Key Features
Description
Applications
Copyright © Peak Scientific Instruments Ltd 2020 — Solaris XE Data sheet Rev 4 RSID 2297 EN —06/02/2020
99-999999-9999 Service Kit :Part Number :
Your local gas generation partner
Peak Scientific’s Solaris XE generator has been engineered using membrane technology to meet the gas delivery requirements for labs using LC-MS, Compact MS or multiple ELSDs. Solaris XE can operate with varying flow rates (up to 35L/min of high purity nitrogen), purity (95-99.5%) and outlet pressure can be adjusted down from 116psi, offering a flexible solution for a variety of applications. The Solaris XE range has been engineered to provide nitrogen to laboratories that utilize an external source of compressed air, which meets a minimum quality grade of ISO 8573-1:2010 Class 1.4.1, and its compact, space-saving chassis allows it to be placed on a benchtop, taking up minimal space in the lab.
Nitrogen up to 35L/min, up to 116 psi @ 95-99.5% purity
Variable outlet flow, pressure and purity Compact, stackable* system maximising the use of valuable laboratory floor or bench space Colour indicative LED lighting for easy status awareness Manufactured and tested to highest spec in UK, CE/FCC certified Safer and more convenient than pressurized cylinders, dewars or bulk storage 12 month on-site comprehensive warranty Backed by [Peak Protected] global on premise servicing and technical support *units stacked on top of Solaris must not exceed 100kg/220.4lbs
LC-MS, Compact MS, Multiple ELSDs, TLD readers
Solaris XENitrogen generator for LC-MS or multiple ELSDs
Peak Scientific UKTel: +44 (0)141 812 8100Fax: +44 (0)141 812 8200
Peak Scientific North America
Tel: +1 866 647 1649Fax: +1 978 608 9503
For a full list of our worldwide office locations, please visit:
Web: www.peakscientific.comEmail: [email protected]
Peak Scientific’s Quality Management System conforms to: ISO:9001:2008
Product Certifications
Copyright 2020 Peak Scientific - Rev - Solaris XE Data sheet Rev 4 RSID 2297 EN - 06/02/20
Technical Specifications Solaris XE
Max Flow Rate up to 35 L/min
Max Pressure up to 116 psi / 8 bar (user adjustable)
Purity 95-99.5 %*
Gas Outlets 1 x 1/4 “ BSPP
Min/Max Air Inlet Pressure 80 psi/5.5 bar - 145 psi/10 bar
Min Inlet Air Quality ISO8573 - 1:2010 Class 1.4.1
Contaminants Phthalate & BHT Free
Suspended Liquids None
Operating Temperature 5°C - 30°C / 41°F - °F
Electrical Requirements Universal100 / 230 VAC / 50 / 60 Hz / 0.03 A
Power Consumption 6.9 Watts
Heat Output 25 BTU/Hr
Generator Dimensions (HxWxD) 156 x 343 x 650 mm6.1 x 13.5 x 25.6’’
Generator Weight 15.1 kg / 33.3 lbs
Max Weight Bearing Capability 100 kg / 220.5 lbs
Solaris Compressor compatible NO
*depending on inlet air pressure and outlet flow
Ordering Information
Part Number 3300671
Annual Service www.peakscientific.com/ordering
FPPM Maintenance Plan 3300671-FPPM
Complete Maintenance Plan 3300671-CSP
Supported Maintenance Plan 3300671-SMP
Peak Scientific gas generators define the benchmark in reliability, convenience and performance in laboratories around the world, and come backed by a 12 month warranty. Beyond this period however you can ensure that your investment continues to be [Protected] by our comprehensive generator care cover. Our world-class aftercare support packages deliver a program of scheduled preventative maintenance whilst giving you the reassurance of instant access to worldwide technical support and priority on-site response in the untimely event of a breakdown.
Description
Copyright © Peak Scientific Instruments Ltd 2020 — Solaris XE Waters Data sheet Rev 5 RSID 2950 EN —06/02/2020
99-999999-9999 Service Kit :Part Number :
Your local gas generation partner
Applications
Peak Scientific’s Solaris XE nitrogen generator offers a reliable, compressor-free gas solution that can comfortably meet the demands of most mass spectrometers within Waters’ portfolio of instruments. Using our tried and tested membrane technology, Solaris XE can operate with varying flow rates (up to 35L/min of high purity nitrogen), purity (95-99.5%) and outlet pressure can be adjusted down from 116psi.
Solaris XE for Waters has been designed in collaboration with Waters and is the only compressor-free nitrogen generator available which is tested and approved by Waters for use with the ACQUITY QDa Detector. Solaris XE is fitted with alignment blocks specifically designed to enable the generator to be stacked underneath the ACQUITY QDa in most modular stacking configurations*, providing a compact and discreet space-saving solution within the lab. Alternatively this unit can be wall mounted or free standing on a lab bench.
The Solaris XE has been engineered to provide nitrogen to laboratories that utilize an external source of compressed air, which meets a minimum quality grade of ISO 8573-1:2010 Class 1.4.1.
*Please ask your local sales representative for advice on supported stacking configurations.
) Nitrogen up to 35L/min, up to 116 psi @ 95-99.5% purity
) Tested and approved by Waters
) Compact, wall-mountable and stackable* with Waters ACQUITY QDa, maximizing the use of valuable laboratory floor or bench space
) Variable outlet flow, pressure and purity
) Colour indicative LED lighting for easy system status awareness
) Manufactured and tested to highest spec in UK, CE/FCC/CSA certified
) Safer and more convenient than pressurized cylinders, dewars or bulk storage
) 12 month on-site comprehensive warranty
) Backed by [Peak Protected] global on premise servicing and technical support
) Waters ACQUITY QDa Detector
) A list of instruments that are compatible with Solaris XE can be found on our Waters application list.
Solaris XE for Waters™Dedicated nitrogen solution for ACQUITY™ QDa™
343 mm / 13.5”
650 mm / 25.6”
156 mm
/ 6.2”
Peak Scientific UKTel: +44 (0)141 812 8100Fax: +44 (0)141 812 8200
Peak Scientific North America
Tel: +1 866 647 1649Fax: +1 978 608 9503
For a full list of our worldwide office locations, please visit:
Web: www.peakscientific.comEmail: [email protected]
Peak Scientific’s Quality Management System conforms to: ISO:9001:2008
Product Certifications
Copyright 2020 Peak Scientific - Rev - Solaris XE Waters Data sheet Rev 5 RSID 2950 EN - 06/02/20
Technical Specifications Solaris XE for Waters
Max Flow Rate up to 35 L/min
Max Pressure up to 116 psi / 8 bar (user adjustable)
Purity 95% - 99.5%*
Gas Outlets 1 x 1/4 “ BSPP
Min/Max Air Inlet Pressure 80 psi/5.5 bar - 145 psi/10 bar (116 psi/8 bar required for QDa)
Min Inlet Air Quality ISO8573-1:2010 Class 1.4.1**
Contaminants Phthalate & BHT Free
Suspended Liquids None
Operating Temperature 5°C - 30°C / 41°F - °F
Electrical Requirements Universal 100 / 230 VAC / 50 / 60 Hz / 0.03 A
Power Consumption 6.9 Watts
Heat Output 25 BTU/Hr
Generator Dimensions (HxWxD) 156 x 343 x 650 mm6.2 x 13.5 x 25.6’’
Generator Weight 15.1 kg / 33.3 lbs
Max Weight Bearing Capability 100 kg / 220.5 lbs
Solaris Compressor compatible NO
*depending on inlet air pressure and outlet flow
** This unit should be connected to an air supply that, as a minimum, meets ISO8573-1:2010 Class 1.4.1
Ordering Information
Part Number 3300670
Annual Service www.peakscientific.com/ordering
FPPM Maintenance Plan 3300671-FPPM
Complete Maintenance Plan 3300671-CSP
Supported Maintenance Plan 3300671-SMP
Peak Scientific gas generators define the benchmark in reliability, convenience and performance in laboratories around the world, and come backed by a 12 month warranty. Beyond this period however you can ensure that your investment continues to be [Protected] by our comprehensive generator care cover. Our world-class aftercare support packages deliver a program of scheduled preventative maintenance whilst giving you the reassurance of instant access to worldwide technical support and priority on-site response in the untimely event of a breakdown.
Description
Copyright 2018 Peak Scientific - Rev - Solaris Compressor Data sheet Rev 3 RSID 1609 EN - 29/11/18
99-999999-9999 Service Kit :Part Number :
Your local gas generation partner
Key Features
Applications
Solaris Air Compressor Dimensions
Peak Scientific’s Solaris Air Compressor generator has been engineered and designed to support the Solaris 10 Nitrogen generator for labs without an external supply of high quality air. Available in both a 230v and 110v option this unit is suitable for all regions. Solaris can be placed on a benchtop and stacked with the nitrogen generator on top to contain gas supply in a single system whilst saving valuable laboratory workspace.
Solaris has been developed to provide an on-site gas supply of high purity nitrogen for ELSD (Evaporative Light Scattering Detector) for labs using HPLC (High Performance Liquid Chromatography) or alternatively to supply Compact Mass Spectrometer instruments.
