8/10/2019 emg lecture
1/26
7th LectureDimitar Stefanov
8/10/2019 emg lecture
2/26
RecappingThree types electrodes are used for sensing of EMG signals:
1. indwelling (intramuscular) electrodes (single fiber electrodes, monopolar
electrodes, concentric electrodes)
2. Wire electrodes
3. surface electrodesnon-invasive recordings
Potential of surface electrode (V)
Differential voltage waveform
8/10/2019 emg lecture
3/26
Velocity of propagation of the m.a.p.4 m/s
There is a delaybetween the EMG and muscle contraction (30-80 milliseconds).
In case of isometric muscle tension, a linear dependencybetween the muscle tensionand the rectified EMG output is observed.
Important parameters of the EMG amplifiers:1. Gain and dynamic range
2. Input impedance
3. Frequency response4. Common mode rejection.
EMG signal:
contains certain level of noises
has specific spectral densityfunction.
Fatigue
(1) If we assume that the EMG is stimulation rate remains constant then the muscle
tension deceases in case of fatigue.
(2) The shape of the m.a.p. is altered in case of fatigue.(3) tremor occurs.
8/10/2019 emg lecture
4/26
Problem with the electrodes: polarization
The electric conductivity of the body involves ions as charge carrier.
Electrodes can be considered as electrical conductors in contact with the aqueous
ionic solutions of the body.The interaction between electrons in the electrodes and ions in the body can affect
the EMG signal
Half -cell potential(HCP) is called thepotential differencebetween the metal of
the electrode and the bulk of the electrolyte.
HCP depends on the ionic concentration
HCP can be measured when no electric current flows between an
electrode and the electrolyte
8/10/2019 emg lecture
5/26
Problem with the electrodes: polarization
Polarizationarises in case when current flows between the electrodeand the solution.
Perfectly polarizable electrodesno actual current crosses the
electrode- electrolyte interface
Nonpolarized electrodesallow the current to pass freely in
electrode-electrolyte interface.
Silversilver chloride electrode (Ag/AgCl)it possesses characteristics which
are similar to a perfect nonpolarizable electrode.
8/10/2019 emg lecture
6/26
AgCl film
insulated lead wire
Ag metal
Ag lead wiresintered Ag and AgCl
(Ag and AgCl powder mechanically pressed)
greater mechanical stability
Silversilver chloride electrodes
Low noise electrodes
8/10/2019 emg lecture
7/26
Equivalent circuit of a biopotential electrode
Ehchalf-cell potential
Rdand Cdrepresent the impedance associated with the electrode-electrolyte
interfaceRsseries resistance.
Biopotential electrode impedance
as a function of frequency
8/10/2019 emg lecture
8/26
EMG amplifiers
Amplifier gainthe ratio of the output voltage to the input voltage
Surface EMG electrodes- maximum amplitude of 5 mV peak-to-peak
Indwelling electrodesamplitude of up to 10 mV
Single m.a.p. electrodesamplitude of 100 mV
Amplitudes of the EMG signal :
Noise level of the amplifieris the amplitude of the higher frequency random
signal on the output of the amplifier when the electrodes are shorten together.
Noise level of the amplifier should not exceed 50 mV,
(preferably 20mV).
8/10/2019 emg lecture
9/26
Input impedance of an amplifier of biosignals
The resistance of the electrode-skin interface depends on:
thickness of the skin layer,
the cleaning of the skin prior to the attachment of the electrodes,the area of the electrode surface,
temperature.
Electrode pastedecreases the resistance
between the electrode
and the skin.
8/10/2019 emg lecture
10/26
Input impedance of an amplifier of biosignals
The capacitancebetween the electrode and the skin causes
frequency distortions.
EMG amplifiers should possess high input resistance
8/10/2019 emg lecture
11/26
Frequency response of the EMG amplifier
Frequency bandwidth
All frequencies presentin the EMG should be
amplified at one and the
same level.
Bandwidththe difference between upper cutoff frequencyf2and the lower cutoff
frequencyf1.The gain of the amplifier atf1andf2is 0.707 from the gain of the gain in the mid-
frequency region (half-power).
Amplifier gain:
Example: linear gain 1000, or 60 dB; gain at the cutoff frequencies
57 dB (3dB less than that atthe mid-frequencies).
