64
ZEISS IOLMaster 700 Replacing assumptions with measurements

ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

  • Upload
    others

  • View
    9

  • Download
    0

Embed Size (px)

Citation preview

Page 1: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

ZEISS IOLMaster 700Replacing assumptions with measurements

Page 2: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

2Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

Agenda

1

2

3

4

5

6

Total Keratometry (TK)

GUI - Measurement, Analyze

GUI - IOL Calculation

Print & Export Options

Websites

Introduction: SW 1.80 & Total Keratometry (TK)

Page 3: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

3Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

Agenda

1

2

3

4

5

Total Keratometry (TK)

GUI - Measurement, Analyze

GUI - IOL Calculation

Print & Export Options

6 Websites

Introduction: SW 1.80 & Total Keratometry (TK)

Page 4: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

4Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

Total Keratometry (TK)Replacing assumptions with measurements

• Total Keratometry(TK®) is a new measurement that combines telecentric keratometry and SWEPT Source OCT technology for the assessment of anterior and posterior corneal curvature.

• The purpose of TK is to replace standard keratometry, aiming to help to reduce outliers and improve refractive outcomes of IOL calculation in cataract surgery.

• TK is ULIB-compatible, therefore existing standard formulas and IOL constants may be applied.

Picture source: Carl Zeiss Meditec media database

Page 5: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

5Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

Total Keratometry (TK)Replacing assumptions with measurements

With Total Keratometry (TK)

• The posterior corneal surface can now be directly measured with SS-OCT and used in formulas onboard the ZEISS IOLMaster and Z CALC, without changing your clinical workflow.

• Thus, it is no longer necessary to use assumptions or nomograms of the posterior cornea.

This improves the results of toric IOL power calculations: 14% more patients within +/– 0.5D cylinder (Haigis-T*).

Source: Total Keratometry Compendium, Fabian and Wehner 2018*Retrospective post-hoc analysis of 145 normal cataract eyes implanted with aspheric IOL, 6 weeks post-op.

Picture source: Carl Zeiss Meditec media database

Page 6: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

6Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

TK Release

SW 1.80TK

(optional license)

Barrett TK formulas(Requires Barrett Suite

License AND TK License)

ZEISS IOLMaster 700SW 1.80 & Total Keratometry (TK) Overview

Page 7: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

7Carl Zeiss Meditec AG, Oliver Klaproth & Kim Nguyen

Contents of ZEISS IOLMaster Software 1.80 UpdateOverview – Part 1

• Measurement• Enhanced Fixation / AL Check Scan

• Analyse• Display pupil offset (CW-Chord aka angle kappa)

• IOL Calculation• Target Refraction• Change of LVC Status• Support of non-constant toric IOL ranges

• Settings• Flexible print & export options• Reference Image

• Bugfixes• Minor bug fixes

Picture source: Carl Zeiss Meditec media database

Page 8: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

8Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

Contents of Total Keratometry (optional license) Overview – Part 2

• Total Keratometry (TK)

• Added new measurement values:

o Total Keratometry: TK (TK1, TK2, ∆TK, axis)o Posterior Corneal Surface: PCS (PK1, PK2, ∆PK, axis)

• New Barrett TK formulas exclusive to ZEISS IOLMaster 700

• The Barrett TK Universal II and Barrett TK Toric formulas allows the

use of actual posterior corneal surface values for IOL power calculation.

• Only available if Barrett Suite AND Total Keratometry licenses

are both activated.

Picture source: Carl Zeiss Meditec media database

Page 9: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

9Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

Total Keratometry (TK)Replacing assumptions with measurements

ZEISS IOLMaster 700, no hardware or operating change

Proven Telecentric Keratometry and SWEPT Source OCT

Use with trusted formulas on the ZEISS IOLMaster 700 and existing IOL constants

Two new Barrett TK formulas for non-toric and toric IOL

Improves toric and non-toric IOL power calculation

Page 10: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

10Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

Agenda

1

2

3

4

5

6

Total Keratometry (TK)

GUI - Measurement, Analyze

GUI - IOL Calculation

Print & Export Options

Websites

Introduction: SW 1.80 & Total Keratometry (TK)

Page 11: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

11Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

BackgroundPosterior corneal surface has an impact on total corneal power

• EG. Koch et al.: The anterior cornea axis can be considerably different to the posterior cornea axis, resulting in different total corneal astigmatism.

