Coudé Lab Instrument Interfaces BOULDERMarch 18, 2009
Facility Services Provided
S & O Building- Vacuum- Nitrogen (gas)
Utility building- Power- Chilled Water- Networking- Control Cables- Compressed Air- Telephone
Services Provided toCoudé Floor
• Power– 480 3Ф, 120, 120UPS
• Coolant– constant temperature
• Compressed Air– 100 PSI, 10 SCFM w/dew point -40C
• Vacuum– 100 SCFM
• Nitrogen (gaseous)– 0 – 5 L/m
• Currently we are not planning compressed He
Ground level
• Power• Coolant• Compressed air• Vacuum• Nitrogen, dry
Coudé Azimuth Wrap-up
Coudé Azimuth Wrap-up to Cable-ways
Mezzanine Level•Cable-ways from wrap-up
Coudé level
Umbilical of cabling (coolant, control) •below removable flooring•connected to interface panels•provide power, coolant, and control to instruments on coudé floor
Coudé Floor
Removable floor panels
Cable access through steal floor
Instrument Electronics Rack Location
•Instrument racks 38U •3 platform sides of floor•Cooled •Cabling out of top•Cable ways to openings in floor
•Signals to LIC(Science Instruments & WFC)
Mounting Instrument Tables on Coudé Floor
•Table Support Beams attached to major coudé floor supports•Isolation pads
Networks Provided
• 2.1.4.1 Communication; Ethernet– Provides 1000BASE-SX, 802.3z-1998,– Ethernet connection onto facility communication LAN– Multi-mode 62.5/125, Duplex SC type connector
• 2.1.4.2 Data; Fibre Channel ,– Provides 1000BASE-SX, 802.3z-1998, – Ethernet connection onto facility data LAN– Multi-mode 62.5/125, Duplex SC type connector
• 2.1.4.3 Safety; Ethernet, – Provides 100BASE-SX, 802.3-2008,– Ethernet connection onto facility safety LAN– Multi-mode 62.5/125, Duplex SC type connector
• 2.1.4.4 ATST Synchrobus;– IRIG-B– Fiber pair (duplex SC), Single-mode fiber
Network Cables
•Fabric used– Copper CAT5e– Fiber Multi-mode pairs, single-mode pairs
•All Networks are full duplex; therefore, – Fiber connections are made through use of
SC (duplex) connectors– Copper utilizes RJ-45 connectors
Control Cables
• Four defined types– 24V control signal
• 16 AGW, stranded, shielded, PVC, energy-chain rated
• limit switch, proximity switched, relay control, e-stop
– Non-defined, twisted pair• 16 AGW, stranded, twisted pair, shielded,
energy-chain rated• Differential digital signals (TTL), differential
analog– Coaxial 50
• RF, audio, single highly-modulated signal (clock)– Coaxial 75
• Video
Coudé level
Facility services to Instrument Lab•Power: 480 3Ф, 208 3Ф, 120UPS•Coolant: constant temp•Compressed air•Vacuum
Utility Chase •Fiber•Phone•Control