Mass spectrometry

Embed Size (px)

Citation preview

Page 1: Mass spectrometry

LOGO

Mass Spectrometry

Mussarat Jabeen

Page 2: Mass spectrometry

Mass Spectrometry

•Powerful analytical technique•Smallest scale•Destructive technique•Useful for identification of species

According to the IUPAC (International Union of Pure and Applied Chemistry), it is the branch of science dealing with all aspects of mass spectroscopes and results obtained with these instruments.

Page 3: Mass spectrometry

Mass Spectrometry

Brief History of Mass Spectrometry

Nobel prize pioneers

Mass spectrometer

Contents

Structural analysis and Fragmentation Patterns

interpretation of mass spectrum

Applications of mass spectrometry

Page 4: Mass spectrometry

Mass Spectrometry

1897

1919

1934

1966

J.J. Thomson. Discovered electrons by cathode rays experiment. Nobel prize in 1906.

Francis Aston recognized 1st mass spectrometer and measure z/m of ionic compounds.

First double focusing magnetic analyzer was invented by Johnson and Neil.

Munson and Field described chemical ionization.

Brief History of Mass Spectrometry

Page 5: Mass spectrometry

Mass Spectrometry

1968

1975

1985

1989

Electrospray Ionization was invented by Dole, Mack and friends.

Atmospheric Pressure Chemical Ionization (APCI) was developed by Carroll and others.

F. Hillenkamp, M.Karas and co-workers describe and coin the term matrix assisted laser desorption ionization (MALDI).

w. Paul discovered the ion trap technique.

Page 6: Mass spectrometry

Mass Spectrometry

Joseph John Thomson1906 Nobel Prize for Physics(theoretical and experimental investigations on the conduction of electricity by gases)

Francis William Aston1922 Nobel Prize for Chemistry(mass spectrograph, of isotopes, in a large number of non-radioactive elements)

Wolfgang Paul1989 Nobel Prize for Physics(for the development of the ion trap technique)

John Bennet Fenn2002 Nobel Prize for Chemistry(for the development of Soft Desorption ionization Method)

Koichi Tanaka2002 Nobel Prize for Chemistry(mass spectrometric analyses of biological macromolecules)

Nobel prize pioneers

Page 7: Mass spectrometry

Mass Spectrometry

Mass spectrometer

Page 8: Mass spectrometry

Mass Spectrometry

Understanding Mass Spectrometry

In a mass spectrometer, the same thing is happening, except it's atoms and molecules that are being deflected, and it's electric or magnetic fields causing the deflection. It's also happening in a cabinet that can be as small as a microwave or as large as a chest freezer.

Page 9: Mass spectrometry

Mass Spectrometry

Mass spectrometer is similar to a prism.

In the prism, light is separated into its component wavelengths which are then detected with an optical receptor, such as visualization. Similarly, in a mass spectrometer the generated ions are separated in the mass analyzer, digitized and detected by an ion detector.

Page 10: Mass spectrometry

Mass Spectrometry

Basic Components of Mass Spectrometer

Four basic components• Sample inlet• Ionization source• Mass analyzer• Ion detector

Page 11: Mass spectrometry

Mass spectrometer

Sample Introduction Techniques

Initial pressure of sample is 760 mmHg or ~10-6 torr

• Direct Insertion (commonly used in MALDI)

• Direct infusion or injection (commonly used in ESI)

Two techniques

Page 12: Mass spectrometry

Mass spectrometer

Direct Insertion sample introduction technique

very simple techniqueSample is placed on a prob and inserted into ionization source and then subjected to any number of desorption processes, such as laser desorption or direct heating, to facilitate vaporization and ionization.

