Page 1: anger RADIATION A L E R T TM EXP · 2020. 6. 2. · Size Outputs Anti-Saturation Power Requirements Includes Limited Warranty R angerRADIATION A L E R T TM EXP ® INTERNATIONAL Weight

of alpha, beta, gamma, and x-rays. The digital readout is displayed with a red count light, and a beeper sounds with each count detected. Other features include an adjustable timer, and selectable alert. The free Observer USB Software allows you to set alarms, select preferences, and download your collected data. The Bluetooth and Observer BLE enables saving, sharing, and automation of surveys and RadResponder compatibility.

DisplayGraphic LCD with BacklightAlert Set RangemR/hr .001 - 100 and CPM 1 - 350,000Count Light

Audio IndicatorInternally mounted beeper. 70db @ 1m.

35Sulfur (S35), 90Strontium (Sr/y90), 137Cesium (Cs137), 32Phosphorus (P32), 14Carbon (C14), 131Iodine (I131), 60Cobalt (Co60), and Alpha

DetectorRAP-RS2 Probe - External Halogen-quenched, uncompensated GM tube with thin mica window, 1.4-2.0 mg/cm2window is 45 mm (1.75 in.). Operating RangemR/hr - 1 to 50 1 - 175,000 CPMµSv/hr - 1 to 500 0.1 to 3000 CPSTotal/Timer - 1 to 9,999,000 countsAccuracy mR/hr ±10% with NIST calibration, ±15% factoryµSv/hr ±10% with NIST calibration, ±15% factoryEnergy Sensitivity3340 CPM/mR/hr referenced to Cs137��������� Alpha down to 2 MeV. ��������������� �����������V; typical detection efficiency at 1 MeV is approx. 25%���������������� ���������������������������� window. ��������������������������!���"125 is .02 μCi at contact.

RADIATION A L E R T TM

RangerEXP

5

4

3

2

10

101 102 103 104

ENERGY RESPONSE

EVI

TA

LE

R

PHOTON ENERGY (keV)

CO

UN

TR

AT

E

End WindowSide Wall

a ß y x

P.O. Box 39, 436 Farm Rd. Summertown, TN 38483 1-800-293-5759 | Fax: 931-964-3564www.seintl.com | [email protected]

®

INTERNATIONAL

Tube

aaaaaaaa aaaaaaaaaaa aaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaNORM

Page 2: anger RADIATION A L E R T TM EXP · 2020. 6. 2. · Size Outputs Anti-Saturation Power Requirements Includes Limited Warranty R angerRADIATION A L E R T TM EXP ® INTERNATIONAL Weight

Size

Outputs

Anti-Saturation

Power Requirements

Includes

Limited Warranty

RADIATION A L E R T TM

RangerEXP

®

INTERNATIONAL

Weight570 g (1.25 lbs.)

Temperature Range-10° to +50°C (14° to 122°F)Humidity Range10% to 70% Non-Condensing