EASYPAT 1600 PAT TESTER
MARTINDALEELECTRIC
Trusted by professionals
INSTRUCTIONS
16th Edition Testers
All-in-ones
Calibration Equipment
Continuity Testers
Electricians kits
Full Calibration & Repair Service
Fuse Finders
Digital Clamp Meters
Digital Multimeters
Microwave Leakage Detectors
Motor Maintenance Equipment
Non-trip loop testers
Pat testers & Accessories
Phase rotation
Proving units
Socket Testers
Thermometers & Probes
Test Leads
Voltage Indicators
and
Specialist Metrohm testers (4 & 5kV)
Specialist Drummond testers
http://www.pattesters.co.uk/8/10/2019 Martindale Epat1600 Manual 59
2/8
2 15
SAFETY INFORMATION: Always read before proceeding.
WARNING
These instructions contain information and warnings that are necessary forsafe operation and maintenance of the EasyPat. It is recommended that you
read the instructions carefully and ensure the contents are understood.Failure to understand the instructions and comply with warnings andinstructions herein can result in serious injury, damage or even death.
In order to avoid the danger of electrical shock, it is important that propersafety measures are taken when working with voltages exceeding 30V ACrms, 42V AC peak or 60V DC. Never exceed the maximum allowable inputlevel for each function and range. Refer to the specifications for maximuminputs. Never touch exposed wiring, connections or live circuits.
The EasyPat must only be used in conditions and for the purpose which ithas been constructed. Attention should be paid to safety instructions,technical specifications and use of the EastPat in dry surroundings.
Always inspect your meter, test leads and accessories for any sign ofdamage before use. If any abnormal conditions exist (e.g: broken test leads,cracked case, display not reading, etc.), do not attempt to use it. Do not
expose it to direct sunlight, excessive temperature or moisture.
Keep these instructions for future reference. Updated instructions andproduct information are available at: www.martindale-electric.co.uk
SYMBOLS:
Equipment complies with relevant EU Directives
AC (Alternating Current)
Ground
Direct Current
Equipment protected by Double Insulation (Class II)
Caution - refer to accompanying documents
Caution - risk of electric shock
8/10/2019 Martindale Epat1600 Manual 59
3/8
14 3
CONTENTS
1 Introduction
1.1 Description
1.2 Earth / Continuity (8A and 25A)
1.3 Earth / Continuity (100mA)
1.4 Insulation1.5 Fuse
1.6 Unpacking & Inspection
1.7 Spares & Accessories
2 Specifications
2.1 Electrical Specification
2.2 Mechanical Specification
2.3 Environmental Specification
3 Operation
3.1 Earth Continuity (8A and 25A)
3.2 Earth Continuity (100mA)
3.3 Insulation
3.4 Fuse
3.5 Lead Test
4 Maintenance4.1 Cleaning4.2 Calibration
4.3 Repair & Service
4.4 Storage Conditions
4.5 Warranty
USER NOTES
8/10/2019 Martindale Epat1600 Manual 59
4/8
8/10/2019 Martindale Epat1600 Manual 59
5/8
12 5
Fig. 2 Display of reading (1.00)
The instrument can be connected to a 230V supply and can test 230V or 110V
appliances.
Appliances to be tested should be connected into the appropriate test socket. If the
appliance has an on/off switch, it should be set to the on position to ensure that theappliance is fully tested.
If the value for the measured parameter is outside the range of the EasyPat 1600
then the display will show overrange. This is indicated on the display by a single 1.
with no following zeros (blank). Figure 3 shows the earth continuity display if the
measured value is greater than 1.99.
Fig. 3 Overrange Indication
NB
On the earth bond tests, overrange means above 1.99 and is a fail.
On the insulation test it means above 19.9M and is a pass.On the fuse test it means the appliance is drawing more than a few mA andis a pass.
3.5 Lead Test
Ensure the earth wander lead is connected to the EasyPat. The crocodile clip
however should not be used and must not be connected to the lead under test
nor any earth path.
For IEC LeadsPlug the lead into the appropriate EasyPat mains socket. Plug the IEC connector
into the EasyPat IEC socket.
For Extension Leads
Plug the lead into the appropriate EasyPat mains socket. Use the short 230V IEC
lead EX-332 (or optional 110V LEAD EX-331) to connect the extension lead
outlet to the EasyPat IEC socket.
Perform earth continuity and insulation tests as 3.1 & 3.3.
If earth and insulation tests have successfully passed, unplug the lead under test
from the EasyPat 1600 mains socket, leaving the IEC connector plugged into the
lead tester. Connect the lead under test to a live mains socket. Check the
illumination of Lamp 1 & Lamp 2 against the table printed on the EasyPat front
panel.
8/10/2019 Martindale Epat1600 Manual 59
6/8
8/10/2019 Martindale Epat1600 Manual 59
7/8
8/10/2019 Martindale Epat1600 Manual 59
8/8
8 9
SOCKETS
Mains: 230V 13A to BS 1363110V 16A to BS4343
Leadtest: IEC320
LAMPS
Over Temperature: Red LED which illuminates when the temperature limit
for the instrument has been exceeded. LED
extinguishes when unit cools.
Lamp 1/ Lamp 2: Red LEDs which illuminate to indicate lead polarity for
extension lead test facility.
FUSES
Plug: 13A 1HBC (Ceramic)Internal: 3.15A (F) 5x20mm, HBC (Ceramic)
Internal fuse is not user replaceable.
SAFETY
EMC: Meets BS EN 50081-1BS EN 50082-1
LVD: Meets BS EN 61010-1
2.2 Mechanical Specification
CASE
Size: 330x263x144mmMaterial: ABS/PolycarbonateColour: Yellow/ClearWeight: 4kg nominal
2.3 Environmental Specification
TEMPERATURE
Operating: 0 C to 35CStorage: -10 C to 50C
This instrument has been designed to be used in a clean dry environment.
Do not use the instrument outdoors in wet conditions.
2. SPECIFICATIONS
2.1 Electrical Specification
Supply Voltage: 230V10% 50/60Hz
Power Consumption: 10/220VA
EARTH CONTINUITY TEST (8A & 25A)
Test Voltage: 6V AC nominal with no loadTest Current: 25A AC nom @ 0.1 (25A)
8A AC nom @ 0.1 (8A)
Display Range: 0 - 1.99Accuracy of Indication: 10% of reading 2 digits
EARTH CONTINUITY TEST (100mA)
Test Voltage: 130mV DC nominal open circuitTest Current: 100mA DC nominal constant currentRange: 0 - 1.99Accuracy of Indication: 10% of reading 2 digit
INSULATION TEST
Test Voltage: 500V DC -0% +20% at 0.5MShort Circuit Current: 1.5mA DC nominalDisplay Range: 0 - 19.9M
Accuracy Of Indication: 5% of reading 1 digit
FUSE TEST
Test Voltage: 6V AC nominal
Over range threshold: 10k nominalZero range threshold: 800k nominal
LEADS
Mains: 1.7M fixed lead, with a 13A moulded plugEarth Continuity: 3M long, detachable lead, heavy duty crocodile/alligator clip.230V IEC Adaptor: 13A BS1363 plug to IEC320 connector, 230mm long.