Page 1: Spectrophotometer(2) Prof.Dr. Moustafa M. Mohamed

Spectrophotometer(2)

Prof.Dr. Moustafa M. Mohamed

Page 2: Spectrophotometer(2) Prof.Dr. Moustafa M. Mohamed

Spectrophotometer

•The spectrophotometer measures light absorption by a liquid substance at various

wavelengths.

•The components of an unknown material can be determined or the concentration of a

number of known substances can be measured.

Page 3: Spectrophotometer(2) Prof.Dr. Moustafa M. Mohamed
Page 4: Spectrophotometer(2) Prof.Dr. Moustafa M. Mohamed

•A monochrometer uses a diffraction grating or prism to disperse the light from the lamp (slit S1).

•The light is broken into its spectral components as it arises from slit S2 and falls on the sample in the

cuvette,•The angle of the diffraction grating determines light

wavelength.•The mirror reduces equipment size.•

Page 5: Spectrophotometer(2) Prof.Dr. Moustafa M. Mohamed

•light output, photodetector sensitivity, and sample substance absorbance changes with

wavelength, and this necessitates zero calibration for each wavelength measurement.

•The double-beam spectrophotometer accomplishes this automatically by beam path

switching (sample to reference) via a mechanical shutter or rotating mirror.

Page 6: Spectrophotometer(2) Prof.Dr. Moustafa M. Mohamed

Maintenance

• Maintenance includes:• Calibration adjustment and

replacement of light source bulbs and photo detectors.

• Mechanically rotating assembles (mirrors, diffraction grating) will occasionally malfunction.

• The electronics is very reliable.

Page 7: Spectrophotometer(2) Prof.Dr. Moustafa M. Mohamed

General Operating Instructions

Page 8: Spectrophotometer(2) Prof.Dr. Moustafa M. Mohamed

Turn on Spectrophotometer

•Turn the Power Switch Clockwise

• Allow the spectrophotometer to warm up for at least fifteen minutes to stabilize the source and detector before proceeding.

Page 9: Spectrophotometer(2) Prof.Dr. Moustafa M. Mohamed

Set the Display Mode to TRANSMITTANCE

•Press the MODE CONTROL

•Until the LED beside TRANSMITTANCE is lit

Page 10: Spectrophotometer(2) Prof.Dr. Moustafa M. Mohamed

Adjust the display to 0% TRANSMITTANCE

•With the Zero Control Knob.

Page 11: Spectrophotometer(2) Prof.Dr. Moustafa M. Mohamed

Set the desired WAVELENGTH•By Adjusting the Wavelength

Control

Page 12: Spectrophotometer(2) Prof.Dr. Moustafa M. Mohamed

Prepare Blank

•Fill clean cell with solvent and wipe the cell with a tissue to remove liquid droplets, dust, and finger prints.

•Place the cell firmly into the sample compartment, aligning the guide mark on the cell with the guide mark at the front of the

sample compartment.

Page 13: Spectrophotometer(2) Prof.Dr. Moustafa M. Mohamed

Adjust Transmittance to 100%

•Close the sample compartment Lid

•Adjust the Transmittance knob until the display reads 100.

Page 14: Spectrophotometer(2) Prof.Dr. Moustafa M. Mohamed

Check Absorbance•Press Mode Control Switch

until LED besides ABSORBANCE is lit.•The display should read 0.000.

Page 15: Spectrophotometer(2) Prof.Dr. Moustafa M. Mohamed

The machine response changes EVERYTIME you change the

WAVELENGTH

You must reset the 100% Transmittance EVERYTIME you

change the WAVELENGTH

Page 16: Spectrophotometer(2) Prof.Dr. Moustafa M. Mohamed

Making Sample Measurements

•Rinse the sample cell twice with a small amount of solution.

•Wipe cell with tissue and place in the sample compartment.

Page 17: Spectrophotometer(2) Prof.Dr. Moustafa M. Mohamed

Read the Absorbance

•Align the guide marks and close the lid.

•Read the absorbance on the display.•Remove the cell from sample compartment.

Page 18: Spectrophotometer(2) Prof.Dr. Moustafa M. Mohamed

If you must change the WAVELENGTH, you must

reset the BLANK ABSORBANCE at the new

wavelength

Page 19: Spectrophotometer(2) Prof.Dr. Moustafa M. Mohamed

Can You Identify the Parts?

#1

#3

#8

#6

#7

#4

#5

#2

Page 20: Spectrophotometer(2) Prof.Dr. Moustafa M. Mohamed
Page 21: Spectrophotometer(2) Prof.Dr. Moustafa M. Mohamed