Page 13: Mass spectrometry

Mass Spectrometry

Direct infusion or injection sample introduction technique

Frequently used due to high efficientlyUsed in coupling techniques like GC-MS and HPLC-MS

Page 14: Mass spectrometry

Mass Spectrometry

Ionization Methods used in Mass spectrometry

Commonly used

• Protonation• Deprotonation• Cationization• Transfer of a charged molecule to the gas phase• Electron ejection• Electron capture

Page 15: Mass spectrometry

Mass Spectrometry

Protonation

Formation of positive ions by the addition of a protonUsed for basic compounds like amines, peptidesUsed in MALDI, APCI and ESI

Page 16: Mass spectrometry

Mass Spectrometry

Deprotonation

Give net negative charge of 1- by removal of one proton Used for acidic species like phenols, carboxylic acid, sulfonic acid etc.Used in MALDI, APCI and ESI

Page 17: Mass spectrometry

Mass Spectrometry

Cationization

produces a charged complex by non-covalently adding a positively charged ion like alkali metal ion or ammonium ion to a neutral molecule.Used for CarbohydratesUsed in MALDI, APCI and ESI

Page 18: Mass spectrometry

Mass Spectrometry

Transfer of a charged molecule to the gas phase

Cation from solution to gasUsed in MALDI or ESI

Page 19: Mass spectrometry

Mass Spectrometry

Electron ejection

Electron is ejected to give positive ionUsually for non-polar compounds with low molecular weights like anthracene.

Page 20: Mass spectrometry

Mass Spectrometry

Electron capture

a net negative charge of 1- is achieved with the absorption or capture of an electron.Used for halogenated compounds

Page 21: Mass spectrometry

Mass Spectrometry

Ionization Sources in mass spectrometer

• Electrospray Ionization (ESI)• Nanoelectrospray Ionization (NanoESI)• Atmospheric Pressure Chemical Ionization (APCI)• Atmospheric pressure photoionization (APPI)• Matrix-assisted laser desorption/ionization mass

spectrometry (MALDI-MS) • Fast Atom Bombardment (FAB)• Electron Ionization (EI)• Chemical Ionization (CI) • Thermal ionization (TI)

Page 22: Mass spectrometry

Mass Spectrometry

Ionization Sources

Hard ionization sources Soft ionization sources

leave excess energy in molecule and produced stable fragments which is not further fragarmented

Little excess energy in molecule and produced unstable fragments which are again fragmented.

Types of

Page 23: Mass spectrometry

Mass Spectrometry

Electrospray Ionization (ESI)

For example peptides, proteins, carbohydrates, small oligonucleotides, synthetic polymers, and lipids

The sample solution is sprayed from a region of the strong electric field at the tip of a metal nozzle maintained at a potential of anywhere from 700 V to 5000 V. The nozzle (or needle) to which the potential is applied serves to disperse the solution into a fine spray of charged droplets. Either dry gas, heat, or both are applied to the droplets at atmospheric pressure thus causing the solvent to evaporate from each droplet

Page 24: Mass spectrometry

Mass Spectrometry

Nanoelectrospray Ionization (NanoESI)

where the spray needle has been made very small and is positioned close to the entrance to the mass analyzer. The end result of this rather simple adjustment is increased efficiency, which includes a reduction in the amount of sample needed.

• Very sensitive • very low flow rates• Very small droplet size (~5µ)

Page 25: Mass spectrometry

Mass Spectrometry

Atmospheric Pressure Chemical Ionization (APCI)

the liquid effluent of APCI is introduced directly into the ionization source. However, the similarity stops there. The droplets are not charged and the APCI source contains a heated vaporizer, which facilitates rapid desolvation/vaporization of the droplets. Vaporized sample molecules are carried through an ion-molecule reaction region at atmospheric pressure.

Page 26: Mass spectrometry

Mass Spectrometry

Atmospheric pressure photoionization (APPI)

it generates ions directly from solution with relatively low background and is capable of analyzing relatively nonpolar compounds.

APPI vaporized sample passes through ultra-violet light.APPI is much more sensitive than ESI or APCI.