) Stackable with Solaris 10 nitrogen generator for ELSD and Compact Mass Specs
) Designed specifically for ELSD and Compact Mass Specs
) Compact, stackable system maximising the use of valuable laboratory floor or bench space
) Manufactured and tested to highest spec in UK, CE/FCC certified
) 12 month on-site comprehensive warranty
) Backed by [Peak Protected] global on premise servicing and technical
) Can supply 1 x Solaris 10 Nitrogen
Solaris Air Compressor Recommended Air Compressor for Solaris 10 Nitrogen generator
406 mm
540 mm 416 mm
Peak Scientific UKTel: +44 (0)141 812 8100Fax: +44 (0)141 812 8200
Peak Scientific North America
Tel: +1 866 647 1649Fax: +1 978 608 9503
For a full list of our worldwide office locations, please visit:
Web: www.peakscientific.comEmail: [email protected]
Peak Scientific’s Quality Management System conforms to: ISO:9001:2008
Copyright 2018 Peak Scientific - Rev - Solaris Compressor Data sheet Rev 3 RSID 1609 EN - 29/11/18
Product Certifications
Technical Specifications Air Compressor 110v Air Compressor 230v
Flow Rate Suitable for 1 X Solaris 10 nitrogen combinations
Max Pressure 120 psi (8.27 bar)
Gas Outlets 1 x 1/4" BSPP
Operating Temperature 5°C - 35°C / 41°F - 95°F
Electrical Requirements 100V+/- 10% 50/60Hz 5.4A 115V+/-10% 60Hz 5.3A 230V+/- 10% 50/60Hz 2.8A
Power Consumption 594 Watts 670 Watts 1150 Watts
Heat Output 2420 BTU/hr
Dimensions 416 mm x 540 mm x 406 mm / 16.3” x 21.2” x 15.8”
Weight 41.5 Kg / 91.5 lbs
Ordering Information
Part Number 65-0004 65-0003
Annual Service Kit 08-9454
Complete Maintenance Plan 09-3010
Peak Scientific gas generators define the benchmark in reliability, convenience and performance in laboratories around the world, and come backed by a 12 month warranty. Beyond this period however you can ensure that you’re investment continues to be [Protected] by our comprehensive generator care cover. Our world-class aftercare support packages deliver a program of scheduled preventative maintenance whilst giving you the reassurance of instant access to worldwide technical support and priority on-site response in the untimely event of a breakdown.
Key Features
Solaris 10 Dimensions
Description
Applications
99-999999-9999 Service Kit :Part Number :
Your local gas generation partner
Copyright © Peak Scientific Instruments Ltd 2019 — Solaris 10 Data sheet Rev 7 RSID 1605 EN —30/09/2019
Peak Scientific’s Solaris 10 nitrogen generator has been engineered using membrane technology to meet the gas delivery requirements for labs using ELSD with their High Performance Liquid Chromatography, Compact Mass Spectrometer or for TLD reader instruments. Solaris 10 can operate with varying flow rate (up to 10L/min of high purity nitrogen), purity (up to 99.5%) and outlet pressure is adjustable to 100psi max offering a flexible solution for labs.
With a compact and space-saving chassis, Solaris 10 can be placed on a benchtop and paired with an additional air compressor unit providing air to the system for labs without an in-house air supply or who wish to contain their gas supply in a single system.
) Nitrogen up to 10 L/min, up to 100 psi and up to 99.5% purity
) Variable outlet flow, pressure and purity
) Compact, stackable system maximising the use of valuable laboratory floor or bench space
) Optional Solaris Air Compressor stackable module
) Colour indicative LED lighting for easy status awareness
) Manufactured and tested to highest spec in UK, CE/FCC certified
) Safer and more convenient than pressurized cylinders, dewars or bulk storage
) 12 month on-site comprehensive warranty
) Backed by [Peak Protected] global on premise servicing and technical
) Up to 2 x ELSD instruments
) 1 x Compact Mass Spectrometer
) TLD Readers
Solaris 10Nitrogen generator for ELSD and Compact Mass Spectrometer
156 mm
540 mm416 mm
Peak Scientific UKTel: +44 (0)141 812 8100Fax: +44 (0)141 812 8200
Peak Scientific North America
Tel: +1 866 647 1649Fax: +1 978 608 9503
For a full list of our worldwide office locations, please visit:
Web: www.peakscientific.comEmail: [email protected]
Peak Scientific’s Quality Management System conforms to: ISO:9001:2008
Product Certifications
Copyright © Peak Scientific Instruments Ltd 2019 — Solaris 10 Data sheet Rev 7 RSID 1605 EN —30/09/2019
Technical Specifications Solaris 10
Max Flow Rate up to 10 l/min (input dependent)
Max Pressure up to 100 psi / 6.9 bar (user adjustable)
Purity up to 99.5%*
Gas Outlets 1 x 1/4 " BSPP
Min/Max Air Inlet Pressure 80 psi/ 5.5 bar - 130 psi/ 8.9 bar
Min Air Inlet Flow 30 l/min
Min Inlet Air Quality ISO8573 - 1:2010 Class 1.4.1
Contaminants Phthalate & BHT Free
Suspended Liquids None
Operating Temperature 5°C - 35°C / 41°F - 95 °F
Electrical Requirements Universal 100-230 VAC / 50/60 Hz / 0.03A
Power Consumption 6.9 Watts
Heat Output 25 BTU/Hr
Generator Dimensions H x W x D 156 x 416 x 540 mm / 6.1 x 16.4 x 21.3”
Generator Weight 12 kg / 26.4 lbs
*depending on inlet air pressure and outlet flow
Ordering Information
Part Number 65-0001
Annual Service 08-9013
Standard Maintenance Plan 09-3010
Complete Maintenance Plan 09-3010
Solaris Air Compressor 110v 230v Annual Service Kit
Part Number 65-0004 65-0003 08-9454
Peak Scientific gas generators define the benchmark in reliability, convenience and performance in laboratories around the world, and come backed by a 12 month warranty. Beyond this period however you can ensure that you’re investment continues to be [Protected] by our comprehensive generator care cover. Our world-class aftercare support packages deliver a program of scheduled preventative maintenance whilst giving you the reassurance of instant access to worldwide technical support and priority on-site response in the untimely event of a breakdown.
Pressure Swing Adsorption (PSA) has been employed for
decades and is the preferred choice for generating Ultra
High Purity Nitrogen. The technology is self- regenerating,
offering you a reliable and durable source of Ultra High
Purity Nitrogen with little need for attention.
We can make our systems available with or without an
internal air compressor.
PEAK SCIENTIFIC. GO WITH THE FLOW.
Nitrogen Generators NG250(A) - 5000(A)
Copyright 2013 Peak Scientific
Technical Specifications
Ordering Information
Annual Service Kit NG250 - 4000 08-4723
Annual Service Kit NG250A - 4000A 08-4704
Standard Maintenance Plan 09-3110
Complete Maintenance Plan 09-3010
Website: www.peakscientific.com Email: [email protected]
Part Number -excl. air compressor Part Number - incl. air compressor
UK EU US UK EU US
NG250 10-2022 10-2020 10-2010 10-1010 10-1020 10-1030
NG600 10-2070 10-2071 10-2072 10-1060 10-1070 10-1080
NG1000 10-2110 10-2111 10-2130 10-1110 10-1120 10-1130
NG2000 10-2160 10-2170 10-2180 10-1160 10-1170 10-1180
NG3000 10-2210 10-2211 10-2230 10-1210 10-1220 10-1230
NG4000 10-2260 10-2270 10-2280 10-1260 10-1270 10-1280
NG5000 N/A 10-1292 10-1295 10-1291
NG250(A) NG600(A) NG1000(A) NG2000(A) NG3000(A) NG4000(A) NG5000 NG5000A
UHP Nitrogen 250 cc/min 600 cc/min
1000 cc/min
2000 cc/min 3000 cc/min 4000 cc/min 5000 cc/min
Max Outlet Pressure5.5 bar /
80 psi5.5 bar /
80 psi5.5 bar /
80 psi5.5 bar /
80 psi5.5 bar /
80 psi5.5 bar /
80 psi5.5 bar /
80 psi5.5 bar /
80 psi
Purity (N2) 99.9995% 99.9995% 99.9995% 99.9995% 99.9995% 99.9995% 99.9995% 99.998%
Pressure Dewpoint -70ºC / -94ºF
Gas Outlet 1x 1/4” BSP Female
Electrical Requirements
- units without air compressor 230v 50Hz 0.5A / 110v 60Hz 1.2A
- units with air compressor 230v 50Hz 2.2A / 110v 60Hz 5.2A 230v 50Hz 3.6A / 110v 60Hz 8.0A
Power Consumption
- units without air compressor 230v - 115 watts / 110v - 132 watts
- units with air compressor 230v - 506 watts / 110v - 572 watts 230v - 828 watts / 110v - 880 watts
Dimensions (cm/ ins) WxDxH 43 x 41 x 62 / 17 x 16 x 25 43 x 41 x 126 / 17 x 16 x 50
Weight (kg / lbs)
- units without air compressor 33 / 73 47 / 104 57 / 126 77 / 170 80 / 176
- units with air compressor 43.5 / 96 43 / 95 45 / 99 65 / 143 88 / 194 88 / 194
The NG Range of Nitrogen Generator is designed specifically for use as make
up and carrier gas for GC Applications, which require ultra high purity nitrogen
for operation. The NG Range is also suitable for ICP, DSC, TGA, TOD, TMA and
DMA.
