Targeted and untargeted
metabolic profiling by
incorporating scanning FAIMS
into LC-MS
Kayleigh Arthur
Introduction
• LC-MS is a highly used technique for untargeted profiling
analyses
• Incorporation of scanning ion mobility with LC-MS
• Why choose FAIMS over IMS (DT or TW)?
• Fast FAIMS scanning achievable with miniaturised
FAIMS device
• Full scan FAIMS within time of a UHPLC peak
• Approach applicable to a range of mass spectrometers
Untargeted profiling – IMS
Valentine et al, J. Proteome. Res., 2006, 5, 2977-2984 Harry et al, J. Chromatogr. B, 2008, 871, 357-361
Untargeted profiling – FAIMS
Beach et al, Anal. Chem., 2011, 83, 9107-9113
No LC
Canterbury et al, Anal Chem, 2008, 80, 6888-6897
LC run time = 120 mins
FAIMS scan time = 2-3 s
Creese et al, J Am Soc Mass Spectrom, 2013, 24, 431-443
ETD
CID
6 FAIMS conditions
Peptide IDs missed
with scanning mode
Scanning can miss ions if
the top of the LC peak
does not coincide with CV
for transmission
FAIMS vs TWIMS
• Direct comparison
• FAIMS covers a greater proportion of the analytical space
• FAIMS covers across the CF range at all m/z
• TWIMS shows a correlation between m/z and bin number o Bin number increases as
m/z increases
FAIMS-MS TWIMS-MS
Set-up
Modes of operation
Acquisition of
nested data
sets
Apply approach
to human urine
Incorporate into
Omics workflow
How? FAIMS/LC-MS Synchronisation
Agilent 1200
series LC
Contact
closure board
fitted to LC
binary pump
FAIMS
control unit
Agilent 6230 TOF MS
ESI Source
Owlstone
Miniaturised
FAIMS chip
Computer Contact
closure
interface
Static
Fixed Dispersion
Field
Fixed Compensation
Field
Scanning
Fixed Dispersion
Field
Scanning Compensation Field on chromatographic
peak timescale
LC-FAIMS-MS Modes of Operation
Application to Biological Matrices
FAIMS-MS Optimisation of FAIMS
DF and CF
Targeted
LC-FAIMS-MS
Isobaric separation, reduction in chemical noise, in-source CID
Untargeted
LC-FAIMS-MS Feature determination
Scanning LC-FAIMS-MS – how?
Chromatographic
Peak
Compensation
Field Scan
11 CFs per s
Mass Spectra
1 per CF
LC peak width
All CFs scanning
Human Urine TIC – Scanning approach
5 x10
1
2
3
4
5
6
7
Counts vs. Retention Time (min)
6.8 7 7.2 7.4 7.6 7.8 8 8.2
6 x10
0.0
0.5
1.0
1.5
2.0
Counts vs. Retention Time (min)
0.5 1 1.5 2 2.5 3 3.5 4 4.5 5 5.5 6 6.5 7 7.5 8 8.5 9 9.5
Human Urine TIC – Scanning approach
• CF deconvolution
into individual
channels
• CF adds another
dimension of
separation
0
1000
2000
3000
4000
5000
6000
-1 0 1 2 3 4
Pe
ak A
rea
Compensation Field (Td)
Urine
Standards
0
100000
200000
300000
400000
500000
600000
-1 0 1 2 3 4
Pe
ak A
rea
Compensation Field (Td)
Urine
Standards
Urinary Creatinine
[M+H]+
m/z 114.066
[M+Na]+
m/z 136.048
Isobaric Separation
4 x10
0
1
2
5.212
3 x10
0
1
2
5.217
3 x10
0
0.5
1 5.203
Counts vs. Retention Time (min) 4.4 4.6 4.8 5 5.2 5.4 5.6 5.8 6
LC-MS
FAIMS off
LC-FAIMS-MS
CF 0.08 Td
LC-FAIMS-MS
CF 2.04 Td
3 x10
0
1
2
3
Counts vs. Retention Time (min) 5.1 5.2 5.3
0
500
1000
1500
2000
-1 0 1 2 3 4
LC
Pe
ak
Are
a
FAIMS Compensation Field (Td)
LC-FAIMS-MS
FAIMS off = 1 feature
FAIMS on = 2 features
m/z 207.11
RT 5.21 min
Reduction in Chemical Noise and
Interferences
3 x10
0
2
4
3 x10
0
1
2
Counts vs. Retention Time (min) 1 2 3 4 5 6 7 8 9
m/z 331.21, RT 4.24 min
LC-MS
FAIMS off
LC-FAIMS-MS
CF 0.08 Td FAIMS on
S:N = 124.1
FAIMS off = 0 features, FAIMS on = 1 features
FAIMS off
S:N = 2.7
In-source CID vs FAIMS-selected CID LC-MS LC-FAIMS-MS
0
2
4
181.08
4037
124.05
2279 105.05
1614 91.05
1225 141.07
893
Counts vs. Mass-to-Charge (m/z)
100 120 140 160 180
x103
0
2
181.07
26158
121.04
4305 141.07
3514 169.97
1489 95.06
1073
x104 Precursor ion
0
2
4
181.08
468
124.05
180
Counts vs. Mass-to-Charge (m/z)
100 120 140 160 180
x102
0
5
181.07
534
x102
Fragment
Precursor ion
Fra
g 2
00 V
F
rag
350 V
Fra
g 2
00 V
F
rag
35
0 V
? ? ?
FAIMS-IN-SOURCE COLLISION INDUCED DISSOCIATION
FISCID
Conclusions
• Acquisition of LC-FAIMS-MS nested data sets on timescale of UHPLC peak for the first time
• Increase in peak capacity using LC-FAIMS-MS in comparison to LC-MS
• Higher level of orthogonality with m/z / RT than IMS
• Separation of isobaric and co-eluting analytes
• DF + CF as additional identifiers
• Can be integrated into non-targeted omics workflows
• Applicable to a range of mass spectrometers