Oxi 3310
Operating manual
Dissolved oxygen (D.O.) meter
USB output
25.0
7.92°C
15.11.200808:37
O2
mg/lMenu
[Sal]
ba75804e06 03/2012
Oxi 3310
Accuracy whengoing to press
The use of advanced technology and the high quality standard of our instruments are the result of a continuous development. This may result in differences between this operating manual and your meter. Also, we cannot guarantee that there are absolutely no errors in this manual. Therefore, we are sure you will understand that we cannot accept any legal claims resulting from the data, figures or descriptions.
Copyright © Weilheim 2009, WTW GmbHReproduction in whole - or even in part - is prohibited without the express written permission of WTW GmbH, Weilheim.Printed in Germany.
ba75804e06 03/2012
Oxi 3310 Contents
Oxi 3310 - Contents
1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51.1 Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61.2 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71.3 Socket field . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92.1 Authorized use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102.2 General safety instructions . . . . . . . . . . . . . . . . . . . . . . . 10
3 Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 133.1 Scope of delivery. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 133.2 Initial commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.2.1 Inserting the batteries. . . . . . . . . . . . . . . . . . . . . 133.2.2 Switching on the meter. . . . . . . . . . . . . . . . . . . . 143.2.3 Setting the date and time . . . . . . . . . . . . . . . . . . 14
4 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 154.1 Switching on the meter . . . . . . . . . . . . . . . . . . . . . . . . . . 154.2 General operating principles . . . . . . . . . . . . . . . . . . . . . . 16
4.2.1 Operating modes . . . . . . . . . . . . . . . . . . . . . . . . 164.2.2 Navigation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 174.2.3 Example 1 on navigation: Setting the language. 194.2.4 Example 2 on navigation: Setting the date and
time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 214.3 Sensor-independent settings . . . . . . . . . . . . . . . . . . . . . 23
4.3.1 System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 234.3.2 Data storage. . . . . . . . . . . . . . . . . . . . . . . . . . . . 244.3.3 Automatic Stability control . . . . . . . . . . . . . . . . . 24
4.4 Dissolved oxygen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 254.4.1 General information . . . . . . . . . . . . . . . . . . . . . . 254.4.2 Measuring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 264.4.3 Settings for D.O. sensors
(menu or measurement and calibration settings) 284.4.4 D.O. calibration . . . . . . . . . . . . . . . . . . . . . . . . . 304.4.5 Displaying calibration records . . . . . . . . . . . . . . 33
4.5 Data storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 354.5.1 Manual storage . . . . . . . . . . . . . . . . . . . . . . . . . 364.5.2 Automatic storage at intervals . . . . . . . . . . . . 374.5.3 Displaying and editing the measurement data
storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 394.5.4 Erasing the measurement data storage . . . . . . . 41
4.6 Transmitting data (USB interface). . . . . . . . . . . . . . . . . . 42
3ba75804e06 03/2012
Contents Oxi 3310
4.6.1 Options for data transmission . . . . . . . . . . . . . . . 424.6.2 Connecting a PC . . . . . . . . . . . . . . . . . . . . . . . . 434.6.3 Operation with MultiLab pilot . . . . . . . . . . . . . . . 43
4.7 Reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 444.7.1 Resetting the measurement settings . . . . . . . . . 444.7.2 Resetting the system settings. . . . . . . . . . . . . . . 45
5 Maintenance, cleaning, disposal . . . . . . . . . . . . . . . . . 475.1 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
5.1.1 Replacing the batteries . . . . . . . . . . . . . . . . . . . . 475.2 Cleaning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 485.3 Packing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 485.4 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
6 What to do if... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
7 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 517.1 General data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 517.2 Measuring ranges, resolution, accuracy . . . . . . . . . . . . . 52
8 Lists . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Appendix: Firmware update. . . . . . . . . . . . . . . . . . . . . 59
4 ba75804e06 03/2012
Oxi 3310 Overview
1 Overview
The Oxi 3310 compact precision dissolved oxygen (D.O.) meter enables you to perform D.O. measurements quickly and reliably.
The Oxi 3310 provides the maximum degree of operating comfort, reliability and measuring certainty for all applications.The proven OxiCal calibration procedure supports you in your work with the D.O. meter.
The USB interface can be used for data transmission to a PC and for software updates of the meter.
1 Keypad
2 Display
3 Socket field
USB output
25.0
7.92°C
15.11.200808:37
O2
mg/lMenu
1
2
3
[Sal]
5ba75804e06 03/2012
Overview Oxi 3310
1.1 Keypad
In this operating manual, keys are indicated by brackets <..> . The key symbol (e.g. <ENTER>) generally indicates a short keystroke (under 2 sec) in this operating manual. A long keystroke (approx. 2 sec) is indicated by the underscore behind the key symbol (e.g. <ENTER_>).
ENTER
M
Oxi 3310
F1 F2
ARCAL
STO RCL
<F1>:<F1_>:<F2>:<F2_>:
Softkeys providing situation dependent functions, e.g.: <F1>/[Menü]: Opens the menu for measurement settings<F1_>/[Menü]: Opens the menu for system settings
<On/Off>: Switches the meter on or off
<M>: Selects the measured parameter
<CAL>: <CAL_>:
Calls up the calibration procedure Displays the calibration data
<STO>:<STO_>:
Saves a measured value manuallyOpens the menu for the automatic save function
<RCL>: <RCL_>:
Displays the manually stored measured valuesDisplays the automatically stored measured values
<▲>: Increments values, scrolls
<▼>: Decrements values, scrolls
F1
F2
CAL
STO
RCL
6 ba75804e06 03/2012
Oxi 3310 Overview
1.2 Display
Function displayindicators
<ENTER>:
<ENTER_>:
Opens the menu for measurement settings / confirms entriesOpens the menu for system settings
<AR> Freezes the measured value (HOLD function)Switches the AutoRead measurement on or off
E N T E R
AR
1 Status information
2 Measured value (with unit)
3 Measured parameter
4 Sensor symbol (calibration evaluation, calibration interval)
5 Measured temperature (with unit)
6 Status line
7 Softkeys and date + time
3
2
4
5
6
7
1 HOLD AR
25.07.92
°C
O2
mg/l
22.09.200808:00Menü USB-Ausgabe
Error An error occurred during calibration
AR Stability control (AutoRead) is active
HOLD Measured value is frozen (<AR> key)
7ba75804e06 03/2012
Overview Oxi 3310
1.3 Socket field
CautionOnly connect D.O. sensors to the meter that cannot return any voltages or currents that are not allowed (> SELV and > current circuit with current limiting). Almost all customary D.O. sensors fulfill these conditions.