Features & Benefits:
Flexible – Available with or without internal air compressor
Purity – Ultra high purity suitable for GC, ICP, DSC, TGA, TOD, TMA and DMA
Single Source Solution – One generator can cater for the requirements of multiple applications
Reliable – Self re- generating CMS column
Simple Installation – Generator designed as plug & play system
Economical – More cost effective than any other gas supply method
Convenient – Gas on demand, no health hazards, no need to worry about running out of gas
Nitrogen Generators NG250(A) - 5000(A)
Peak Scientific’s Quality Management
System conforms to: ISO:9001:2008
Peak Scientific UK
Tel: +44 (0)141 812 8100
Fax: +44 (0)141 812 8200
Peak Scientific North America
Tel: +1 866 647 1649
Fax: +1 978 608 9503
For a full list of our worldwide office locations, please visit:
Web:
www.peakscientific.com/contact
2007
Product Accreditations:
Infinity XE 50X0 SeriesUser Manual
Copyright © Peak Scientific Instruments Ltd 2019 — Infinity XE 50X0 Series User Manual Rev 1 RSID 2750 EN —27/11/2019
Contents
Change History 3
How to use this Manual 3
Introduction 4
Warranties and Liabilities 5
Warranty & Liability Coverage 5
Safety Notices 7
Symbols 7
Safety Notice to Users 7
Declaration of Conformity 8
Environmental Declaration 9
Technical Specification 10
Unpacking 12
Fittings Kit Contents 13
Installation 14
Generator Environment 14
Generator Overview 15
General Dimensions 15
Rear Connections 16
Inlet Air Quality 16
Electrical Connection 17
Electrical Panel Access 17
Normal Operation 18
On Demand Gas 18
Eco Mode 18
Commissioning 19
Tubing Lengths 19
Service Requirements 20
Service Schedule 20
Peak Protected 21
Cleaning 22
Safely Isolated Condition 22
Safe State After Repair 22
Troubleshooting 23
Page 2
Page 3
Change HistoryRev Comment Name Date
How to use this Manual
This manual is intended for end users and has been written as a reference document where you can skip to the relevant information.
Users can refer to the contents page to find the relevant information.
Please review each of the following sections carefully.
Thank you for selecting Peak Scientific to meet your gas generation needs, and should you require any further assistance or support please do not hesitate to contact Peak Scientific or the Peak Partner from which you purchased your generator.
Page 4
Introduction
The Infinity XE 50 Series delivers performance and value beyond expectations, offering a pure and reliable source of gas. These perpetual flow nitrogen generators are suitable for a variety of applications in your laboratory.
These models provide a source of nitrogen gas with other features including:
• Single Source Solution - One generator can cater for the requirement of multiple applications & multiple laboratories on site
• Durable - Few moving parts with minimum maintenance requirements
• Quiet in operation – Silent whilst operating
• Mobile – Flexibility to position the generator where required
• Simple installation – Generator designed as a plug and play system
• Economical – More cost effective than any other gas supply method
• Convenient – Gas on demand, no hazards, no need to worry about running out of gas
• Eco-mode – Energy efficient eco-mode ensures low running costs
The Infinity XE 50 Series generators are based on proven technology, and selectively remove oxygen, moisture and other gases to leave clean, dry, phthalate free nitrogen.
To ensure these Generator models meet our high expectations with regards to reliability and performance, we have tested them extensively at our manufacturing plant and with end users around the world to ensure reliability and longevity of the system.
Page 5
Warranties and LiabilitiesWarranty & Liability Coverage
1. Peak warrants that, subject to the provisions in this statement, purchased Peak generators, whether purchased directly from Peak or indirectly via an approved, certified and trained distributor or partner (referred to hereafter as a “Peak Partner”) will comply in all material respects with any specifications referred to in your customer order confirmation and, subject to installation and operational guidelines being followed as described in applicable product manuals, shall be free from any defects in quality of materials or workmanship for a period of one year from the date of installation, provided this takes place within 3 months of factory dispatch.
2. Where the purchased generator is from the Precision Hydrogen series, Peak further warrants that, subject to installation and operational guidelines being followed as described in applicable product manuals, the hydrogen cell shall be free from any defects in quality of materials or workmanship for a total period of three years (inclusive of warranty period specified in clause 1) from date of installation, provided this takes place within 3 months of factory dispatch.
3. Where the purchased generator is from the i-Flow 6000 series, Peak further warrants that, subject to installation and operational guidelines being followed as described in applicable product manuals, the generator shall be free from any defects in quality of materials or workmanship for a total period of two years (inclusive of warranty period specified in clause 1) from the date of installation, provided this takes place within 3 months of factory dispatch and the following provisions have also been met: a. you must purchase a service plan, ensuring the generator is serviced by Peak or a Peak Partner on or before the end of the first 12 months of your ownership, and serviced at least once during each subsequent 12 month period thereafter; b. the generator (and any associated equipment) must have been commissioned by Peak or a Peak Partner; c. the feed air or inlet air supply to the generator must comply with ISO 8573-1:2010 Class 1.2.1 at all times; d. your air compressor, dryer, filtration and oil removal systems must be deemed suitable for use by Peak or a Peak Partner, and must be changed and serviced regularly, in line with the equipment manufacturer’s recommended guidelines; and e. any generator failure or fault that is deemed to have been caused by the failure of any upstream equipment, component, part or system (such as air compressor, air treatment or filtration) will be excluded from the warranty described herein.
4. Peak also warrants that any replacement parts whether purchased (directly from Peak, or via a Peak Partner) or supplied as part of any remedial action undertaken in line with the provisions of clauses 12 and 13, shall be free from any defects in quality of materials or workmanship for a period of 180 days from the date of factory dispatch, provided its installation is performed by Peak or a Peak Partner.
5. This warranty does not exclude Peak’s liability in respect of any claim for death or personal injury to any person, in so far as such can be attributed to negligence or breach of duty of care directly resulting from failure of Peak to comply with the provisions in clauses 1, 2, 3 & 4. Exclusions & Limitations
6. This warranty does not cover: a. damage, deterioration or malfunction resulting from an alteration or modification to a generator which has not been carried out by Peak or a Peak Partner; b. damage, deterioration or malfunction resulting from what Peak reasonably believes to be abuse, or misuse of a generator by you or any third party; c. liability for accident or neglect (other than pursuant to clause 5); d. maintenance or repairs which have not been carried out by Peak or a Peak Partner; e. operation of a generator or exposure of a generator to environmental conditions that fall out-with operational guidelines as specified in the applicable product user manual; and f. lightning, power surges or any other acts of God or nature.
7. This warranty is non-transferrable. Only the original owner of the generator may benefit from the terms within this statement.
Page 6
8. Peak shall not be liable in respect of any claim made for costs, damages, losses or expenses (whether consequential, direct, indirect or otherwise) or in any respect howsoever arising including, but not limited to, liability from accident or negligence (other than pursuant to clause 5) that may be suffered by you or any third party.
9. No person or entity is authorised to change the terms and conditions outlined in this warranty statement in any respect, or to create any additional obligations or liabilities for any party involved.
10. This warranty statement supersedes any and all prior warranty agreements between the parties and constitutes the complete, final and exclusive understanding of the parties with respect to the subject matter. All prior negotiations, representations, or promises, whether oral or written, of either party shall be deemed to have been merged herein.
11. If any part of this warranty statement is invalidated, for whatever reason, such part will be deleted and the rest shall remain unaffected, continuing to be in full force and effect. Delivery of Warranty Service
12. Subject to clause 13, and: a. Peak being notified by you, within the duration of the applicable warranty period, of any defect that you think is subject to any warranty valid under clauses 1, 2, 3 or 4; and b. Peak being permitted to inspect the generators, parts and their installation (along with any relevant packaging) Peak shall at its option repair or replace defective generators or parts (including, if necessary, any moving parts and irrespective of runtime). No additional charges will apply, for parts or delivery and, where applicable, labour or travel. Peak will endeavour to deliver this service within 3 working days of your notification.
13. Where, in Peak’s reasonable opinion, a defect is subject to an exclusion described in clause 6, Peak reserves the right to charge for parts or delivery and, where applicable, you may also be charged by Peak for call out, labour or travel in respect of any repair or replacement which you authorize Peak to carry out.
WARNING
WARNING
WARNING
WARNING
CAUTION
Page 7
Safety NoticesPeak Scientific Instruments cannot anticipate every possible circumstance which may represent a potential hazard. The warnings detailed within this manual refer to the most likely potential hazards, but by definition cannot be all inclusive. If the user employs an operating procedure, item of equipment or a method of working which is not specifically recommended by Peak Scientific, the user must ensure that the equipment will not be damaged or become hazardous to persons or property.
Symbols
This manual uses the following symbols to highlight specific areas important to the safe and proper use of the generator.
A WARNING notice denotes a hazard. It calls attention to an operating procedure, process or similar, which if not correctly performed or adhered to, could cause personal injury or in the worst case death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood or met.
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, process or similar, which if not correctly performed or adhered to, could cause damage to the generator or the application. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood or met.
Caution, risk of electric shock. Ensure power to the generator has been removed before proceeding.