Page 27: Mass spectrometry

Mass Spectrometry

Matrix-assisted laser desorption/ionization mass spectrometry (MALDI-MS)

the analyte is first co-crystallized with a large molar excess of a matrix compound, usually a UV-absorbing weak organic acid. Irradiation of this analyte-matrix mixture by a laser results in the vaporization of the matrix, which carries the analyte with it. The matrix plays a key role in this technique. The co-crystallized sample molecules also vaporize, but without having to directly absorb energy from the laser. Molecules sensitive to the laser light are therefore protected from direct UV laser excitation.

Page 28: Mass spectrometry

Mass Spectrometry

Fast Atom Bombardment (FAB)Immobilized matrix is bombarded with a fast beam of Argon or Xenon atoms. Charged sample ions are ejected from the matrix and extracted into the mass analyzers

Used for large compounds with low volatility (eg peptides, proteins, carbohydrates)Solid or liquid sample is mixed with a non-volatile matrix (eg glycerol, crown ethers, nitrobenzyl alcohol)

Page 29: Mass spectrometry

Mass Spectrometry

Electron Ionization (EI)

Energetic process a heated filament emits electrons which are accelerated by a potential difference of usually 70eV into the sample chamber. Ionization of the sample occurs by removal of an electron from the molecule thus generating a positively charged ion with one unpaired electron.

• Produces M+.radical cation giving molecular weight• Produces abundant fragment ions

Page 30: Mass spectrometry

Mass Spectrometry

Chemical Ionization (CI)

process is initiated with a reagent gas such as methane, isobutane, or ammonia, which is ionized by electron impact. High gas pressure in the ionization source is required for the reaction between the reagent gas ions and reagent gas neutrals.

possible mechanism

Reagent (R) + e- → R+ + 2 e-

R+ + RH → RH+ + RRH+ + Analyte (A) → AH+ + R

biologically important molecules (sugars, amino acids, lipids etc.).

Page 31: Mass spectrometry

Mass Spectrometry

Thermal ionization (TI)

Samples are deposited on rhenium or tantalum filament and then carefully evaporated and sent to mass analyzer.

used to • quantify toxic trace elements in foods.• measurement of stable isotope ratio of

inorganic elements.

Page 32: Mass spectrometry

Mass Spectrometry

Accuracy

The range over which a mass spectrometer analyzer can operate.

A measure of how well a mass spectrometer separates ions of different mass

is a measure of how close the value obtained is to the true value. The accuracy varies dramatically from analyzer to analyzer depending on the analyzer type and resolution.

Mass Range Resolution Scan Speed

Mass AnalyzerProperties of mass Analyzer

Analyzers are scanned with a regular cycle time from low to high m/z or vice versa.

Page 33: Mass spectrometry

Mass Spectrometry

Mass Analyzer

• Quadrupoles• Quadrupole Ion Trap• Linear Ion Trap• Double-Focusing Magnetic Sector• Quadrupole Time-of-Flight Tandem MS• Quadrupole Time-of-Flight MS

Page 34: Mass spectrometry

Mass Spectrometry

Quadrupoles -ions travel parallel to four rods- opposite pairs of rods have rapidly alternating potentials (AC)- ions try to follow alternating field in helical trajectories- stable path only for one m/z value for each field frequency

Smalll and low cost Rmax ~ 500Harder to push heavy molecule - m/zmax < 2000

Page 35: Mass spectrometry

Mass Spectrometry

Quadrupole Ion TrapThe quadrupole ion trap typically consists of a ring electrode and two hyperbolic endcap electrodes. The motion of the ions induced by the electric field on these electrodes allows ions to be trapped or ejected from the ion trap. In the normal mode, the radio frequency is scanned to resonantly excite and therefore eject ions through small holes in the endcap to a detector. As the RF is scanned to higher frequencies, higher m/z ions are excited, ejected, and detected.