Connectors:
1 D.O. sensor
2 USB B (device) interface
3 Service interface
3 1 2
8 ba75804e06 03/2012
Oxi 3310 Safety
2 Safety
This operating manual contains basic instructions that you must follow during the commissioning, operation and maintenance of the meter. Consequently, all responsible personnel must read this operating manual before working with the meter. The operating manual must always be available within thevicinity of the meter.
Target group The meter was developed for work in the field and in the laboratory. Thus, we assume that, as a result of their professional training and experience, the operators will know the necessary safety precautions to take when handling chemicals.
Safety instructions Safety instructions in this operating manual are indicated by the warning symbol (triangle) in the left column. The signal word (e.g. "Caution") indicates the level of danger:
Warningindicates instructions that must be followed precisely in order to avoid possibly great dangers to personnel.
Cautionindicates instructions that must be followed precisely in order to avoid the possibility of slight injuries or damage to the meter or the environment.
Further notesNoteindicates notes that draw your attention to special features.
Noteindicates cross-references to other documents, e.g. operating manuals.
9ba75804e06 03/2012
Safety Oxi 3310
2.1 Authorized use
The authorized use of the meter consists exclusively of the measurement of dissolved oxygen and temperature of liquid media in a field and laboratory environment. The technical specifications as given in chapter 7 TECHNICAL DATA must be observed. Only the operation and running of the meter according to the instructions given in this operating manual is authorized. Any other use is considered unauthorized.
2.2 General safety instructions
This meter is constructed and tested in compliance with the IEC 1010 safety regulations for electronic measuring instruments. It left the factory in a safe and secure technical condition.
Function andoperational safety
The smooth functioning and operational safety of the meter can only be guaranteed if the generally applicable safety measures and the specific safety instructions in this operating manual are followed during operation.
The smooth functioning and operational safety of the meter can only be guaranteed under the environmental conditions that are specified in chapter 7 TECHNICAL DATA.
If the meter was transported from a cold environment to a warm environment, the formation of condensate can lead to the faulty functioning of the meter. In this event, wait until the temperature of the meter reaches room temperature before putting the meter back into operation.
CautionThe meter is only allowed to be opened by authorized personnel.
10 ba75804e06 03/2012
Oxi 3310 Safety
Safe operation If safe operation is no longer possible, the meter must be taken out of service and secured against inadvertent operation!Safe operation is no longer possible if the meter:
has been damaged in transport
has been stored under adverse conditions for a lengthy period of time
is visibly damaged
no longer operates as described in this manual.
If you are in any doubt, please contact the supplier of the meter.
Obligations of thepurchaser
The purchaser of this meter must ensure that the following laws and guidelines are observed when using dangerous substances:
EEC directives for protective labor legislation
National protective labor legislation
Safety regulations
Safety datasheets of the chemical manufacturers.
CautionIn addition to the safety instructions mentioned here, also follow the safety instructions of the sensors used.The operating manuals of the sensors are available on the supplied CD and on the Internet under www.WTW.com.
11ba75804e06 03/2012
Safety Oxi 3310
12 ba75804e06 03/2012
Oxi 3310 Commissioning
3 Commissioning
3.1 Scope of delivery
D.O. meter Oxi 3310
4 batteries 1.5 V Mignon type AA
USB cable
Short instructions
CD-ROM with detailed operating manual
3.2 Initial commissioning
Perform the following activities:
Insert the supplied batteries
Switch on the meter
Set the date and time
3.2.1 Inserting the batteries
1 Unscrew the two screws (1) on the underside of the meter.
2 Open the battery compartment (2) on the underside of the meter.
2
1
13ba75804e06 03/2012
Commissioning Oxi 3310
NoteAlternatively, you can also use Ni-MH rechargeable batteries (type Mignon AA). In order to charge the batteries, an external charging device is required.
CautionMake sure that the poles of the batteries are positioned correctly.The ± signs on the batteries must correspond to the ± signs in the battery compartment.
3.2.2 Switching on the meter
NoteThe meter has an energy saving feature to avoid unnecessary battery depletion. The energy saving feature switches off the measuring instrument if no key is pressed during the adjusted interval. (How to set the switch-off interval, see section 4.3.1).
3.2.3 Setting the date and time
3 Place four batteries (type Mignon AA) in the battery compartment.
4 Close the battery compartment (2) and tighten the screws (1).
1 Press the <On/Off> key.The meter performs a self-test. The display shows the manufacturer's logo while the self-test is being performed.Subsequently, the meter switches to the measuring mode (measured value display).
1 See section 4.2.4
14 ba75804e06 03/2012
Oxi 3310 Operation
4 Operation
4.1 Switching on the meter
Switching on Press the <On/Off> key.The meter performs a self-test. The display shows the manufacturer's logo while the self-test is being performed.The measured value display appears.
Switching off Press the <On/Off> key.
Automatic switch-offfunction
The instrument has an automatic switch-off function in order to save the batteries (see section 4.3.1). The automatic switch-off function switches off the measuring instrument if no key is pressed for an adjustable period.
The automatic switch-off function is not active
if the meter is supplied with power via a connected USB cable
if the Automatic data storage function is active, or with automatic data transmission
Display illumination The meter automatically switches off the display illumination if no key is pressed for 30 seconds. The illumination is switched on with the next keystroke again.
You can also generally switch the display illumination on or off (see section 4.3.1).
O2
7.92 mg/l
25.1 °C
22.09.200808:00Menu USB output
15ba75804e06 03/2012
Operation Oxi 3310
4.2 General operating principles
This section contains basic information on the operation of the Oxi 3310.
Operating elements,display
An overview of the operating elements and the display is given in section 1.1 and section 1.2.
Operating modes,navigation
An overview of the operating modes and navigation of the Oxi 3310 is given in section 4.2.1 and section 4.2.2.
4.2.1 Operating modes
The meter has the following operating modes:
MeasuringThe measurement data of the connected sensor is shown in the measured value display
CalibrationThe course of a calibration with calibration information, functions and settings is displayed
Data storage The meter stores measurement data manually or automatically
Transmitting dataThe measuring instrument transmits measurement data and calibration records to the USB interface automatically or manually.
SettingThe system menu or a sensor menu with submenus, settings and functions is displayed
16 ba75804e06 03/2012
Oxi 3310 Operation
4.2.2 Navigation
Measured value display In the measured value display, you can
Open the menu for calibration and measurement settings with <F1> (short keystroke)
Open the Storage & config menu with the sensor-independent settings with <F1_> (long keystroke (approx. 2 s).
Change the display in the measurement window by pressing <M> (e.g. D.O. concentration −> D.O. saturation index −> D.O. partial pressure −>).
Menus and dialogs The menus for settings and dialogs in procedures contain further submenus. The selection is made with the <▲><▼> keys. The current selection is displayed with a frame.