Safety Notice to Users
These instructions must be read thoroughly and understood before installation and operation of your Peak Generator. Use of the generator in a manner not specified by Peak Scientific MAY impair the SAFETY provided by the equipment.
When handling, operating or carrying out any maintenance, personnel must employ safe engineering practices and observe all relevant local health and safety requirements and regulations. The attention of UK users is drawn to the Health and Safety at Work Act 1974, and the Institute of Electrical Engineers regulations.
If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment maybe impaired.
Page 8
Declaration of Conformity
We Peak Scientific Instruments Ltd.
Of Fountain Crescent, Inchinnan, Renfrewshire, PA4 9RE
Declare that:
Equipment: Nitrogen Gas Generator
Models: Infinity XE 50 Series
To which this declaration relates, is in conformity with the applicable EC directives, harmonized standards, and other normative requirements.
• Low Voltage Directive 2014/35/EU EN 61010-1: 2010 Electrical Equipment for measurement, control and laboratory use.
• Electromagnetic Compatibility Directive 2014/30/EU EN 61326-1: 2013 Electrical Equipment for measurement, control and laboratory use.
All evaluation, testing and certification issued by:
York EMC Services
Unit 1, Grangemouth Technology Park
Earls Road
Grangemouth
UK
FK3 8UZ
Signed: Name: Chris Pugh
Date: 22/06/2017 Position: Chief Technology Officer
Page 9
Environmental Declaration
We Peak Scientific Instruments Ltd.
Of Fountain Crescent, Inchinnan, Renfrewshire, PA4 9RE
Declare that:
Equipment: Nitrogen Gas Generator
Models: Infinity XE 50 Series
Is fully compliant with the following Directives
• 2012/19/EU WEEE (Waste of Electrical and Electronic Equipment)
• 2011/65/EU RoHS 2 (Restriction of Hazardous Substance)
Peak Scientific Instruments Ltd fully complies with its obligations towards the European WEEE (Waste of Electrical and Electronic Equipment) Directive 2012/19/EU. These obligations are being met within the B2B compliance group.
Peak Scientific Instruments Ltd has developed all reasonable ‘due diligence’ controls to ensure that our products comply with the principles and requirements of the European recast RoHS (Restriction of Hazardous Substances) Directive 2011/65/EU. Similar directives in the United States and China, for example, have also been captured within this program.
Where a specific certificate of compliance is required, this can be requested, on a product serial number basis, directly from Peak Scientific Instruments Ltd, by contacting us through our website on www.peakscientific.com
Signed: Name: Chris Pugh
Date: 22/06/2017 Position: Chief Technology Officer
Page 10
Technical SpecificationInfinity XE 5010 & 5020Environment
5010 5020
Minimum Operating Ambient Temperature 5°C (41°F)Maximum Operating Ambient Temperature 30°C (86°F)Maximum Relative Humidity 80%
* Maximum safety ambient temperature 30°C** When taken out of storage the generator should be allowed to acclimatize at room temperature for a
minimum of 3 hours before operation.
Inlet Conditions
Minimum Air Inlet Pressure 4.1 bar (60 psi)Maximum Air Inlet Pressure 9.9 bar (145 psi)Minimum Air Inlet Flow Dependant on required flow and purityMinimum Air Quality ISO 8573-1:2010 Class [1.4.1]
Generator Outlets
Maximum Pressure Drop (Inlet-Outlet) 0.7bar (10psi)Maximum Outlet Flow (95% Nitrogen) 130 L/min 260 L/minOutlet Nitrogen Purity 95 - 99.5%Start-Up Time 30 minutesParticles <0.01µmPressure Gauges 2
Electrical Requirements
Voltage 100-230VFrequency 60 Hz / 50 HzCurrent 0.4A at 100V / 0.174A at 230VInput Connection C20 PlugPower Cord (Supplied) C19 socket to local connection (13A min 16A Max)Pollution Degree 2Insulation Category I
General
Dimensions cm (inches) WxDxH 50 x 50 x 99 (19.6 x 19.6 x 39.0)Generator Weight Kg (lbs) 50.4 kg (111 lb) 54.6 kg (120 lb)Shipping Weight Kg (lbs) 67 kg (148 lb) 72 kg (159 lb)
Page 11
Technical SpecificationInfinity XE 5030 & 5040Environment
5030 5040
Minimum Operating Ambient Temperature 5°C (41°F)Maximum Operating Ambient Temperature 30°C (86°F)Maximum Relative Humidity 80%
* Maximum safety ambient temperature 30°C** When taken out of storage the generator should be allowed to acclimatize at room temperature for a
minimum of 3 hours before operation.
Inlet Conditions
Minimum Air Inlet Pressure 4.1 bar (60 psi)Maximum Air Inlet Pressure 9.9 bar (145 psi)Minimum Air Inlet Flow Dependant on required flow and purityMinimum Air Quality ISO 8573-1:2010 Class [1.4.1]
Generator Outlets
Maximum Pressure Drop (Inlet-Outlet) 1.4bar (20psi)Maximum Outlet Flow (95% Nitrogen) 390 L/min 520 L/minOutlet Nitrogen Purity 95 - 99.5%Start-Up Time 30 minutesParticles <0.01µmPressure Gauges 2
Electrical Requirements
Voltage 100-230VFrequency 60 Hz / 50 HzCurrent 0.4A at 100V / 0.174A at 230VInput Connection C20 PlugPower Cord (Supplied) C19 socket to local connection (13A min 16A Max)Pollution Degree 2Insulation Category I
General
Dimensions cm (inches) WxDxH 50 x 50 x 99 (19.6 x 19.6 x 39.0)Generator Weight Kg (lbs) 58.8 kg (130 lb) 63 kg (138 lb)Shipping Weight Kg (lbs) 77 kg (170 lb) 82 kg (180 lb)
Page 12
UnpackingAlthough Peak Scientific takes every precaution with safe transit and packaging, it is advisable to fully inspect the unit for any sign of transit damage.
Check ‘SHOCKWATCH’ and ‘TIP-N-TELL’ labels for signs of rough handling prior to unpacking
Any damage should be reported immediately to the carrier and Peak Scientific or the Peak Partner from where the unit was purchased.
Follow the unpacking instructions posted on the side of the crate. It will require two people to remove the unit from the shipping crate and to manoeuvre the generator to the desired location.
Please save the product packaging for storage or future shipment of the generator.
Note: Included with the generator is a “Fittings Kit” containing mains power leads for UK, EU & US and also all the required fittings and warranty registration card. Be careful not to discard these with the packaging.
Page 12
Page 13
Fittings Kit ContentsSupplied in the Fittings Kit are all the fittings required to connect the generator to the application. The contents of the Fittings Kit are as follows:
1. 6mm PE Tubing x 3m
2. 12mm PE Tubing x 6m
3. 6mm Push-Fit Fitting x 1
4. 12mm Push-Fit Fitting x 2*
5. 16mm Push-Fit Fitting x 1**
6. Tee fitting x 1
7. Shut-off Valve x 1
8. UK Mains Power Cable x 1
9. EU Mains Power Cable x 1
10. JP Mains Power Cable x 1
11. US Mains Power Cable 230v x 1
12. US Mains Power Cable 110v x 1
13. 4mm Hex Key x 1
14. 5mm Hex Key x 1
15. 8mm Hex Key x 1
All of the generators output ports are located on the output panel at the rear of the unit.
* x1 with 5030 & 5040
** 5030 and 5040 only
Page 14
InstallationGenerator Environment
The generator is designed for indoor use only. It should be installed adjacent to the application(s) it is supplying. If this is not convenient then the unit can be sited elsewhere, however, consideration should be made of the lengths of pipe runs as pressure drops can result from extended runs of pipe.
Performance of the generator is affected by ambient conditions. Note should also be taken to the proximity of Air Conditioning outlets. These can sometimes give rise to “pockets” of air with high relative humidity. Operation of the unit within such a pocket could adversely affect its performance. Consideration should also be given to the air flow around the unit. It is recommended that an air gap of 75mm (3”) should be maintained down both sides and at the rear of the unit. Please refer to the drawing below for the general dimensions of the unit.
Minimum Operating Ambient Temperature: 5 °C (41 °F)
Maximum Operating Ambient Temperature: 30 °C (86 °F)
Page 15
Generator OverviewGeneral Dimensions
The generator must always be placed on a flat, level surface. Failure to do so will affect the performance of the generator.
WARNING
500 mm / 19.6” 500 mm / 19.6”
99
0 m
m / 39
"
Page 16
Rear Connections
Inlet Air QualityThe Nitrogen Generator should be connected to a ISO 8573-1:2010 (1:4:1) source of compressed air. A minimum pressure of 60 psig is required for efficient operation of the Generator. Any doubts as to the suitability of your compressed air supply should be referred to the factory for advice.
A MAXIMUM OIL VAPOUR CONTENT OF 0.01 mg per m^3 is PERMISSIBLE
Air Inlet and Nitrogen Outlet connections are at the rear of the generator in the middle right side of the panel. The Air inlet connection is the right port, 12mm Push-Fit Fitting (5010,5020) 16mm Push-Fit Fitting (5030,5040), Outlet port is 12mm Push-Fit Fitting
12mm Push-Fit Fitting (5010,5020) 16mm Push-Fit Fitting (5030,5040)
Mains Cable
6mm Push-Fit FittingTee Fitting/Shut-Off Valve/12mm Push-Fit Fitting
Page 17
WARNING
Electrical ConnectionConnect the generator to an appropriate 85-264 volt single-phase supply, refer to the generator serial plate for input specification and ensure your supply matches the requirements.
