Transcript
Page 1: Metrohm Autolab - ecochemie.nl · Metrohm Autolab website. • Install the Edit Serial Number application. • Switch on the instrument, connect the instrument to the computer and

Autolab Service Document # 008 – Replacing the USB DIO Decoder

Metrohm Autolab Service Document # 008

Replacing the USB DIO Decoder

Table of Contents

1. Scope of delivery .............................................................................................. 2 2. Replacing the USB DIO Decoder on Autolab Series 7 instruments ..................... 3

2.1. Removal of the USB DIO Decoder ................................................................ 3 2.2. Mounting the flat cable ............................................................................... 4 2.3. Mounting the new USB DIO Decoder .......................................................... 5

3. Replacing the USB DIO Decoder on Autolab Series 8 instruments ..................... 8 3.1. Removal of the USB DIO Decoder ................................................................ 8 3.2. Mounting the flat cable ............................................................................. 10

4. Changing the Autolab serial number in the USB DIO Decoder ........................ 10

Description: this document describes the procedure for replacing the USB DIO Decoder of the Series 7 and Series 8 instruments.

Warning

The instructions specified in this document require ESD protections to be worn at all times to avoid damaging the electronic components.

Page 2: Metrohm Autolab - ecochemie.nl · Metrohm Autolab website. • Install the Edit Serial Number application. • Switch on the instrument, connect the instrument to the computer and

Autolab Service Document # 008 – Replacing the USB DIO Decoder

1. Scope of delivery

The new USB DIO Decoder (see Figure 1) is supplied with an additional connector PCB, labeled UDD Bus Converter (see Figure 2).

Figure 1 – The USB DIO Decoder

Figure 2 – The UDD Bus Converter PCB

The UDD Bus Converter PCB is required for interfacing the new USB DIO Decoder fitted with a high density connector to the flat cable of the Autolab, if applicable. When this is required, press the metal connector of the UDD Bus Converter PCB into the connector located on the bottom of the USB DIO Decoder, as shown in Figure 3. The UDD Bus Converter PCB only fits in one way.

Page 3: Metrohm Autolab - ecochemie.nl · Metrohm Autolab website. • Install the Edit Serial Number application. • Switch on the instrument, connect the instrument to the computer and

Autolab Service Document # 008 – Replacing the USB DIO Decoder

Figure 3 – Insert the UDD Bus Converter into the bottom connector of the USB DIO Decoder

2. Replacing the USB DIO Decoder on Autolab Series 7 instruments

The Autolab Series 7 instruments produced after the year 2003 are fitted with a USB DIO Decoder. This section only applies to this type of instruments.

To open the instrument, please refer to the Install new module in a 7-series instrument.

2.1. Removal of the USB DIO Decoder

After removing the Autolab frame from the Autolab cabinet (see document Install new module in a 7-series instrument), remove the flat cable and the green phoenix connector from the USB DIO Decoder (see Figure 4).

Page 4: Metrohm Autolab - ecochemie.nl · Metrohm Autolab website. • Install the Edit Serial Number application. • Switch on the instrument, connect the instrument to the computer and

Autolab Service Document # 008 – Replacing the USB DIO Decoder

Figure 4 - Remove the flat cable and the phoenix connector

Remove the four M4x10 Phillips screws which fix the USB DIO Decoder to the Autolab rear panel. You can now take the USB DIO Decoder out of the frame.

Keep the lock-washers together (7x).

2.2. Mounting the flat cable

Insert the flat cable into the boxed header connector located on the UDD Bus Converter. The flat cable should be positioned firmly into the UDD Bus Converter connector (see Figure 5).

Figure 5 – The flat cable must be mounted firmly into the UDD Bus Converter connector

Page 5: Metrohm Autolab - ecochemie.nl · Metrohm Autolab website. • Install the Edit Serial Number application. • Switch on the instrument, connect the instrument to the computer and

Autolab Service Document # 008 – Replacing the USB DIO Decoder

2.3. Mounting the new USB DIO Decoder

Put two screws M4x10 from the rear into the back panel, below the USB-ports (see Figure 6 and Figure 7).

Figure 6 – Insert two screws below the USB ports

Figure 7 – Place two lock-washers on each screw before mounting the USB DIO Decoder

Warning

For each screw, place two lock-washers (on the inside).

Page 6: Metrohm Autolab - ecochemie.nl · Metrohm Autolab website. • Install the Edit Serial Number application. • Switch on the instrument, connect the instrument to the computer and

Autolab Service Document # 008 – Replacing the USB DIO Decoder

Put the USB DIO Decoder over the screws M4x10, and tighten the screws lightly (see Figure 8).

Figure 8 - Slide the USB DIO Decoder over the bottom screws and tighten these slightly

Put two screws M4x10 from the rear into the rear panel, above DIO P1 (see Figure 9).

Figure 9 – Insert two screws above the P1 port

Place two lock-washers (on the inside) onto the left screw (see Figure 10). Place one lock-washer (on the inside) onto the right screw (see Figure 10).