Page 36: Mass spectrometry

Mass Spectrometry

Linear Ion Trap

The linear ion trap differs from the 3D ion trap as it confines ions along the axis of a quadrupole mass analyzer using a two-dimensional (2D) radio frequency (RF) field with potentials applied to end electrodes. The primary advantage to the linear trap over the 3D trap is the larger analyzer volume lends itself to a greater dynamic ranges and an improved range of quantitative analysis.

Page 37: Mass spectrometry

Mass Spectrometry

Double-Focusing Magnetic Sectorthe ions are accelerated into a magnetic field using an electric field. A charged particle traveling through a magnetic field will travel in a circular motion with a radius that depends on the speed of the ion, the magnetic field strength, and the ion’s m/z. A mass spectrum is obtained by scanning the magnetic field and monitoring ions as they strike a fixed point detector.

Page 38: Mass spectrometry

Mass Spectrometry

Quadrupole Time-of-Flight Tandem MS

Time-of-flight analysis is based on accelerating a group of ions to a detector where all of the ions are given the same amount of energy through an accelerating potential. Because the ions have the same energy, but a different mass, the lighter ions reach the detector first because of their greater velocity, while the heavier ions take longer due to their heavier masses and lower velocity. Hence, the analyzer is called time-of-flight because the mass is determined from the ions’ time of arrival. Mass, charge, and kinetic energy of the ion all play a part in the arrival time at the detector.

Page 39: Mass spectrometry

Mass Spectrometry

Quadrupole Time-of-Flight MS

Quadrupole-TOF mass analyzers are typically coupled to electrospray ionization sources and more recently they have been successfully coupled to MALDI. It has high efficiency, sensitivity, and accuracy as compared to Quadrupole and TOF analyzer.

Page 40: Mass spectrometry

Mass Spectrometry

Photomultiplier Conversion Dynode

Faraday Cup

Array Detector

Charge (or Inductive) Detector

Electron Multiplier

Detectors used in mass spectrometer

Page 41: Mass spectrometry

Mass Spectrometry

Faraday CupA Faraday cup involves an ion striking the dynode (BeO, GaP, or CsSb) surface which causes secondary electrons to be ejected. This temporary electron emission induces a positive charge on the detector and therefore a current of electrons flowing toward the detector.

not particularly sensitiveoffering limited amplification of signalis tolerant of relatively high pressure.

– Ions are accelerated toward a grounded “collector electrode”– As ions strike the surface, electrons flow to neutralize charge, producing a small current that can be externally amplified.– Size of this current is related to # of ions in– No internal gain → less sensitive

Page 42: Mass spectrometry

Mass Spectrometry

Photomultiplier Conversion Dynode

is not as commonly Life limit is high as compared to others.

the secondary electrons strike a phosphorus screen instead of a dynode. The phosphorus screen releases photons which are detected by the photomultiplier. Photomultipliers also operate like the electron multiplier where the striking of the photon on scintillating surface results in the release of electrons that are then amplified using the cascading principle.

Page 43: Mass spectrometry

Mass Spectrometry

Array Detector

detects ions according to their different m/z, has been typically used on magnetic sector mass analyzers.

The primary advantage of this approach is that, over a small mass range, scanning is not necessary and therefore sensitivity is improved.

Page 44: Mass spectrometry

Mass Spectrometry

Charge (or Inductive) Detector

Charge detectors simply recognize a moving charged particle (an ion) through the induction of a current on the plate as the ion moves past

Detection is independent of ion size.

Page 45: Mass spectrometry

Mass spectrometer

Electron Multiplier

• Most important part• made up of a series (12

to 24) of aluminum oxide (Al2O3) dynodes

• Used for increasing potential

Ions strike the first dynode surface causing an emission of electrons. These electrons are then attracted to the next dynode held at a higher potential and therefore more secondary electrons are generated.

Page 46: Mass spectrometry

Mass spectrometer

Vacuum in the Mass SpectrometerAll mass spectrometers need a vacuum to allow ions to reach the detector without colliding with other gaseous molecules or atoms. If such collisions did occur, the instrument would suffer from reduced resolution and sensitivity.

