2-Way Manifold
3-Way Manifold
IPPB Manifold
Neb-T Valve
In-Line Valve
Mapleson & Control Valve
Compoments and Accessories
˙
˙
˙
˙
˙
˙
˙
Manifold and One-Way Valve
REF DESCRIPTIONS
7VMF11 Durable 2 Way Manifold
7VMF12 Durable 2 Way Manifold w/7.6F Port
REF DESCRIPTIONS
71053 Dispo. 2 Way Manifold w/Elbow Drive Port
71033 Dispo. 2 Way Manifold w/Straight Drive Port
Inhale
Patient
Pressure & Flow Rate: < 4.52 cmH20 @ 60 LPM (7VMF11)
7.6F standard connector
Meets ISO 5356-1 standard
Meets ISO 10993-1 compatibility
The 2-Way Manifold is used to regulate the flow for the
expiratory of the dual limb breathing circuits. The control
line from the ventilator controls the opening of the valve to
enable air to flow out accordingly.
˙
˙
˙
˙
M-01
22/15mm 22/15mm
4M
7.6F
22/15mm 22M
4M
Control line
Exhale
www.GaleMed.com
REF DESCRIPTIONS
7VWF21 Durable 2 Way Manifold w/Water Trap
REF DESCRIPTIONS
7VWP21 Dispo. 2 Way Manifold w/CS Water Trap
7WSV02 Dispo. Suction Tube 3M, w/Angle Port
Control line
Patient
Machine exhale
M-02
22F
22/15mm
4M
22F
22/15mm
4M
Water Trap Volume:50 ml
Pressure & Flow Rate: < 3.31 cm H20 @ 60 LPM (7VWF21)
Meets ISO 5356-1 standard
Meets ISO 10993-1 compatibility
The 2-Way Manifold with a water trap is attached to the
expiratory end of the dual limb breathing circuit to regulate
the expiratory valve. The water trap is used to collect
condensation in the circuit to prevent it from damaging the
ventilator or flowing backing to the patient.
CS (closed system) Water Trap provides the convenience
of removing water from the water trap without having to
remove the water jar. This also helps to minimize the risk
of cross-contamination to ensure clinician and patient
safety. Simply insert the suction tube into the suction port
of the CS Water Trap to remove water.
˙
˙
˙
˙
CS Water Trap Volume:55 ml
Pressure & Flow Rate: < 3.02 cm H20 @ 60 LPM (7VWP21)
Meets ISO 5356-1 standard
Meets ISO 10993-1 compatibility
˙
˙
˙
˙
REF DESCRIPTIONS
71054 Durable 3 Way Manifold
REF DESCRIPTIONS
P1TP2-0001 Dispo. 3 Way Manifold w/Swivel Exhale Port
The 3-Way Manifold is attached to the single limb
breathing circuit to regulate the expiratory valve. The
control line from the ventilator controls the opening of the
valve to enable air to flow in and out accordingly. This line
intermittently provides the pressure needed to close the
expiratory valve during inspiration, which blocks gas
outflow during the application of inspiratory positive
pressure. This is done to ensure the patient receives
optimal gas exchange conditions.
The Disposable 3-Way Manifold with Exhale Port also
provides a standard 22mm port for the monitoring or
collection of expired gases through a scavenging bag.
M-03
4M
22B
22F
22B
Control line
Patient
Exhale
Inhale
22/15mm 22/15mm
4M
4M barb 7.6F
Pressure & Flow Rate: < 1.63 cmH20 @ 60 LPM
Meets ISO 5356-1 standard
Meets ISO 10993-1 compatibility
˙
˙
˙
Pressure & Flow Rate: < 5.0 cmH20 @ 60 LPM
Meets ISO 5356-1 standard
Meets ISO 10993-1 compatibility
˙
˙
˙
www.GaleMed.com
REF DESCRIPTIONS
71047 3 Way Manifold w/Swivel Exhale Cap
71050 3 Way Manifold w/Swivel Exhale Cap & 6M Port
REF DESCRIPTIONS
71063 3 Way Manifold w/22M Exhale Port
71067 3 Way Manifold w/22M Exhale Port & 6M port
REF DESCRIPTIONS
71064 3 Way Manifold w/30M Exhale Port
71065 3 Way Manifold w/30M Exhale Port & 6M port
Inhale
Control line
Patient
Exhale Pressure line Hanger
M-04
22M 30M
22M
4M
22M
22M
22M 4M
22M Cap
22M 4M
Pressure & Flow Rate: < 3.7 cmH20 @ 60 LPM
Meets ISO 5356-1 standard
Meets ISO 10993-1 compatibility
The (parallel 3-way) manifold is attached to the single limb
breathing circuit to regulate the expiratory valve. The control
line from the ventilator controls the opening of the valve to
enable air to flow in and out accordingly. This line
intermittently provides the pressure needed to close the
expiratory valve during inspiration, which blocks gas outflow
during the application of inspiratory positive pressure.
