GRAYWOLFTM
SENSING SOLUTIONS12 Cambridge Drive Trumbull, CT 06611Ph. 203-416-0005Fax: 203-416-0002Email: [email protected]
Website: www.WolfSense.com
Harnessing the Power of Mobile Computers f o r A d v a n c e d E n v i r o n m e n t a l M e a s u r e m e n t s
6
DIRECTSENSE IAQ
I ndoor A i r Qua l i t y Mon i to r
DIRECTSENSE
IAQ
GrayWolf is changing the way that environmental measurements are taken. The DirectSense IAQ harnesses leading-edge sensor
technology with the power and performance of mobile computing: your choice of Pocket PC or Notebook PC platforms.
DIRECTSENSETM IAQINDOOR AIR QUALITY MONITOR
• Organizes ALL survey info; data, text, calibration history and more...
• Tap the screen for educational info: sen-sor tips, IAQ documents and guidelines
• Measures CO2, CO, %RH, Temperature and more...
• Options for VOCs, Toxic Gases, Particulates, Airspeed and more...
• Extremely easy to use
Comes with 18 IAQ Pocket Word forms (mold checklist, etc.) for field inputWolfSenseTM PC desktop software is included for data review, cut & paste report generation Optional WolfSense ARG Advanced Report Generator software for automated, ultra-efficient, detailed IAQ reports
•
•
•
GrayWolf provides Pocket PCs, Tablet PCs, Ultra-Mini PCs and Notebooks that are ideal for the application – or use your own.
DirectSense probes can also plug directly into a Windows XP/Vista PC via a Serial/USB cable and
GrayWolf’s WolfSense LAP software.
E x a m p l e P o c k e t P C S c r e e n s
Examp le No tebook Sc reens
Advanced Environmental Measurements Made Easy
GrayWolf DirectSense Monitors are an essential component of a superior IAQ program. Conduct baseline surveys. React immediately to complaints. Test conditions before and after building modifications. Optimize energy settings.Display measurements in realtime. Access “sensor tips”, IAQ documents and guidelines on-screen. Store “snapshot” read-ings. “Trend” log over hours/days/weeks on preset intervals. Document each measurement location with text, graphic, even audio information. Also attach Pocket Word field forms, photos, time-stamped event notes and other info to data files.
Help Files
On-board Docs
Field (Word) Form Graphic Note
Realtime
Sensor Tips
Graphic Note Attached
Text Note
FieldForm(s)Attached
Audio Note Attached
Event Notes
Cal Info
Rea l t ime Sc reen
R u g g e d i z e d P o c k e t P C
M i n i - N o t e b o o k P CTrend Log
S t a n d a r d P o c k e t P C
Trend Log
A second probe may be connected simultaneously via an optional AD-Y adapter or PCC-20Y case. All parameters display at the same time and log to the same data file.
IQ-410
✔ Display ✔ Notate ✔ Data-Log Highly accurate, extremely fast response sensors for Carbon Dioxide, Carbon Monoxide, Relative Humidity and Temperature are perfect for both quick walk-through surveys and for long-term trending.
GrayWolf offers additional probes to expand your measurement capabilities.
TG-501NO, SO2, EtO, CI2, NO2, H2S, HCN, O2, NH3, O3, and more. Up to five sen-sors plus ºF / ºC in one probe
AS-201Air velocity probe (telescoping)
AS-202AAir Velocity probe (telescoping and 90º articulating tip)
CO2
CO %RH
Temperature
DEWPOINT, WETBULB, ABSOLUTE HUMIDITY SPECIFIC HUMIDITY,
HUMIDITY RATIO
With an ACC-SCCF serial adapter, add particulate measurements or differential pressure to concurrently display and log on the Mobile PC.
More probes, sensors and software features are on the way. Expand your DirectSense system as your measurement needs grow and as legislative requirements change.
VOCsCO2
CO%RHTemperature
✔ Report✔ Organize
PC-GW66 Channel Particle Counter
PM-201Particulate Concentration
ZEPHYRDifferential Pressure
✔ Download
All of the IQ-410 capability plus a ppb range PID sensor for VOC measurements: Track VOCs over time, bloodhound VOC sources or determine optimum locations and times to take air samples for lab analysis. A very powerful screening tool.
IQ-610
An additional specifi c gas sensor slot is available on the IQ-610Choose from:
OzoneAmmoniaChlorineHydrogen Sulfi deNitrogen DioxideNitric OxideSulfur DioxideOxygenEthylene OxideHCN, HCl, H2, SiH3, AsH3, PH3, ClO2, CCl2O, B2H6, HF or F2
Optional WolfSense ARG Advanced Report Generator; IAQ edition, takes report production to an entirely different level. A software wizard walks you through report template set-up. Whether you want 5 page reports or 250 pagers, once templates are set up just dump data to effi ciently produce highly professional reports. Make use of a range of included text blocks (Disclaimers, Survey Strategies, Results & Conclusions, etc.), edit them and/or add your own.Template elements that may be selected include:• Recurring Text (for every report)• Prompting Text (author selects from a list of text blocks)• Parameter by Parameter Data Tables • Parameter Info (including Health Effects, Typical Background Levels, Government & Industry Guidelines, etc.)• Location by Location Data and Notes (collected on-site)• Auto-Generated Data Graphs• Equipment List• Calibration Data...and much moreOutput to Word or RTF format for easy editing and printing.
