1
An in vitro pilot study of the abutment stability during loadi ng in virgin and
fatigue-loaded conical dental implants using synchr otron-based radiography
T. Rack, S. Zabler, A. Rack, H. Riesemeier, K. Nelson
Keywords: X-ray imaging; dental implants; digital radiography; titanium, X-ray phase-
contrast.
ABSTRACT
The implant-abutment connection of a two-piece dental implant depicts a complex
micro-mechanical behavior: a micro-gap is evident at the implant-abutment interface,
even in the virgin state. Its width varies when external mechanical load is applied.
High resolution synchrotron-based radiography in combination with hard X-ray phase
contrast to image this gap and estimate its size is deployed. The study involves
commercially available implants with different internal conical implant-abutment
connections. A comparison between virgin implants and the fatigue loaded implants
underlines that fatigue extends the size of the micro-gap and increases the ability of
micromovement. The cone angle of the connection influences the stability of the
abutment. Cyclic loading at medium force (120 N) hence induces plastic deformation
of the titanium implant.
1. Introduction
Two-piece dental implants, commonly used, feature a mating zone termed implant–
abutment connection (IAC). The mating zone utilized in all two-piece implants can be
subdivided into two principles: a horizontal butt-joint connection or a conical
connection.1 Research concerning the IAC has focused on its geometric precision, 2,3
its mechanical stability 4-9 and the risk of leakage at the implant-abutment interface. 10-16 Conical connections have been proposed to improve the resistance against
mechanical failure of the IAC.6,8 Technical complications concerning the
abutment/abutment screw or the implant occur in 5 - 27 % of the implants placed with
a doubling of the failure rate after 5 years in function.17-19
IAC´s are screwed joints that are exposed to dynamic loading with axial forces up to
450 N 20 and extra-axial forces with angles of up to 90 °. 21,22 Various scenarios to
simulate long-term dynamic loading for the evaluation of fatigue and fracture strength
of an implant have mounted to the DIN ISO Standard 14801:2003.23
2
To date the impact of fatigue loading on the implant-abutment interface of conical
IAC´s has only been investigated by quantifying and visualizing the size of the micro-
gap in unloaded specimens using synchrotron-based micro-tomography.24 It has
shown that there is no surface contact but only focal contact of the assembled
components leading to a micro-gap even in assembled virgin implant-abutment
connections that have not experienced any load application. 24,25
For in vitro visualization of a micro-gap at the IAC of implants with conical mating
surfaces monochromatic hard X-ray synchrotron radiography can be employed.24,26
In comparison to conventional radiography based on x-ray tubes, images recorded
with hard synchrotron light feature higher spatial resolution and material contrast. 27,28
Using phase-contrast radiography even sub-micrometer gaps can be detected
whereby their size is quantified through comparison with numerical simulations 26
allowing besides visualization the quantification of the micro-gap in fatigue loaded
conical implant-abutment connections during static load application. This study
investigates and compares the effect of fatigue loading on the implant-abutment
interface of conical implants with varying conus length and angle.
2. Materials and methods
2.1. Dental implant and test stand
All implants and abutments were purchased from the manufacturers. A pair of
implants with abutments from three different dental implant systems (Table 1) with
conical IAC were assembled and screw-tightened with a system-specific torque and
ratchet. According to EN ISO Standard 14801:2003 individually fabricated steel-balls
(Ø 10 mm) were perforated on one side and glued to the abutments with a two-
component Methyl-Methacrylate-based adhesive (X60; HBM Germany, Darmstadt).
The implant–abutment assemblies were embedded in 15 mm brass cylinders
(Fraunhofer Institut Werkstoffmechanik, Freiburg, Germany) according to EN ISO
Standard 14801:2003 25. One implant-abutment assembly from each pair was fatigue
loaded according to the ISO Standard 14801:2003 by the Fraunhofer Institute,
Freiburg in a dry environment (5 x 106 cycles, 12-120 N amplitude, R=0.1 force ratio
and f=15 Hz frequency at 30° inclination) prior to the radiographic measurements.