SubmenusThe name of the submenu is displayed at the upper edge of the frame. Submenus are opened by confirming with <ENTER>. Example:
SettingsSettings are indicated by a colon. The current setting is displayed on the right-hand side. The setting mode is opened with <ENTER>. Subsequently, the setting can be changed with <▲><▼> and <ENTER>. Example:
General
InterfaceClockService information
Reset
System
22.09.200808:00Back
Language: English Beep: OffIllumination: OnContrast: 48 %
Switchoff time: 30 min
General
22.09.200808:00Back
17ba75804e06 03/2012
Operation Oxi 3310
FunctionsFunctions are designated by the name of the function. They are immediately carried out by confirming with <ENTER>. Example: Display the Calibration record function.
Messages Information is marked by the i symbol. It cannot be selected. Example:
NoteThe principles of navigation are explained in the two following sections by reference of examples:
Setting the language (section 4.2.3)
Setting the date and time (see section 4.2.4).
Calibration recordCalibration data storageCalibration interval: 150 d
Comparison meas.
O2
22.09.200808:00Back
Calibration
Temperature unit °CSal correction OnSalinity 21.7
Stability control OnReset
i Air pressure = 941 mbar
O2
22.09.200808:00Back
18 ba75804e06 03/2012
Oxi 3310 Operation
4.2.3 Example 1 on navigation: Setting the language
1 Press the <On/Off> key.The measured value display appears.The instrument is in the measuring mode.
2 Using <F1_>/[Menu] open the Storage & config menu.The instrument is in the setting mode.
3 Select the System submenu with <▲><▼>. The current selection is displayed with a frame.
4 Open the System submenu with <ENTER>.
O2
7.92 mg/l
25.1 °C
22.09.200808:00Menu USB output
SystemData storage
Storage & config
22.09.200808:00
Back
19ba75804e06 03/2012
Operation Oxi 3310
5 Select the General submenu with <▲><▼>. The current selection is displayed with a frame.
6 Open the General submenu with <ENTER>.
7 Open the setting mode for the Language with <ENTER>.
8 Select the required language with <▲><▼>.
GeneralInterfaceClock
Service informationReset
System
22.09.200808:00
Back
Language: EnglishBeep: OffIllumination: On
Contrast: 48 %Switchoff time: 30 min
General
22.09.200808:00Back
Language: EnglishBeep: OffIllumination: On
Contrast: 48 %Switchoff time: 30 min
General
22.09.200808:00Back
20 ba75804e06 03/2012
Oxi 3310 Operation
4.2.4 Example 2 on navigation: Setting the date and time
The measuring instrument has a clock with a date function. The date and time are indicated in the status line of the measured value display. When storing measured values and calibrating, the current date and time are automatically stored as well.
The correct setting of the date and time and date format is important for the following functions and displays:
Current date and time
Calibration date
Identification of stored measured values.
Therefore, check the time at regular intervals.
NoteAfter a fall of the supply voltage (empty batteries), the date and time are reset to 01.01.2008, 00:00 hours.
Setting the date, timeand date format
The date format can be switched from the display of day, month, year (dd.mm.yyyy) to the display of month, day, year (mm/dd/yyyy or mm.dd.yyyy).
9 Confirm the setting with <ENTER>. The meter switches to the measuring mode.The selected language is active.
1 In the measured value display:Using <F1_>/[Menu], open the Storage & config menu.The instrument is in the setting mode.
2 Select and confirm the System / Clock menu with <▲><▼> and <ENTER>.The setting menu for the date and time opens up.
3 Select and confirm the Time menu with <▲><▼> and <ENTER>.The hours are highlighted.
21ba75804e06 03/2012
Operation Oxi 3310
4 Change and confirm the setting with <▲><▼> and <ENTER>.The minutes are highlighted.
5 Change and confirm the setting with <▲><▼> and <ENTER>.The seconds are highlighted.
6 Change and confirm the setting with <▲><▼> and <ENTER>.The time is set.
7 If necessary, set the Date and Date format. The setting is made similarly to that of the time.
8 To make further settings, switch to the next higher menu level with [Back]<F1>.orSwitch to the measured value display with <M>. The instrument is in the measuring mode.
Date format: dd.mm.yyyy
Date: 30.10.2008Time: 14:53:40
Clock
22.09.200808:00Back
22 ba75804e06 03/2012
Oxi 3310 Operation
4.3 Sensor-independent settings
The Storage & config menu comprises the following settings:
System (see section 4.3.1).
Data storage (see section 4.3.2)
4.3.1 System
Overview The following sensor-independent meter characteristics can be adjusted in the Storage & config/System menu:
Menu language
Beep on keystroke
Illumination
Display contrast
Interval of the automatic switch-off function
Data interface
Clock and date function
Reset of all sensor-independent system settings to the default condition
Settings To open the Storage & config menu, press the <F1_>[Menu] key in the measured value display. After completing the settings, switch to the measured value display with <M>.
Menu item Setting Description
System / General / Language
DeutschEnglish(further)
Select the menu language
System / General / Beep
OnOff
Switch on/off the beep on keystroke
System / General / Illumination
AutoOnOff
Switching the display illumination on/off
System / General / Contrast
0 ... 100 % Changing the display contrast
System / General / Switchoff time
10 min ... 24 h Adjust the switch-off time
System / Interface / Baud rate
1200, 2400, 4800, 9600, 19200
Baud rate of the data interface
23ba75804e06 03/2012
Operation Oxi 3310
4.3.2 Data storage
This menu contains all functions to display, edit and erase stored measured values and calibration records.
NoteDetailed information on the storage functions of the Oxi 3310 is given in section 4.5.
4.3.3 Automatic Stability control
The function, automatic Stability control (AutoRead) continually checks the stability of the measurement signal. The stability has a considerable impact on the reproducibility of measured values.
You can activate or switch off the automatic Stability control function (see section 4.4.3).
The measured parameter flashes on the display
as soon as the measured value is outside the stability range
if you switch over between the measured parameters with <M>.
if the automatic Stability control is switched off.
System / Interface / Output format
ASCIICSV
Output format for data transmissionFor details, see section 4.6
System / Interface / Decimal separator
Dot (xx.x)Comma (xx,x)
Decimal separator
System / Interface / Output header
Output of a header for Output format: CSV
System / Clock TimeDateDate format
Settings of time and date. For details, see section 4.2.4
System / Service information
Hardware version and software version of the meter are displayed.
System / Reset - Resets the system settings to the default values. For details, see section 4.7.2
Menu item Setting Description
24 ba75804e06 03/2012
Oxi 3310 Operation
4.4 Dissolved oxygen
4.4.1 General information
You can measure the following parameters:
D.O. concentration
D.O. saturation index ("D.O. saturation")
D.O. partial pressure
D.O. measurements with the Oxi 3310 can be carried out with the CellOx 325 or DurOx 325 D.O. sensor. The measuring instrument automatically recognizes the type of the connected D.O. sensor.