D. D. Koch, S. F. Ali, M. P. Weikert, M. Shirayama, R. Jenkins, und L. Wang, „Contribution of posterior corneal astigmatism to total corneal astigmatism“, J Cataract Refract Surg, Bd. 38, Nr. 12, S. 2080–2087, Dez. 2012

Page 12: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

12Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

How do we handle this today?Existing methods are not entirely satisfactory

Models and nomograms are based on statistical measurements and might not

compensate for outliers

Incorporate IOL power adjustment nomograms into formulas

>Bild<

Comprehensive eye scanners

Other devices provide posterior measurements, but are not always compatible with trusted formulas

and constants

Page 13: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

13Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

Total Keratometry MeasurementNo change to existing workflow

No change tomeasurement

workflow< 45 sec!

Page 14: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

14Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

Total Keratometry MeasurementEnabled by patented Cornea-to-Retina Scan based on SS-OCT

Po

steri

or

K

Page 15: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

15Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

Total Keratometry MeasurementMeasurement principle

Page 16: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

16Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

Total Keratometry MeasurementMeasuring total corneal power

Page 17: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

17Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

Measuring the Posterior Corneal Surface with the ZEISS IOLMaster 700Enhanced biometric measurement for individualized surgical planning

• This first description of posterior corneal astigmatism measurement by the ZEISS IOLMaster 700

• Found the average magnitude of posterior corneal astigmatism and proportion of vertical orientation of steep axis was lower than previous estimates.

• The ZEISS IOLMaster 700 appears capable of providing enhanced biometric measurement for individualized surgical planning.

LaHood BR, Goggin M. Measurement of Posterior Corneal Astigmatism by the IOLMaster 700. J Refract Surg 2018;34(5):331–36.

Page 18: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

18Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

Remarkable IOL Calculation Results ObservedImprove toric IOL calculation

Source: Total Keratometry CompendiumFabian and Wehner 2018* *Retrospective post-hoc analysis of 145 normal cataract eyes implanted with aspheric IOL, 6 weeks post-op.

Improve toricIOL calculation.

Page 19: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

19Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

Total Keratometry Post-Laser Vision Correction?First post-SMILE cases promising

Ganesh, Sri. Outcomes of post-SMILE Cataract Surgery with multifocal IOL implantation. CRSTE 2018.

“Using the IOLMaster 700 for biometric measurements, including Total Keratometry and IOL calculation, and by choosing the AT LISA tri 839MP IOL, we were able to achieve excellent

refractive and functional outcomes.”

Page 20: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

20Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

What Do Doctors Say?

Graham Barrett, MD, Perth, Australia

“I am very impressed with the results I have obtained with the IOLMaster 700.”

“TK has the potential to reduce refractive surprises to a minimum.”

Han Bor Fam, MD, Singapore

“For toric and non-toric, The TK value improves the overall outcome.”

“It tightens consistency, it reduces the outliers and enhances the overall outcome.”

Page 21: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

21Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

Agenda

1

2

3

4

5

6

Total Keratometry (TK)

GUI - Measurement, Analyze

GUI - IOL Calculation

Print & Export Options

Websites

Introduction: SW 1.80 & Total Keratometry (TK)

Page 22: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

22Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

ZEISS IOLMaster 700 SW 1.80Customized user Interface for your personal preference

Same GUI(grey interface

optional)

Page 23: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

23Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

ZEISS IOLMaster 700 SW 1.80Enhanced Fixation / AL Check Scan throughout the entire measurement process

• Axial length measurement is taken from different time points in the measurement process and will be compared to further ensure the patient was fixating.

1. Anterior segment OCT scan (ie. during fine alignment scan)

• Axial length measurement #1 taken

2. Retina OCT scan (ie. during fixation check scan)

• Axial length measurement #2 taken

Page 24: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

24Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

ZEISS IOLMaster 700 SW 1.80 & TKEvaluate your measurements with intuitive quality checks

Keratometry (K)

Total Keratometry (TK)

Page 25: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

25Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

ZEISS IOLMaster 700 SW 1.80 & TKEasily evaluate quality of your measurements

Page 26: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

26Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

ZEISS IOLMaster 700 SW 1.80 & TKConveniently view all keratometry data on one screen – K, TK & PK

Keratometry (K)

Total Keratometry (TK)

Posterior Keratometry (PK)

Page 27: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

27Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

ZEISS IOLMaster 700 SW 1.80 & TKConveniently view all keratometry data on one screen – K, TK & PK

Check individual

cylinder and axis

Page 28: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

28Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

• The IOLMaster is the only optical biometer on the market that uses a smart optical configuration that allows telecentric and thus distance-independent keratometry.

o ‘Motion independent’ keratometryo Independent of focus (easy & comfortable to use)o Direct radius measurement (constant spot distance)o Visual assessment of dots for tear film integrity

• Robust, precise and repeatable keratometry measurements

• NOW, the IOLM combines unique telecentric keratometry measurement of the anterior corneal surface with measurement of the posterior corneal surface in order to calculate Total Keratometry (TK).

