[email protected] Work Instruction File: PAI-0584-U Revision 1.00 Page: 1 of 8 Number of local copies: Version: 1.00 Written: 2015-11-05 11mhaag Date of print: Uncontrolled copy if printed Approved: 2015-12-07 11ttoth Contact: Released: Replace HiBan Mounted (HBM) This Work Instruction applies to the platforms and processes listed below: Health and safety Extranet: extranet.sonova.com/unitron/ Table of contents 1 Quality Control In ........................................ 3 2 Open the Device ......................................... 3 3 Remove HiBan Antenna .............................. 4 4 Mount HiBan Antenna................................. 5 5 Quality Control Out ..................................... 7 6 Revision History .......................................... 8 WH30 Wiring Drilling & Chemicals: For information about safety, handling and storage, refer to your local MSDS! Green Lacquer Silicone 734 Isopropanol Attention Grinding and beveling work: ESD Protection: Soldering Work: We recommend an adequate extraction system with a protective glass! Electrostatic discharge (ESD) may causes damage to electronic components! Electrostatic discharge (ESD) may cause damage to electronic components! 330° Celsius = 626° Fahrenheit Wear safety glasses, if protective shield is not available Wear dust mask, if extraction system is not available Do not touch any components if you are not grounded Use lead-free solder. Solder temperature 330°C (+/-15°C). Reference time for heating: 1 second. Glasses Mask Isopropanol Extraction Temperature 330°C F 626 Green Lacquer Silicon 734

Work nstruction - · PDF [email protected] Work nstruction File: PAI-0584-U Revision 1.00 Page: 1 of 8 Number of local copies: Version: 1.00 Written: 2015-11-05 11mhaag

Embed Size (px)

Citation preview

Page 1: Work nstruction - · PDF filetechnical.support@unitron.com Work nstruction File: PAI-0584-U Revision 1.00 Page: 1 of 8 Number of local copies: Version: 1.00 Written: 2015-11-05 11mhaag

[email protected]

Work Instruction

File: PAI-0584-U Revision 1.00Page: 1 of 8 Number of local copies:

Ver

sion

: 1.

00 Written: 2015-11-05 11mhaag

Date of print: Uncontrolled copy if printed Approved: 2015-12-07 11ttoth

Contact: Released:

Replace HiBan Mounted (HBM)

This Work Instruction applies to the platforms and processes listed below:

Hea

lth a

nd s

afet

y

Extranet: extranet.sonova.com/unitron/

Table of contents1 Quality Control In ........................................ 3

2 Open the Device ......................................... 3

3 Remove HiBan Antenna .............................. 4

4 Mount HiBan Antenna ................................. 5

5 Quality Control Out ..................................... 7

6 Revision History .......................................... 8

WH30Wiring

Drilling &

Chemicals:

For information about safety, handling and storage, refer to your local MSDS!

Green Lacquer

Silicone 734

Isopropanol

Attention

Grinding and beveling work:

ESD Protection:

Soldering Work:

We recommend an adequate extraction system witha protective glass!

Electrostatic discharge (ESD) may causes damage to electronic components!

Electrostatic discharge (ESD) may cause damage to electronic components!

330° Celsius = 626° Fahrenheit

• Wear safety glasses, if protective shield is not available

• Wear dust mask, if extraction system is not available

Do not touch any components if you are not grounded

• Use lead-free solder.

• Solder temperature 330°C (+/-15°C).

• Reference time for heating: 1 second.

