8
General Specifications <<Contents>> <<Index>> WN100 Wireless Sound Level Meter Yokogawa Electric Corporation 2-9-32, Nakacho, Musashino-shi, Tokyo, 180-8750 Japan GS 01W07D01-01EN GS 01W07D01-01EN ©Copyright June 2019 4th Edition: July 2021 (KP) GENERAL WN100 is a battery operated wireless sound level meter designed as a field device used under harsh environment or explosive atmosphere, and it is suitable for measurement of large sound at noisy area such as an offshore platform or a process plant to protect workers from noise exposure. In combination with the noise map software WN30 which is also provided by Yokogawa, WN100 enables to create a dynamic online noise map. This product is based on the industrial automation wireless communication standard ISA100.11a of the International Society of Automation (ISA). ISA100.11a is approved as an international standard (IEC 62734) by International Electrotechnical Commission (IEC). FEATURES Sound Level Measurement and Spectral Analysis WN100 measures three frequency-weighted sound level parameters (LAeq,T, LCeq,T, LCpeak) in accordance with IEC 61672. Spectral analysis is provided as 1/3 octave band data. LAeq,T, LCeq,T , LCpeak and 1/3 octave band data are transmitted to the host system with wireless communication. Long Life Battery Operation WN100 can continuously measure sound level for 2 years without replacing batteries. Antivibration stage Dedicated antivibration stage improves low level sound measurement by reducing a mechanical vibration effect from an installation environment. STANDARD SPECIFICATIONS FUNCTIONAL SPECIFICATIONS FREQUENCY-WEIGHTED SOUND PRESSURE LEVEL Measurement Parameters: LAeq,T (Time-averaged, A-weighted sound level) LCeq,T (Time-averaged, C-weighted sound level) LCpeak (C-weighted peak sound level) Level Range: From typical noise floor to maximum level Sinusoidal electrical signal of 1 kHz LAeq,T: 60 to 140 dB LCeq,T: 70 to 140 dB LCpeak: 80 to 143 dB Linear Operating Range: In accordance with IEC 61672 class 2 Sinusoidal electrical signal of 1 kHz LAeq,T: 75 to 140 dB LCeq,T: 80 to 140 dB Measurement Accuracy: In accordance with IEC 61672 class 2 Averaged response from all directions in a horizontal plane by sinusoidal acoustical signals LAeq,T: Within acceptance limits of A-weightings LCeq,T: Within acceptance limits of C-weightings SPECTRAL ANALYSIS 1/3 Octave Band Center Frequencies: 25 Hz to 16 kHz (29 bands) Level Range: From typical noise floor to maximum level 25 Hz to 8 kHz bands: 60 to 140 dB 10 kHz, 12.5 kHz bands: 65 to 140 dB 16 kHz band: 70 to 140 dB

WN100 Wireless Sound Level Meter

  • Upload
    others

  • View
    11

  • Download
    0

Embed Size (px)

Citation preview

GeneralSpecifi cations

<<Contents>> <<Index>>

WN100Wireless Sound Level Meter

Yokogawa Electric Corporation2-9-32, Nakacho, Musashino-shi, Tokyo, 180-8750 Japan

GS 01W07D01-01EN

GS 01W07D01-01EN©Copyright June 2019

4th Edition: July 2021 (KP)

GENERALWN100 is a battery operated wireless sound level meter designed as a fi eld device used under harsh environment or explosive atmosphere, and it is suitable for measurement of large sound at noisy area such as an off shore platform or a process plant to protect workers from noise exposure.In combination with the noise map software WN30 which is also provided by Yokogawa, WN100 enables to create a dynamic online noise map.This product is based on the industrial automation wireless communication standard ISA100.11a of the International Society of Automation (ISA). ISA100.11a is approved as an international standard (IEC 62734) by International Electrotechnical Commission (IEC).

FEATURES● Sound Level Measurement and Spectral

AnalysisWN100 measures three frequency-weighted sound level parameters (LAeq,T, LCeq,T, LCpeak) in accordance with IEC 61672.Spectral analysis is provided as 1/3 octave band data.LAeq,T, LCeq,T , LCpeak and 1/3 octave band data are transmitted to the host system with wireless communication.

● Long Life Battery OperationWN100 can continuously measure sound level for 2 years without replacing batteries.

● Antivibration stageDedicated antivibration stage improves low level sound measurement by reducing a mechanical vibration eff ect from an installation environment.