Telecentric KeratometryUnique optical system to the ZEISS IOLMaster

Page 29: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

29Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

IOLM

70

0 • 3-zoneTelecentricKeratometry

• 18 points

• *1.5mm• *2.5mm• *3.5mm

• SE, K1, K2, ∆K, axis• TSE, TK1, TK2, ∆TK, axis• PSE, PK1, PK2, ∆PK, axis

3-zone Telecentric KeratometryFor improved repeatability & reproducibility

• Average of all 18 points are used

o “3-zone consistency” (Improved repeatability and reproducibility due to improved consistency checks – less outliers)

o “3 zone correlation” (additional correlation in 3 zones for improved repeatability)

* Approximate values, depending on corneal radii

Page 30: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

30Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

Chang-Waring Chord (Angle Kappa)Evaluate the axis symmetry of the eye

• The IOLMaster offers the option to view thepupil offset in:

• Cartesian coordinates (x,y) or• Polar coordinates (CW-Chord)

• Easily set preference in Advanced settings > Parameters, units > Chang-Waring Chord

Page 31: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

31Carl Zeiss Meditec AG, Oliver Klaproth & Kim Nguyen

Angle alpha Angle kappa (CW-Chord)

• Angle alpha is the difference between the center of the limbus and the visual axis.

• Angle kappa is the difference between the center of the pupil and the visual axis.

Angle Alpha and Angle KappaEvaluate the axis symmetry of the eye

Page 32: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

32Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

Chang-Waring Chord (Angle Kappa)Evaluate pre-op for MIOL implantation – Optimize IOL centration

Px = -0,1 mm

Py = + 0,5 mm CWC = 0,5 mm @ 101°

Page 33: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

33Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

Chang-Waring Chord (Angle Kappa)Evaluate pre-op for MIOL implantation – Screen patients suitability

• Jack Holladay – „Pay attention to the Angla Kappa!“

• The dimension between the visual axis and the pupil is called the angla kappa (or chord mu / CWC).

• The proper place to center a diffractive, premium aspheric, or toric IOL is halfway between the visual axis and thecenter of the pupil.

• If that value is greater than 0.6mm, patients with diffractive lenses will have halos and glare.

Jack T. Holladay, MD, Hidden Figures, Cataract & Refractive Surgery Today Europe, April 2018, pg 78-80

Page 34: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

34Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

Agenda

1

2

3

4

5

6

Total Keratometry (TK)

GUI - Measurement, Analyze

GUI - IOL Calculation

Print & Export Options

Websites

Introduction: SW 1.80 & Total Keratometry (TK)

Page 35: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

35Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

Target refraction

Keratometry & Total Keratometry

TK values

LVC Status

ZEISS IOLMaster 700 - IOL Calculation screenNew features to improve workflow efficiency

Page 36: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

36Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

Target refractionSetting target refraction individually for OD & OS to improve workflow

Workflow improvement

Page 37: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

37Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

Target refractionIdentify changes in target refraction with appropriate warning messages

Appropriatewarning

messages

Page 38: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

38Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

Change of LVC statusNo need to re-measure the patient!

• Ability to change LVC status post-acquisiton.

o Only Barrett Suite (ie. Barrett True-K) will calculate forpost-RK patients (withstandard Ks).

Page 39: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

39Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

Total Keratometry & Post-LVC FormulasExpand IOL calculation for your post-LVC patients

• TK measurement values can be used to calculate post myopic LASIK and PRK with Haigis and Holladay II formulas (ie. formulas that do not consider Ks for the ELP prediction).

• TK measurement values cannot be used with LVC formulas such as Haigis-L and Barrett True-K.

• ZEISS IOLMaster 700 will block any calculations that are not yet available.

o EG. Barrett TK calculation is not possible using LVC ‘RK’.

o EG. Barrett TK Toric calculation is not possible using LVC ‘LASIK/LASEK/PRK’.

