21
XIS Calibration at XIS Calibration at Osaka Osaka K. Hayashida, K. Torii, M. N K. Hayashida, K. Torii, M. N amiki, T. Shiroshoji, amiki, T. Shiroshoji, M. Shoji, S. Kastuda, H. M. Shoji, S. Kastuda, H. Tsunemi Tsunemi (Osaka University) (Osaka University) T. Kohmura (Kougakuin Univer T. Kohmura (Kougakuin Univer sity) sity)

XIS Calibration at Osaka K. Hayashida, K. Torii, M. Namiki, T. Shiroshoji, M. Shoji, S. Kastuda, H. Tsunemi (Osaka University) T. Kohmura (Kougakuin University)

Embed Size (px)

Citation preview

Page 1: XIS Calibration at Osaka K. Hayashida, K. Torii, M. Namiki, T. Shiroshoji, M. Shoji, S. Kastuda, H. Tsunemi (Osaka University) T. Kohmura (Kougakuin University)

XIS Calibration at OsakaXIS Calibration at Osaka

K. Hayashida, K. Torii, M. Namiki, K. Hayashida, K. Torii, M. Namiki, T. Shiroshoji, M. Shoji, S. Kastuda, T. Shiroshoji, M. Shoji, S. Kastuda,

H. Tsunemi H. Tsunemi

(Osaka University) (Osaka University)

T. Kohmura (Kougakuin University)T. Kohmura (Kougakuin University)

Page 2: XIS Calibration at Osaka K. Hayashida, K. Torii, M. Namiki, T. Shiroshoji, M. Shoji, S. Kastuda, H. Tsunemi (Osaka University) T. Kohmura (Kougakuin University)

Calibration Item at Osaka Calibration Item at Osaka Ex<2.2keVEx<2.2keV

Quantumn Efficiency as a function of Quantumn Efficiency as a function of energyenergyXANES at O-K, Si-KXANES at O-K, Si-K

Ex-PH relation (gain)Ex-PH relation (gain)Ex-FWHM relation (energy resolution)Ex-FWHM relation (energy resolution)Response Profile to Single Energy IncResponse Profile to Single Energy Inc

idence idence

Page 3: XIS Calibration at Osaka K. Hayashida, K. Torii, M. Namiki, T. Shiroshoji, M. Shoji, S. Kastuda, H. Tsunemi (Osaka University) T. Kohmura (Kougakuin University)

Quantumn Efficiency MeasurementQuantumn Efficiency Measurement Relative Efficiencies of FM-S0,S1,S2,S3, and XIRelative Efficiencies of FM-S0,S1,S2,S3, and XI

S-EU are measured by irradiating X-rays from thS-EU are measured by irradiating X-rays from the spectrometer to whole the CCD area. e spectrometer to whole the CCD area. Generator beam current is always monitoredGenerator beam current is always monitored Si Photodiode is sometimes used to measure the 0Si Photodiode is sometimes used to measure the 0thth o o

rder light.rder light.

XIS-EU was cross-calibrated to a Gas PC on 20XIS-EU was cross-calibrated to a Gas PC on 2003Dec. XIS-FM are not installed in the chamber 03Dec. XIS-FM are not installed in the chamber with the Gas PC simultaneously.with the Gas PC simultaneously.

The gas PC was calibrated in 2000, and calibratThe gas PC was calibrated in 2000, and calibrated again through the slant incident method in 20ed again through the slant incident method in 2004 January. 04 January.

Page 4: XIS Calibration at Osaka K. Hayashida, K. Torii, M. Namiki, T. Shiroshoji, M. Shoji, S. Kastuda, H. Tsunemi (Osaka University) T. Kohmura (Kougakuin University)

Detector Chamber

Manson Soft X-ray Generator

Hetrick Spectrometer

Calibration Facility in Osaka Clean Calibration Facility in Osaka Clean Room Room

Page 5: XIS Calibration at Osaka K. Hayashida, K. Torii, M. Namiki, T. Shiroshoji, M. Shoji, S. Kastuda, H. Tsunemi (Osaka University) T. Kohmura (Kougakuin University)

XIS measurement without slitXIS measurement without slitNo Gas PC inside the chamberNo Gas PC inside the chamber

Page 6: XIS Calibration at Osaka K. Hayashida, K. Torii, M. Namiki, T. Shiroshoji, M. Shoji, S. Kastuda, H. Tsunemi (Osaka University) T. Kohmura (Kougakuin University)

X-rays to whole the CCD areaX-rays to whole the CCD area

Dispersion Direction

Page 7: XIS Calibration at Osaka K. Hayashida, K. Torii, M. Namiki, T. Shiroshoji, M. Shoji, S. Kastuda, H. Tsunemi (Osaka University) T. Kohmura (Kougakuin University)

Slant Incidence MethodSlant Incidence Method

No need to use other reference counters. / One way to make a reference counter.