If the appropriate power cord is not supplied; a new plug, rated to at least 12 amps, can be fitted by a qualified electrician.
This unit is classified as SAFETY CLASS 1. THIS UNIT MUST BE EARTHED. Before connecting the unit to the mains supply, please check the information on the serial plate. The mains supply must be of the stated AC voltage and frequency.
EARTH/GROUND (E):- Green & Yellow or Green
LIVE (L):- Brown or Black
Neutral (N):- Blue or White
Electrical Panel Access
Changing settings within this enclosure may adversely affect the operation of the generator. This enclosure should not be accessed unless by or under direct instruction from a trained individual.
WARNING
Page 18
Normal OperationThe Infinity XE 50 Series Gas Generators are designed specifically to minimize operator involvement. Given that the systems are installed as described in earlier sections and is serviced in accordance with the specified maintenance recommendations (see Service Requirements), then it should simply be a matter of turning the Generator on when it is required.
The Generator will automatically produce the factory set flow and pressure as detailed in the Technical Specifications.
On Demand GasThe generator will produce nitrogen on demand. If the application is operating and requires a gas flow, the system it is connected to will supply this to suit the requirements of the application. If the application requirement for gas stops, the system will also stop. If the demand from the application starts again, the system will detect the demand for gas and will automatically start again to suit the demand.
Eco ModeWhen demand from the instrument is below 1.2L/min the Infinity XE generator will stop demanding air from the external compressor. If demand from the application starts again, the system will detect the demand for gas and will automatically start up again.
Page 19
CommissioningWith the generator installed as previously described, connect the application to the Nitrogen output port and open the air supply to the unit. On reaching operational pressure the generator will produce Nitrogen at the customer-specified purity. The design of the generator is such that it will deliver up to the rated output flow of Nitrogen, determined by demand of the consuming equipment. Rated flow, purity and pressure are customer-specific and dependent on inlet conditions being met as detailed on the individual generator
The design of the generator is that it will deliver variable purity dependant on the inlet pressure and outlet flow rate. Outlet pressure may be adjusted by the pressure regulator on the rear of the unit, allowing for up to 20 psi pressure drop across the system. Should the demand for the nitrogen stop, the generator will automatically go into "eco-mode" (refer to page 18), shutting off the air inlet. Once the demand restarts the generator automaticcaly resumes producing Nitrogen. The generator is protected from over pressurising.
Tubing LengthsThe diameter of the tubing which will be connected to the gas outlet is important and is determined by the length of tubing required. Failure to follow these recommendations could lead to pressure between generator and application.
> 10 - 40 meters: Use 12/10 (12mm O/D, 10mm I/D) Tubing.
> 40 metres: Please contact Peak Scientific with the relevant distance andwe will calculate the flow resistance and the tubing size required.
The imperial equivalents are: 12/10 = 1/2” O/D, 3/8” I/D.
Air input tubing for 5030 & 5040 is not supplied. 16mm OD, 12mm ID PTFE Tubing is reccomended.
CAUTION
Page 20
Service RequirementsService SchedulePurchase Interval Component Visit
12 Months Infinity XE 50 Series Annual Maintenance Kit www.peakscientific.com/ordering
Page 21
Peak ProtectedWith Peak Scientific you invest in not only a product but peace of mind. With a network of certified Peak engineers stationed throughout the globe, Peak’s rapid response team are never far away and our commitment is to keep your generator running day in, day out, protecting your laboratory workflow.
[Peak Protected] can provide…
Installation A dedicated Peak engineer will visit
your lab to install and setup your generator
IQ/OQCertified assurance for applications requiring documented qualification
Complete planSwift response by a Peak Service
Engineer within 72 hours & planned preventative maintenance
SparesGenuine Peak parts with express
delivery, ensuring optimal performance and lifetime
Premium ProtectedGuaranteed rapid on-site response
within 24 hours & planned preventative maintenance
Technical Support HotlineAround the clock support by phone or online with our global technical
helpdesk
To find out more about protecting your investment visit: www.peakscientific.com/protected
Page 22
CleaningClean the outside of the generator only using warm soapy water and a clean damp cloth. Ensure all excess fluid is thoroughly removed from the cloth prior to use.
Cleaning should only be undertaken with the power switched off and the power cord removed from the rear of the generator.
Under no circumstances should any solvents or abrasive cleaning solutions be used as these can contain fumes that could be harmful to the generator.
Care should be taken with Leak Detections Liquids.
Safely Isolated ConditionThe unit is in a safely isolated condition when it is disconnected from its application and fully de-pressurised. Directions for isolating the Generator are shown below.
Failure to place the Generator in a safely isolated condition when instructed to do so may lead to personal injury or injury to others and even death.
1. Switch generator off.2. Disconnect from air supply.3. Ensure the output pressure gauge reads zero. (If gauge does not fall to zero, open
the manual ball valve at the outlet port, to allow trapped gas to escape.)4. Disconnect from the application.
Safe State After RepairTo guarantee the generator is in a safe state after a repair, please ensure the following conditions are met;
• All earth leads have been reconnected. Earth leads should be connected to the earth tabs located on the door panels, as shown below.
• The electricity cable has been correctly refitted. The mains cable should be plugged into the IEC power inlet located on the rear panel of the generator. This unit is classified as SAFETY CLASS 1. THIS UNIT MUST BE EARTHED.
CAUTION
CAUTION
WARNING
Page 23
TroubleshootingProblem Possible Solution
The Generator will not switch on and the power switch does not illuminate.
• Ensure power cord is plugged into the Generator and that the power socket is turned on.
• Check the fuse in the power cord plug.
• Contact your service provider.
The Generator will not switch on but the power switch is illuminated.
• Disconnect power cord from the rear of the Generator. Open the left hand panel are check that the circuit breaker is turned on (switch in the up position). Reconnect power cord.
• Contact your service provider.
The output pressure has dropped below its original set point.
• Check the pressure regulator at the back of the unit has not been adjusted.
• Check to see if the inlet pressure gauge has dropped below the set air supply pressure.
• If so, check your air supply does not have a pressure drop over the length of tubing and has no leaks.
• Contact your service provider.
Page 24
Go Online or Complete and ReturnYou can register for your FREE 12 month Warranty with ease online at www.peakscientific/protected.
Alternatively, you can send the completed form to Peak Scientific by post or email at [email protected].
Product Warranty RegistrationContact name
Email address
Company
Address
City/town
Postcode
Country
Telephone
Generator serial #
Model type
Installation date
Do you still use an alternative gas solution i.e. cylinders or bulk liquid?
Yes No
What gas requirements do you have in your lab? Hydrogen Nitrogen Zero Air
Extend your cover withPeak Scientific offer comprehensive gas generator after sales support packages. Peak [Protected] aftercare support can guarantee an on-site response within 72 hours*, genuine parts from our ISO9001 approved factory and a 95% first-time fix rate. See our enclosed Peak [Protected] leaflet for further information.
Important!You have 1 month to register your Peak Scientific product from the date of installation. Once registered the warranty will be honoured for a period of 12 months. If you wish to defer the installation of your generator, you must notify Peak Scientific immediately by emailing [email protected]. For generators that remain unregistered after 1 month from the shipment date, the warranty will be considered active from the date of factory dispatch.
* Complete Plan only
Go Online or Complete and ReturnWe know that registering any of your recently purchased products is not the first thing on your mind- but it is very important to both of us. Not all warranties are alike and Peak Scientific stand out against other gas suppliers as we offer a comprehensive, quick response, on-site warranty. This means that in the very unlikely case that your gas generator develops a fault we have rapid support teams on-hand around the world who are able to come to your lab and get you back up and running in no time.
Register for your comprehensive 12 month on-site warranty with ease online at www.peakscientific.com/protected.
Alternatively, you can send the completed form to Peak Scientific by post or email at [email protected].
Important!You have 1 month to register your Peak Scientific product from the date of installation. Once registered the warranty will be honoured for a period of 12 months. If you wish to defer the installation of your generator, you must notify Peak Scientific immediately by emailing [email protected]. For generators that remain unregistered after 1 month from the shipment date, the warranty will be considered active from the date of factory dispatch.
Page 25
Page 26
Page 27
Peak ScientificFountain CrescentInchinnan Business ParkInchinnanPA4 9REScotland, UKTel: +44 141 812 8100
For further information on any of our generator products please contact [email protected]
Peak Scientific has highly trained, fully certified Field Service Engineers located in over 20 countries
across every continent around the world. This allows us to provide an industry-leading rapid response
service to our customers. With [Peak Protected], your laboratory’s productivity becomes our top
priority.