Page 47: Mass spectrometry

Mass spectrometer

Structural analysis and Fragmentation Patterns

• Molecular ion (Parent ion)• Fragmentation peaks• Base peak• Isotopic peaks.

Mass spectrum

Graph of ion intensity (relative abundance) along x-axis versus mass-to-charge ratio (m/z) (unitsdaltons, Da) along Y-axis

Page 48: Mass spectrometry

Mass spectrometer

Molecular ion (Parent ion)

the peak corresponding to the mol wt of the compound

The peak of an ion formed from the original molecule by electron ionization, by the loss of an electron, or by addition or removal of an anion or cation and also known as parent peak, radical peak.

Page 49: Mass spectrometry

Mass spectrometer

Fragmentation peaks

The peaks observed by fragments of compounds.

The molecular ions are energetically unstable, and some of them will break up into smaller pieces. The simplest case is that a molecular ion breaks into two parts - one of which is another positive ion, and the other is an uncharged free radical.

The uncharged free radical won't produce a line on the mass spectrum. Only charged particles will be accelerated, deflected and detected by the mass spectrometer. These uncharged particles will simply get lost in the machine - eventually, they get removed by the vacuum pump.

Page 50: Mass spectrometry

Mass Spectrometry

Base peakThe most intense (tallest) peak in a mass spectrum, due to the most abundant ion. Not to be confused with molecular ion: base peaks are not always molecular ion and molecular ion are not always base peaks.

Page 51: Mass spectrometry

Mass Spectrometry

Fragmentation Patterns

By using fragmentation pattern we can easily study the structure of a compound.

Stevenson’s Rule Homolytic bond cleavage Heterolytic fragmentation Alpha cleavage Beta-cleavage Inductive cleavage Retro Diels-Alder Cleavage McLafferty rearrangement Ortho effect Onimum Reaction CO Elimination

Page 52: Mass spectrometry

Mass Spectrometry

Stevenson’s RuleThe most probable fragmentation is the one that leaves the positive charge on the fragment with the lowest ionization energy

In other words, fragmentation processes that lead to the formation of more stable ions are favored over processes that lead to less-stable ions.

Cleavages that lead to the formation of more stable carbocations are favored. When the loss of more than one possible radical is possible, a corollary to Stevenson’s Rule is that the largest alkyl radical to be lost preferentially.

Page 53: Mass spectrometry

Mass Spectrometry

Homolytic bond cleavage

A type of ion fragmentation in which a bond is broken by the transfer of one electron from the bond to the charged atom, the other electron remaining on its starting atom. The movement of one electron is signified by a fishhook arrow. The fragmentation of a ketone is shown in the figure.

Page 54: Mass spectrometry

Mass Spectrometry

Heterolytic bond cleavage

type of ion fragmentation in which a bond is broken by the transfer of a pair of electrons from the bond to the charged atom

The movement of 2 electrons is signified by a double-barbed arrow and also referred to as charge-induced fragmentation.

Page 55: Mass spectrometry

Mass Spectrometry

Alpha cleavage

For example in alcohols, aliphatic ethers, aromatic ethers, cyclic compounds and aromatic ketones etc.

Alpha cleavage occurs on α-bonds adjacent to heteroatoms (N, O, and S). Charge is stabilized by heteroatom. Occurs only once in a fragmentation (cation formed is too stable to fragment further)

Page 56: Mass spectrometry

Mass Spectrometry

Beta-cleavage

Fission of a bond two removed from a heteroatom or functional group, producing a radical and an ion. Also written as β-cleavage. For example allylic fragmentation.

Page 57: Mass spectrometry

Mass Spectrometry

Inductive cleavage

If an electron pair is completely transferred towards a centre of positive charge as a result of the inductive effect, shown schematically by the use of a double-headed arrow, then the ion will fragment by inductive cleavage. The figure illustrates this for a radical cation ether.