The 3-Way Manifold comes with a hanger that can be
attached to a hanger arm to help position or keep the
breathing circuit in place.
˙
˙
˙
All product meet these info:
REF DESCRIPTIONS
71051 IPPB Manifold
REF DESCRIPTIONS
7VMP41 IPPB Manifold w/Nebulizer
71052 Splitter for IPPB Manifold
Control line
Patient
Inhale
Splitter
Exhale
Nebulizer drive line
M-05
22M
22M
4M
22M
6M
22M
22M
22M
4M
18F
Pressure & Flow Rate: < 3.57cmH20 @ 60 LPM (71051) Meets ISO 5356-1 standard Meets ISO 10993-1 compatibility
The Intermittent Positive Pressure Breathing (IPPB) manifold is used to help treat atelectasis and improve airway functions with a nebulizer that delivers aerosolized medication. The IPPB manifold features a 22M port to connect ventilated gas to the patient and a 18M port for nebulizer attachment. The nebulizer drive line connects to the nebulizer via a 6M port, which controls the pressure setting to deliver optimal medication and particle size. It also features a swivel 22m port that directs exhaled gas away from the patient and therapist.
˙ ˙ ˙
www.GaleMed.com
REF DESCRIPTIONS
7VLD31 Neb-T Valve, 22F, 18F, 22/15
7VLD32 Neb-T Valve, 22F, 18F, 22M
7VLD33 Neb-T Valve, 22M, 18F, 22M
REF DESCRIPTIONS
7VLD21 Neb-T Valve, 15F, 18F, 15M
REF DESCRIPTIONS
7VLD11 Neb-T Valve, 10F, 18F, 10M
18F
22/15mm
or
22M
22F
or
22M
15M
15F
10M
10F
Patient Inhale
Exhale
Nebulizer drive line
M-06
Pressure & Flow Rate: < 0.29 cmH20 @ 60 LPM (7VLD31-33) < 0.50 cmH20 @ 60 LPM (7VLD21) < 3.65 cmH20 @ 60 LPM (7VLD11) Meets ISO 5356-1 standard Meets ISO 10993-1 compatibility
The Neb-T Valve employs a T-shaped valve for use in support of a nebulizer within a breathing circuit. With the use of the Neb-T valve, patients are able to receive aerosol therapy without interrupting ventilation. The Neb-T Valve can be used with or without a nebulizer.; the valve automatically closes itself when a nebulizer is not used and provides only ventilated gas. However, when a Nebulizer is locked into the valve, the mechanism allows the valve to open to deliver both gas and aerosol medication.