The brass cylinders with the implant–abutment assemblies were tightened into an
individually fabricated steel test stand as described, 25 and a static extra-axial force
3
was applied. The constant force was thus applied at an standard angle of 30°
(nominal 200 N) and 90° (nominal 30 N and 100 N) wi th respect to the implant axis
onto the ball. Prior to loading each sample was measured without load. The force
application was monitored using a digital force gauge model SH-500 [PCE-group
OHG (serial No. 5808062790)]. 25 For the fatigue loaded specimen the static force
was applied collinear to the fatigue force-vector which has been marked by a notch in
the brass cylinder.
2.2 Synchrotron radiography with absorption and inline x-ray phase-contrast
High-resolution radiographs were acquired at the BAMline on the BESSY-II light
source (Helmholtz Zentrum Berlin, Germany) (Rack et al., 2008). To penetrate the 3 -
4 mm Titanium, a photon energy of 50 keV was selected through the setting of a
double-multilayer monochromator. Radiographic projection images were recorded by
an indirect detector employing an effective pixel size of 0.45 µm (1.7 mm x 1.1 mm
field of view). Radiography of each specimen was achieved with a propagation
distance between sample and detector of either 4.5 or 74 cm. Owing to the limited
field of view of a high-resolution indirect X-ray pixel detector, only a region of interest
of the specimen is investigated. 25 As the length of the implant-abutment interface
varies between 0.7- 2.3 mm and because the height of the synchrotron beam is
limited as well, generally two images were acquired for each side of each implant and
later merged in order to illustrate the complete gap. Despite the small pixel size the
spatial detector resolution was around 4 µm in absorption mode and ~14 µm in
phase-contrast, due to smearing by the demagnified X-ray source. The raw images
were corrected for the camera's dark current then normalized by flatfield images. Line
scans across the IAC were extracted with the software ImageJ (National Institute of
Mental Health, Bethesda, Maryland, USA). To ensure a good signal and to
compensate for photon noise, a line width of 200 pixels (~80 µm) was chosen, each
profile having a length of 0.4 mm.
2.3 Quantitative characterization of the micro-gap at the IAC
Recent work by Zabler et al. has shown that micro-gaps at the IAC can be detected
and quantified down to a gap width of the order of 0.1 µm by phase-contrast X-ray
4
radiography. Subsequently estimating the exact width is possible by further matching
line-profiles across the gap to interference patterns obtained through numerical
forward simulations of the partially coherent x-ray propagation across a virtual IAC. 26
For gaps larger than 5 µm – with the setup described above – it is also possible to
determine the gap width from x-ray absorption radiographs where they show up as a
discontinuity in attenuation profiles.
Results
Measurement of the geometric dimensions of the conical IAC´s including the length
of the mating zone of the implant and abutment and the angle of the conus was
performed using the radiographic images (Table 1). The region of interest of the data
acquisition within the IAC is shown in Figure 1 and the results are listed in Table 2.
Load condition: 0 N
The virgin implant-abutment interface of each system were radiographed prior to
static load application, they showed a micro-gap varying between 0.1 and 11 µm.
The fatigue loaded samples showed a considerably larger gap between implant and
abutment (0.1 - 31 µm). As depicted in Figure 2b the most significant changes were
seen in the system Ankylos plus (ANP02). The gap in position A of ANP02 increased
by a factor 100x compared to the virgin implant (ANP01)(Figure 2a), namely from
0.32 µm up to 31 µm, and a fissure is visible in the upper edges (A and AF) of the
implant (Figure 2c).
Ankylos c/x IAC demonstrated a decreased micro-gap after fatigue loading. After
fatigue loading (AN02) the maximum micro-gap value was smaller (4 µm) compared
to the sample of the virgin implant (AN01) with a maximum gap size of 11 µm.
The size of the micro-gap of the fatigue loaded BoneLevel IAC (ST02: 0.8 µm)
compared to the virgin IAC (ST01: 0.4 µm) increased 100 %.
Load condition: 30 N, 90°
Static load of 30 N was applied perpendicular (90°) to the implant axis and the
resulting gap was measured. The implants of the Ankylos c/x system (AN01 and
AN02) showed an increase of their micro-gap on the side of load application up to 12
µm in the coronal portion (AF) of the IAC and 0.1 µm in the most apical part of the
IAC (BF).