WarningWhen connecting an earthed PC, measurements cannot be performed in earthed media as incorrect values would result.The USB interface is not galvanically isolated.
Temperaturemeasurement
The CellOx 325 and DurOx 325 D.O. sensors have an integrated temperature sensor.
Preparatory activities Perform the following preparatory activities when you want to measure:
NoteIncorrect calibration of D.O. sensors will result in incorrect measured values. Calibrate at regular intervals.
1 Connect the D.O. sensor to the meter. The D.O. measuring window is displayed.
2 Calibrate or check the meter with the sensor.
25ba75804e06 03/2012
Operation Oxi 3310
4.4.2 Measuring
You can carry out D.O. measurements as follows:
Selecting thedisplayed
measured parameter
You can switch between the following displays with <M>:
D.O. concentration [mg/l]
D.O. saturation [%]
D.O. partial pressure [mbar].
Salinity correction When measuring the concentration of solutions with a salt content of more than 1 g/l, a salinity correction is required. For this, you have to measure and input the salinity of the measured medium first. When the salinity correction is switched on, the [Sal] indicator is displayed in the measuring window.
NoteYou can switch the salinity correction on or off and enter the salinity in the menu for calibration and measurement settings (see section 4.4.3).
Freezing the measuredvalue (HOLD function)
With the HOLD function, you can freeze the current measured value. The displayed measured value stops changing until you switch the HOLD function off.
NoteIf the HOLD function is active, you can, e.g. start a manual measurement with stability control.
1 Perform the preparatory activities according to section 4.4.1.
2 Immerse the D.O. sensor in the test sample.
O2
7.92 mg/l
25.1 °C[Sal]
22.09.200808:00Menu USB output
1 Freeze the measured value with <AR>.The [HOLD] status indicator is displayed.
26 ba75804e06 03/2012
Oxi 3310 Operation
Stability control(AutoRead )
The stability control function (AutoRead) continually checks the stability of the measurement signal. The stability has a considerable impact on the reproducibility of measured values.
The measured parameter flashes on the display
as soon as the measured value is outside the stability range
when the automatic Stability control is switched off.
You can start the Stability control function manually at any time, irrespective of the setting for automatic Stability control (see page 24) in the Measurement menu.
NoteYou can prematurely terminate the Stability control function manually with <ENTER> at any time. If the Stability control function is prematurely terminated, the current measurement data are output to the interface without the AutoRead info.
Criteria for a stablemeasured value
The Stability control function checks whether the measured values are stable within the monitored time interval.
2 Release the frozen measured value again with <AR>.The HOLD function is switched off.The [HOLD] status display disappears.
1 Freeze the measured value with <AR>.The [HOLD] status indicator is displayed.The HOLD function is active.
2 Using <ENTER>, activate the Stability control function manually.The [AR] status indicator appears while the measured value is assessed as not stable. A progress bar is displayed and the display of the measured parameter flashes.The [HOLD][AR] status indicator appears as soon as a stable measured value is recognized. The current measurement data is output to the interface. Measurement data meeting the stability control criterion is marked by AR.
3 Using <ENTER>, start a further measurement with Stability control.orRelease the frozen measured value again with <AR>.The display switches to the measured value display. The [AR][HOLD] status display disappears.
27ba75804e06 03/2012
Operation Oxi 3310
The minimum duration until a measured value is assessed as stable is the monitored time interval. The actual duration is mostly longer.
4.4.3 Settings for D.O. sensors (menu or measurement and calibration settings)
Overview The following settings are possible for D.O. sensors:
Salinity correction
Salinity (salinity equivalent)
Calibration interval
Comparison meas.
Automatic Stability control
Settings The settings are available in the menu for measurement and calibration settings. To open the settings, display the required parameter in the measured value display and press the <F1>/[menu] or <ENTER> key. After completing the settings, switch to the measured value display with <M>.
Measured parameter Time interval Stability in the time interval
D.O. concentration 20 seconds Δ : better than 0.05 mg/l
D.O. saturation 20 seconds Δ : better than 0.6 %
D.O. partial pressure 20 seconds Δ : Better than 1.2 mbar
Temperature 15 seconds Δ T (° C): Better than 0.02
Menu item Possible setting
Description
Calibration /Calibration record
- Displays the calibration record of the last calibration.
Calibration /Calibration interval
1 ... 999 d Calibration interval for the D.O. sensor (in days).The meter reminds you to calibrate regularly by the flashing sensor symbol in the measuring window.
28 ba75804e06 03/2012
Oxi 3310 Operation
Calibration /Comparison meas.
On Off
Enables to adjust the measured value with the aid of a comparison measurement, e.g. Winkler titration.For details, see section 4.4.4.
Sal correction On Off
Manual salt content correction for concentration measurements.
Salinity 0.0 ... 70.0 Salinity or salinity equivalent for the salt content correction.This function is only available for concentration measurements if the manual salt content correction is switched on.
Stability control OnOff
Switches on or off the automatic stability control during measurement (see section 4.3.3)
Temperature unit °C °F
Temperature unit, degrees Celsius or degrees Fahrenheit.All temperatures are displayed with the selected unit.
Reset - Resets all sensor settings to the delivery condition (see section 4.7.1).
Menu item Possible setting
Description
29ba75804e06 03/2012
Operation Oxi 3310
4.4.4 D.O. calibration
Why calibrate? D.O. sensors age. This changes the slope of the D.O. sensor. Calibration determines the current slope of the sensor and stores this value in the instrument.
When to calibrate? After connecting another D.O. sensor
When the sensor symbol flashes (after the calibration interval has expired).
Calibration datasets The Oxi 3310 administrates two sets of calibration data:
Sensors of different types can be calibrated separately from each other. When one sensor type is calibrated, the calibration data of the other type remains stored. The Oxi 3310 recognizes the type of the connected sensor and automatically uses the correct calibration data.
Calibration procedures With the Oxi 3310, 2 calibration procedures are available:
Calibration in water vapor-saturated air. Use an OxiCal® air calibration vessel for calibration.
Calibration via a comparison measurement (e.g. Winkler titration according to DIN EN 25813 or ISO 5813). At the same time, the relative slope is adapted to the comparison measurement by a correction factor. When the correction factor is active, the [Factor] indicator appears in the measuring window.
Stability control(AutoRead)
In calibration, the Stability control function (AutoRead) is automatically activated.
Display calibration dataand output to interface
You can have the data of the last calibration displayed (see section 4.4.5). Subsequently, you can transmit the displayed calibration data to the interface, e. g. to a PC, with the <F2>[USB output] key.