˙
˙ ˙
18F
18F
REF DESCRIPTIONS
73016 Parallel Wye with Valve
REF DESCRIPTIONS
73005 Bi-Wye with Valve
REF DESCRIPTIONS
71024 Trachea T with Valve
Pressure & Flow Rate: ≦ 2.77 cmH2O @ 60LPM (Inhale)
≦ 3.11 cmH2O @ 60LPM (Exhale)
Meets ISO–5356-1 standard
Meets ISO-10993-1 compatibility
˙
˙
˙
0.00
0.50
1.00
1.50
2.00
2.50
3.00
3.50
20LPM 30LPM 40LPM 50LPM 60LPM Air flow (LPM)
Res
ista
nce
(cm
H2O
)
Inhale
Exhale
0
0.5
1
1.5
2
2.5
3
3.5
20LPM 30LPM 40LPM 50LPM 60LPM Air flow (LPM)
Re
sist
an
ce (
cmH
2O)
71024
0
0.5
1
1.5
2
2.5
3
3.5
4
4.5
5
5.5
20LPM 30LPM 40LPM 50LPM 60LPM Air flow( LPM )
Res
ista
nce
(cm
H2O
)
Inhale
Exhale
Pressure & Flow Rate: ≦ 4.43 cmH2O @ 60LPM (Inhale)
≦ 4.8 cmH2O @ 60LPM (Exhale)
Meets ISO–5356-1 standard
Meets ISO-10993-1 compatibility
˙
˙
˙
Pressure & Flow Rate: ≦ 3.19 cmH2O @ 60LPM
Meets ISO–5356-1 standard
Meets ISO-10993-1 compatibility
˙
˙
˙
22M
22/15mm
22M
22F 22M
22/15mm
M-07
22M 22/15mm
22M
www.GaleMed.com
REF DESCRIPTIONS
71070 22M Valve Insert
71071 19M Valve Insert
71091 Straight with One-Way Valve 22F to 22M
71092 Straight with One-Way Valve 22M to 22F
REF DESCRIPTIONS
075-0018 EMT Valve
Pressure & Flow Rate: ≦ 3.18 cmH2O @ 60LPM (71070)
Pressure & Flow Rate: ≦ 9.56 cmH2O @ 20LPM (71071)
Pressure & Flow Rate: ≦ 4.69 cmH2O @ 20LPM (71091)
Pressure & Flow Rate: ≦ 4.16 cmH2O @ 20LPM (71092)
Meets ISO–5356-1 standard
Meets ISO-10993-1 compatibility
Pressure & Flow Rate: ≦ 5.2 cmH2O @ 60LPM
Meets ISO–5356-1:2004 standard
Meets ISO-10993-1:1997 compatibility
˙
˙
˙
19M 22M
M-08
22F 22M
22M 22F
71091
71092
Application:
0
1
2
3
4
5
15LPM 30LPM 50 LPM 60LPM Air flow (LPM)
Res
ista
nce
(cm
H2O
)
71091
0
0.5
1
1.5
2
2.5
3
3.5
4
4.5
15LPM 30LPM 50 LPM 60LPM Air flow (LPM)
Res
ista
nce
(cm
H2O
)
71092
22M
22M
18F
˙
˙
˙
˙
˙
˙
22/15mm for patient side or air source. Connector 22/15mm conforms to ISO-5356-1. Luer connector for pressure sensing.
REF DESCRIPTIONS
71034 Release Control Valve
15M for breathing bag or 15F breathing circuit.
12mmOD connector for scavenge port.
Available in autoclavable or single patient use meterial.
REF DESCRIPTIONS
70165 Tail-end Control Valve
70029 Tail-end Control Valve
M-09
Application:
Application:
˙ ˙ ˙
22/15mm
22/15mm
Close to Open
Closed Half-open Open
˙
˙
˙ 12M
15M
The Release Control Valve is used to control or adjust the pressure in breathing circuit or hyperinflation system which can build up during procedures. The valve opening can be manually adjusted to enable pressure release. The Release Control Valve also comes with a luer port for pressure monitoring.
The Tail-End Control Valve is used in a Jackson Rees system whereby it is connected between the gas supply source and the reservoir. The open-close valve adjusts the pressure level of the system.
www.GaleMed.com
REF DESCRIPTIONS
003-0103-01 Silicone valve,
Green
˙ 35mm Dia.
M-10
REF DESCRIPTIONS
70042-02 Silicone valve w/tip,
Green
˙ 34.6mm Dia. / T 1.8mm
REF DESCRIPTIONS
3020-3 Silicone Accessory
˙ 20mm Dia.
REF DESCRIPTIONS
003-0218-01 Diaphragm valve
˙ 20mm Dia.
REF DESCRIPTIONS
003-0259-01 Check Valve
˙ 9 mm Dia.