5
An augmentation of the gap size in fatigue loaded IAC (ANP02) compared to the IAC
of the virgin implant (ANP01) was seen in the system Ankylos plus (ANP).
Regardless of their pretreatment the gap increased tenfold when a load of 30 N was
applied with a maximum of 32 µm.
The behaviour of the micro-gap of the virgin BoneLevel implant ST01 during
horizontal loading has been described.25 When horizontal load is applied the fatigue-
loaded sample (ST02) a 40 % increase of the size of the micro-gap up to 1.8 µm can
be detected. Noticeable is that the micro-gap shows a parallel opening with almost
the same value in the coronal (AF) and the apical (BF) portion of the IAC.
Load condition: 100 N, 90°
At 100 N of static load 90° to the implant axis the tendency of the gap width
perpetuates. Ankylos c/x (AN01, AN02) doubles its gap size to 28 µm when load is
increased from 30 N to 100 N. Again this is independent of the pretreatment of the
implants. Along the gap no point contact between the abutment and the implant can
be seen. The IAC opposite of the side of load application shows a complete closure
of the gap in the coronal region (A) but an opening at the apex (B) with 4.6 µm in
AN01 and 2.5 µm in AN02.
The gap of the fatigue-loaded IAC of Ankylos plus (ANP02) shows an angular micro-
gap formation of 36 µm in position AF and a point contact in BF (Figure 3a). Whereas
the virgin implant shows no point contact with an angular micro-gap opening of 18.5
µm in region AF and 0.1 µm in BF. A point contact can be seen at the most apical
part of the loading side (BF) and at the top of the contralateral side (A) of load
application only in the fatigue loaded implant (ANP02).
In the BoneLevel system the micro-gap opens almost parallel with a size of 10 -
14 µm (ST01) and 13 - 18 µm (ST02) from the apical to the coronal part of the IAC
(Figure 3b).
Load condition: 200 N, 30°
Load of 200 N applied in a 30° angle to the virgin IAC Ankylos c/x (AN01) showed a
gap opening at position AF to 24 µm, which was almost the same in the fatigue
loaded implant (AN02: 22 µm). But the gap value on the side opposite to the side of
force application of the fatigue-loaded sample (A) is fourfold higher (AN01: 0.7 µm;
6
AN02: 3 µm). No point contact between the abutment and implant can be seen in
either sample anywhere in the radiographs of the IAC (Fig. 4a).
Ankylos plus (ANP02) shows approximately a threefold enhancement of the gap in
position AF (9 µm) in respect to the same position in ANP01 (25 µm) with no point
contact between the implant and abutment in any radiograph.
The IAC of BoneLevel is less sensitive to force applied in a 30° angle then to
horizontal force application (Fig. 4b). The micro-gap size varies between 0.3 µm in
the virgin implant (ST01) to 1 µm (ST02). Again the opening of the gap is almost
parallel (see values position AF and BF in Table 2). Fatigue loading doubles the gap
size in BoneLevel and a point contact was not seen at any measurement site.
Discussion
The results of this study show that a micro-gap exists in the investigated conical
implant-abutment connections with or without load. Furthermore, fatigue loading
increases this micro-gap in the systems evaluated.
Prior to the evaluation of the IAC under load application, the IAC´s were
radiographed as received. All implant-abutment assemblies showed a continuous
micro-gap with a minimum of 0.1 µm, with one implant-abutment assembly (AN)
showing an exceptionally large micro-gap of up to 11 µm, this has also been verified
previously using synchrotron based radio- and tomography. 24,25 The quantification of
submicrometer gaps can be performed using synchrotron phase contrast radiography
and numerical forward simulations of the optical Fresnel propagation. This allows the
measurements of the microgap width down to 0.1 µm. 26
A leakage-proof interface with surface contact between abutment and implant has
long been proven not to exist.10 This is due to the fact that the machining of metal
parts leaves a certain surface roughness and surface geometry of the mating zone.
