470KHz RF LESION GENERATOR
Product Name : ThermoConModel No : TC-50
TC-50’s Basic Accessories
TC-50’s strong points
1) One RF generator but Various applications
2) Very easy to use
3) Very safe to use
4) Automatic output power control
5) Automatic temperature control
6) Automatic working time control
7) Cost effective
1) Various applications
TC-50’s Applications
Low Power Lesion
Maximum power 8W
Temperature range 60 to 90ºC
Lesion time 30 to 180 seconds
High Power Lesion
Maximum power 50WTemperature range 60 to 100ºC
Lesion time 1 to 30 minutes
High power applications
Uterine myoma - Tumor Coagulation and Necrosis
Breast tumor - Tumor Coagulation and Necrosis
Thyroid - Tumor Coagulation and Necrosis
Low power applications
Pain Management – Pain reduction by nerve block
Masseter muscle reduction by nerve block
Calf reduction by nerve coagulation
2) Very easy to use
How to use in high power lesion
Impedance:
to check impedance
Lesion:
to set up temperature
and working time
Start working
How to use in low power lesion
Impedance:
to check impedance
Stimulation:
to find motor nerve
Continuous Lesion:
to set up temperature
and working time
3) Very safe to use
Check electrode warning
Low power lesion mode High power lesion mode
BUT! Electrode for high power has been connected
BUT! Electrode for low power has been connected
4) Automatic output power control
5) Automatic temperature control
6) Automatic working time control
▪ In High Power mode
▪ Target temperature: 100ºC
▪ Working time: 10minutes
▪ At the beginning, automatically power UP to temperature UP!
▪ Stay power to Stay temperature!
▪ Stop working automatically!
TC-50’s RFA mechanism
Temperature sensor probe
Cannula electrode
Ablation area
Ablation range
▪ Exposure: 30mm
▪ Temperature: 90ºC
Temperature sensor probes
Cannula electrodes