To discuss Peak Protected generator cover and payment options speak to your local Peak
Representative or for further information contact: [email protected]
Copyright © Peak Scientific Instruments Ltd 2019 — Infinity XE 50X0 Series User Manual Rev 1 RSID 2750 EN —27/11/2019
Service Kit :
Copyright 2019 Peak Scientifi c - Rev - Infi nity XE 5011-5041 Data sheet � Rev 7� RSID 1974 � EN - 09/01/19
Your local gas generation partner
Part Number :
Description
See Reverse See Reverse
Key Features
Applications
Producing between 10 and 520 L/min of high purity nitrogen gas, Infi nity XE 50 series gas generators can comfortably supply multiple laboratory instruments. O� ering variable N2 purities (95- 99.5%) with a built in purity fail-safe, you can now e� ectively meet the demands of numerous applications such as ELSD, NMR & Glovebox, as well as LC-MS. With no moving parts or mechanical sound, our Infi nity range is easy to service and requires minimal maintenance.
Peak Scientifi c's Infi nity XE series generators have been engineered to provide nitrogen to laboratories that utilize an external source of compressed air, which meets a minimum quality grade of ISO 8573-1:2010 Class 1.4.1. Where required, Peak can provide assistance on external compressors & pre/post fi ltration as part of delivering a complete solution for your needs.
Infi nity XE 5011-5041Single nitrogen source for multiple applications
) Flow rates ranging from 10-520 L/min*
) Variable N2 purities, ranging from 95-99.5%*
) Maximum output pressure of 135 psi
) Built upon tried & tested membrane technology
) Compact, space saving design
) Purity safeguard, warning & auto-shutdown
* Dependent on inlet pressure, inlet/outlet fl ow & inlet air quality
) LC-MS
) ELSD
) Sample Evaporators
) NMR
) FT/MS
) Glovebox
INFINITY XE 5011-5041 Dimensions
500.00mm 500.00mm
995.00mm
Peak Scientifi c UKTel: +44 (0)141 812 8100Fax: +44 (0)141 812 8200
Peak Scientifi c North America
Tel: +1 866 647 1649Fax: +1 978 608 9503
For a full list of our worldwide o� ce locations, please visit:
Web: www.peakscientifi c.comEmail: marketing@peakscientifi c.com
Product Certifi cationsPeak Scientifi c’s Quality Management System conforms to: ISO:9001:2008
Copyright 2019 Peak Scientifi c - Rev - Infi nity XE 5011-5041 Data sheet � Rev 7� RSID 1974 � EN - 09/01/19
Technical Specifi cations 5011 5021 5031 5041
Flow Rate Min - Max (L/min)* 10-130 20-260 30-390 40-520
Purity N2 Min - Max * 95%-99.5%
Max Outlet Pressure 135 psi
Inlet Air Quality Requirement ISO 8753-1:2010 (1:4:1)
Inlet Air Pressure Requirement 60-145 psi (depending on outlet pressure requirements)
Voltage 100-230 VAC +/- 10%
Frequency 50/60 Hz
Current 0.4A @ 100V / 0.174A @ 230V
Input ConnectionW C20 Plug
Power Cord (Supplied) C19 Socket to local connection
Power Consumption 40 Watts
Operating Temperature 5°C - 30°C / 41°F - 86°F
Pollution Degree 2
Installation Category II
Dimensions mm (inches) W x D x H 500 (19.7) x 500 (19.7) x 995 (39.2)
Weight kg (lbs) 75kg (165lbs) 80kg (176lbs) 85kg (187lbs) 90kg (198lbs)
Shipping weight kg (lbs) 91kg (200lbs) 96kg (211lbs) 101kg (222lbs) 106kg (233lbs)
*Dependant on inlet pressure, inlet/outlet fl ow and Inlet Air Quality
Ordering Information
Part Number 3300189 3300190 3300191 3300192
Annual Service Kit
Visit: www.peakscientifi c.com/orderingFixed Price Preventative Maintenance
Complete/Standard Maintenance Plan
Peak Scientifi c gas generators defi ne the benchmark in reliability, convenience and performance in laboratories around the world, and come backed by a 12 month warranty. Beyond this period however you can ensure that you’re investment continues to be [Protected] by our comprehensive generator care cover.
Our world-class aftercare support packages deliver a program of scheduled preventative maintenance whilst giving you the reassurance of instant access to worldwide technical support and priority on-site response in the untimely event of a breakdown.
Key Features
i-FlowLab 6XX2 Dimensions
Description
Applications
Copyright 2018 Peak Scientific - Rev - i-FlowLab 6XX2 datasheet Rev 3 RSID 1269 EN - 10/07/18
Service Kit :Part Number :
760 mm (29.92 inch) 160 mm(6.29 inch)
500.00 mm (19.68 inch)1738 m
m (68.42 Inch)
) Real-time monitoring of purity (in ppm) provided by an integrated Oxygen Analyzer
) Modular design allowing for columns to be added as your requirements grow
) Sustainable on-site gas generation, no need for storage of high volumes of bulk gas supply
) Consistent gas purity, so no fluctuations as can be experienced with bulk gas storage
) Minimal maintenance, only scheduled preventative maintenance required to ensure long life and performance of the system
) A safer solution with no storing of highly pressurized gas on-site
) Reduce carbon footprint by removing the requirement for bulk gas supply delivery
) Comprehensive installation including compressed air supply, pipework, pre-filtration system, nitrogen generator and storage
) Multiple LC-MS instruments
) Sample Concentrators
) Whole lab solution
) FT-IR/Glove Boxes/TOC
) Larger scale laboratories
i-FlowLab nitrogen generators have been engineered by Peak Scientific gas generation specialists to meet the supply needs of laboratories which require a large capacity, on-site nitrogen supply. Harnessing the latest gas purification technologies i-FlowLab 6XX2 removes oxygen from air, delivering nitrogen at flow rates from 40 l/min to 989 l/min. Available in over 100 pre-configured flow and purity specifications, i-FlowLab is modular with the option to add columns to increase the flow rate as your nitrogen requirements grow.
The 6XX2 range measures generator purity output as oxygen content in ppm, as opposed to its 6XX1 counterpart which measures oxygen content as a percentage. Integrated with the compressed air pre-filtration system, PureAir, ensures the highest quality nitrogen output by cleaning air before it enters the i-FlowLab generator.
Contact us for detailsSee back page
i-FlowLab 6XX2 Scalable, high-flow, nitrogen generation solution for laboratories
Peak HQTel: +44 (0)141 812 8100Fax: +44 (0)141 812 8200
Peak North America
Tel: +1 866 647 1649Fax: +1 978 608 9503
For a full list of our worldwide office locations, please visit:
Web: www.peakscientific.comEmail: [email protected]
Peak Scientific’s Quality Management System conforms to: ISO:9001:2008
Copyright 2018 Peak Scientific - Rev - i-FlowLab 6XX2 datasheet Rev 3 RSID 1269 EN - 10/07/18
Product Certifications
Technical Specifications LPM *
Product Name 6012 6022 6032 6042 6052 6062 6072 6082 6092 6102
Oxygen Content 5ppm 30 62 95 113 137 173 199 225 248 278
Oxygen Content 10ppm 40 80 120 150 182 230 265 300 330 370
Oxygen Content 50ppm 58 116 174 230 285 345 400 455 510 565
Oxygen Content 100ppm 65 135 202 270 335 405 462 532 591 644
Oxygen Content 500ppm 89 190 283 362 452 545 634 724 812 900
Oxygen Content 1000ppm 105 205 305 406 508 600 700 795 890 989
Voltage 100-240 VAC +/- 10%
Frequency 50/60 Hz
Current 2.0 - 1.0 Amp
Input Connection C20 Plug
Power Cord (Supplied) C19 Socket to local connection
Power Consumption 250 Watts
Operating Temperature 5°C - 50°C / 41°F - 122°F
Pollution Degree 2
Installation Category II
Width mm (inch) 500 (19.68)
Height mm (inch) 1738 (68.42)
Depth mm (inch)760
(29.92)
920
(36.22)
1080
(42.52)
1240
(42.52)
1400
(55.12)
1560
(61.42)
1720
(67.72)
1880
(74.02)
2040
(80.31)
2200
(86.61)
Weight kg (lbs)197
(433)
282
(620)
367
(807)
452
(994)
537
(1181)
622
(1368)
707
(1555)
792
(1742)
877
(1929)
962
(2116)
Shipping weight kg (lbs)277
(609)
364
(801)
452
(992)
538
(1184)
625
(1375)
712
(1566)
799
(1758)
886
(1949)
973
(2141)
1060
(2333)
Noise Level 59dBa @ 1m
*Performance data is based on 7 bar (G) inlet air pressure and 20 - 25 deg C ambient temperature.
(Flow reference conditions, 20 deg C, 1013 millibar (a), 0% Relative Humidity)
Ordering Information
Part Number 20-6012 20-6022 20-6032 20-6042 20-6052 20-6062 20-6072 20-6082 20-6092 20-6102
Service Kit 08-0318 08-0319
Complete Plan 09-3707 09-3708
Peak Scientific gas generators define the benchmark in reliability, convenience and performance in laboratories around the world, and come backed by a 12 month warranty. Beyond this period however you can ensure that you’re investment continues to be [Protected] by our comprehensive generator care cover. Our world-class aftercare support packages deliver a program of scheduled preventative maintenance whilst giving you the reassurance of instant access to worldwide technical support and priority on-site response in the untimely event of a breakdown.
