – total 72.5%
• Aluminium 70%
• Steel, zink, brass 2.5%
Polymeric material – total 8%
• PA (Polyamide)
• POM (Polyoxymethylene)
• SI (Silicone)
• TPE (Thermoplastic elastomer)
• PUR (Polyurethane)
• ABS (Acrylicnitrilebutadienstyrene)
• EPDM (Ethylenepropylenedienemonomer)
• PTFE (Polytetrafluoroethylene)
• FPM (Fluororubber)
• NBR (Nitrilerubber)
• PP (Polypropylene)
• PVC (Polyvinyl chloride)
• PS (Polystyrene)
Electronics – total 19.5%
• Battery modules Nickel-Metal Hydride
• Printed circuit boards, cables etc.
Others – very small amounts
• Filter paper of fibre glass
Product End-of-Life Disassembly Instructions SERVO-s Ventilator System
MTS-07-0010 3
To open:1. Release screw.
2. Pull out cover.
3. Press latch and pull off cover.
4. Remove screws.
5. Lift off User Interface.
�
�
�
�
�
SERVO-s Ventilator System Product End-of-Life Disassembly Instructions
4 MTS-07-0010
To open:1. Disconnect cable.
2. Remove six screws and separate the unit.
�
Other electronicsPC board
GlassPC board
� �
Product End-of-Life Disassembly Instructions SERVO-s Ventilator System
MTS-07-0010 5
To disassemble:1. Remove screws.
�
Otherelectronics
PC board
�
�
�
�
SERVO-s Ventilator System Product End-of-Life Disassembly Instructions
6 MTS-07-0010
To disassemble:1. Remove screws.
2. Release screws.
3. Pull off cover
Nickel-MetalHydride battery
Lithium batteryon PC boards
PC board
PC board
{Other electronics
Other electronics
Sealed unit,containscaustic limeand lead
�
�
�
Otherelectronics
�
�
Product End-of-Life Disassembly Instructions SERVO-s Ventilator System
MTS-07-0010 7
PC board
Otherelectronics
Otherelectronics
To disassemble:1. Remove screws.
�
PC board
�
PC board
�
�
PC board
Maquet Critical Care ABSE-171 95 Solna, SwedenPhone: +46 8 730 73 00www.maquet.com/criticalcare
For local contact:Please visit our websitewww.maquet.com
GETINGE GROUP is a leading global provider of equipment and systems that contribute to quality enhancement and cost efficiency within healthcare and life sciences. Equipment, services and tech-nologies are supplied under the brands ARJO for patient hygiene, patient handling and wound care. GETINGE for infection control and prevention within healthcare and life science and MAQUET for Surgical Workplaces, Cardiopulmonary and Critical Care.
© M
aque
t C
ritic
al C
are
AB
, 20
07.
All
right
s re
serv
ed.
MA
QU
ET
rese
rves
the
rig
ht t
o m
odify
the
des
ign
and
sp
ecifi
catio
ns c
onta
ined
her
ein
with
out
prio
r no
tice.
Doc
umen
t N
o.:
MTS
-07-
0010
. M
ay 2
007.
Pric
e gr
oup:
1. P
rinte
d in
Sw
eden
.
PC board
Otherelectronics
PC board
To disassemble:1. Remove screws.
�
