41
DEFIBRILLATOR

Defibrillator (ppt)

Embed Size (px)

DESCRIPTION

 

Citation preview

  • 1. 1. Define defibrillation. 2. Describe Need and history of defibrillation. 3. Describe the principle and mechanism of defibrillation. 4. Types and classes of defibrillator 5. Describe the Automated external defibrillator 6. Identify the precautions and risk 7. Troubleshooting of defibrillator

2. Definition: Defibrillation is a process in which an electronic device sends an electric shock to the heart to stop an extremely rapid, irregular heartbeat, and restore the normal heart rhythm. Defibrillation is a common treatment for life threatening cardiac dysrhythmias, ventricular fibrillation, and pulse less ventricular tachycardia. 3. Ventricular fibrillation is a serious cardiac emergency resulting from asynchronous contraction of the heart muscles. Due to ventricular fibrillation, there is an irregular rapid heart rhythm.Fig. Ventricular fibrillationFig. Normal heart beat 4. Ventricular fibrillation can be converted into a more efficient rhythm by applying a high energy shock to the heart. This sudden surge across the heart causes all muscle fibres to contract simultaneously. Possibly, the fibres may then respond to normal physiological pace making pulses. The instrument for administering the shock is called a DEFIBRILLATOR. 5. Defibrillation is performed to correct lifethreatening fibrillations of the heart, which could result in cardiac arrest. It should be performed immediately after identifying that the patient is experiencing a cardiac emergency, has no pulse, and is unresponsive. 6. Defibrillation was invented in by Prevost and Batelli, two Italian physiologists. They discovered that electric shocks could convert ventricular fibrillation to sinus rhythm in dogs.The first case of a human life saved by defibrillation was reported by Beck in 1947 . 7. Energy storage capacitor is charged at relatively slow rate from AC line. Energy stored in capacitor is then delivered at a relatively rapid rate to chest of the patient. Simple arrangement involve the discharge of capacitor energy through the patients own resistance. 8. Fig: Schematic diagram of a defibrillator 9. The discharge resistance which the patient represents as purely ohmic resistance of 50 to 100 approximately for a typical electrode size of 80cm2. This particular waveform Fig is called Lown waveform. The pulse width of this waveform is generally 10 ms. 10. current (amps)defibrillation occursno defibrillationpulse duration 11. minimum defibrillation energy occurs for pulse durations of 3 - 10 ms (for most pulse shapes). pulse amplitude in tens of amperes (few thousand volts). 12. operator selects energy delivered: 50-360 joules, depends on: intrinsic characteristics of patient patients disease duration of arrhythmia patients age type of arrhythmia (more energy required for v. fib.) 13. Fibrillations cause the heart to stop pumping blood, leading to brain damage. Defibrillators deliver a brief electric shock to the heart, which enables the heart's natural pacemaker to regain control and establish a normal heart rhythm. 14. Higher voltages are required for external defibrillationthan for internal defibrillation. A corrective shock of 750-800 volts is applied within atenth of a second. That is the same voltage as 500-533 no of AA batteries! 15. Electrical patternECG tracing 16. Occulsion of thecoronary artery leads to ischemia Ischemia leads to infarct which causes interruption of normal cardiac conduction Infarct = VF/VT 17. Ventricular FibrillationVentricular Tachycardia 18. Types of Defibrillator electrodes:a) Spoon shaped electrode Applied directly to the heart. b) Paddle type electrode Applied against the chest wall c) Pad type electrode Applied directly on chest wall 19. fig: Electrodes used in defibrillator (a) a spoon shaped internal electrode that is applied directly to the heart. (b) a paddle type electrode applied against the anterior chest wall. 20. Fig.- Pad electrode 21. Anterior electrode padApex electrode padFig: anterior apex scheme of electrode placement 22. Monophasic pulse or waveformBi-phasic pulse or waveform 23. There are two general classes of waveforms: a) mono-phasic waveformEnergy delivered in one direction through the patients hearta) Biphasic waveform Energy delivered in both direction through the patients heart 24. Fig:- Generation of bi-phasic waveform 25. The biphasic waveform is preferred over monophasic waveform to defibrillate. Why????? A monophasic type, give a high-energy shock, up to 360 to 400 joules due to which increased cardiac injury and in burns the chest around the shock pad sites. A biphasic type, give two sequential lowerenergy shocks of 120 - 200 joules, with each shock moving in an opposite polarity between the pads. 26. InternalExternal28 27. Internal defibrillatora) Electrodes placed directly to the hearte.g..-Pacemakerb) External defibrillator Electrodes placed directly on the hearte.g..-AED 28. For each minute elapsing between onset of ventricular fibrillation and first defibrillation, survival decreases by 10%. defibrillators should be portable, battery operated, small size. energy in defibrillators usually stored in large capacitors. total energy stored in capacitor:WC1 CVC2 2Vc = capacitor voltage 29. standby power supplychargedischargegatepatientswitch is under operator controlenergy storagetiming circuitryapplies shock about 20 ms after QRS complex, avoids T-waveECG monitor 30. AED is a portable electronic device that automatically diagnoses the ventricular fibrillation in a patient. Automatic refers to the ability to autonomously analyse the patient's condition. AED is a type of external defibrillation process. 31. AEDs require self-adhesive electrodes instead of hand held paddles. The AED uses voice prompts, lights and text tell the rescuer what steps have to take next.messages to 32. Turned on or opened AED. AED will instruct the user to:- Connect the electrodes (pads) to the patient. Avoid touching the patient to avoid false readings by the unit. The AED examine the electrical output from the heart and determine the patient is in a shock able rhythm or not 33. When device determined that shock is warranted, it will charge its internal capacitor in preparation to deliver the shock. When charged, the device instructs the user to ensure no one is touching the victim and then to press a red button to deliver the shock. Many AED units have an 'event memory' which store the ECG of the patient along with details of the time the unit was activated and the number and strength of any shocks delivered. 34. The paddles used in the procedure should not be placed: on a woman's breasts over an internal pacemaker patients. Before the paddle is used, a gel must be applied to the patient's skin 35. Skin burns from the defibrillator paddles are the most common complication of defibrillation. Other risks include injury to the heart muscle, abnormal heart rhythms, and blood clots. 36. Attach the external and internal paddles if the monitor reads, "No paddles." Check to ensure that the leads are securely attached if the monitor reads, "No leads. Connect the unit to AC power if the message reads, "Low battery." Verify that the Energy Select control settings are correct if the defibrillator does not charge. 37. Change the electrodes and make sure that the electrodes adapter cable is properly connected if you receive a message of "PACER FAILURE." Restart the pacer. Close the recorder door and the paper roll if the monitor message reads, "Check recorder. 38. Willis A Tacker, External Defibrillators, in The Biomedical Engineering Handbook, J. Bronzino (ed) CRC Press, 1995. www.google.com en.wikipedia.org/wiki/Defibrillation http://www.slideworld.org/viewslides.aspx/defibrillator