While the geometries generated by designers in a technical drawing generally are
smooth and straight, the real surface texture or topography of the workpiece is to a
large extent determined by the formation of burrs and surface irregularities from the
final milling process, therefore a surface contact is highly improbable in machined
parts.29 The imprecision of the implant-abutment connection has been documented in
in vitro experiments 2,3 and the method to visualize the incongruence between the
7
implant and abutment in the mating zone has been described.5 One virgin sample
presented an exceptionally large micro-gap in the coronal portion of the implant-
abutment interface prior to any load application, this might be based on the fact that
the part is slightly out of specification; to elucidate the degree of fabrication tolerance
the examination of more samples would be necessary.
Conical implant-abutment connections have become popular, as they have been
proposed to have an enhanced mechanical stability during loading. 30 The
mechanical stability has been calculated and shown through FE-Analysis 31 or
measurements of the displacement of the abutment. 6,32 To date no data exist on the
relation of the cone angle, the angle of load application and the mechanical stability
of the abutment. This study shows that the flat conical connection (16°) shows higher
stability (increase in micro-gap is smaller) to load applied in a 30° angle than the 5.7°
conus, suggesting that the angle of the conical implant surface is decisive for the
resistance to the application of load. A conical implant-abutment connection based
on a 16° angled conus (ST) cannot be considered a s elf-locking connection, its
mechanical features are rather similar to a butt-joint connection. This is concordance
to the findings of a current study which shows that butt-joint connections seem to
have a higher load bearing capacity against load applied in a 30° angle compared to
conical connections with a smaller conus angle. 33 The length of the conus does not
seem to influence the degree of micro-motion; it rather determines the mode of
micro-gap formation. The annulling of the abutment when external force is applied is
stopped by the abutment contacting the implant. The location of this contact depends
on the angle of the conus and the extension of the abutment into the implant. In the
flat angled implant this contact zone is below the mating zone (not captured within
this study) leading to an almost parallel gap opening in the mating zone. Cyclic
loading does not only decrease the mechanical stability of the implant-abutment
connection, but also induces wear of the material. This has been shown in studies
comparing different fatigue loading parameters 34,35 and materials used 36. Individual
parameters influence the wear behavior of the implant but to date this has not
induced a change of the ISO Standard currently used to certify dental implants.
Within this study fatigue loading has been performed by the Fraunhofer Institute in a
standardized set-up in a dry environment and with a frequency of 15 Hz as proposed
in the ISO Standard 14801:2003. When comparing the radiograph of the virgin and
the fatigue-loaded smaller-diameter implant a massive plastic deformation of the
8
titanium and a fissure/crack at the implant-shoulder is evident besides an increase of
the size of the micro-gap. The dimensions of the implant wall (0.4 mm) in this
reduced diameter (3.5 mm) pure titanium (Grade 4) implant appears to be
inadequate for cycling loading at a medium force (120 N). Plastic deformation of this
implant type has been described to occur at 349 N in an experimental investigation
under static overload at a diameter of 4.5 mm, suggesting that the thickness of the
implant wall might also be of importance for the load-bearing capacity. 33
Synchrotron-based micro-tomography of IAC´s after fatigue loading has shown that
there is essential material wear with plastic deformation of the titanium at the implant-
abutment interface 24 at 120 N; albeit other reports suggesting only an elastic
deformation at forces up to 600 N. 8,9 A lower frequency of load application is known
to be a decisive factor of earlier fatigue failure of the implant, within this study the
frequency was at medium range with excessive wear of the titanium implants as
documented elsewhere. 35,24
The fatigue loading protocol (120 N at an 30° angle ) used within this study is a
standard all implants need to pass to be certified. It comprises unidirectional loading
on implants glued into a brass cylinder, which is not a physiological situation;
multidirectional diverse forces would be more realistic on implants embedded in a
material that has a similar elastic modulus as bone. As only two implants of each
implant-abutment design has been used a statistical analyses cannot be performed
within this pilot study. Currently a study has been initiated with a higher number of
specimens and multi-directional load application to assess data in a more In-vivo-like
situation. The data demonstrated within this study give an insight into the behavior of
conical implant-abutment connections applying the ISO Standard 14801:2003.
Conclusion
This study demonstrates a detrimental impact of cyclic loading at medium force (120
N) on small-diameter implants. It also shows that the investigated conical IAC´s show
a continuous micro-gap regardless of their design and the amount of force applied.
The mechanical resistance of the abutment to micromotion seems to be related to
the cone angle of the implant-abutment connection.