Key Features
i-FlowLab 6XX1 Dimensions
Description
Applications
Copyright 2020 Peak Scientific - Rev - i-FlowLab 6XX1 datasheet Rev 4 RSID 1268 EN - 07/10/20
Service Kit :Part Number :
760 mm (29.92 inch) 160 mm(6.29 inch)
500.00 mm (19.68 inch)1738 m
m (68.42 Inch)
) Oxygen Anaylzer provides continuous real-time monitoring of purity produced in %
) Modular design allowing for columns to be added as your requirements grow
) Sustainable on-site gas generation, no need for storage of high volumes of bulk gas supply
) Consistent gas purity, so no fluctuations as can be experienced with bulk gas storage
) Minimal maintenance, only scheduled preventative maintenance required to ensure long life and performance of the system
) A safer solution with no storing of highly pressurized gas on-site
) Remote monitoring, Live product status information at your fingertips, wherever you are
) Reduce carbon footprint by removing the requirement for bulk gas supply delivery
) Comprehensive installation including compressed air supply, pipework, pre-filtration system, nitrogen generator and storage
) Multiple LC-MS instruments
) Sample Concentrators
) Whole lab solution
) FT-IR/Glove Boxes/TOC
) Larger scale laboratories
i-FlowLab nitrogen generators have been engineered by Peak
Scientific gas generation specialists to meet the supply needs of
laboratories which require a large capacity, on-site nitrogen supply.
Harnessing the latest gas purification technologies i-FlowLab 6XX1
removes oxygen from air, delivering nitrogen at flow rates from 155 l/min
to 3402 l/min. Available in over 100 pre-configured flow and purity
specifications, i-FlowLab is modular with the option to add columns
to increase the flow rate as your nitrogen requirements grow.
The 6XX1 range measures generator purity output as oxygen
content in percentage, as opposed to its 6XX2 counterpart which
measures oxygen content in ppm. Integrated with the compressed
air pre-filtration system, PureAir, ensures the highest quality nitrogen
output by cleaning air before it enters the i-FlowLab generator.
Contact us for detailsSee back page
i-FlowLab 6XX1 Scalable, high-flow, nitrogen generation solution for laboratories
Peak HQTel: +44 (0)141 812 8100Fax: +44 (0)141 812 8200
Peak North America
Tel: +1 866 647 1649Fax: +1 978 608 9503
For a full list of our worldwide office locations, please visit:
Web: www.peakscientific.comEmail: [email protected]
Peak Scientific’s Quality Management System conforms to: ISO:9001:2008
Copyright 2020 Peak Scientific - Rev - i-FlowLab 6XX1 datasheet Rev 4 RSID 1268 EN - 07/10/20
Product Certifications
Technical Specifications LPM *
Product Name 6011 6021 6031 6041 6051 6061 6071 6081 6091 6101
Oxygen Content - 0.50% 155 305 450 585 730 874 1015 1160 1300 1450
Oxygen Content - 1% 190 380 530 725 885 1062 1238 1415 1574 1748
Oxygen Content - 2% 245 490 665 845 1065 1280 1480 1685 1885 2090
Oxygen Content - 3% 305 578 810 1025 1278 1574 1815 2065 2330 2580
Oxygen Content - 4% 362 651 915 1115 1420 1704 1988 2272 2556 2840
Oxygen Content - 5% 402 750 1100 1385 1649 2045 2478 2778 3045 3402
Voltage 100-240 VAC +/- 10%
Frequency 50/60 Hz
Current 2.0 - 1.0 Amp
Input Connection C20 Plug
Power Cord (Supplied) C19 Socket to local connection
Power Consumption 250 Watts
Operating Temperature 5°C - 50°C / 41°F - 122°F
Pollution Degree 2
Installation Category II
Width mm (inch) 500 (19.68)
Height mm (inch) 1738 (68.42)
Depth mm (inch)760
(29.92)
920
(36.22)
1080
(42.52)
1240
(42.52)
1400
(55.12)
1560
(61.42)
1720
(67.72)
1880
(74.02)
2040
(80.31)
2200
(86.61)
Weight kg (lbs)197
(433)
282
(620)
367
(807)
452
(994)
537
(1181)
622
(1368)
707
(1555)
792
(1742)
877
(1929)
962
(2116)
Shipping weight kg (lbs)277
(609)
364
(801)
452
(992)
538
(1184)
625
(1375)
712
(1566)
799
(1758)
886
(1949)
973
(2141)
1060
(2333)
Noise Level 59dBa @ 1m
*Performance data is based on 7 bar (G) inlet air pressure and 20 - 25 deg C ambient temperature.
(Flow reference conditions, 20 deg C, 1013 millibar (a), 0% Relative Humidity)
Ordering Information
Part Number 20-6011 20-6021 20-6031 20-6041 20-6051 20-6061 20-6071 20-6081 20-6091 20-6101
Service Kit 08-0318 08-0319
Complete Plan 09-3705 09-3706
Peak Scientific gas generators define the benchmark in reliability, convenience and performance in laboratories around the
world, and come backed by a 12 month warranty. Beyond this period however you can ensure that you’re investment continues
to be [Protected] by our comprehensive generator care cover.
Our world-class aftercare support packages deliver a program of scheduled preventative maintenance whilst giving you the
reassurance of instant access to worldwide technical support and priority on-site response in the untimely event of a breakdown.
Key Features
Applications
Copyright © Peak Scientifi c Instruments Ltd 2020 — Genius XE Data sheet Rev 7 RSID 2463 EN —10/03/2020
99-999999-9999 Service Kit :Part Number :Your local gas generation partner
Description
Inspired by the success of our best-selling Genius line of nitrogen
gas generators for LC-MS, Genius XE Nitrogen is a cutting-edge
evolution combining advanced technology with refined and robust
engineering. With two models - XE 35 (up to 35 L/min) and XE
70 (up to 70 L/min) - Genius XE Nitrogen provides a premium
standalone nitrogen solution for high performance LC-MS and
other mission-critical laboratory applications where performance
and reliability are paramount.
Featuring Multi-Stage Purification™ and next-generation
integrated compressors with Electronic Compressor Optimization™ (ECO) technology, Genius XE delivers factory
certifiable purity up to 99.5% on-demand, 24/7 with a convenient
fixed annual service interval.
) Variable fl ow up to 70 LPM
) Variable pressure up to 116 psi
) Multi-Stage Purifi cation™ producing analytical grade nitrogen gas up to 99.5% purity
) 2 year comprehensive manufacturer’s warranty*
) Touch-screen full colour user interface for ease of operation
) ECO (Electronic Compressor Optimisation™) technology for low energy consumption and compressor durability
) Next-generation high performance premium compressors, engineered exclusively for Genius XE
) Smaller & quieter than previous generation of Genius
) On-board service scheduling and system diagnostics
) Additional compressor capacity for lower fl ow applications (XE 70 only)
On-demand nitrogen source for:
) LC-MS including new high-sensitivity instruments
) ELSD
) Small sample evaporators
) Other laboratory applications requiring nitrogen
Genius XE Nitrogen High performance nitrogen generator for LC-MS
710 mm / 28”570 mm / 22.5”
650
mm
/ 2
5.6
”10
00
mm
/ 3
9.4
”
Peak Scientifi c’s Quality Management System conforms to: ISO 9001:2015
Product CertificationsPeak Scientifi c UKTel: +44 (0)141 812 8100Fax: +44 (0)141 812 8200
Peak Scientifi c North America
Tel: +1 866 647 1649Fax: +1 978 608 9503
For a full list of our worldwide o� ce locations, please visit:www.peakscientifi c.com/o� ces
Web: www.peakscientifi c.comEmail: discover@peakscientifi c.com
Copyright 2020 Peak Scientifi c - Rev - Genius XE Data sheet Rev 7 RSID 2463 EN - 10/03/20
Technical Specifi cations Genius XE 35 Genius XE 70
Max Flow Rate up to 35 L/min up to 70 L/min
Max Pressure 116 psi (8 bar) 116 psi (8 bar)
Gas Outlets 1 x ¼” BSPP 1 x ¼” BSPP
Purity** 95-99.5% 95-99.5%
Max Relative Humidity 80% RH @ 31°C 80% RH @ 31°C
Max Altitude 3000m 3000m
Particles < 0.01 µm < 0.01 µm
Phthalates Phthalate & BHT Free Phthalate & BHT Free
Suspended Liquids None None
Non-Methane Hydrocarbon Content < 1 ppm NMHC < 1 ppm NMHC
Operating Temperature 5°C (41°F) to 35°C (95°F) 5°C (41°F) to 35°C (95°F)
Electrical Requirements 120VAC, 60Hz, 12A or 230VAC, 50/60Hz, 8A 230VAC, 50/60Hz, 11.3A
Power Consumption 960 VA (120V)/1265 VA (230V) 2530 VA
Heat Output 4118 BTU 8785 BTU
Noise Level† 56 dB 59 dB
Generator Dimensions (H x W x D) 650 x 570 x 710 mm25.6 x 22.5 x 28 ”
1000 x 570 x 710 mm39.4 x 22.5 x 28 ”
Generator Weight 92 kg / 202 lbs 147 kg / 325 lbs
Ordering Information
Part Number 3300252 (230VAC) /3300807 (120VAC) 3300253
Annual Servicevisit: www.peakscientifi c.com/ordering
Complete Maintenance Plan
** Factory certifi ed purity with respect to O2 content. Actual purity varies in relation to fl ow (min. 95% at maximum fl ow output)† Noise level expressed as SPL (Sound Pressure Level) measured at 1m from source in a reverberant chamber
Peak Scientific gas generators define the benchmark in reliability, convenience and performance in laboratories around the
world, and come backed by a 2 year warranty*. Beyond this period however you can ensure that your investment continues to
be [Protected] by our comprehensive generator care cover.
