BMTS 353 1
Chapter 5
Part Three
Blood Pressure and flowby Ibrhim AlMohimeed
12/3/2013
BMTS 353 2
Blood Flow
12/3/2013
• Blood flow helps to understand basic physiological
processes and e.g. the dissolution of a medicine into the
body.
• Blood flow and changes in blood volume, are usually
correlated with concentration of nutrients and other
substance in the blood.
• Also, Blood Flow measurement reflects the
concentration of O2.
BMTS 353 3
Cont. Blood Flow
12/3/2013
Normal blood flow velocity 0,5 m/s 1 m/s (Systolic, large vessel)Normal blood flow velocity 0,5 m/s 1 m/s (Systolic, large vessel)
BMTS 353 4
Blood Flow Measurement
12/3/2013
Blood Pressure
Non-invasive (Indirect) Method of Blood Flow Measurement
1
BMTS 353 5
Ultrasonic Doppler Method
12/3/2013
Blood Pressure• The blood cells in the fluid
reflects the ultrasound signal
with a shift in the ultrasonic
frequency due to its movement.
• In the recent years ultrasound contrast agents have been used in
order to increase the echoes.
c
vff cd 2
f = 2 – 10 MHzc
c = 1500 - 1600 m/s (1540 m/s)f = 1,3 – 13 kHzd
BMTS 353 6
Cont. Ultrasonic Doppler Method
12/3/2013
Blood Pressure
CW DOPPLER PULSED DOPPLER
Range ambiguity
Low flow cannot be detected
No minimum range
Simpler hardware
Minimum range
Accuracy
No minimum flow
(Maximum flow) x (range)= limited
The ultrasound Doppler device can be either a continuous wave or a pulsed Doppler
BMTS 353 7
Laser Doppler Flowmetry
12/3/2013
Blood Pressure• The principle of measurement is the
same as with ultrasound Doppler.
• The laser parameter may have e.g.
the following properties:
5 mW
He-Ne-laser
632,8 nm wavelength
• The method is used for capillary
(microvascular) blood flow
measurements.
BMTS 353 8
Plethysmography Method(Strain Gage)
12/3/2013
Blood PressurePlethysmography means the methods
for recording volume changes of an
organ or a body part.
• Strain gage is made of silicone rubber tubes, which are filled with
conductive liquid (e.g. mercury) whose impedance changes with
volume.
• Venous occlusion cuff is inflated to 40 – 50 mmHg. In this way
there will be the arterial inflow into the limb but no venous outflow.
BMTS 353 9
Plethysmography Method(Electric-Impedance)
12/3/2013
Blood Pressure• Different tissues in a body have a different resistivity. Blood is one
of the best conductors in a body.
• A constant current is applied
via skin electrodes.
• The change in the impedance
is measured.
• The accuracy is often poor.
BMTS 353 10
Plethysmography Method(Photoelectric)
12/3/2013
Blood Pressure• A beam of IR-light is directed to
the part of the tissue which is to
be measured for blood flow (e.g.
a finger or ear lobe).
• The blood flow modulates the attenuated / reflected light which is
recorded.
• The light that is transmitted / reflected is collected with a photo
detector.
A B
Poor measure for changes in volumeVery sensitive to motion artefacts
Method is simple Heart rate is clearly seen
BMTS 353 11
Blood Flow Measurement
12/3/2013
Blood Pressure
Invasive (direct) Method of Blood Flow Measurement
2
BMTS 353 12
Indicator Dilution Methods(Dye Dilution)
12/3/2013
Blood Pressure
• A bolus of indicator, a colored dye (indocyanine green),
is rapidly injected in to the vessel.
• The concentration is measured in the downstream
• The blood is drawn through a colorimetric cuvette and
the concentration is measured using the principle of
absorption photometry.
BMTS 353 13
Indicator Dilution Methods (Thermal Dilution)
12/3/2013
Blood Pressure
• A bolus of chilled saline solution is injected into the
blood circulation system (right atrium).
• This causes decrease in the artery temperature.
• Catheter-tip probes are used to measure the change in
tempreture.
BMTS 353 14