WolfSense ARG
WolfSenseTM PC
Example WolfSense ARG
Screens
Example VentCal Screens
Air Ratio Tracer Decay Equilibrium Tracer
Example WolfSense PC Screens
Example Printed ARG Report
Field Notes, Photos, Cal Info, etc are
attached to data files
VentCalTM
WolfSense PC data transfer software, supplied as standard, allows readings logged on the mobile PC to be reviewed on desktop PCs*. Simply export to Excel, or utilize WolfSense PC’s intuitive menu to produce graphs from trend data in 2D or 3D, with data point markers, add comments, adjust colors, auto-include event note lines, alarm lines, etc. for a highly professional presentation. Data is displayed with attached notes (text, drawing, calibration, etc.) for easy cut and paste into a Word report. Photos, graphs and other fi les may be attached for exceptional organization and documentation.*GrayWolf lets you make as many copies of this software as your organization needs.
The optional VentCal™ software module walks the user through CO2 tracer calculations of outdoor air ventilation rates utilizing your choice of 3 tracer gas techniques: • Air - Ratio• Equilibrium Tracer• Tracer DecayDisplay results as:• Air changes/Hour• Volume Airfl ow/Occupant• Volume Airfl ow/Floor AreaProduces an HTML report for easy printing.
SPECIFICATIONS: IQ-410, IQ-610Carbon Dioxide: (NDIR) Range: 0 to 10,000 ppm Accuracy: ± 3% rdg ± 50 ppm
Carbon Monoxide: (electrochemical) Range: 0 to 500 ppm Accuracy: ±2ppm <50ppm, ±3%rdg >50ppm
VOCs: (PID) IQ-610 only Range: 20 to 20,000 ppb (user selectable units: ppb, ppm, μg/m3, mg/m3) Resolution 1 ppb, L.O.D. <20 ppb
Relative Humidity: (capacitive) Range: 0 to 100 %RH. Accuracy: ±2%RH <80%RH (±3%RH>80%RH)
Temperature: (Pt100) Range: 15° to 160°F (-10° to +70°C) Accuracy: ±0.3°CResponse Time: All sensors exhibit 90% response <1 minute
Probe Dimensions : 2in. (5cm) dia. x 12.5in. (30cm) lgth. Weight (probe): w/batteries 1lb. 10oz. (0.7kg) Power (probe): 2 x D cells; typical battery life 130+ hrs. or use optional ACC-ADY-P AC adapter
• The DirectSense IAQ Standard, DS IAQ Professional and DS IAQ Master are complete kits that include your choice of Pocket PC or Notebook computer and come with the IQ-410 or IQ-610 IAQ probe, WolfSense
IAQ application software, WolfSense PC desktop data transfer soft-ware, hardshell security case, probe and mobile PC pouches and much more. Contact your representative or GrayWolf for details.
• Portable printers, VOC calibration kits, air pumps, extension cables, dual probe adapters, dual probe security cases, serial adapters (for simultaneous connection of particulate or pressure meters to mobile PCs) and other accessories are available from GrayWolf.
CA-RH1 %RH Cal Kit
▼
PCC-10 Security Case▲
CA-GS12 CO2 and CO Gas Cal Kit
▲
PCC-PH01 PPC Case and
PCC-05IQ Probe Pouch
(shown w/ PM-201
particulate concentration
meter)
▼
PCC-20Y RuggedPPC housing ▼
(computer battery life varies by make and model)
PC Software Req’s: Windows 2000/XP/Vista or MS.NET 1.1 or higher 30MB hard disc (min.) spaceGRAYWOLF SOFTWARE IS 21 CFR PART 11 COMPLIANT.ALL SPECIFICATIONS ARE TO SUBJECT TO CHANGE WITHOUT FURTHER NOTICE.
OCTOBER 2007
ACC-TPD1 Tripod with ACC-MNT1 PPC Mount(addACC-TF2fitting for dual probes)
CA-HD2Gas Calibration
Hood
▼
▲
www.WolfSense.com
Email: [email protected]
GRAYWOLF SENSING SOLUTIONSTuamgraney Industrial Estate Unit 8Tuamgraney, Co Clare, IrelandPh. (353)61921736Fax: (353)61921528
GRAYWOLFTM SENSING SOLUTIONS12 Cambridge DriveTrumbull, CT 06611 USAPh. 203-416-0005Fax: 203-416-0002
ACC-KEY2B fabric keyboard (bluetooth) accommodates field notation
▲
10.6 eV PIDs do not respond to VOCs with ionization potentials >10.6, such as ethane, methane or formaldehyde. However, they do respond to the vast majority of VOCs. Standard calibration is to Isobutylene. Users may calibrate to alternative VOCs.
DIRECT SENSETM
IAQI n d o o r A i r Q u a l i t y M o n i t o r
a d v a n c e d e n v i r o n m e n t a l m e a s u r e m e n t s