Our world-class aftercare support packages deliver a program of scheduled preventative maintenance whilst giving you
the reassurance of instant access to worldwide technical support and priority on-site response in the untimely event of a
breakdown.
*Year 2 of warranty subject to generator being serviced at end of year 1 by a Peak-approved agent in accordance with fi xed annual maintenance schedule.
For full terms and conditions visit www.peakscientifi c.com/warranty-statement
Key Features
Applications
Copyright © Peak Scientifi c Instruments Ltd 2020 — Genius XE Data sheet Rev 7 RSID 2463 EN —10/03/2020
99-999999-9999 Service Kit :Part Number :Your local gas generation partner
Description
Inspired by the success of our best-selling Genius line of nitrogen
gas generators for LC-MS, Genius XE Nitrogen is a cutting-edge
evolution combining advanced technology with refined and robust
engineering. With two models - XE 35 (up to 35 L/min) and XE
70 (up to 70 L/min) - Genius XE Nitrogen provides a premium
standalone nitrogen solution for high performance LC-MS and
other mission-critical laboratory applications where performance
and reliability are paramount.
Featuring Multi-Stage Purification™ and next-generation
integrated compressors with Electronic Compressor Optimization™ (ECO) technology, Genius XE delivers factory
certifiable purity up to 99.5% on-demand, 24/7 with a convenient
fixed annual service interval.
) Variable fl ow up to 70 LPM
) Variable pressure up to 116 psi
) Multi-Stage Purifi cation™ producing analytical grade nitrogen gas up to 99.5% purity
) 2 year comprehensive manufacturer’s warranty*
) Touch-screen full colour user interface for ease of operation
) ECO (Electronic Compressor Optimisation™) technology for low energy consumption and compressor durability
) Next-generation high performance premium compressors, engineered exclusively for Genius XE
) Smaller & quieter than previous generation of Genius
) On-board service scheduling and system diagnostics
) Additional compressor capacity for lower fl ow applications (XE 70 only)
On-demand nitrogen source for:
) LC-MS including new high-sensitivity instruments
) ELSD
) Small sample evaporators
) Other laboratory applications requiring nitrogen
Genius XE Nitrogen High performance nitrogen generator for LC-MS
710 mm / 28”570 mm / 22.5”
650
mm
/ 2
5.6
”10
00
mm
/ 3
9.4
”
Peak Scientifi c’s Quality Management System conforms to: ISO 9001:2015
Product CertificationsPeak Scientifi c UKTel: +44 (0)141 812 8100Fax: +44 (0)141 812 8200
Peak Scientifi c North America
Tel: +1 866 647 1649Fax: +1 978 608 9503
For a full list of our worldwide o� ce locations, please visit:www.peakscientifi c.com/o� ces
Web: www.peakscientifi c.comEmail: discover@peakscientifi c.com
Copyright 2020 Peak Scientifi c - Rev - Genius XE Data sheet Rev 7 RSID 2463 EN - 10/03/20
Technical Specifi cations Genius XE 35 Genius XE 70
Max Flow Rate up to 35 L/min up to 70 L/min
Max Pressure 116 psi (8 bar) 116 psi (8 bar)
Gas Outlets 1 x ¼” BSPP 1 x ¼” BSPP
Purity** 95-99.5% 95-99.5%
Max Relative Humidity 80% RH @ 31°C 80% RH @ 31°C
Max Altitude 3000m 3000m
Particles < 0.01 µm < 0.01 µm
Phthalates Phthalate & BHT Free Phthalate & BHT Free
Suspended Liquids None None
Non-Methane Hydrocarbon Content < 1 ppm NMHC < 1 ppm NMHC
Operating Temperature 5°C (41°F) to 35°C (95°F) 5°C (41°F) to 35°C (95°F)
Electrical Requirements 120VAC, 60Hz, 12A or 230VAC, 50/60Hz, 8A 230VAC, 50/60Hz, 11.3A
Power Consumption 960 VA (120V)/1265 VA (230V) 2530 VA
Heat Output 4118 BTU 8785 BTU
Noise Level† 56 dB 59 dB
Generator Dimensions (H x W x D) 650 x 570 x 710 mm25.6 x 22.5 x 28 ”
1000 x 570 x 710 mm39.4 x 22.5 x 28 ”
Generator Weight 92 kg / 202 lbs 147 kg / 325 lbs
Ordering Information
Part Number 3300252 (230VAC) /3300807 (120VAC) 3300253
Annual Servicevisit: www.peakscientifi c.com/ordering
Complete Maintenance Plan
** Factory certifi ed purity with respect to O2 content. Actual purity varies in relation to fl ow (min. 95% at maximum fl ow output)† Noise level expressed as SPL (Sound Pressure Level) measured at 1m from source in a reverberant chamber
Peak Scientific gas generators define the benchmark in reliability, convenience and performance in laboratories around the
world, and come backed by a 2 year warranty*. Beyond this period however you can ensure that your investment continues to
be [Protected] by our comprehensive generator care cover.
Our world-class aftercare support packages deliver a program of scheduled preventative maintenance whilst giving you
the reassurance of instant access to worldwide technical support and priority on-site response in the untimely event of a
breakdown.
*Year 2 of warranty subject to generator being serviced at end of year 1 by a Peak-approved agent in accordance with fi xed annual maintenance schedule.
For full terms and conditions visit www.peakscientifi c.com/warranty-statement
Your local gas generation partner
Service Kit :Part Number :
Key Features
Copyright 2015 Peak Scientific - Corona 1010 Data Sheet Rev 7 RSID 1287 EN - 10/10/19
Description
81-4000 08-9013
Applications
Corona 1010 Nitrogen Generator
) Suitable for applications requiring high purity nitrogen
) Available with or without external compressor module
) Dedicated solution for Thermo Scientific’s Corona and Vanquish instruments
) Tried and tested membrane technology delivers high purity nitrogen
) Gas is supplied on demand so generator works to your schedule
) The most cost-effective method of gas supply
) ThermoFisher Scientific Corona Veo
) ThermoFisher Scientific Vanquish CAD
Peak’s Corona 1010 model has been developed to meet the demands of labs requiring high purity nitrogen for their applications.
Suitable primarily for Charged Aerosol Detectors, the Corona 1010 generator delivers a high-purity, filtered nitrogen supply for applications requiring flows of up to 5l/min at 80psi.
With a space-saving design, the Corona 1010 can fit seamlessly into any laboratory. Engineered with an optional, dedicated compressor module for customers with no suitable supply of in-house compressed air, the Corona system provides a flexible solution for labs with differing configurations.
Corona 1010 shown in a stack with Corona Air Compressor.
156 mm
417 mm540 mm
406 mm
Peak Scientific UKTel: +44 (0)141 812 8100Fax: +44 (0)141 812 8200
Peak Scientific North America
Tel: +1 866 647 1649Fax: +1 978 608 9503
For a full list of our worldwide office locations, please visit:
Web: www.peakscientific.comEmail: [email protected]
Peak Scientific’s Quality Management System conforms to: ISO:9001:2008
Copyright 2015 Peak Scientific - Corona 1010 Data Sheet Rev 7 RSID 1287 EN - 10/10/19
Product Certifications
Peak Scientific gas generators define the benchmark in reliability, convenience and performance in laboratories around the world, and come backed by a 12 month warranty. Beyond this period however you can ensure that you’re investment continues to be [Protected] by our comprehensive generator care cover. Our world-class aftercare support packages deliver a program of scheduled preventative maintenance whilst giving you the reassurance of instant access to worldwide technical support and priority on-site response in the untimely event of a breakdown.
Technical Specifications Corona 1010
Max Flow Rate 5 lpm
Max Pressure 80 psi / 5.5 bar
Purity <2% 02 @ 5 L/min
Gas Outlets 1 x 1/4 " BSPP
Min Air Inlet Pressure 7.6 bar / 110 psi
Min Air Inlet Flow 30 lpm
Min Inlet Air Quality ISO8573 - 1:2010 Class 1.4.1
Phthalates None
Suspended Liquids None
Start-Up Time For Purity 5 minutes
Operating Temperature 5°C - 35°C / 41°F - 95 °F
Electrical Requirements 100-230V 50/60 HZ 0.01 A
Power Consumption 10 Watts
Heat Output Up to 140 BTU/Hr
Generator Dimensions (WxDxH) 156 x 417 x 540 mm /6.2 x 16.5 x 21.3’’
Generator Weight 12 Kg (26.5 lbs)
Ordering Information
Part Number 81-4000
Annual Service 08-9013
Standard Maintenance Plan 09 - 3110
Complete Maintenance Plan 09 - 3010
Corona Air Compressor 230V 110VPart Number 85-0555 85-1555
Annual Service 08-8352
Generator Dimensions (WxDxH) 406 x 417 x 540 mm / 16 x 16.5 x 21.3’’
Generator Weight 41.5 Kg (91.5 lbs)