Page 58: Mass spectrometry

Mass Spectrometry

Retro Diels-Alder Cleavage

A multicentered ion fragmentation which is the reverse of the classical Diels-Alder reaction employed in organic synthesis that forms a cyclic alkene by the cycloaddition of a substituted diene and a conjugated diene. In the retro reaction, a cyclic alkene radical cation fragments to form either a diene and an alkene radical cation or a diene radical cation and an alkene. Depending on the substituents present in the original molecule, the more stable radical cation will dominate.

Page 59: Mass spectrometry

Mass Spectrometry

McLafferty rearrangement

An ion fragmentation characterised by a rearrangement within a six-membered ring system. The most usual configuration is for a radical cation formed by EI to undergo the transfer of a γ- hydrogen atom to the ionisation site through a ring system as shown here.

The distonic radical cation so formed can break up by radical-site-induced (α), or charged site-induced fragmentation as shown in the figure. For example ketones, carboxylic acid and esters.

Page 60: Mass spectrometry

Mass Spectrometry

Ortho effect

The interaction between substituents oriented ortho, as opposed to para and meta, to each other on a ring system, can create specific fragmentation pathways. This permits the distinction between these isomeric species. The diagram shows a case in which only the ortho isomer can undergo the rearrangement.

Page 61: Mass spectrometry

Mass Spectrometry

Onium Reaction

Mostly observed in cationic fragments containing a heteroatom as charge carrier, e.g.oxonium, ammonium, phosphonium and sulphonium ions.

The onium reaction is not limited to alkyl substituents acyl groups can also undergo the onium reaction

Onium Ion: A hypervalent species containing a non-metallic element such as the methonium ion CH5+. It includes ions such as oxonium, phosphonium, and nitronium ions.

Page 62: Mass spectrometry

Mass Spectrometry

CO Elimination

If there is more than one CO group present sequential elimination of all COgroups is possible.From carbonyl compounds CO elimination reaction takes place like in aldehyde, ketones and phenols etc

Cyclic unsaturated carbonyl compounds and cationic carbonyl fragmentswhich resulted from an a-cleavage tend to eliminate CO .

Page 63: Mass spectrometry

Mass Spectrometry

Rules for interpretation of mass spectrum

• DBR Calculations• Nitrogen Rule• Isotopic effect

Page 64: Mass spectrometry

Mass Spectrometry

Double bond or ring calculations tell us about how many rings or double bonds are present in a compound.DBR= C-H/2+N/2+1C= number of carbon atomsH= number of hydrogen atomsN= number of nitogen atoms

DBR Calculations

Page 65: Mass spectrometry

Mass Spectrometry

Nitrogen Rule• If a compound contains an even number of nitrogen atoms (or no

nitrogen atoms), its molecular ion will appear at an even mass number.

• If, however, a compound contains an odd number of nitrogen atoms, then its molecular ion will appear at an odd mass value.• This rule is very useful for determining the nitrogen content of an unknown compound.

Page 66: Mass spectrometry

Mass Spectrometry

Isotopic effect

Page 67: Mass spectrometry

Mass Spectrometry

Mass spectra (examples)

AlkanesStrong M+ (but intensity decreases with an increase of branches.Carbon-carbon bond cleavageloss of CH units in series: M-14, M-28, M-42 etc

Page 68: Mass spectrometry

Mass Spectrometry

Alkanes

Page 69: Mass spectrometry

Mass Spectrometry

Cycloalkanes

Strong M+, strong base peak at M-28 (loss of ethene)A series of peaks: M-15, M-28, M-43 etcMethyl, ethyl, propyl with an additional hydrogen give peaks

Page 70: Mass spectrometry

Mass Spectrometry

Alkenes Strong M+Fragmentation ion has formula CnH2n+ and CnH2n-1-Cleavage A series of peaks: M-15, M-29, M-43, M-57 etc