NoteThe calibration record is automatically transmitted to the interface after calibrating.
Set 1 for the type, "CellOx": – CellOx 325
Set 2 for the type, "DurOx": – DurOx 325
30 ba75804e06 03/2012
Oxi 3310 Operation
Sample record:
Calibration evaluation After the calibration, the measuring instrument automatically evaluates the current status of the calibration. The evaluation appears on the display and in the calibration record.
Calibration inwater vapor
saturated air(air calibration vessel)
For this calibration procedure, the Comparison meas. setting must be set to Off in the Calibration menu.
Proceed as follows to calibrate the instrument:
NoteThe sponge in the air calibration vessel must be moist (not wet). Leave the sensor in the air calibration vessel for a time long enough to adjust.
31.10.2008 16:55:12Oxi 3310Ser. no. 08502113
CALIBRATION O231.10.2008 16:13:33
S = 0.88 25.0 °CSensor +++
Display Calibration record Relative slope
+++ S = 0.8 ... 1.25
++ S = 0.7 ... 0.8
+ S = 0.6 ... 0.7
Error Error S < 0.6 or S > 1.25
Eliminate the error according to chapter 6 WHAT TO DO IF...
1 Connect the D.O. sensor to the meter.
2 Put the D.O. sensor into the air calibration vessel.
31ba75804e06 03/2012
Operation Oxi 3310
Calibration by means ofa Comparison meas.
For this calibration procedure, the Comparison meas. setting must be set to On in the Calibration menu.
NoteBefore calibrating via a comparison measurement, the sensor should be calibrated in the air calibration vessel.
Proceed as follows to calibrate the instrument:
3 Start the calibration with <CAL>.The last calibration data (relative slope) is displayed.
4 Start the measurement with <ENTER>.The measured value is checked for stability (stability control).The [AR] status indicator is displayed. The measured parameter flashes.
5 Wait for the end of the AutoRead measurement or accept the calibration value with <ENTER>.The calibration record is displayed and output to the interface.
6 Using <F1>/[Continue] or <ENTER>, switch to the measured value display.
O2
0.8825.1 °C
OxiCal
22.09.200808:00
1 Connect the D.O. sensor to the measuring instrument.
2 Immerse the D.O. sensor in the reference solution.
3 Start the calibration with <CAL>.
32 ba75804e06 03/2012
Oxi 3310 Operation
4.4.5 Displaying calibration records
The calibration data can be displayed and then output to the interface.
Displaying thecalibration record
The calibration record of the last calibration is to be found under the menu item, Calibration / Calibration record. To open it in the measured value display, press the <CAL_> key.
4 Start the measurement with <ENTER>.The measured value is checked for stability (stability control).The [AR] status indicator is displayed. The measured parameter flashes.
5 Wait for the end of the AutoRead measurement or accept the calibration value with <ENTER>.The factor that was set last is displayed.
6 Using <▲> <▼>, set the correction factor to adjust the displayed concentration value to the nominal value (value of the comparison measurement). Subsequently, accept the correction factor with <ENTER>.The measuring instrument switches to the measured value display. The status display [Factor] is active.
O2
7.92 mg/l
25.1 °C
OxiComp
22.09.200808:00
O2
7.92 mg/l
0.973OxiComp
22.09.200808:00
33ba75804e06 03/2012
Operation Oxi 3310
The calibration records of the last calibrations (up to 10) are available in the menu, <F1>/[Menu] / Calibration / Calibration data storage and in the menu, <F1_>/[Menu] / Storage & config/Data storage / Calibra-tion data storage.
Sample
Menu item Setting/function
Description
Calibration / Calibration data storage /Display
or
Data storage / Calibration data storage /Display
- Displays the calibration record.
Further options: Scroll through the
calibration records with <▲><▼>.
Output the displayed calibration record to the interface with <F2>/[USB output].
Quit the display with <F1>/[Back] or <ENTER>.
Switch directly to the measured value display with <M>.
Calibration / Calibration data storage / Output to USBor
Data storage / Calibration data storage /Output to USB
- Outputs the calibration records to the interface.
31.10.2008 16:55:12Oxi 3310Ser. no. 08502113
CALIBRATION O231.10.2008 16:13:33
S = 0.88 25.0 °CSensor +++
etc...
34 ba75804e06 03/2012
Oxi 3310 Operation
4.5 Data storage
You can transmit measured values (datasets) to the data storage:
Manual storage (see section 4.5.1)
Automatic storage at intervals (see section 4.5.2)
Each storage process transmits the current dataset to the interface at the same time.
Measurement dataset A complete dataset consists of:
ID number
Date/time
Measured value of the connected sensor
Measured temperature value of the connected sensor
AutoRead info: AR appears with the measured value if the AutoRead criterion was met while storing (stable measured value). Otherwise, the AR display is missing.
Calibration evaluation: +++, ++, +, -, or no evaluation
Storage locations The Oxi 3310 meter has two measurement data memories. The measured values recorded either manually or automatic are stored separately in individual measurement data memories.
Storage Maximum number of datasets
Manual data storage 200
Automatic data storage 5000
35ba75804e06 03/2012
Operation Oxi 3310
4.5.1 Manual storage
You can transmit a measurement dataset to the data storage as follows.
The dataset is at the same time output to the interface:
If the storage is full The following window appears if all 200 storage locations are occupied:
You have the following options:
To erase the entire storage, confirm Yes.
To cancel the storage process and switch to the measured value display, confirm No. Then you can e.g. transmit the stored data to a PC (see section 4.5.3) and subsequently erase the storage (see section 4.5.4).
1 Press the <STO> key shortly.The menu for manual storage appears.
2 If necessary, change and confirm the ID number (1 ... 10000) with <▲><▼> and <ENTER>.The dataset is stored. The meter switches to the measured value display.
30.10.2008 11:24:16pH 7.000 24.8 °C AR +++
ID number: 1Continue
Manual data storage 4 From 200
22.11.201008:00Back
Data storage full. Erase?
YesNo
Warning
22.09.200808:00Back
36 ba75804e06 03/2012
Oxi 3310 Operation
4.5.2 Automatic storage at intervals
The storage interval (Interval) determines the time interval between automatic storage processes. Each storage process transmits the current dataset to the interface at the same time.
Configuring theautomatic storage
function
Settings You can configure the automatic storage function with the following settings:
1 Press the <STO_> key.The menu for automatic storage appears.
Adjusted entire storage duration
Max. available storage duration
Graphical display of the storage usage
ID number 1Interval 30 sDuration 180 min
Continue 0d03h00min
0 1d17h33min
Automatic data storage
22.09.200808:00
Back
Menu item Possible setting
Description
ID number 1 ... 10000 ID number for the dataset series.