In the simplest case, in which all the X-rays stop in the detection layer

d)exp()0( 0 II

)cos/dexp()( 0 IIX-rays

Dead Lasyer

Detection Layer

θ

d

))(/)0(ln(cos1

cos1d

II

Dead layer thickness is

Even if detection layer thickness is finite, we can measure the dead layer thickness by using the ratio of I(0)/I() of various energies.

Hayashida et al., 2003, SPIE4851

Page 8: XIS Calibration at Osaka K. Hayashida, K. Torii, M. Namiki, T. Shiroshoji, M. Shoji, S. Kastuda, H. Tsunemi (Osaka University) T. Kohmura (Kougakuin University)

X-rays

Slant Incidence Method: Experiment with HPK FI CCD camera

Page 9: XIS Calibration at Osaka K. Hayashida, K. Torii, M. Namiki, T. Shiroshoji, M. Shoji, S. Kastuda, H. Tsunemi (Osaka University) T. Kohmura (Kougakuin University)

0.4keV ~1keV

1keV ~ 2keV

O-Kα W-Mα

I(0)

I(45)

I(0)

I(45)

Hayashida et al., 2003,SPIE4851

HPK FI-CCDHPK FI-CCD

Page 10: XIS Calibration at Osaka K. Hayashida, K. Torii, M. Namiki, T. Shiroshoji, M. Shoji, S. Kastuda, H. Tsunemi (Osaka University) T. Kohmura (Kougakuin University)

Energy [keV]

I(45

)/I(

0)

Ratio Ratio I(45I(45°°)/I(0)/I(0°°))

SiO2(SiO2(m)m) Si(Si(m)m) Depletion Depletion Layer (Layer (m)m)

MeasuredMeasured 1.731.73±0.05±0.05 0.520.52±0.06±0.06 8.298.29±0.39±0.39

DesignDesign ~2~2 ~0.4~0.4

*) HPK FI CCD used in this experimentwas old model.

Hayashida et al., 2003,SPIE4851

Page 11: XIS Calibration at Osaka K. Hayashida, K. Torii, M. Namiki, T. Shiroshoji, M. Shoji, S. Kastuda, H. Tsunemi (Osaka University) T. Kohmura (Kougakuin University)

CCD raw frame dataCCD raw frame data

Experiment with HPK BI-CCDExperiment with HPK BI-CCD

Energy Low⇔High

O Kα

O Kα Fe 55

Page 12: XIS Calibration at Osaka K. Hayashida, K. Torii, M. Namiki, T. Shiroshoji, M. Shoji, S. Kastuda, H. Tsunemi (Osaka University) T. Kohmura (Kougakuin University)

0°8mm

0°0mm

30°8mm

30°0mm 45°0mm

45°8mm

O-K1st order

C-K

O-K 2nd order

BI-CCD

Page 13: XIS Calibration at Osaka K. Hayashida, K. Torii, M. Namiki, T. Shiroshoji, M. Shoji, S. Kastuda, H. Tsunemi (Osaka University) T. Kohmura (Kougakuin University)

X-raysX-rays

Slant Incident Method: Application Slant Incident Method: Application to Gas PCto Gas PC

We determined to use the Gas PC as the reference counter

Page 14: XIS Calibration at Osaka K. Hayashida, K. Torii, M. Namiki, T. Shiroshoji, M. Shoji, S. Kastuda, H. Tsunemi (Osaka University) T. Kohmura (Kougakuin University)

Cou

nts

PH (ch)

0° 30° 45°

Gas PC Spectra for Different Gas PC Spectra for Different Incident AngleIncident Angle

Page 15: XIS Calibration at Osaka K. Hayashida, K. Torii, M. Namiki, T. Shiroshoji, M. Shoji, S. Kastuda, H. Tsunemi (Osaka University) T. Kohmura (Kougakuin University)

  Window   (C3H6)

H2O

CC33HH66 [ [m]m] 0.974±0.0760.974±0.076

HH22O [O [m]m] 0.318±0.0600.318±0.060

Dead layer [Dead layer [m]m] 70.9±5.770.9±5.7

Gas Detection Layer

Gas dead layer beneath the window is introduced to fit the data.