8/10/2019 emg lecture
12/26
The EMG amplifier should amplify equally all EMG frequency components.
Most of the EMG signalsare concentrated in the band
between 20 and 200 Hz.
8/10/2019 emg lecture
13/26
Recommended range of the EMG amplifiers:
from 10 Hz to 1000 Hzwhen the signal is collected with surface
electrodes;
from 20 Hz to 2000 Hzwhen the signal is collected withindwelling electrodes.
Interferences:
Hum from power line (60 Hz in the USA and 50 Hz in Europe)-in the middle of the EMG spectrum
Movement artifactstheir frequency lies in the 0 to 10 Hz range
dont cause big problems
Noise from low quality cabling systemsinterfere with thebaseline of the EMG signal; can be eliminated by good low
frequency filtering (by setting of f1 to about 20 Hz).
8/10/2019 emg lecture
14/26
8/10/2019 emg lecture
15/26
Single-ended amplifier
Differential amplifier
A perfect subtraction
never occurs.
8/10/2019 emg lecture
16/26
Common mode rejection ratio (CMRR)
CMRR is measured in dB.
In good quality EMG amplifiers CMRR should be 10,000 (80 dB) or higher.
8/10/2019 emg lecture
17/26
Processing of EMG
Example:
1. Half of full-wave rectification (absolute value)
2. Linear envelope (low-pass filtering of the rectified signal)
main decision here is the choice of the low pass filter!
3. Integration of the signal from (2)over the period of the musclecontractionarea under the curve
4. Integration of the signal from (2) for a fixed time, reset to zero,
and repeating the integration cyclesuch scheme represents the
trend of the EMG amplitude with time
5. Integration of the signal from (2) to a present level, reset to zero,
and repeating the integration cyclerepresents the level of the
muscle activity (high or low muscle activity).
8/10/2019 emg lecture
18/26
Diagram of several common EMG processing systems and the processing results
8/10/2019 emg lecture
19/26
Biopotential amplifiers
Basic amplifier requirements:1. The physiological process to be monitored should not be influenced in any way
by the amplifier
2. The measured signal should be not distorted
3. The amplifier should provide the best possible separation of signal and
interferences
4. The amplifier should offer protection of the patient from any hazard and electric
shock
5. The amplifier should be protected against damages due to high input voltages.
The input signal to the amplifier consists of 5 components:
1. Desired biopotential
2. Undesired biopotentials
3. A power line interference signal and its harmonics
4. Interference signals generated by the tissue-electrode interface
5. Noise.
8/10/2019 emg lecture
20/26
Block diagram of a biopotential amplifier
FET transistorsGalvanic
decoupling of
the patient
Motion artifactsthe contact between the electrode and the tissue changes
during the relative motions between the electrodes and the tissue.
Measures for decreasing the motion artifacts:
High input resistance of the amplifier
Usage of non-polarized electrodes (Ag/AgCl)
Reduction of the source impedance by usage of electrode gel.
Artifactsdue to electric and magnetic fieldsExample.
8/10/2019 emg lecture
21/26
Amplitude/frequency characteristics of the
bioamplifiers used in different applications
S i l i i hi h b il h bi i l lifi
8/10/2019 emg lecture
22/26
Special circuits which built the biopotential amplifier
Instrumentation amplifiers
DCinstrumentation amplifiers
8/10/2019 emg lecture
23/26
AC instrumentation amplifiers
AC amplifiers eliminate the electrode offset potential, permit high gain and
permits higher CMRR.
The capacitors between the electrodes and the input stage of the amplifier causecharging effects from the input bias current.
8/10/2019 emg lecture
24/26
Isolation amplifier
Isolation is realized in the following technologies:Transformer isolation
Opto-isolation.
Isolation provides a complete galvanic separation between the input stage
(patient) and the other part of the measure equipment.
8/10/2019 emg lecture
25/26
Surge protection of the bioamplifiers
Protection of the amplifier from damage due to surge input potentials.
Diodes
Zener diodes
Gas-discharge tubes
8/10/2019 emg lecture
26/26
Input guarding
Instrumentation
amplifier providing
input guarding
Technique for increase both the input impedance of the amplifier of biopotentials and
the CMRR
Dr iven-r ight-leg circui t
reducing common-mode
interference