Silicone valve with tip Silicone valve
Diaphragm valve
Silicone accessory Check valve
M-11
Cap for 4M/4B Port Cap for 6M Port
Cap for Luer Port (non-threaded) Cap for 6F Port
Cap for 7.6F Port
Elbow with 6B Port
Elbow with 8.5M Port Elbow with 4F Port
Elbow with Luer Port Straight with Luer Port
Elbow with 4B Port
Cap for Luer Port
REF DESCRIPTIONS
78036 Durable Cap
REF DESCRIPTIONS
78031 Durable Cap
REF DESCRIPTION
78006 Dispo. Green Cap
002-
0479-01
Dispo. White Luer
Cap w/ Strap
REF DESCRIPTION
78032 Durable Cap
REF DESCRIPTION
78033 Durable Cap
78016 Disposable Cap
REF DESCRIPTION
L0040F Durable Elbow
REF DESCRIPTION
72217 Disposable Elbow
REF DESCRIPTION
72233 Disposable Elbow
REF DESCRIPTION
L0120C Disposable Elbow
7.6M
6B
REF DESCRIPTION
S0050F
Durable Straight 001-
0797-11
7.6M
Luer
REF DESCRIPTION
L0050F Durable Elbow
7.6M
Luer
7.6M
4F
7.6M
4B
7.6M
8.5M
REF DESCRIPTION
002-
0459-01
Dispo. Green w/
Strap
78010 Dispo. White Luer
Cap w/ Strap
002-
0183-01
Durable White Luer
Cap w/ Strap
www.GaleMed.com
Manifold and Valve Control Port Monitoring port Exhale Diverter
REF Descriptions Patient Machine or exp.
4M L4M 7.6F L6B L4B Cap 22M 30M
7VMF11 2 Way Durable Manifold 22/15 22/15 Ⅴ
7VMF12 2 Way Durable Manifold w/7.6F port 22/15 22/15 Ⅴ Ⅴ
7VWF21 2 Way Durable Manifold w/Watertrap 22F 22/15 Ⅴ
71033 2 Way Disposable Manifold 22/15 22M Ⅴ
71053 2 Way Disposable Manifold 22/15 22M Ⅴ
7VWP21 2 Way Dispo. Manifold w/CS Watertrap 22F 22/15 Ⅴ
7WSV02 Suction Tube for CS Watertrap, 3M
71054 3 Way Durable Manifold 22/15 22/15 Ⅴ Ⅴ Ⅴ
P1TP2-0001 3 Way Dispo. Manifold, w/Swivel Exhale Port 22F 22B Ⅴ
71047 3 Way Disposable Manifold w/Swivel Cap 22M 22M Ⅴ Ⅴ
71050 3 Way Disposable Manifold w/Swivel Cap 22M 22M Ⅴ Ⅴ Ⅴ
71063 3 Way Disposable Manifold w/22M port 22M 22M Ⅴ Ⅴ
71067 3 Way Disposable Manifold w/22M port 22M 22M Ⅴ Ⅴ Ⅴ
71064 3 Way Disposable Manifold w/30M port 22M 22M Ⅴ Ⅴ
71065 3 Way Disposable Manifold w/30M port 22M 22M Ⅴ Ⅴ Ⅴ
71051 IPPB Manifold (18F for Nebulizer) 22M 22M Ⅴ Ⅴ
7VWP41 IPPB Manifold w/Nebulizer 22M 22M Ⅴ
71052 Splitter for IPPB Manifold
Neb-T Valve
REF Descriptions Patient Machine Nebulizer 18F
7VLD11 Neb-T Valve, 10M 10F 10M Ⅴ
7VLD21 Neb-T Valve, 15M 15F 15M Ⅴ
7VLD31 Neb-T Valve, 22M 22F 22/15 Ⅴ
7VLD32 Neb-T Valve, 22M 22F 22M Ⅴ
7VLD33 Neb-T Valve, 22M 22M 22M Ⅴ
In-line Valve
REF Descriptions Patient In Out Applications
73005 Bi-Wye Adaptor w/Valve 22/15 22M 22M Isolated Cirucit
73016 Swivel Wye Adaptor w/Valve 22/15 22M 22M Isolated Cirucit
71024 T Adaptor w/Valve, 22F, 22M, T-22/15
71070 In-line Valve 22M One Way
71071 In-line Vlave 19M One Way
71091 One Way Straight Adaptor, 22F to 22M 22F 22M One Way
71092 One Way Straight Adaptor, 22M to 22F 22M 22F One Way
075-0018 EMT Valve, 22M, 22M, T-18F EMT
Control Valve
REF Descriptions Patient Machine Luer Cap
71034 Release Control Valve 22/15 22/15 Ⅴ Ⅴ
70165 Durable Tail-end Valve for Jackson Ree 15M 12M
70029 Dispo. Tail-end Valve for Jackson Ree 15M 12M
M-12
Components
REF Descriptions Diameter Tip
70042-02 Silicone valve w/tip, Green 34.6 mm 1.8 mm
003-0103-01 Silicone valve, Green 35 mm
003-0218-01 Diaphragm main 20 mm
3020-3 Silicone Accessory 20 mm tip
003-0259-01 Check Valve 9 mm
Accessories
REF Descriptions 4M Luer 6M 6F 7.6F
78036 Silicone Cap for 4M Ⅴ
78031 Silicone Cap for 6M Ⅴ
78032 Silicone Cap for 6F Ⅴ
78033 Silicone Cap for 7.6F Ⅴ
78010 Disposable Tread Cap for Luer Ⅴ
002-0459-41 Disposable Cap for Luer / white PE Ⅴ
78006 Diposable Cap for Luer Ⅴ
78016 Disposable Cap for 7.6F, +Type Ⅴ