9
Acknowledgement
We would like to thank Dr. -Ing. M. Lengauer for the discussion and MDT Juergen
Mehrhof for the sample preparation. T. Rack is a recipient of a doctorate grant from
the Camlog Foundation. This study was partially funded by the German Research
Association (DFG Nel 656/1-1 and ZA 656/1-1).
10
1. Binon P. Implants and components: entering the new millennium. Int. J. Oral
Maxillofac. Implant 2000; 15: 76–95.
2. Semper W, Kraft S, Krüger T, Nelson K. Theoretical considerations: implant
positional index design. J Dent Res 2009; 88: 725–730.
3. Semper W, Kraft S, Mehrhof J, Nelson K. Impact of abutment rotation and
angulation on marginal fit: theoretical considerations. Int J Oral Maxillofac Impl
2010;25(4):752-8.
4. Binon PP, McHugh MJ The effect of eliminating implant/abutment rotational
misfit on screw joint stability. Int J Prosthodont 1996;9(6):511-9.
5. Cibirka RM, Nelson SK, Lang BR, Rueggeberg FA. Examination of the
implant-abutment interface after fatigue testing. J Prosthet Dent
2001;85(3):268-75.
6. Norton MR. In vitro evaluation of the strength of the conical implant-to-
abutment joint in two commercially available implant systems. J Prosthet Dent
2000;83(5):567-71.
7. Cehreli MC, Akça K, Iplikçioğlu H, Sahin S. Dynamic fatigue resistance of
implant-abutment junction in an internally notched morse-taper oral implant:
influence of abutment design. Clin Oral Impl Res 2004;15(4):459-65.
8. Khraisat A, Baqain ZH, Smadi L, Nomura S, Miyakawa O, Elnasser Z.
Abutment rotational displacement of external hexagon implant system under
lateral cyclic loading. Clin Impl Dent Relat Res 2006;8(2):96-9.
9. Steinebrunner L, Wolfart S, Ludwig K, Kern M. Implant-abutment interface
design affects fatigue and fracture strength of implants. Clin Oral Impl Res
2008;19: 1276-1284.
10. Jansen VK, Conrads G, Richter EJ. Microbial leakage and marginal fit of the
implant-abutment interface. Int J Oral Maxillofac Implant 1997;12: 527–540.
11. Gross M, Abramovich I, Weiss EI. Microleakage at the abutment-implant
interface of osseointegrated implants: a comparative study. Int J Oral
Maxillofac Impl 1999;14(1):94-100.
12. Steinebrunner L, Wolfart S, Bösmann K, Kern M. In vitro evaluation of
bacterial leakage along the implant-abutment interface of different implant
systems. Int J Oral Maxillofac Impl 2005;20, 875–881.
11
13. Dibart S, Warbington M, Su MF, Skobe Z. In vitro evaluation of the implant-
abutment bacterial seal: the locking taper system. Int J Oral Maxillofac Impl
2005;20(5):732-7.
14. Coelho PG, Sudack P, Suzuki M, Kurtz KS, Romanos GE, Silva NR. In vitro
evaluation of the implant abutment connection sealing capability of different
implant systems. J Oral Rehab 2008;35(12):917-24.
15. Harder S, Dimaczek B, Açil Y, Terheyden H, Freitag-Wolf S, Kern M.
Molecular leakage at implant-abutment connection--in vitro investigation of
tightness of internal conical implant-abutment connections against endotoxin
penetration. Clin Oral Investig 2010;14(4):427-32.
16. Aloise JP, Curcio R, Laporta MZ, Rossi L, da Silva AM, Rapoport A. Microbial
leakage through the implant-abutment interface of Morse taper implants in
vitro. Clin Oral Impl Res 2010;21(3):328-35.
17. Nedir R, Bischof M, Szmukler-Moncler S, Belser UC, Samson J. Prosthetic
complications with dental implants: from an up-to-8-year experience in private
practice. Int J Oral Maxillofac Impl 2006;21(6):919-28.
18. Jung RE, Pjetursson BE, Glauser R, Zembic A, Zwahlen M, Lang NP. A
systematic review of the 5-year survival and complication rates of implant-
supported single crowns. Clin Oral Impl Res 2008;19(2):119-30.