Interval 1 s, 5 s, 10 s, 30 s, 1 min, 5 min, 10 min, 15 min, 30 min, 60 min
Storage interval.
The lower limit of the storage interval can be restricted by the number of free storage locations.The upper limit is restricted by the storage duration.
37ba75804e06 03/2012
Operation Oxi 3310
Starting the automaticstorage function
To start the automatic storage function, select Continue with <▲><▼> and confirm with <ENTER>. The measuring instrument switches to the measured value display.
The active automatic storage function can be recognized by the progress bar in the status line. The progress bar indicates the remaining storage duration.
NoteIf the automatic storage function is activated, only the following keys are active: Softkeys, <M>, <STO_> and <On/Off>. The other keys and the automatic switch-off function are deactivated.
Energy saving mode([Eco mode])
If the automatic storing function is active, the meter provides an energy saving mode ([Eco mode]) to avoid unnecessary energy consumption. The energy saving mode switches off functions of the meter that are not required for the automatic storage of measurement data (such as the display). By pressing any key the energy saving mode is switched off again.
Duration 1 min ... x min Storage duration.Specifies after which time the automatic storage should be terminated.
The lower limit of the storage duration is restricted by the storage interval.The upper limit is restricted by the number of free storage locations.
Menu item Possible setting
Description
O2
7.92 mg/l
25.0 °C
0d03h00min[Sal]
22.09.200808:00Eco mode USB output
Remaining storage duration
Graphical display of the storage duration
38 ba75804e06 03/2012
Oxi 3310 Operation
Terminating theautomatic storage
function prematurely
Proceed as follows to switch off the automatic storage function before the adjusted storage duration has expired:
4.5.3 Displaying and editing the measurement data storage
The contents of the manual or automatic measurement data storage can be shown on the display.
Each of the measurement data memories has a function to erase the entire contents.
The contents of the manual or automatic measurement data storage can be shown on the display and output to the interface.
Editing the data storage The storage is edited in the menu, Storage & config/ Data storage. To open the Storage & config menu, press the <F1_>[Menu] key in the measured value display. Open the manual or automatic storage directly with the <RCL> or <RCL_> key.
NoteThe settings are explained here using the manual storage as an example. The same settings and functions are available for the automatic storage.
1 Press the <STO_> key.The following window appears.
2 Using <▲><▼>, select Yes and confirm with <ENTER>.The measuring instrument switches to the measured value display.The automatic storage function is terminated.
Stop automatic storage?
YesNo
Warning
22.09.200808:00Back
39ba75804e06 03/2012
Operation Oxi 3310
Settings
Display presentationof a dataset
Menu item Setting/function
Description
Data storage / Manual data storage / Display
- Displays all measurement datasets page by page.
Further options: Scroll through the
datasets with <▲><▼>.
Output the displayed dataset to the interface with <F2>/[USB output].
Quit the display with <F1>/[Back].
Data storage / Manual data storage / Erase
- Erases the entire manual measurement data storage.
Note:All calibration data remains stored when this action is performed.
Data storage / Manual data storage / Output to USB
- Outputs all stored measurement data to the interface.
Manual data storage 3 of 64
30.10.2008 11:24:16 ID number: 1
O2 = 7.92 mg/l 25.1 °C AR +++Sal = 6.5
22.09.200808:00Back USB output
40 ba75804e06 03/2012
Oxi 3310 Operation
Sample
Quitting the display To quit the display of stored measurement datasets, you have the following options:
Switch directly to the measured value display with <M>.
Quit the display and move to the next higher menu level with <F1>/[Back].
4.5.4 Erasing the measurement data storage
How to erase the measurement data storage is described in section 4.5.3 DISPLAYING AND EDITING THE MEASUREMENT DATA STORAGE.
31.10.2008 09:27:20Oxi 3310 Ser. No. 12345678
ID number 1O2 = 7.92 mg/l 25.0 °C AR +++Sal = 6.5________________________________________
31.10.2008 09:56:24Oxi 3310 Ser. No. 12345678
ID number 1O2 8.14 mg/l 25.0 °C AR +++
41ba75804e06 03/2012
Operation Oxi 3310
4.6 Transmitting data (USB interface)
4.6.1 Options for data transmission
Via the USB interface you can transmit data to a PC. The following table shows which data are transmitted to the interface in which way:
NoteThe following rule applies: With the exception of the menus, shortly pressing the <F2>/[USB output] key generally outputs the display contents to the interface (displayed measured values, measurement datasets, calibration records).
Data Control Operation / descriptionCurrent measured valuesof all connected D.O. sensors
Manual With <F2>/[USB output].
Simultaneously with every manual storage process (see section 4.5.1).
Automatic, at intervals
With <F2_>/[USB output]. Then you can set the transmission interval.
Note:For operation with MultiLab pilot: set the Send ID option to No (see section 4.6.3).
Simultaneously with every automatic storage process (see section 4.5.2).
Stored measured values
Manual Displayed dataset with <F2_>[USB output] after calling up from the storage.
All datasets with the Output to USB function.
For details, see section 4.5.3.Calibration records
Manual Calibration record with <F2_>[USB output].
For details, see section 4.6.Automatic At the end of a calibration
procedure.
42 ba75804e06 03/2012
Oxi 3310 Operation
4.6.2 Connecting a PC
Connect the Oxi 3310 to the PC via the USB interface.
WarningThe USB interface is not galvanically isolated. When a grounded PC is connected, measurements cannot be performed in grounded media as incorrect values would result.
Installation of the USBdriver on the PC
System requirements of the PC for installation of the USB driver:
PC with Pentium processor or higher with at least one free USB connection and CD-ROM drive
Windows 2000, XP, Vista.
4.6.3 Operation with MultiLab pilot
With the aid of the MultiLab pilot software, you can record and evaluate measuring data with a PC. The data transmission takes place after the meter is connected to the USB interface of the PC.
NoteMore detailed information can be found in the MultiLab pilot operating manual.
NoteTo display the measurement data of the Oxi 3310 correctly on the PC, the Oxi 3310 has to be set up for communication with the MultiLab pilot. To do so, set the Send ID option to No in the menu for automatic storage (<F2_>/[USB output]).
1 Insert the supplied installation CD in the CD drive of your PC.
2 Install the driver from the CD.Follow the Windows installation instructions as necessary.
3 Connect the Oxi 3310 to the PC via the USB interface.The meter is listed as a virtual COM interface among the connections in the Windows instrument manager.
43ba75804e06 03/2012
Operation Oxi 3310
4.7 Reset
You can reset (initialize) all sensor settings and sensor-independent settings separately from each other.