Glasses Mask

Isopropanol

Extraction

Temperature

330°C F 626

Green

Lacquer

Silicon

734

53bhynes
Typewritten Text
2016-10-24 53Info Control
53bhynes
Typewritten Text
53bhynes
Typewritten Text
53bhynes
Typewritten Text
53bhynes
Typewritten Text
53bhynes
Typewritten Text
Page 2: Work nstruction - · PDF filetechnical.support@unitron.com Work nstruction File: PAI-0584-U Revision 1.00 Page: 1 of 8 Number of local copies: Version: 1.00 Written: 2015-11-05 11mhaag

File: PAI-0584-U Version: 1.00

Title: Work Instruction Replace HiBan Mounted (HBM) Page: 2 / 8

Extranet: extranet.sonova.com/unitron/

Back

Attention

Unitron HQ delivers recommended standard equipment, tools and material, however local use of equivalent equipment, tools and material is allowed after approval / release by Unitron HQ.

Attention

Equipment:

Tools:

Materials:

Unitron HQ delivers recommended standard equipment, tools and material, however local use of equivalent equipment, tools and material is allowed after approval / release by Unitron HQ.

Fun-Tak Putty 021-5011

HiBan antenna with silicone tube 047-0108

Isopropanol 020-0007

Lacquer, green, 2ml 021-0393

Sandpaper, 320 grit 021-0527

Silicone 734, transparent 021-0204

Solder wire, Stannol Flowtin 0.23mm 021-1004

Tape medical 021-0646

Tape; single-sided, yellow; 3 x 2.5mm 021-0383

Microscope

Soldering station 026-1215

Flat nose pliers 026-0020

Holder for scalpel 026-0025

Razor blade holder 026-0036

Razor blades 026-0037

Sandpaper holder conical 026-0592

Sandpaper holder cylindrical 026-0593

Scalpel blade pointed 026-0026

Scalpel blade round 026-0027

Tweezers for soldering work 026-0032Cleaning tip (Q-tip) 021-0301Wireless Functional Test Device(WFT) 090-0002

WFT Distance Guage Board 026-0890

Cotton swabs cosmetic 021-0523

Page 3: Work nstruction - · PDF filetechnical.support@unitron.com Work nstruction File: PAI-0584-U Revision 1.00 Page: 1 of 8 Number of local copies: Version: 1.00 Written: 2015-11-05 11mhaag

Extranet: extranet.sonova.com/unitron/

Back

File: PAI-0584-U Version: 1.00

Title: Work Instruction Replace HiBan Mounted (HBM) Page: 3 / 8

• Check that you have the correct instrument and parts.

1 Quality Control In

2 Open the Device

Cover the microphone openings with adhesive tape. Medical

Tape

Tweezers

Cover the receiver opening with adhesive tape.

Insert a battery into the instrument (you can use discharged batteries as well).Sand off the lacquer from the entire shell with 320 grit sandpaper.

Max. rotating speed 10’000 rpm.

Use a razor blade and carefully apply pressure on the concha side of the instrument. The battery will block the razor blade from cutting the wires accidentally.

Take care not to damage the serial number.

Take care not to damage the shell.Take care not to damage the T-coil or other inside components and wires.

Attention

Attention

Medical

Tape

Tweezers

Battery

Battery

Sandpaper

Razor Blade

Page 4: Work nstruction - · PDF filetechnical.support@unitron.com Work nstruction File: PAI-0584-U Revision 1.00 Page: 1 of 8 Number of local copies: Version: 1.00 Written: 2015-11-05 11mhaag

Extranet: extranet.sonova.com/unitron/

Back

File: PAI-0584-U Version: 1.00

Title: Work Instruction Replace HiBan Mounted (HBM) Page: 4 / 8

Push the receiver tubing into the shell by using tweezers.

Tweezers

Separate the electronics from the shell.

Tweezers

3 Remove HiBan Antenna

Fix the hybrid with Fun-Tak putty. Turn the hybrid, so that the yellow tape points to the top.

Remove the yellow tape.

Unsolder the two HiBan litzwires from the hybrid.