Gas Dead Layer

Xray

Simple Model cannot reproduce the Simple Model cannot reproduce the data points of the Ratio data points of the Ratio

Page 16: XIS Calibration at Osaka K. Hayashida, K. Torii, M. Namiki, T. Shiroshoji, M. Shoji, S. Kastuda, H. Tsunemi (Osaka University) T. Kohmura (Kougakuin University)

0.01

0.1

1

0.2 0.4 0.6 0.8 1

モデル:斜入射較正法モデル:窓の透過率測定

Energy(keV)

Eff

icie

ncy

0.01

0.1

1

0.2 0.4 0.6 0.8 1

モデル:斜入射較正法モデル:窓の透過率測定

Energy(keV)

Eff

icie

ncy

0.01

0.1

1

0.2 0.4 0.6 0.8 1

モデル:斜入射較正法モデル:窓の透過率測定

Energy(keV)

Eff

icie

ncy

The transmission of the PC window was also measured in 2000 by on/off method.

Two independent methods agree well above 0.6keV

Gas dead layer is only measurable with the slant incident method.

Model by Slant Incident

Model by Window Tranmission

Detection Efficiency of PCDetection Efficiency of PC

Page 17: XIS Calibration at Osaka K. Hayashida, K. Torii, M. Namiki, T. Shiroshoji, M. Shoji, S. Kastuda, H. Tsunemi (Osaka University) T. Kohmura (Kougakuin University)

XIS-EU and Gas PC cross XIS-EU and Gas PC cross calibration using slit calibration using slit

Page 18: XIS Calibration at Osaka K. Hayashida, K. Torii, M. Namiki, T. Shiroshoji, M. Shoji, S. Kastuda, H. Tsunemi (Osaka University) T. Kohmura (Kougakuin University)

X-rays through Slit ( ~1mm)X-rays through Slit ( ~1mm)

Dispersion Direction

Page 19: XIS Calibration at Osaka K. Hayashida, K. Torii, M. Namiki, T. Shiroshoji, M. Shoji, S. Kastuda, H. Tsunemi (Osaka University) T. Kohmura (Kougakuin University)

PC (0.525keV) XIS-EU (0.525keV)

PC Spectra and CCD Spectra PC Spectra and CCD Spectra

Page 20: XIS Calibration at Osaka K. Hayashida, K. Torii, M. Namiki, T. Shiroshoji, M. Shoji, S. Kastuda, H. Tsunemi (Osaka University) T. Kohmura (Kougakuin University)

0

0.1

1

0.2 0.4 0.6 0.8 1Energy(keV)

データ点モデル

Eff

icie

ncy

0

0.1

1

0.2 0.4 0.6 0.8 1Energy(keV)

データ点モデル

Eff

icie

ncy

0

0.1

1

0.2 0.4 0.6 0.8 1Energy(keV)

データ点モデルデータ点モデル

Eff

icie

ncy

SiOSiO22 [ [m]m] 0.57±0.050.57±0.05

Si [Si [m]m] 0.30±0.030.30±0.03

BPSG of EU ~200nm

cf ~70nm for FM

QE of XIS-EUQE of XIS-EU

New Structure model is employed.Depletion Layer 65m is assumed.

cf Mark-san reports SiO2~0.35-0.39mSi ~0.25-0.31m

Page 21: XIS Calibration at Osaka K. Hayashida, K. Torii, M. Namiki, T. Shiroshoji, M. Shoji, S. Kastuda, H. Tsunemi (Osaka University) T. Kohmura (Kougakuin University)

ScheduleSchedule XIS-EU and PC cross calXIS-EU and PC cross cal XIS-EUXIS-EU XIS-FMS2 (Noise Problem)XIS-FMS2 (Noise Problem) PC slant incidence methodPC slant incidence method XIS-FMS3XIS-FMS3 XIS-FMS2 againXIS-FMS2 again XIS-EUXIS-EU XIS-FMS0 (Door Open Problem)XIS-FMS0 (Door Open Problem) XIS-FMS1 XIS-FMS1 XIS-FMS0 (TCE card problem)XIS-FMS0 (TCE card problem) TCE card replacementTCE card replacement XIS-FMS0XIS-FMS0 XIS-FMS1XIS-FMS1 XIS-EUXIS-EU Grating SA->SX for Si-edge XANESGrating SA->SX for Si-edge XANES XIS-EUXIS-EU  (SX) (SX) XIS-FMS0 (SX)? XIS-FMS0 (SX)?