Page 71: Mass spectrometry

Mass Spectrometry

Alkynes Strong M+Strong base peak at M-1 peak due to the loss of terminal hydrogenAlpha cleavage

Page 72: Mass spectrometry

Mass Spectrometry

Aromatic Hydrocarbons

Strong M+Loss of hydrogen gives base peakMcLafferty rearrangement Formation of benzyl cation or tropylium ion

Page 73: Mass spectrometry

Mass Spectrometry

Alcohols M+ weak or absentLoss of alkyl group via a-cleavageDehydration (loss of water) gives peak at M-18

Page 74: Mass spectrometry

Mass Spectrometry

PhenolsStrong M+M-1 due to hydrogen eliminationM-28 due to loss of COM-29 due to loss of HCO (formyl radical)

Page 75: Mass spectrometry

Mass Spectrometry

EthersM+ weak but observableLoss of alkyl radical due to a-cleavageB-cleavage( formation of carbocation fragments through loss of alkoxy radicals)C-O bond cleavage next to double bondPeaks at M-31, M-45, M-59 etc

Page 76: Mass spectrometry

Mass Spectrometry

Aldehyde M+ weak, but observable (aliphatic)Aliphatic : M-29, M-43 etcMcLafferty rearrangement is common gives the base peakA-cleavageB-cleavage

Page 77: Mass spectrometry

Mass Spectrometry

M+ strong (aromatic)Aromatic: M-1 (loss of hydrogen)M-29 (loss of HCO)McLafferty rearrangement is commonA-cleavageB-cleavage

Aldehyde

Page 78: Mass spectrometry

Mass Spectrometry

Ketones Strong M+A series of peaks M-15, M-29, M-43 etcLoss of alkyl group attached to the carbonyl group by a-cleavageFormation of acylium ion (RCO+)McLafferty rearrangement

Page 79: Mass spectrometry

Mass Spectrometry

EstersM+ weak but generally observableLoss of alkyl group attached to the carbonyl group by a-cleavageFormation of acylium ion (RCO+)McLafferty rearrangementAcyl portion of ester OR+Methyl esters: M-31 due to loss of OCH3Higher esters: M-32, M-45, M-46, M-59, M-60, M-73 etc

Page 80: Mass spectrometry

Mass Spectrometry

Carboxylic acidsAliphatic carboxylic acids:M+ weak but observableA-cleavage on either side of C=OM-17 due to loss of OHM-45 due to loss of COOHMcLafferty rearrangement gives base peak

Page 81: Mass spectrometry

Mass Spectrometry

Aromatic carboxylic acids:M+ StrongA-cleavage on either side of C=OM-17 due to loss of OHM-18 due to loss of HOHM-45 due to loss of COOHMcLafferty rearrangement gives base peak

Page 82: Mass spectrometry

Mass Spectrometry

AminesM+ weak or absentNitrogen rule obeyA-cleavage

Page 83: Mass spectrometry

Mass Spectrometry

Nitriles

M+ weak but observableM-1 visible peak due to loss of termiminal hydrogen

Page 84: Mass spectrometry

Mass Spectrometry

Nitro Compounds

M+ seldom observedLoss of NO+ give visible peakLoss of NO2+ give peak

Page 85: Mass spectrometry

Mass Spectrometry

Alkyl chloride and alkyl bromidesStrong M+2 peakFor Cl M/M+2 = 3:1F or Br M/M+2 = 1:1A-cleavageLoss of Cl or BrLoss of HCl or HBr

Page 86: Mass spectrometry

Mass Spectrometry

Alkyl chloride

Page 87: Mass spectrometry

Mass Spectrometry

Applications of Mass Spectrometry

The technique has both quantitative and qualitative uses. These include identifying unknown compounds, determining the isotopic composition of elements in a molecule, and determining the structure of a compound by observing its fragmentation. Followings are the main applications

Toxicity of Toothpastes Measuring nanoparticle size Pharmacokinetics Protein characterization Space exploration Isotope dating and tracking Molecular weight Bonding Reaction mechanism

Page 88: Mass spectrometry

Mass Spectrometry

Toxicity of ToothpastesDEG (diethylene glycol) which is a toxic chemical and usually present in Chinese toothpastes.

