Upload
wilfred-higgins
View
221
Download
1
Tags:
Embed Size (px)
Citation preview
Purpose of Anesthesia Units ◦ Supplies a set mixture of gases and anesthetic
vapors to the patient ◦ Helps the clinician control patient breath rates,
volumes, and pressures ◦ Monitors gas concentrations from the anesthesia
machine ◦ Provides patient monitoring ◦ Includes safety features to help protect the
patient ◦ Removes excess gases from the breathing system
and out of the anesthetizing location
Anesthesia unit
Anesthetizing Locations ◦ An area used to administer any flammable or
nonflammable anesthetic agent such as the operating room, delivery room, emergency room and anesthetizing room
◦ Requirements for a nonflammable anesthetizing area
◦ Areas must be clearly labeled “RESTRICTED TO NONFLAMMABLE ANESTHETICS”
Anesthesia unit
◦ Environments factors Humidity must be maintained at a minimum of
35% Ventilation requirements
Used to remove bacteria, anesthetic gases, and maintain proper humidity
A positive pressure is maintained within the anesthetizing area
Temperature Decreases metabolic rate of the patient Decreases blood loss Decreases infection by minimizing bacteria growth
Anesthesia unit
Clinical Aspects of Anesthesia ◦ Definitions
Anesthesia – the loss of the normal perception of pain
Anesthetic – a substance which produces anesthesia
◦ Purpose of anesthesia – to remove the patient’s sensitivity to pain during surgical procedures
Anesthesia unit
◦ Types of anesthesia Local anesthesia
Applied only to the area of treatment Patient remains conscious
General anesthesia Applied to the entire body Usually results in loss of consciousness Administered one of two ways o Intravenous – the anesthetic is injected into the blood
stream o Inhalation – the patient breathes a gas mixture containing
the anesthetic agent
Anesthesia unit
Inhalation Anesthetics ◦ Delivered by inhalation ◦ May be gases or volatile liquids ◦ All common anesthetics in use today are non-
explosive and nonflammable ◦ Minimum alveolar concentration level (MAC)
Concentration of any anesthetic agent in which 50% of the population will be anesthetized
The MAC level must be exceeded to ensure proper anesthesia
Anesthesia unit
Nitruous It is colorless and odorless Nitrous oxide is a gas at ambient temperature and
pressure Kept as a liquid under pressure Does not provide significant muscle relaxation MAC = 105% Frequently used in combination with the more potent
volatile agents
Anesthesia unit
◦ Halogenated anesthetic agents Causes rapid, shallow breathing Relaxes skeletal muscles Halothane
Detrimental effect on rubber goods and some metals MAC = 0.77%
Isoflurane Several side effects and is not popular MAC = 1.77%
Anesthesia unit
Enflurane Few side effects MAC = 1.3%
Deslurane Will not evaporate at room temperature Requires a special heated vaporizer Low solubility in blood and body tissues causes a very
rapid wash-in and wash-out MAC = 7.25%
Aevoflurane Low solubility in blood and body tissues MAC = 1.7%
Anesthesia unit
Stages of General Anesthesia ◦ Induction – start of administration of the
anesthesia agents (during induction it is very important to remain as calm as possible)
◦ Excitement – the patient is delirious and physiologically unstable
◦ Surgical Plane – the patient is insensible to pain. Breathing is steady and automatic
◦ Danger – the patient is in danger of cardiopulmonary arrest
Anesthesia unit
Guidelines set by the manufacturer must be strictly followed ◦ Recommendations before attempting
maintenance of this type of equipment All test equipment and ventilators are not the
same Ventilator Operation is Usually
Unsupervised ◦ The calibrations are extremely critical to the
support of the patient's health and/or life
General Calibration Procedures for an Anesthesia unit
Anesthesia Unit Operation is Usually Supervised ◦ The calibrations are extremely critical to the
support of the patients health and/or life◦ Tolerances
Are not as narrow as the ventilator due to the constant supervision of the patient
Still very real because of the drugs that are used A unit’s reaction to the changes in the patient’s
condition Can cause under or over anesthetic dosage Can lead to damage or death
General Calibration Procedures for an Anesthesia unit
◦ Tolerances Are narrower than most other devices in the
hospital (ex. +/- 1% instead of +/-3%) A units reaction to the changes in the patient's
condition Can cause under or over ventilation Can lead to injury or death
General Calibration Procedures for an Anesthesia unit
Test Equipment ◦ Calibrations are only as accurate as the
calibration of the test equipment Calibration Items
◦ Pressure delivery ◦ Pressure sensing
Peak Mean Peep
General Calibration Procedures for an Anesthesia unit
◦ Overpressure ◦ Volume delivered ◦ Volume sensing ◦ Breath rate ◦ Oxygen blending◦ Assist sensitivity ◦ Safety devices ◦ Alarm limits
General Calibration Procedures for an Anesthesia unit
Anesthesia unit special items of concern◦ Safety devices
Proportioning system Monitors the nitrous oxide flow to the oxygen flow to
maintain <3:1 ratio Designed to prevent oxygen concentration from falling
below 25% The kick-in point will o limit nitrous oxide flow from increasing if nitrous oxide is
increased above 3:1 o increase oxygen flow if nitrous oxide is increased above 3:1
Below 25% oxygen is considered hypoxic decrease nitrous oxide flow if oxygen is decreased above
3:1
General Calibration Procedures for an Anesthesia unit
◦ Secondary gas shutoff Monitors the oxygen source pressure Controls nitrous oxide and air flows (secondary gases) Will either o Decrease flow of secondary gases in proportion to oxygen
pressure o Shutoff flow instantly at a preset pressure
Either way the secondary gases will be completely cut-off before oxygen is depleted
Vaporizer interlocksUsed on multiple vaporizer systems Prevents more than one vaporizer from being turned
on at a time
General Calibration Procedures for an Anesthesia unit
◦ Alarm limits◦ Common/fresh gas concentrations
Ensure minimum oxygen content cannot be adjusted to hypoxic levels
VaporizerCalibration verification can only be performed Use a Rankin gas analyzer If verification falls outside of limits then vaporizer will
be sent to manufacturer for repair or replacement
General Calibration Procedures for an Anesthesia unit