-1Adjust_Center_Beam.fm17. April 2012, 12:08 DRAFT
1.1 Center beam adjustment
After the generator has been mounted to the system and is placed into operation, the center beamneeds to be adjusted, before the collimator is installed.
1.1.1 Required tools for center beam adjustment
In order to perform this adjustment, a center beam test device is required (91116).Aditionally a pre collimator should be used for radiation protection during adjustment.(65699)
Fig. 1-1 Center beam test device and pre collimator
7/23/2019 ZIEHM C-ARM BEAM ADJ..pdf
http://slidepdf.com/reader/full/ziehm-c-arm-beam-adjpdf 2/6
-2 Adjust_Center_Beam.fm17. April 2012, 12:08 DRAFT
1.1.2 Setup for center beam adjustment
In order to do the center beam adjustment, the center beam test device has to be put on the gen-erators output window. Insert the pre collimator first, in order to provide radiation protection (beamwill be limited)
Fig. 1-2 Setup for center beam adjustment
1.1.3 Procedure of center beam adjustment
Center beam adjustment is done in two steps:
• First: find the correct position of the center beam test device
• Second: adjust the generator position to get the image of the center beam test device into thecenter of the screen
Note
Horizontal and vertical image reversal (mirroring) and Image rotation must be inac-tive in order to provide a correct readout of the image on the screen.
7/23/2019 ZIEHM C-ARM BEAM ADJ..pdf
http://slidepdf.com/reader/full/ziehm-c-arm-beam-adjpdf 3/6
-3Adjust_Center_Beam.fm17. April 2012, 12:08 DRAFT
1.1.3.1 Correct position of the center beam test device.
By shifting the position of the center beam test device, a concentric shape of the center beam testdevice has to appear on the screen. This concentric shape may appear uncentered to the screen.
Fig. 1-3 Shifting the center beam test device
Fig. 1-4 wrong and correct position of the center beam test device
Warning
Center beam adjustment is done while screening.
Take care for radiation protection.
wrong correct
7/23/2019 ZIEHM C-ARM BEAM ADJ..pdf
http://slidepdf.com/reader/full/ziehm-c-arm-beam-adjpdf 4/6
-4 Adjust_Center_Beam.fm17. April 2012, 12:08 DRAFT
1.1.3.2 Adjusting the generator position
The generator position can be adjusted in two directions:
− by the camshaft in order to shift the image of the center beam test device left and right
− by the generator attachment in order to shift the image of the center beam test device upand down.
1.1.3.2.1 Adjusting camshaft
Fig. 1-5 elements for adjusting the camshaft
In order to adjust the camshaft, the countering nut has to be loose.
By adjusting the generator camshaft, the concentric shape has to be centered to the image.Only a left to right centering is possible by this adjustment.
Fix the coutering nut tightly after the adjustment is done (46Nm). Use two 17mm ring wrenches,one for tightening the countering nut and one for keeping the adjustment in position.
7/23/2019 ZIEHM C-ARM BEAM ADJ..pdf
http://slidepdf.com/reader/full/ziehm-c-arm-beam-adjpdf 5/6
-5Adjust_Center_Beam.fm17. April 2012, 12:08 DRAFT
Fig. 1-6 Adjusting generator camshaft
Fig. 1-7 Effect of camshaft adjustment
Warning
Center beam adjustment is done while screening.
Take care for radiation protection.
7/23/2019 ZIEHM C-ARM BEAM ADJ..pdf
http://slidepdf.com/reader/full/ziehm-c-arm-beam-adjpdf 6/6
-6 Adjust_Center_Beam.fm17. April 2012, 12:08 DRAFT
1.1.3.2.2 Adjusting Generator attachment
If there is a top down deviation, this is adjusted by turning the generator attachment.In order to to this, the fixation bolts have to be losened. (8mm allen key required)Then the generator can be adjusted in a small range.Fix the bolts tightly when the correct position is found (46Nm) .
Fig. 1-8 Adjusting generator attachment
In the image above generator is removed in order to make the attachment visible.For adjustment, the generator is not removed.
Fig. 1-9 Effect of generator attachment adjustment
Center beam adjustment is finished if the concentric shape appears centered to the screen.
Warning
Center beam adjustment is done while screening.
Take care for radiation protection.
fixation bolts