Measuring nanoparticle sizeMass spectrometry is used to measure nanoparticle size like platinum nanoparticles which is used as catalyst.Once size of a sphere is measured, its density is also calculated.PharmacokineticsPharmacokinetics is often studied using mass spectrometry because of the complex nature of the matrix (often blood or urine) and the need for high sensitivity to observe low dose and long time point data.

Page 89: Mass spectrometry

Mass Spectrometry

Protein characterizationMass spectrometry is an important emerging method for the characterization of proteins. The two primary methods for ionization of whole proteins are electrospray ionization (ESI) and (MALDI).

Space explorationMass spectrometers are also widely used in space missions to measure the composition of plasmas. For example, the Cassini spacecraft carries the Cassini Plasma Spectrometer (CAPS),[44] which measures the mass of ions in Saturn's magnetosphere.

Isotope dating and trackingMass spectrometry is also used to determine the isotopic composition of elements within a sample. Differences in mass among isotopes of an element are very small, and the less abundant isotopes of an element are typically very rare, so a very sensitive instrument is required. These instruments, sometimes referred to as isotope ratio mass spectrometers (IR-MS).

Page 90: Mass spectrometry

Mass Spectrometry

Molecular weightMolecular weight can be determined by mass spectrometry.Actual number of carbons, hydrogen, oxygen etcBy using relative intensities(peak hight), we can easily calculated the actual numbers of C,H,O etc atoms.

BondingBonding can be studied by fragmentation patterns for example, beta cleavage is possible only if double bonds or heteroatom present.

Reaction mechanismMass spectrometry is best technique to study reaction mechanism and intermediates produced in reaction, for example, in carboxylic acid and alcohols a peak at M-18 indicates that water is produced.

Page 91: Mass spectrometry

Mass Spectrometry

Determination of Elements

Bulk materials such as steel or refractory metals, elements are determined by low-resolution glow-discharge mass spectrometry. High-resolution GDMS has been used to study semiconductor materials. GDMS is considered virtually free of matrix effects.

Detection limits in ICPMS as in Table

Page 92: Mass spectrometry

Mass Spectrometry

Species Analysis

Heavy metals in the environment are stored in complexes with humic acids, can be converted by microbes in different complexes, and can be transported in live animals and humans. This applies to many elements such as lead, mercury, arsenic, astatine, tin and platinum

For example, tin and lead alkylates established in soil, water or muscle tissue by GC / MS after exhaustive alkylation or thermal spray, and ICP-LC/MS API methods.

Page 93: Mass spectrometry

Mass Spectrometry

Environmental Chemistry

the analysis of trace elements and compounds in environmental samples like air, water, soil etc because of its detection power, specificity and structural analysis functionsGenerally, sample preparation is at least one type of chromatography coupled with MS either offline or online like GCMS

Page 94: Mass spectrometry

Mass Spectrometry

References

Dictionary of Mass Spectrometry, A.I. Mallet and S. Down, 2009 Introduction to spectroscopy, Donald L. PaviaHand book of spectroscopic data, B.D.Mistry.Comprehensive analytical chemistry. . Handbook of Spectroscopy, by G. Gauglitz and T. Vo-Dinh Instant notes of Analytical chemistry, D.Kealey.Modern Analytical Chemistry, David Harvey.The Basics of Spectroscopy, David.W.Ball.Encyclopedia of Analytical Chemistry Applications, Theory and

Instrumentation Edited by R.A.MeyersHandbook of Analytical Techniques edited by Helmut Giinzler and Alex Williams 1st Edition 2001Encyclopedia of Spectroscopy and Spectrometry part 2(M-Z) Edited By

john C. lindon, George E. Tranter and John L. Holmes