STANDARD SPECIFICATIONS FUNCTIONAL SPECIFICATIONS

FREQUENCY-WEIGHTED SOUND PRESSURE LEVELMeasurement Parameters:

LAeq,T (Time-averaged, A-weighted sound level)LCeq,T (Time-averaged, C-weighted sound level)LCpeak (C-weighted peak sound level)

Level Range:From typical noise fl oor to maximum levelSinusoidal electrical signal of 1 kHz

LAeq,T: 60 to 140 dBLCeq,T: 70 to 140 dBLCpeak: 80 to 143 dB

Linear Operating Range:In accordance with IEC 61672 class 2Sinusoidal electrical signal of 1 kHz

LAeq,T: 75 to 140 dBLCeq,T: 80 to 140 dB

Measurement Accuracy:In accordance with IEC 61672 class 2Averaged response from all directions in a horizontal plane by sinusoidal acoustical signals

LAeq,T: Within acceptance limits of A-weightingsLCeq,T: Within acceptance limits of C-weightings

SPECTRAL ANALYSIS1/3 Octave Band Center Frequencies:

25 Hz to 16 kHz (29 bands)Level Range:From typical noise fl oor to maximum level

25 Hz to 8 kHz bands: 60 to 140 dB10 kHz, 12.5 kHz bands: 65 to 140 dB16 kHz band: 70 to 140 dB

2

All Rights Reserved. Copyright © 2019, Yokogawa Electric Corporation

<<Contents>> <<Index>>

GS 01W07D01-01EN July 2, 2021-00

Measurement Accuracy:Averaged response from all directions in a horizontal plane by sinusoidal acoustical signals

40 Hz to 2 kHz bands: ±3 dB25 Hz, 31.5 Hz, 2.5 kHz to 16 kHz bands: ±5 dB

Measurement Time Interval (Measurement Period):10 to 600 s

Self Diagnostics:Electronics failure, configuration error, battery alarm, wireless communication alarm and over-range error for process variables.

POWER SUPPLY SPECIFICATIONSBattery Pack:The dedicated battery pack with primary lithium-thionyl chloride batteries is replaceable in hazardous area. • Rated voltage: 7.2 V • Rated capacity: 19 AhTypical battery life is 2 years at 1 minute measurement interval under the following conditions.• Ambient temperature: 23±2°C• Sound level measurement: LAeq,T, LCeq,T , LCpeak only

WIRELESS SPECIFICATIONSCommunication Protocol:

ISA100.11a (IEEE802.15.4)Data Rate:

250 kbpsFrequency:

2400-2483.5 MHz license free ISM bandRadio security:

AES 128 bit RF Transmitter Power:

Less than 10.0 dBm (EIRP)Antenna:

+2 dBi Omni directional monopole typeUse only the dedicated +2dBi antenna. Extension by a remote antenna cable is not available.

INSTALLATION ENVIRONMENTAmbient Temperature Limits:

Operating: -20 to 70°C Storage: -25 to 70°C

Ambient Humidity Limits:Operating: 0 to 100%RH (non-condensation) Storage: 0 to 90%RH (non-condensation)

Vibration Resistance:Mechanical vibration from an installed environment significantly affects to the sound level measurement.In combination with the dedicated antivibration stage, LAeq,T can be measured down to around 65 dB under the vibration of 4 m/s2 rms at a typical offshore platform.

Horizontal Attitude:Tilt angle limit of the anti-vibration stage: ±6°

REFERENCE CONDITIONSReference Ambient Temperature:

23°CReference Ambient Humidity:

50%RHReference Frequency:

1 kHzReference Sound Pressure Level:

114 dB

REGULATORY COMPLIANCE STATEMENTSPlease confirm that an installation region fulfills an applicable standard. If additional regulatory information and approvals are required, contact a Yokogawa representative.

Telecommunication compliant countries:EU countries, Australia, New Zealand, Thailand, Malaysia, Brazil, Saudi Arabia, Japan

CE Conformity:RoHS Directive:

EN IEC 63000ATEX Directive:

See “OPTIONAL SPECIFICATIONS (For Explosion Protected Types)”

RE Directive:Safety: EN61010-1 (Indoor/Outdoor use),

EN61010-2-030, EN62479EMC: EN 301 489-1, EN 301 489-17, EN61326-1

Class A, Table 2, EN61326-2-3, EN55011Radio Spectrum: EN 300 328

PHYSICAL SPECIFICATIONSHousing:

Low copper cast aluminum alloyCoating of housing:

Urethane curing type polyester resin powder coating Mint-green paint (Munsell 5.6 BG 3.3/2.9 or its equivalent)

Degrees of Protection:IP55This rating applies when the antenna connector is properly tightened and the all-weather protector is properly used.

