Transcript

Water analysisDIT-M, DIT-L photometers and DIT-IR interface module

DIT-L photometerThe DIT-L compact photometer quickly analyses the concen-tration of chlorine, chlorine dioxide, chlorite or ozone as well as the pH value in water treatment. High operating conveni-ence, ergonomic design, compact dimensions and safe han-dling make this unit essential for water analysis. Long-term stable reagents.

Features• Languageneutraluserinterface.

• Analysisisexecutedwith2interferencefiltersandlong-term stable LEDs as light sources without moving parts.

• QuickaccesstofavouritemethodsthankstoScrollMemory.

• Upto16datasetscanbesaved.• DatatransfertoaPCoraprinterispossiblewithaninfrared

interface via the optional DIT-IR module.

Applications• Drinkingwatertreatment• Swimmingpoolandbathingwatertreatment

DIT-M photometerTheDIT-Mphotometerisastate-of-the-artmeasuringunitthat combines the mobility of a portable photometer with the characteristics of a modern laboratory photometer. Simplehandlingensuresreliableandquickphotometricmeasurementofupto15parametersinwatertreatment.Long-term stable reagents are used.

Features• Multilingualplain-textoperatorpromptingmakeshan-

dling easy. The user can select the following languages: English,German,French,Spanish,Italian,PortugueseandPolish.

• TheDIT-Mphotometeroperateswith6interferencefiltersand long-term stable LEDs as light sources without moving parts.

• Upto1000datasetscanbesaved.

• DatatransfertoaPCoraprinterispossiblewithaninfra-red interface via the optional DIT-IR module.

Applications• Drinkingwatertreatment

• Swimmingpoolandbathingwatertreatment

• Industrialwatertreatment

DIT-IR infrared interface moduleDIT-IRtransfersdatafromaDIT-MoraDIT-Lphotometerviainfraredsignaltoaninterface.DIT-IRcanbeconnectedtoaPC(USB)oraserialprinter.

DIT-M, DIT-L, DIT-IR

gRunDfos WATeR TReATMenT

Thenam

eGrundfos,theGrundfoslogo

,andbeth

inkinnov

ateareregistered

tradem

arksowned

byGrundfosHoldingA/S

orGrundfosA/S

,Den

mark.Allrigh

tsreserve

dw

orldwide.

95727

8900713

GRunDFos Holding A/sPoul Due Jensens Vej 7DK-8850 BjerringbroTel: +45 87 50 14 00www.grundfos.com

Technical data DIT-MDisplay Graphic display

optics 6LEDs,6interferencefilters(IF),photosensor,transparentsample chamber; Wavelengthaccuracy:±1nm; Wavelength ranges:λ1=530nm (IF ∆λ =5nm),λ2 = 560nm (IF ∆λ =5nm), λ3=610nm (IF ∆λ =6nm),λ4 = 430nm (IF ∆λ =5nm), λ5=580nm (IF ∆λ =5nm),λ6 = 660nm (IF ∆λ =5nm)

Photometric accuracy

2%FS(FullScale,T=20°C-25°C), measured with standard solutions

operation Acidandsolventresistanttouch-sensitivekeypadwithacoustic signal

Power supply 4batteries(typeAA/LR6);lifetimeapprox.3500tests

Auto oFF 20min.afterlastfunction; 30secondsbeforeswitch-offanacousticalsignalisemitted

operating conditions 5-40°Catmax.30-90%rel.humidity(notcondensing)

Interface Data transfer with infrared interface module DIT-IR

Parameters DIT-MAluminium 0.01-0.3 mg/l

Bromine 0.5-13.0 mg/l

Chlorine (free, com-bined, total)

0.01-6.0 mg/l

Chlorine dioxide 0.02-11.0 mg/l

Chloride 0.5-25 mg/l

Chlorite 0.01-6.0 mg/l

Cyanuric acid 2-160 mg/l

Iron 0.02-1.0 mg/l

Fluoride 0.05-2.0 mg/l

Manganese 0.2-4.0 mg/l

ozone 0.02-2.0 mg/l

Phosphate 0.05-4.0 mg/l

Hydrogen peroxide 0.03-3.0 mg/l

pH 6.5-8.4

Acid demand Ks 4.3 0.1-4.0 mmol/l

Technical data DIT-LDisplay LCD,backlitwhenakeyispressed

optics 2LEDs,2interferencefilters(IF),photosensor,transparentsample chamberWavelengthaccuracy:±1nmWavelength ranges:λ1 = 530nm(IF ∆λ =5nm) λ2 = 560nm(IF ∆λ =5nm)

Photometric accuracy

3%FS(FullScale,T=20°C-25°C), measured with standard solutions

operation Acidandsolventresistanttouch-sensitivekeypad

Power supply 4batteries(typeAAA/LR03);lifetimeapprox.5000tests

Auto oFF 10min.afterlastfunction

operating conditions 5-40°Catmax.30-90%rel.humidity(notcondensing)

Interface Data transfer with infrared interface module DIT-IR

Parameters DIT-LChlorine (free, com-bined, total)

0.01-6.0 mg/l

Chlorine dioxide 0.02-11.0 mg/l

Chlorite 0.01-6.0 mg/l

ozone 0.02-2.0 mg/l

pH 6.5-8.4

Photometric measuring principle

Interferencefilter

Photosensor

Light beam Digital display

LED Water sample


Recommended