Confi rm therapy cable is not connected to test plug before pressing ON button.
1 Press ON. Follow audio/visual instructions.
2 Press ANALYZE and stand clear.
3 Press SHOCK when directed to deliver energy.
AED OPERATION
LIFEPAK® 20 Defi brillator/Monitor
Quick Guide for Basic Operation
Note: This Quick Guide for Basic Operation of the LIFEPAK 20 defi brillator/monitor is an abridged reference for basic unit operation.Refer to the LIFEPAK 20 Defi brillator/Monitor Operating Instructions for complete directions for use, indications, contraindications, warnings, precautions, and potential adverse events. Rx only.
LIFEPAK is a registered trademark of Medtronic Emergency Response Systems, Inc. QUIK-COMBO is a trademark of Medtronic Emergency Response Systems, Inc. Medtronic is a registered trademark of Medtronic, Inc.©2005 Medtronic Emergency Response Systems, Inc. MIN 3205874-000 / CAT. 26500-001960
• Press ON.*
• Apply QUIK-COMBO™ therapy electrodes, or hard paddles with conductive gel.
• Select ENERGY SELECT. Adjustup or down as needed.
• Press CHARGE.
• Stand clear.
• Press SHOCK to deliver energy.
Defi brillation
MANUAL OPERATION
• Press ON.
• Attach ECG cable and applyECG electrodes (ensure adequatespacing between ECG and QUIK-COMBO electrodes).
• Apply QUIK-COMBO therapy electrodes.
• Press PACER.
• Press RATE to adjust up or down as needed.
• Press CURRENT and adjust up until electrical capture is achieved.
• Confi rm mechanical capture.
• As needed, Press PAUSEand hold to confi rm underlying rhythm.
Pacing Operation
• Press ON.
• Attach ECG cable and apply ECG electrodes.
• Apply QUIK-COMBO therapy electrodes, or hard paddles with conductive gel.
• Select lead with greatest QRS.
• Press SYNC. Confi rm placement of triangle sense marker ( ) is on R wave.
• Press ENERGY SELECT.Adjust up or down as needed.
• Press CHARGE.
• Stand clear.
• Press and hold SHOCK until energy is delivered.
Synchronized Cardioversion
* To change from AED operation
to manual operation press gray manual tab and open door or press ENERGY SELECT.
For more information please call 800.442.1142 or visit www.medtronic-ers.com • Medtronic Emergency Response Systems • Redmond, WA 98052 USA