Central placement of screw fixation in scaphoid fracture—a computational approach and biomechanical study

Yang Guo MD, FHKCOSAssociate ConsultantDepartment of Hand Surgery, Ji Shui Tan hospital, beijing, China

Page 2: Yang  Guo  MD, FHKCOS Associate Consultant

OPTIMAL position of fixation for scaphoid fractures continues to be debatedCentral placement : supported

by(Tremble,Mccallister,2000)biomechanical studies /clinical evaluations Has been accepted as a prefered method of fixationRecently, some studies :(Luria 2010,2012) biomechanical studies /finite element analysissuggested no difference in stability between central and eccentric placement

The authors did not receive any outside funding or grants in support of their research for or preparation of this work. Neither they nor a member of their immediate families received payments or other benefits.

Page 3: Yang  Guo  MD, FHKCOS Associate Consultant

To be central or Not to be?Our study consisted of 2 parts

Establishment “central” concept computational approach

Biomechanical test central Vs peripheral

Page 4: Yang  Guo  MD, FHKCOS Associate Consultant

1 Establishing “Central”

Page 5: Yang  Guo  MD, FHKCOS Associate Consultant

Shrinkage in a uniform manner in software-- pre-op planning systemScaphoid surface model was extracted.One pixel space was allowed by one time of shrinkage in the programScaphoid model: smaller and smaller

Page 6: Yang  Guo  MD, FHKCOS Associate Consultant

Scahpoid distal pole

Scahpoid proximal pole

Waist portion

Central one third

Further shrinkage of 1.5mm

Central zone

Central one-third zone and central zone

Page 7: Yang  Guo  MD, FHKCOS Associate Consultant

Central zone-geometry

Screw axis

Page 8: Yang  Guo  MD, FHKCOS Associate Consultant

Part 2: Biomechanical Study Design : Types of fixation

Fracture Plane Location of Screw

Group 1 Proximal Proximal Central Zone

Group 2 Proximal Distal Central Zone(eccentric)

Group 3 Distal Distal Central Zone

Group 4 Distal Proximal Central Zone(eccentric)

Page 9: Yang  Guo  MD, FHKCOS Associate Consultant

Biomechanical study : osteotomy design

Page 10: Yang  Guo  MD, FHKCOS Associate Consultant

Group 1 Group 2

Group 3Group 4

Screw axisthread

Fracture plane

Page 11: Yang  Guo  MD, FHKCOS Associate Consultant

Materials and Test machine

Page 12: Yang  Guo  MD, FHKCOS Associate Consultant
Page 13: Yang  Guo  MD, FHKCOS Associate Consultant
Page 14: Yang  Guo  MD, FHKCOS Associate Consultant

Estimation of sample size and power10 of scaphoids , consistent with a previous dataData analysis: Wilcoxon nonparametric analysis Failure load of the specimens (P <0.05).The mode of failure :Screw migration fracture at the screw–bone interface, sliding along the fracture. None of the specimens dislodged from the potting fixtureNone of screw was bent

Page 15: Yang  Guo  MD, FHKCOS Associate Consultant

Stroke-Force curve

Page 16: Yang  Guo  MD, FHKCOS Associate Consultant

ResultsLoad to failure (N)

Dispalcement of fracture ( mm) Significance

Group 1 80.82±15.63 2.3±0.5 0.001

Group 2 58.32±17.18 3.1±0.5

Group 3 76.83±14.54 2.2±0.7 0.166

Group 4 73.38±13.32 2.5±0.6

Page 17: Yang  Guo  MD, FHKCOS Associate Consultant

Brief Disccussion

1. Concept of Central need to be clarified Straight central axis does not exist

2. Central placement : important as long as possible

Page 18: Yang  Guo  MD, FHKCOS Associate Consultant

LimitationsSame with any cadaveric biomechanical studyThe screw itself may contributed resistance ,when it was placed in a more horizontal direction

Saw bone material: Failure load lower than previous tests

We simplified the forces into a single vector and did not perform cyclic loading

Fx plane differed slightly, which affects its stability.