4.7.1 Resetting the measurement settings
NoteThe calibration data are reset to the default settings together with the measuring parameters. Recalibrate after performing a reset.
The following settings for D.O. measurements are reset to the default settings with the Reset function:
The sensor settings are reset under theReset menu item in the menu for calibration and measurement settings. To open the settings, display the required parameter in the measured value display and press the <F1>/[menu] or <ENTER> key.
Setting Default settings
Cal. interval 14 d
Comparison meas. Off
Measured parameter D.O. concentration
Relative slope (SRel) 1.00
Salinity (value) 0.0
Salinity (function) Off
Stability control On
Temperature unit °C
44 ba75804e06 03/2012
Oxi 3310 Operation
4.7.2 Resetting the system settings
The following system settings can be reset to the delivery status:
The system settings are reset in the menu, Storage & config / System / Reset. To open the Storage & config menu, press the <F1_>[Menu] key in the measured value display.
Setting Default settings
Language English
Beep On
Baud rate 4800 Baud
Output format ASCII
Contrast 50 %
Illumination Auto
Switchoff time 1 h
45ba75804e06 03/2012
Operation Oxi 3310
46 ba75804e06 03/2012
Oxi 3310 Maintenance, cleaning, disposal
5 Maintenance, cleaning, disposal
5.1 Maintenance
The only maintenance activity required is replacing the batteries.
NoteSee the relevant operating manuals of the D.O. sensors for instructions on maintenance.
5.1.1 Replacing the batteries
NoteAlternatively, you can also use Ni-MH rechargeable batteries (type Mignon AA). In order to charge the batteries, an external charging device is required.
CautionMake sure that the poles of the batteries are positioned correctly.The ± signs on the batteries must correspond to the ± signs in the battery compartment.
1 Unscrew the two screws (1) on the underside of the meter.
2 Open the battery compartment (2) on the underside of the meter.
3 Remove the four batteries from the battery compartment.
4 Place four new batteries (type Mignon AA) in the battery compartment.
2
1
47ba75804e06 03/2012
Maintenance, cleaning, disposal Oxi 3310
5.2 Cleaning
Occasionally wipe the outside of the measuring instrument with a damp, lint-free cloth. Disinfect the housing with isopropanol as required.
CautionThe housing is made of synthetic material (ABS). Thus, avoid contact with acetone or similar detergents that contain solvents. Remove any splashes immediately.
5.3 Packing
This meter is sent out in a protective transport packing. We recommend: Keep the packing material. The original packing protects the meter against damage during transport.
5.4 Disposal
NoteThis meter contains batteries. Batteries that have been removed must only be disposed of at a recycling facility set up for this purpose or via the retail outlet. It is illegal to dispose of it in household refuse.
5 Close the battery compartment (2) and tighten the screws (1).
48 ba75804e06 03/2012
Oxi 3310 What to do if...
6 What to do if...
Error message,OFL
Error message,Error
Sensor symbol flashes
Display
Meter does not react tokeystroke
You want to know whichsoftware
version is in the meter
Cause Remedy
– Measured value outside the measuring range
– Use a suitable D.O. sensor
Cause Remedy
– D.O. sensor contaminated – Clean D.O. sensor and replace it if necessary
Cause Remedy
– Cleaning interval expired – Recalibrate the measuring system
Cause Remedy
– Batteries almost empty – Replace the batteries (see section 5.1 MAINTENANCE)
Cause Remedy
– Operating condition undefined or EMC load unallowed
– Processor reset:Press the <ENTER> and <On/Off> key simultaneously
Cause Remedy
– E. g., a question by the service department
– Switch on the meter.Open the menu, <F1_>/[Menü] / Speicher & Konfig. / System / Service Information. The instrument data is displayed.
49ba75804e06 03/2012
What to do if... Oxi 3310
50 ba75804e06 03/2012
Oxi 3310 Technical data
7 Technical data
7.1 General data
Dimensions approx. 180 x 80 x 55 mm
Weight approx. 0.4 kg
Mechanical structure Type of protection IP 67
Electrical safety Protective class III
Test certificates CE, cETLus
Ambient conditions
Storage - 25 °C ... + 65 °C
Operation -10 °C ... + 55 °C
Allowable relative humidity
Annual mean: < 75 %30 days/year: 95 %Other days: 85 %
Powersupply
Batteries 4 x 1.5 V alkali-manganese batteries, type AA
Rechargeable batteries
4 x 1,2 V NiMH rechargeable batteries, type AA (no charging function)
Operational life up to 800 h without / 100 h with illumination
USB interface Type USB 1.1USB B (device), data output
Baud rate adjustable: 1200, 2400, 4800, 9600, 19200 Baud
Data bits 8
Stop bits 2
Parity None
Handshake RTS/CTS
Cable length max. 3 m
Guidelinesand norms used
EMC EC directive 2004/108/ECEN 61326-1EN 61000-3-2EN 61000-3-3FCC Class A
Meter safety EC directive 2006/95/ECEN 61010-1
IP protection class EN 60529
51ba75804e06 03/2012
Technical data Oxi 3310
7.2 Measuring ranges, resolution, accuracy
NoteThe accuracy values specified here apply exclusively to the meter. The accuracy of the D.O. sensors has also to be taken into account.
Measuring ranges,resolution
Note: The values quoted in brackets apply especially for the DurOx 325 sensor.
Parameter Measuring range Resolution
D.O. concentration [mg/l]
0 ... 20.00 (0 ... 20.0)0 ... 90.0 (0 ... 90)
0.01 (0.1)0.1 (1)
Saturation [%] 0 ... 200.0 (0 ... 200)0 ... 600
0.1 (1)1
D.O. partial pressure [mbar]
0 ... 200.0 (0 ... 200)0 ... 1250
0.1 (1)1
T [°C] 0 ... 50.0 0.1
Accuracy (± 1 digit) Parameter Accuracy
D.O. concentration [mg/l]
± 0.5 % of measured valueat ambient temperature + 5 °C ... + 30 °C
Saturation [%] ± 0.5% of measured value when measuring in the range of ± 10 K around the calibration temperature
D.O. partial pressure [mbar]
± 0.5 % of measured valueat ambient temperature + 5 °C ... + 30 °C
T [°C] / temperature sensor
NTC 30 ± 0,1
PT 1000 ± 0,1
Correction functions Temperaturecompensation
Accuracy better than 2 % at 0 ... + 40 °C
Salinity correction 0 ... 70.0 SAL
Air pressure correction
Automatic through integrated pressure sensor in the range of 500 ... 1100 mbar
52 ba75804e06 03/2012
Oxi 3310 Technical data
FCC Class A Equipment Statement
Note: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.Changes or modifications not expressly approved by the manufacturer could void the user‘s authority to operate the equipment.