Tweezers

Tweezers

Tweezers

Double Side

Tape

Microscope

Solder

330°

Page 5: Work nstruction - · PDF filetechnical.support@unitron.com Work nstruction File: PAI-0584-U Revision 1.00 Page: 1 of 8 Number of local copies: Version: 1.00 Written: 2015-11-05 11mhaag

Extranet: extranet.sonova.com/unitron/

Back

File: PAI-0584-U Version: 1.00

Title: Work Instruction Replace HiBan Mounted (HBM) Page: 5 / 8

Clean the solder pads with a cotton swap and Isopropanol.

PulloutthedefectiveHiBanantennabyusingflatnosepliers.

Cleaning Tip

Isopropanol

Be careful, HiBan coil core break easily, if applying too much force.Attention

Nose Pliers

Flat

Peel off silicone residuals from faceplate and hole by using tweezers and/or scalpel.

Dispense silicone 734 into the HiBan cavity, Approximately ¼ of the cavity shouldbefilled!

Take a new HiBan antenna and insert the HiBan antenna upright into cavity.

4 Mount HiBan Antenna

Tweezers

Scalpel

Silicon

734

Page 6: Work nstruction - · PDF filetechnical.support@unitron.com Work nstruction File: PAI-0584-U Revision 1.00 Page: 1 of 8 Number of local copies: Version: 1.00 Written: 2015-11-05 11mhaag

Extranet: extranet.sonova.com/unitron/

Back

File: PAI-0584-U Version: 1.00

Title: Work Instruction Replace HiBan Mounted (HBM) Page: 6 / 8

Direction of the HiBan wires should always be directed opposite of the battery door hinge pin.

Hinge

Attention

Wrong

Correct

Inspect amount of silicone.There should be a small amount around the HiBan antenna without adhesive flowingontoneighboringcomponents.Add silicone or wipe off excess silicone if necessary.

Ensure that the antenna is mounted in a right angle to the faceplate surface.Ensure that the HiBan is properly seated, about half of the coil winding should be covered by the faceplate Allow 15 minutes for air drying of silicone adhesive before further processing.

Solder the two Hiban coil wires to the pads.

HiBan antenna with a tilt of more than 5° shall be repositioned. 90°Attention

Tweezers

Solder

330°

Microscope

Page 7: Work nstruction - · PDF filetechnical.support@unitron.com Work nstruction File: PAI-0584-U Revision 1.00 Page: 1 of 8 Number of local copies: Version: 1.00 Written: 2015-11-05 11mhaag

Extranet: extranet.sonova.com/unitron/

Back

File: PAI-0584-U Version: 1.00

Title: Work Instruction Replace HiBan Mounted (HBM) Page: 7 / 8

Coat the solder joints with green lacquer. Let it dry for 2 minutes.

Cover the solder connection by a yellow single sided tape (3 x 2.5mm).

Tweezers

Dosing Tool

Green

Lacquer

Single Side

Tape

Set the WFT device to test mode “P3 HI WFT – ON”.Insert a full battery and connect the receiver to a proper stethoclip.Hold the faceplate in correct position to the distance gauge.A pure tone (buzzing sound) should be heard as a clear sound through the stethoclip without any signal interruptions, perform the test for at least 5 seconds.

Test to be done on a wooden table. No metal or electromagnetic source in a radius of 20 cm.

It’s not allowed to start to measure closer than the minimum distance and move to requested distance!

Power level “M” – Stethoclip with 1 x 4.7kOhm filterPower level “P, HP & SP” – Stethoclip with 3 x 4.7kOhm filter

Attention

Attention

Attention

Perform wireless function test (WFT).A distance gauge and wireless function test device is required to perform the test.

For detailed instruction how to use the WFT device, refer to document:PAI-0544.pdf - Wireless Functional Test Device

5 Quality Control Out

Page 8: Work nstruction - · PDF filetechnical.support@unitron.com Work nstruction File: PAI-0584-U Revision 1.00 Page: 1 of 8 Number of local copies: Version: 1.00 Written: 2015-11-05 11mhaag

File: PAI-0584-U Version: 1.00

Title: Work Instruction Replace HiBan Mounted (HBM) Page: 8 / 8