Weight:7.1 kg (including battery and antivibration stage)5.4 kg (including battery, excluding antivibration stage)

Mounting:2-inch pipe mounting

3<<Contents>> <<Index>>

All Rights Reserved. Copyright © 2019, Yokogawa Electric Corporation GS 01W07D01-01EN July 2, 2021-00

MODEL AND SUFFIX CODESModel Suffix Codes Description

WN100 ����������������������������������������������������������������������������������� Wireless Sound Level Meter

GeneralSpecifications

Output Signal -L ������������������������������������������������������ ISA100 WirelessAmplifier Housing 8 ����������������������������������������������� Cast aluminum alloy with detachable antenna (2 dBi)--- D ���������������������������������������� Always DMicrophone C �������������������������������� Electret condenser microphoneMounting Bracket K ������������������������ 316 SST 2-inch pipe mounting

N ������������������������ NoneAntivibration Stage B ����������������� 316 SST baseplate with dampers

N ����������������� None--- A ��������� Always A

Option codes /Optional specification

OPTIONAL SPECIFICATIONS (For Explosion Protected Types)Item Description Code

ATEX

ATEX Intrinsically safe ApprovalApplicable Standards: EN IEC 60079-0:2018, EN 60079-11:2012, EN 60079-28:2015

Certificate: Presafe 19 ATEX 14009X II 1 G Ex ia op is IIC T4 Ga Amb. Temp. (Tamb): –40 to 70 °C (–40 to 158 °F)

KS27

IECEx

IECEx Intrinsically safe Approval Applicable Standards: IEC 60079-0:2017, IEC 60079-11:2011, IEC 60079-28:2015 Certificate: IECEx PRE 19.0003X Ex ia op is IIC T4 Ga Amb. Temp. (Tamb): –40 to 70 °C (–40 to 158 °F)

SS27

INMETRO

INMETRO Intrinsially safe ApprovalApplicable Standard: ABNT NBR IEC 60079-0:2013 Versão Corrigida 2:2016,

ABNT NBR IEC 60079-11:2013 Versão Corrigida:2017, ABNT NBR IEC 60079-28:2016 Certificate: DEKRA 20.0011X Ex ia op is IIC T4 Ga Amb. Temp. (Tamb): –40 to 70 °C (–40 to 158 °F)

US17

OPTIONAL SPECIFICATIONS (Others)Item Description Code

Non explosion proof Non explosion proof type N11

OPTIONAL ACCESSORIESItem Parts Number Description

Battery pack assembly F9915NQ*1 Battery case, Lithium-thionyl chloride batteries 2 piecesBattery case F9915NK*2 Battery case only (Standard accessories)Batteries F9915NR*3 Lithium-thionyl chloride batteries 2 piecesAntenna*4 F9915KW 2 dBi standard antenna (Standard accessories)Microphone F9095UD*5 Electret condenser microphone (Standard accessories)All-weather protector F9095GK*6 All-weather protector with windscreen (Standard accessories)Windscreen F9095GL*7 Polyurethane foamAntivibration stage F9095GM 316 SST baseplate with dampers

*1: The part number of the battery pack assembly is described as F9915MA in the user’s manual. Please purchase F9915MA as F9915NQ which consists of F9915MA and its instruction manual.

*2: The part number of the battery case is described as F9915NS in the user’s manual. Please purchase F9915NS as F9915NK which consists of F9915NS and its instruction manual.

*3: Alternatively, Tadiran SL-2780/S, TL-5930/S or VITZROCELL SB-D02 batteries can be purchased from your local distributor.*4: Extension by remote antenna cables or use of any other antennas except for F9915KW are not permitted.*5: Alternatively, Ono Sokki MI-1271 can be purchased from your local distributor.*6: Alternatively, Norsonic P4560 can be purchased from your local distributor.*7: Alternatively, Norsonic P4529 can be purchased from your local distributor.