Page 7: Metrohm Autolab - ecochemie.nl · Metrohm Autolab website. • Install the Edit Serial Number application. • Switch on the instrument, connect the instrument to the computer and

Autolab Service Document # 008 – Replacing the USB DIO Decoder

Figure 10 – Place two lock washers onto the left hand side screw and one on the right hand side screw together with the ground wire

Place the (previously removed) ground wire onto the right top screw (see Figure 10).

Tighten all four screws.

Press the green phoenix connector into the matching socket on the USB DIO Decoder (see Figure 11).

Warning

Do not forget the ground wire. This wire must be placed over the right hand side screw holding the USB DIO Decoder.

Page 8: Metrohm Autolab - ecochemie.nl · Metrohm Autolab website. • Install the Edit Serial Number application. • Switch on the instrument, connect the instrument to the computer and

Autolab Service Document # 008 – Replacing the USB DIO Decoder

Figure 11 – Replace the green phoenix connector

Put the Autolab back into the instrument cabinet.

3. Replacing the USB DIO Decoder on Autolab Series 8 instruments

The Autolab Series 8 instruments are fitted with a USB DIO Decoder. This section only applies to this type of instruments. To open the instrument, please refer to the Install new module in a 8-series instrument.

3.1. Removal of the USB DIO Decoder

After removing the cover from the Autolab cabinet (see document Install new module in a 8-series instrument), remove the flat cable and the green phoenix connector from the USB DIO Decoder (see Figure 12).

Page 9: Metrohm Autolab - ecochemie.nl · Metrohm Autolab website. • Install the Edit Serial Number application. • Switch on the instrument, connect the instrument to the computer and

Autolab Service Document # 008 – Replacing the USB DIO Decoder

Figure 12 – Remove the flat cable and the phoenix connector

Remove the four sub-D screws and the M3x6 Philips screws holding the USB DIO Decoder to the Autolab rear panel (see Figure 13).

Figure 13 – Remove the Sub-D and the Philips screw to remove the USB DIO Decoder

Page 10: Metrohm Autolab - ecochemie.nl · Metrohm Autolab website. • Install the Edit Serial Number application. • Switch on the instrument, connect the instrument to the computer and

Autolab Service Document # 008 – Replacing the USB DIO Decoder

You can now take the USB DIO Decoder out of the frame.

3.2. Mounting the flat cable

The UDD Bus Converter PCB is not required for the following instruments:

• PGSTAT302N with serial numbers AUT84440 and higher • PGSTAT128N with serial numbers AUT84460 and higher • PGSTAT100N with serial numbers AUT84500 and higher

Insert the flat cable into the boxed header connector located on the UDD Bus Converter. The flat cable should be positioned firmly into the UDD Bus Converter connector (see Figure 14).

Figure 14 – The flat cable must be mounted firmly into the UDD Bus Converter connector

4. Changing the Autolab serial number in the USB DIO Decoder

After replacing the USB DIO Decoder, it might be necessary to change the Autolab serial number which is programmed in the memory of the USB DIO Decoder.

This can be done with the following procedure:

• Make sure the NOVA software is installed properly.

Note

The M3x6 screw is not present on all USB DIO Decoders.

Page 11: Metrohm Autolab - ecochemie.nl · Metrohm Autolab website. • Install the Edit Serial Number application. • Switch on the instrument, connect the instrument to the computer and

Autolab Service Document # 008 – Replacing the USB DIO Decoder

• Download the Edit Serial Number application from the Service page of the Metrohm Autolab website.

• Install the Edit Serial Number application. • Switch on the instrument, connect the instrument to the computer and start

the Edit Serial Number application.

A window will be displayed (see Figure 15) and the connection with the instrument will be established.

Figure 15 – The Edit Autolab Serial Number application

1. Adjust the parameters of the instrument where required (PGStat type, Serial number and Casing type (see Figure 16).

Page 12: Metrohm Autolab - ecochemie.nl · Metrohm Autolab website. • Install the Edit Serial Number application. • Switch on the instrument, connect the instrument to the computer and

Autolab Service Document # 008 – Replacing the USB DIO Decoder

Figure 16 – Adjust the PGStat type, Serial Number and Casing type

2. Right click the Serial number in the frame on the right-hand side and select the Write to EEPROM option from the context menu to save the changes (see Figure 17).

Note

Set the Casing to (2000 – 2006) for all 7 Series instruments and to (2007 - …) for all 8 Series instruments.

Page 13: Metrohm Autolab - ecochemie.nl · Metrohm Autolab website. • Install the Edit Serial Number application. • Switch on the instrument, connect the instrument to the computer and

Autolab Service Document # 008 – Replacing the USB DIO Decoder

Figure 17 – Saving the changes to the EEPROM

Note

The PGStat type, serial number and Casing type can be found on the sticker on the back plane of the instrument (see Figure 18).

Page 14: Metrohm Autolab - ecochemie.nl · Metrohm Autolab website. • Install the Edit Serial Number application. • Switch on the instrument, connect the instrument to the computer and

Autolab Service Document # 008 – Replacing the USB DIO Decoder

Figure 18 – The Autolab Serial number, PGStat type and Casing type can be found on the rear panel of the Autolab