78019 Disposable Cap for 7.6F, ○ Type Ⅴ
REF Descriptions 4B 4F Luer 6B 8.5M
S0050F Durable Straight (Luer) Ⅴ
L0050F Durable Elbow (4B) Ⅴ
L0120C Disposable Elbow (4.3F) Ⅴ
L0040F Durable Elbow (6B) Ⅴ
72233 Disposable Elbow (4F) Ⅴ
72217 Disposable Elbow (8.5M) Ⅴ
M-13
Code Page
70029 …………………………………………………………................. M-09
70165 …………………………………………………………................. M-09
71024 …………………………………………………………................. M-07
71033 …………………………………………………………................. M-01
71034 …………………………………………………………................. M-09
71047 …………………………………………………………................. M-04
71050 …………………………………………………………................. M-04
71051 …………………………………………………………................. M-05
71052 …………………………………………………………................. M-05
71053 …………………………………………………………................. M-01
71054 …………………………………………………………................. M-03
71063 …………………………………………………………................. M-04
71064 …………………………………………………………................. M-04
71065 …………………………………………………………................. M-04
71067 …………………………………………………………................. M-04
71070 …………………………………………………………................. M-08
71071 …………………………………………………………................. M-08
71091 …………………………………………………………................. M-08
71092 …………………………………………………………................. M-08
72217 …………………………………………………………................. M-11
72233 …………………………………………………………................. M-11
73005 …………………………………………………………................. M-07
73016 …………………………………………………………................. M-07
78006 …………………………………………………………................. M-11
78010 …………………………………………………………................. M-11
78016 …………………………………………………………................. M-11
78019 …………………………………………………………................. M-11
78031 …………………………………………………………................. M-11
78032 …………………………………………………………................. M-11
78033 …………………………………………………………................. M-11
78036 …………………………………………………………................. M-11
7VLD11 …………………………………………………………................. M-11
7VLD21 …………………………………………………………................. M-06
7VLD31 …………………………………………………………................. M-06
7VLD32 …………………………………………………………................. M-06
7VLD33 …………………………………………………………................. M-06
7VMF11 …………………………………………………………................. M-01
7VMF12 …………………………………………………………................. M-01
7VMF21 …………………………………………………………................. M-02
7VMP21 …………………………………………………………................. M-02
7VMP41 …………………………………………………………................. M-05
7WSV02 …………………………………………………………................. M-02
L0040F …………………………………………………………................. M-11
L0050F …………………………………………………………................. M-11
L0120C …………………………………………………………................. M-11
S0050F …………………………………………………………................. M-11
002-0459-41 …………………………………………………………................. M-11
003-0103-01 …………………………………………………………................. M-10
003-0218-01 …………………………………………………………................. M-10
003-0259-01 …………………………………………………………................. M-10
075-0018 …………………………………………………………................. M-08
3020-3 …………………………………………………………................. M-10
70042-02 …………………………………………………………................. M-10
P1TP2-0001 …………………………………………………………................. M-03
GaleMed Corporation
www.GaleMed.com
Headquarter
87, Li-gong 2nd Road, Wu-jia I-Lan, Taiwan 268
Tel:+ 886 3 990 8618
Fax:+ 886 3 990 8245
Sales Department
10F, No. 109, Sec. 6, Min-chuian E. Road, Taipei,
Taiwan 114
Tel:+ 886 2 8791 9348
Fax:+ 886 2 2794 9259
GaleMed Xiamen Co., Ltd
39, Section 3, Haijing East Road, Xiamen, Fujian
Province, China
Tel:+ 86 592 6880588
Fax:+86 592 6880589
099-M
F13-E
01
Copyrig
ht ©
2013 G
ale
Med C
orp
ora
tion. A
ll Rig
hts
Reserv
ed