19. Bozini T, Petridis H, Garefis K, Garefis P. A meta-analysis of prosthodontic
complication rates of implant-supported fixed dental prostheses in edentulous
patients after an observation period of at least 5 years. Int J Oral Maxillofac
Impl 2011;26(2):304-18.
20. Duyck J, Van Oosterwyck H, Vander Sloten J, De Cooman M, Puers R, Naert
I. Magnitude and distribution of occlusal forces on oral implants supporting
fixed prostheses: an in vivo study. Clin Oral Impl Res 2000;11(5):465-75.
21. Mericske-Stern R, Geering AH, Buergin W. Three-dimensional force
measurements on mandibular implants supporting overdentures. Int J Oral
Maxillofac Impl 1992;7: 185–194.
22. Morneburg T, Proeschel PA. Measurement of masticatory forces and implant
loads: a methodologic clinical study. Int J Prosthodont 2002 ;15 : 20–7.
23. Standardization. International Standard ISO 14801 – dentistry – implants-
dynamic fatigue test for endosseous dental implants. Geneva: International
Organization for Standardization; 2007.
12
24. Zabler S, Rack T, Rack A, Nelson K. Fatigue induced deformation of taper
connections in dental titanium implants. Int J Mat Res 2012; 103:207-16
25. Rack A, Rack T, Stiller M, Riesemeier H, Zabler S, Nelson K. Journal of
Synchrotron Radiation 2010;17(2): 289-294
26. Zabler S, Rack T, Rack A, Nelson K. Quantitative studies on inner interfaces
in conical metal joints using hard X-ray inline phase contrast radiography. Rev
Sci Instr ;81(10): 103703
27. Banhart J. Advanced Tomographic methods in material research and
engineering. Oxford University Press; 2008
28. Stock SR. Synchrotron microComputed Tomography of the mature bovine
dentinoenamel junction. J Struct Biol 2008;161(2):144-50.
29. Burrs P. Analysis, control and removal. CIRP Ann Man Tech 2009;58:519–542
30. Bozkaya D, Müftü S. Efficiency considerations for the purely tapered
interference fit (TIF) abutments used in dental implants. J Biomech Eng
2004;126(4):393-401.
31. Akça K, Cehreli MC. Evaluation of the mechanical characteristics of the
implant-abutment complex of a reduced-diameter morse-taper implant. A
nonlinear finite element stress analysis. Clin Oral Impl Res 2003 ;14 : 444–54.
32. Meng JC, Everts JE, Qian F, Gratton DG. Influence of connection geometry on
dynamic micromotion at the implant-abutment interface. Int J Prosthodont
2007;20(6):623-5.
33. Dittmer S, Dittmer M, Kohorst P, Jendras M, Borchers L, Stiesch M. Effect of
implant-abutment connection design on load bearing capacity and failure
mode of implants. J Prosthodont 2011 ;20(7):510-6
34. Cornell KL, Karl M, Kelly RJ. Evaluation of test protocol variables for dental
implant fatigue research. Dent Mat 2009 ; 25 : 1419-25
35. Karl M, Kelly RJ. Influence of loading frequecy on implant failure under cyclic
fatigue conditions. Dent Mat 2009 ; 25 :1426-32
36. Klotz MW, Taylor TD, Goldberg AJ. Wear at the titanium-zirconia implant-
abutment Interface : A Pilot Study. Int J Oral Maxillofac Impl 2011 ; 26 :970-75
13
Table 1 Details of the implant-abutment assemblies used (V: virgin, F: fatigue loaded)
AbbreviationAbbreviationAbbreviationAbbreviation
SystemSystemSystemSystem
ManuManuManuManu----facturerfacturerfacturerfacturer
Standardized Standardized Standardized Standardized characteristics of characteristics of characteristics of characteristics of