4

All Rights Reserved. Copyright © 2019, Yokogawa Electric Corporation

<<Contents>> <<Index>>

GS 01W07D01-01EN

DIMENSIONS● 2-inch pipe mounting with antivibration stage (for horizontal piping)

Unit: mm (approx. inch)

F01.ai

81 (3.19) 88 (3.46) 103 (4.06) 142 (5.59)

Groundterminal Ground

terminal

307

(12.

09)

44

(1.7

3)

16

(0.6

3)

130 (5.12) 2-inch pipe

Terminalcover 34

9 (1

3.74

)

65

(2.56)

94 (3.70)

Ø 1

10 (4

.33)

Ø 1

10 (4

.33)

160

(6.3

0)

116

(4.5

7) 102

(4.0

2)

Mountingbracket

169 (6.65)

150 (5.91)

92 (3.62) 19 (0.75)

110 (4.33)

July 2, 2020-00

5<<Contents>> <<Index>>

All Rights Reserved. Copyright © 2019, Yokogawa Electric Corporation GS 01W07D01-01EN

● 2-inch pipe mounting with antivibration stage (for vertical piping)Unit: mm (approx. inch)

F02.ai

81 (3.19) 88 (3.46) 103 (4.06) 142 (5.59)

Groundterminal Ground

terminal

307

(12.

09)

44

(1.7

3)

16

(0.6

3)

92 (3.62) 19 (0.75)

110 (4.33)

130 (5.12)

2-inch pipe

Terminalcover 34

9 (1

3.74

)

65

(2.56)

Ø 1

10 (4

.33)

Ø 1

10 (4

.33)

160

(6.3

0)

116

(4.5

7)

102

(4.0

2)

Mountingbracket

190 (7.48)

150 (5.91)

July 2, 2020-00

6

All Rights Reserved. Copyright © 2019, Yokogawa Electric Corporation

<<Contents>> <<Index>>

GS 01W07D01-01EN

● 2-inch pipe mounting without antivibration stage (for horizontal piping)Unit: mm (approx. inch)

F03.ai

81 (3.19) 88 (3.46) 103 (4.06) 142 (5.59)

Groundterminal Ground

terminal

307

(12.

09)

16 (0

.63)

110 (4.33)

130 (5.12)

2-inch pipe

Terminalcover 34

9 (1

3.74

)

65

(2.56)

94 (3.70)

169 (6.65)

Ø 1

10 (4

.33)

Ø 1

10 (4

.33)

122

(4.8

0)

116

(4.5

7)

64 (2

.52)

Mountingbracket

92 (3.62) 69 (2.72)

150 (5.91)

July 2, 2020-00

7<<Contents>> <<Index>>

All Rights Reserved. Copyright © 2019, Yokogawa Electric Corporation GS 01W07D01-01EN

● 2-inch pipe mounting without antivibration stage (for vertical piping)Unit: mm (approx. inch)

F04.ai

81 (3.19) 88 (3.46) 103 (4.06) 142 (5.59)

Groundterminal Ground

terminal

307

(12.

09)

16 (0

.63)

92 (3.62)

110 (4.33)

130 (5.12)

2-inch pipe

Terminalcover 34

9 (1

3.74

)

65

(2.56)

190 (7.48)

Ø 1

10 (4

.33)

Ø 1

10 (4

.33)

122

(4.8

0)

116

(4.5

7)

64 (2

.52)

Mountingbracket

69 (2.72)

150 (5.91)

July 2, 2020-00

8

All Rights Reserved. Copyright © 2019, Yokogawa Electric Corporation

<<Contents>> <<Index>>

GS 01W07D01-01EN

INFRARED CONFIGURATIONInfrared port

F05.ai

Subject to change without notice.

<Ordering Information>Specify the following when ordering Model, suffix codes, and option codes

<Related Products General Specifications>Field Wireless System: Refer to GS 01W01A01-01ENField Wireless Management Station YFGW410: Refer to GS 01W02D01-01ENField Wireless Access Point YFGW510: Refer to GS 01W02E01-01ENField Wireless Access Point YFGW520: Refer to GS 01W02E02-01ENFieldMate Versatile Device Management Wizard: Refer to GS 01R01A01-01ENoise Map Software WN30: Refer to GS 01W07E01-01EN

<Trademark>All brand or product names of Yokogawa Electric Corporation in this document are trademarks or registered trademarks of Yokogawa Electric Corporation.All other company brand or product names in this document are trademarks or registered trademarks of their respective holders.In this document, trademarks or registered trademarks are not marked with “™” or “®”.

July 2, 2020-00