Page 40: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

40Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

ZEISS IOLMaster 700 - IOL Calculation screenView and compare TK & K measurement values on one screen

Page 41: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

41Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

Total Keratometry (TK)

ZEISS IOLMaster 700 - IOL Calculation screenView and compare TK & K measurement values on one screen

Keratometry (K)

Page 42: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

42Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

Detailed IOL information screenQuickly view and change IOL selection

Formula used forIOL calculation Easily toggle between

standard K & TK forIOL calculation

Select IOL

View formulaconstants

Change IOL

Page 43: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

43Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

Detailed IOL information screenIdentify selected IOL and formula

Page 44: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

44Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

Preselection of an IOL and export to ZEISS CALLISTO eyeAutomatic transfer target axis and detailed printout

Page 45: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

45Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

ZEISS IOLMaster 700 - IOL Calculation screenNo need to manually enter data to online calculators

• Support of non-constant toric IOL ranges

• EG. Alcon toric IOLs

• Cylinder step sizes• Naming convention (eg. T2, T3,

T4, etc)

Page 46: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

46Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

Select IOL

• Spherical selection (IOL SE)

ZEISS IOLMaster 700 - IOL Calculation screenNo need to manually enter data to online calculators

Page 47: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

47Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

ZEISS IOLMaster 700 - IOL Calculation screenNo need to manually enter data to online calculators

Select IOL

• Toric selection (Cyl & Axis)

• 1 overcorrection (axis flip)• 2 undercorrection (no axis flip)

Page 48: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

48Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

Agenda

1

2

3

4

5

6

Total Keratometry (TK)

GUI - Measurement, Analyze

GUI - IOL Calculation

Print & Export Options

Websites

Introduction: SW 1.80 & Total Keratometry (TK)

Page 49: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

49Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

Print and Export OptionsCustomizable and flexible options to improve workflow efficiency

Set printer & exportsettings in menuwill be applied

Page 50: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

50Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

Print and Export OptionsCustomizable and flexible options to improve workflow efficiency

Print options

will appear

Export options

will appear

Print & Export icons

Page 51: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

51Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

Print and Export OptionsCustomizable and flexible options to improve workflow efficiency

Optional pop-up menu

• Pop-up print/export menu will appearwhen clicking on icon (default settings)

• User can easily change as required

• Eyes to be printed (Selected/Both)

• Formula results to be printed (same asIOLM500)

Page 52: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

52Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

Print and Export OptionsCustomizable and flexible options to improve workflow efficiency

• Optional pop-up menu

• Enable in Advanced settings > Printout > Show print options

• Enable in Advanced settings > Export > Enable PDF export > Display export options

Page 53: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

53Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

ZEISS IOLMaster 700 PrintoutsIntuitive and detailed printouts

• Printouts available:

• IOL Calculation (OD/OS)

• Analyze (OD)

• Analyze (OS)

• Biometric values (OD/OS)

• Corneal values (OD/OS)

• Reference image (OD)

• Reference image (OS)

Picture source: Carl Zeiss Meditec Media Database

NEW!

Page 54: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

54Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

IOLCalculation

(OU)

Picture source: Carl Zeiss Meditec Media Database

TK

Page 55: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

55Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

Analyze(OD/OS)

TK

Picture source: Carl Zeiss Meditec Media Database

Page 56: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

56Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

BiometricValues

(OD/OS)

TK

Picture source: Carl Zeiss Meditec Media Database

Page 57: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

57Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

NEW!CornealValues(OU)

TK

Picture source: Carl Zeiss Meditec Media Database

Page 58: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

58Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

ReferenceImage

(OD/OS)

TK

Picture source: Carl Zeiss Meditec Media Database

Page 59: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

59Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

Reference ImageOptions for report – Implantation axis or steep axis to be printed

Chosen IOL KeratometryTotal Keratometry

3 Blue LinesCALLISTO implantation axis

2 Green LinesTotal Keratometry steep axis

(2 surfaces - anterior & posterior)

1 Purple LineKeratometry steep axis(1 surface - anterior)

Page 60: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

60Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

Reference ImageOptions for report – Implantation axis or steep axis to be printed

Page 61: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

63Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

Online User HelpContext Sensitive Help Menu

Page 62: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

64Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

Agenda

1

2

3

4

5

6

Total Keratometry (TK)

GUI - Measurement, Analyze

GUI - IOL Calculation

Print & Export Options

Websites

Introduction: SW 1.80 & Total Keratometry (TK)

Page 63: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

65Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)

Total Keratometry (TK)Websites

• IOLMaster 700 website www.zeiss.com/IOLMaster

• Total Keratometry website www.zeiss.com/tk

• ZEISS Cataract Community https://cataract-community.zeiss.com/

• IOL Constants ULIB http://ocusoft.de.ulib/c1.htmIOLCon http://iolcon.org/

Picture source: Carl Zeiss Meditec Media Database

Page 64: ZEISS IOLMaster 700 Replacing assumptions with measurements · 2021. 8. 2. · Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK) 4 Total Keratometry (TK) Replacing

66Carl Zeiss Meditec AG, IOLMaster 700 SW 1.80 & Total Keratometry (TK)EN_32_150_0064I