Generator
Frequency KnobFrequency display
Frequency Range
Output
Amplitude (output Level) Knob
-20 db (output attenuate)
Functions
Auto Setup
Move cursor Up/Down or Left/Right
Press and hold toggle deltaV and deltaTPress once to active cursor (broken to dask) and use Cursor to move
ACDC Press one toggle AC/DC look at left bottom corner of display
INV Press and hold to invert the signalPress once GND (see left bottom corner of display)
To ADD the signal Press both key DUAL and CH2 at the same timeLook at the bottom display see + symbol
Time/Division Display
Volt/Division Display
Volt/Division Knob
Time/Division Knob
Generate A signal
• Functions (Sine, Triangle, Square Wave)• Amplitude (A)• Frequency(F)
• On Generator:– Select a function (Sine, Triangle, or Square Wave)– Select frequency range.
– Turn Amplitude knob to mid range.
Setup Amplitude
• On Oscilloscope Side:
1) Find out appropriate Volt/Division.
1- Amplitude / 4 = Estimated Volt/Division.
2- Turn Volts/Division knob and observe Volt/Division display on the screen, stop when Volts/Division = or < then Estimated Volts/Division.
3- Look at the displaying Volts/Division, This is the appropriate Volts/Division
2) Find out the number of Division.
- The Number of Divisions = Amplitude / (Volts/Division)
3) When the setting is done, then do not touch Oscilloscope go to generator.
4) On generator side, turn Amplitude knob.
5) Observe the Oscilloscope screen until the signal display Vp-p to the calculated number of vertical divisions.
Setup Frequency
• On Oscilloscope Side:1) Calculate period (T). T= (1/F)
2)Find out appropriate Time/Division for one period.1- Period(T) / 4 = Estimated Time/Division.2- Turn Time/Division knob and observe Time/Division
display on the screen. , stop when Volt/Division = or < then Estimated Time/Division.
3- Look at the displaying Time/Division, This is the appropriate Time/Division.
3)Find out the number of Division. - The Number of Divisions = T / (displayed
Volts/Division)
4) On generator side, set the Frequency Range to the appropriate range.
5) Adjust the Frequency knob and Observe the Oscilloscope screen until one period(T) gets to the calculated number of horizontal divisions .
Use ADD function to measure the differential voltage with Oscilloscopea) Connect Scope CH#1 to first pointb) Connect Scope CH#2 to second pointc) Adjust both channels to the same Volt/ Divisiond) If you want CH#1 – CH#2 then Press Invert CH#2 ( see at the bottom on
the display C2 get the bar over that means inverte) Press both button DUAL and CH#2 same time ( see a + sign at bottom on display means ADD ok)f) Press CH#2 and use cursor to measure the voltage