implants and abutmentsimplants and abutmentsimplants and abutmentsimplants and abutments
Dimensions ofDimensions ofDimensions ofDimensions of implant and IACimplant and IACimplant and IACimplant and IAC mmmmmmmm //// LLLL mmmmmmmm
Angle of Angle of Angle of Angle of
cone (°)cone (°)cone (°)cone (°)
Length of Length of Length of Length of
mating zone mating zone mating zone mating zone
(mm)(mm)(mm)(mm)
AN01 (V) AN02 (F )
Ankylos c/x Friadent Implant: REF: 17-0546/31010420, LOT: 0020035388 Abutment: REF: 31024140 LOT: 20035853
Implant: 3.5 / 14; IAC: 2.5 / 1.8
5.7
1.82
ANP01 (V) ANP02 (F)
Ankylos plus
Friadent Implant: REF: 31010220, LOT: 20037235 Abutment: REF: 31024180 LOT: 20035821
Implant: 3.5 / 14 IAC: 2.5 / 2.2
5.7
2.01
ST01 (V) ST02 (F)
Bonelevel Straumann AG, Basel, Switzerland
Implant: REF: 021.4114, LOT: G6582 Abutment: REF:
Implant: 4.1 / 14.2 IAC: 3.3 / 0.7
16 0.76
14
022.2202 LOT: F6601
15
Table 2 Size of microgap under different mechanical load conditions
SystemSystemSystemSystem
Gap size in µmGap size in µmGap size in µmGap size in µm
0 N0 N0 N0 N 30 N, 90°30 N, 90°30 N, 90°30 N, 90° 100 N, 90°100 N, 90°100 N, 90°100 N, 90° 200 N, 30°200 N, 30°200 N, 30°200 N, 30°
vvvvirginirginirginirgin fatigue fatigue fatigue fatigue loadedloadedloadedloaded
virginvirginvirginvirgin fatigue fatigue fatigue fatigue loadedloadedloadedloaded
virginvirginvirginvirgin fatigue fatigue fatigue fatigue loadedloadedloadedloaded
virginvirginvirginvirgin fatigue fatigue fatigue fatigue loadedloadedloadedloaded
Ankylos c/x (AN) 11 0.6 4 0.7 0.2 12 0.5 12 0.5 28 1.2 24 0.1 24 0.1 22 0.3 0.1 0.2 0.1 0.9 0.1 0.7 0.2 4.6 0.1 2.5 0.2 0.7 0.1 3 0.1
Ankylos plus (ANP)
0.3 0.2 31 5 0.6 3 1 32 3 18.5 0 36 0.1 9 - 25 0.2 0.3 1 3 1.2 0.2 12 0.3 10 0.1 30 0 4 0.1 - 0.1
Straumann (ST) 0.2 0.2 0.8 0.7 4 1.8 14 18 0.2 0.3 0.3 1 0.4 0.3 0.2 0.7 4 1.7 10 13 0.3 0.6 0.1 1
Assignment of the values in the table with respect to the positions as defined in Fig. 1.
Horizontal loading values for virgin IAC described in Rack et al., 2010
0.1 0.3
0.6 11
BF
AF A
B
16
Figure 1. Schematic illustration of positions of extracted profiles (yellow) with the direction and values of static load application (grey: steel ball glued onto abutment according to DIN ISO 14801; orange: rotation safe abutment torque tightened to the implant body (blue); yellow: area of measurement as described (Zabler et al., 2010).
30° (load 200 N)
A
B
AF
BF
90° (load 30 N, 100 N)
17
18
2a
2b
2c
500 µm
100 µm
19
Figure 2. Radiographs of Ankylos plus IAC without mechanical static load (0 N). I – implant, A – abutment, S – screw. 2a – Ankylos plus (ANP01) before fatigue loading (V). 2b – Ankylos plus (ANP02) after fatigue loading (F), the marked region is depicted in Fig 2.c. 2c – Fissure in the coronal portion of implant (left: original image, right: edge enhanced image).
20
3a
3b
Figure 3. Radiographic images of fatigue loaded Ankylos plus IAC (ANP02) – 3a, and fatigue loaded Straumann BoneLevel (ST02) – 3b under static horizontal load of 100 N. I – implant, A – abutment, S – screw.
500 µm
500 µm
21
4a
b
Figure 4. Radiographic images of virgin Ankylos c/x IAC (AN01) – 4a, and fatigue loaded Straumann BoneLevel (ST02) – 4b under static load of 200 N in a 30 ° angle. I – implant, A – abutment, S – screw.
22