53ba75804e06 03/2012
Technical data Oxi 3310
54 ba75804e06 03/2012
Oxi 3310 Lists
8 Lists
This chapter provides additional information and orientation aids.
Specialist terms The glossary briefly explains the meaning of the specialist terms. However, terms that should already be familiar to the target group are not described here.
Index The index will help you to find the topics that you are looking for.
Glossary
Adjusting To manipulate a measuring system so that the relevant value (e. g. the displayed value) differs as little as possible from the correct value or a value that is regarded as correct, or that the difference remains within the tolerance.
AutoRange Name of the automatic selection of the measuring range.
Calibration Comparing the value from a measuring system (e. g. the displayed value) to the correct value or a value that is regarded as correct. Often, this expression is also used when the measuring system is adjusted at the same time (see adjusting).
D.O. partial pressure Pressure caused by the oxygen in a gas mixture or liquid.
D.O. saturation Short name for the relative D.O. saturation.
Measured parameter The measured parameter is the physical dimension determined by measuring, e. g. pH, conductivity or D.O. concentration.
Measured value The measured value is the special value of a measured parameter to be determined. It is given as a combination of the numerical value and unit (e. g. 3 m; 0.5 s; 5.2 A; 373.15 K).
OxiCal® WTW name for a procedure to calibrate D.O. measuring systems in water vapor saturated air.
Reset Restoring the original condition of all settings of a measuring system.
Resolution Smallest difference between two measured values that can be displayed by a meter.
55ba75804e06 03/2012
Lists Oxi 3310
Salinity The absolute salinity SA of seawater corresponds to the relationship of the mass of dissolved salts to the mass of the solution (in g/Kg). In practice, this dimension cannot be measured directly. Therefore, the practical salinity according to IOT is used for oceanographic monitoring. It is determined by measuring the electrical conductivity.
Salt content General designation for the quantity of salt dissolved in water.
Slope (relative) Designation used by WTW in the D.O. measuring technique. It expresses the relation of the slope value to the value of a theoretical reference sensor of the same construction type.
Stability control Function to control the measured value stability.
Test sample Designation of the test sample ready to be measured. Normally, a test sample is made by processing the original sample. The test sample and original sample are identical if the test sample was not processed.
56 ba75804e06 03/2012
Oxi 3310 Lists
Index
AAir calibration beaker . . . . . . . . . . . . . . . . . 30Authorized use . . . . . . . . . . . . . . . . . . . . . . 10Automatic switch-off . . . . . . . . . . . . . . . . . . 15AutoRead . . . . . . . . . . . . . . . . . . . . . . . . . . 27
BBattery compartment . . . . . . . . . . . . . . 13, 47
CCalibration evaluation . . . . . . . . . . . . . . . . 31Calibration records . . . . . . . . . . . . . . . . . . . 33Comparison measurement (D.O.) . . . . . . . 30Connecting a PC . . . . . . . . . . . . . . . . . . . . 43
DDataset . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35Date and time . . . . . . . . . . . . . . . . . . . . . . . 21Default settings
Measured parameter . . . . . . . . . . . . . . 44System settings . . . . . . . . . . . . . . . . . . 45
EEnergy saving feature . . . . . . . . . . . . . . . . 14Energy saving mode . . . . . . . . . . . . . . . . . 38
FFirmware update . . . . . . . . . . . . . . . . . . . . 59
IInitial commissioning . . . . . . . . . . . . . . 13, 14Initialize . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
KKeys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
MMeasured value display . . . . . . . . . . . . . . . 17Measurement data storage
Edit . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39Erase . . . . . . . . . . . . . . . . . . . . . . . . . . 39Storage locations . . . . . . . . . . . . . . . . . 35
Measurement dataset . . . . . . . . . . . . . . . . 35
Measuring . . . . . . . . . . . . . . . . . . . . . . . . . 26Menu for calibration and measurement
settings . . . . . . . . . . . . . . . . . . . . . . 28Menus (navigation) . . . . . . . . . . . . . . . . . . 17Messages . . . . . . . . . . . . . . . . . . . . . . . . . 18
OOperational safety . . . . . . . . . . . . . . . . . . . 10
PPrecautions . . . . . . . . . . . . . . . . . . . . . . . . . 9Print . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
RReset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
SSafety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Sample display . . . . . . . . . . . . . . . . . . . . . . 7Saving . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
At intervals . . . . . . . . . . . . . . . . . . . . . . 37Automatic . . . . . . . . . . . . . . . . . . . . . . . 37Manual . . . . . . . . . . . . . . . . . . . . . . . . . 36
Scope of delivery . . . . . . . . . . . . . . . . . . . . 13Setting the date . . . . . . . . . . . . . . . . . . . . . 14Setting the time . . . . . . . . . . . . . . . . . . . . . 14Slope relative . . . . . . . . . . . . . . . . . . . . . . . 30Socket field . . . . . . . . . . . . . . . . . . . . . . . . . 8Stability control
Automatic . . . . . . . . . . . . . . . . . . . . . . . 24Storage interval . . . . . . . . . . . . . . . . . . . . . 37
TTemperature measurement . . . . . . . . . . . . 25Transmitting data . . . . . . . . . . . . . . . . . . . . 42Transmitting measured values . . . . . . . . . . 42
57ba75804e06 03/2012
Lists Oxi 3310
58 ba75804e06 03/2012
Oxi 3310 Appendix: Firmware update
Appendix: Firmware update
General information Available firmware updates are provided on the Internet.With the "Firmware Update " program and a PC you can update the firmware of the Oxi 3310 to the newest version.
For the update you have to connect the meter to a PC.
For the update via the USB interface, the following is required:
a free USB interface (virtual COM port) on the PC
the driver for the USB interface (on the enclosed CD-ROM)
the USB cable (included in the scope of delivery of the Oxi 3310).
Program installation
Program start
Firmware update
After switching the meter off and on you can check whether the meter has taken over the new software version (see page 49).
1 Install the downloaded firmware update on a PC.
An update folder is created in the Windows start menu. If an update folder already exists for the meter (or meter type), the new data is displayed there.
2 In the windows start menu, open the update folder and start the firmware update program.
3 Using the USB interface cable, connect the Oxi 3310 to a USB interface (virtual COM port) of the PC.
4 Switch on the Oxi 3310.
5 In the firmware update program, start the update process with OK.
6 Follow the instructions of the firmware update program.During the programming process, a corresponding message and a progress bar (in %) are displayed. The programming process takes approx. three minutes. A terminatory message is displayed after a successful programming process. The firmware update is completed.
7 Disconnect the Oxi 3310 from the PC. The Oxi 3310 is ready for operation again.
59ba75804e06 03/2012
Appendix: Firmware update Oxi 3310
60 ba75804e06 03/2012