What you need to know about the new European IVD Regulation
Sue Spencer Stefan Burde September 2014
2 Copyright © 2014 BSI. All rights reserved.
What you need to know about the new European IVD Regulation
•What is it?
•Why do we need it?
•When is it coming?
•What are the major changes?
• Risk-based IVD Device Classification Approach
• Routes to Conformity
• Clinical Expectations
• Responsibilities of the Manufacturer, Importer and Distributors and In-house Manufacture
9/15/2014
3 Copyright © 2014 BSI. All rights reserved.
Caution
15/09/2014
• The new regulations are draft and subject to change
• Further details will be added later pre and post application through implementing and delegating legislation
4 Copyright © 2014 BSI. All rights reserved.
5 Copyright © 2014 BSI. All rights reserved. Copyright © 2014 BSI. All rights reserved.
Why do we need it?
9/15/2014
6 Copyright © 2014 BSI. All rights reserved.
Impetus for the changes
• Discovery of a 16 year fraud in PIP breast implants using low quality “industrial grade” silicon oil
• Stress test performed by EU Commission
• Determine that changes were needed to improve early detection and prevent this type if incident
• Other high profile vigilance cases with hips, pelvic floor meshes, pacemaker leads, etc.
Outcome
• Short term changes proposed to the system:
• Increased market surveillance
• Additional unannounced visits on top of regular audits
• Identify a person who is responsible for regulatory compliance
7 Copyright © 2014 BSI. All rights reserved.
STEP CHANGE 2010 2014 2016-2018
Joint audits
UAV
CoC audits
COEN - MS
EUDAMED
Electronic submission
UDI
Industry in Spotlight
MDR / IVDR
EU oversight
Clinical trials
Special NBs
and MORE….
8 Copyright © 2014 BSI. All rights reserved.
Advantages to the proposed changes
•Appropriate regulation of the entire IVD market • Current directive focuses on a limited number of devices of particular concern at the time of issue
• New risk-based classification scheme captures the majority of devices, and is applicable to new devices that may come to market in the future
• The level of regulatory scrutiny is commensurate with the relative risk of the device
•Consistent application • Under the current directive, each member state has implemented the
directive slightly differently into national law
• Raises concern about unequal competition
9 Copyright © 2014 BSI. All rights reserved.
Notified Body Code of Conduct
• Mandatory to sign for TEAM-NB members
• Available on website www.team-nb.org
•New version to be approved on paper soon • Frequency of UAV
• Focus of UAV
10 Copyright © 2014 BSI. All rights reserved. Copyright © 2014 BSI. All rights reserved.
What is it?
15/09/2014
11 Copyright © 2014 BSI. All rights reserved.
IVD directive will become a regulation
What’s the difference
• A Directive is agreed by the European Parliament and Council and directs member states to pass national legislation to implement the directive
• A Regulation is a law agreed by the European Parliament and Council that takes effect directly in all member states
Impact of becoming a regulation
• The regulation is intended to result in more consistent application
• The same text is in force in all member states
• Direct entry into force
• No Transposition period
• No need for transposition into national law
• There will be a transition period of 3 years - reduced from 5 in the first draft
• The reduction to 3 years is intended to align with the MDR
• The regulation identifies areas which can be updated in the future using additional implementing acts according to Article 84(3)
12 Copyright © 2014 BSI. All rights reserved.
Structure of the IVDR Chapters 10
Articles 90
Annexes 14
Annex I General Safety and Performance Requirements • Equivalent to the current essential
requirement
• Broadly similar with additional clarification
• New sections for software and requirements for use with mobile platforms
• Requirements for self tests are extended to include near patient testing
Annex II Technical documentation • Significantly more detail regarding the
expectations for technical documentation
Annex III Declaration of Conformity
Annex IV CE marking
Annex V Registration and UDI
Annex VI Requirements for Notified Bodies
Annex VII Classification
Annex VIII Conformity Assessment based on Full QA or Design Examination
Annex IX Conformity Assessment based on Type Examination
Annex X Conformity Assessment based on Production QA
Annex XI Notified Bodies Certificate content
Annex XII Clinical Evidence and Post Market Follow up
Annex XIII Interventional Clinical Performance Studies
Annex XIV Correlation table
15/09/2014
More detailed & more prescriptive; consistent with the proposed Medical Device Regulation
13 Copyright © 2014 BSI. All rights reserved. Copyright © 2014 BSI. All rights reserved.
When is it coming?
15/09/2014
14 Copyright © 2014 BSI. All rights reserved.
Timeline
ENVI Parliament
approved the ENVI report
Plenary Vote
Council Second Reading
Adoption Implementing
Acts Transition Enforced
3 Years possibly 5 years
IVDR must be applied 2018-20
IVDR
2015-6
15 Copyright © 2014 BSI. All rights reserved.
Managing the transition
Entry into Force
NB designation (6 months)
New conformity assessment procedures, including use of clinical evidence (24 months)
Implementing measures (12 months)
Cooperation Between Authorities (12 months)
Unique Device Identification Systems (UDI) (18 months)
New Vigilance Procedures (24 months)
16 Copyright © 2014 BSI. All rights reserved.
Timeline
ENVI Parliament
approved the ENVI report
Plenary Vote
Council Second Reading
Adoption Implementing
Acts Transition Enforced
3 Years possibly 5 years
IVDR must be applied 2018-20
IVDR
2015-6
Copyright © 2014 BSI. All rights reserved.
What are the major changes?
18 Copyright © 2014 BSI. All rights reserved. 18 Copyright © 2014 BSI. All rights reserved.
Risk-based classification approach
15/09/2014
19 Copyright © 2014 BSI. All rights reserved.
Change to scope of IVDR through MDR • ‘medical device’ means any instrument, apparatus, appliance, software, implant, reagent, material or
other article, intended by the manufacturer to be used, alone or in combination, for human beings for one or more of the specific direct or indirect medical purposes of:
• diagnosis, prevention, monitoring, prediction, prognosis, treatment or alleviation of disease,
• diagnosis, monitoring, treatment, alleviation of or compensation for an injury or disability,
• investigation, replacement or modification of the anatomy or of a physiological process or state,
• investigation, replacement or modification of the anatomy or of a physiological process or state,
• disinfection or sterilisation of any of the above-mentioned products,
• providing information concerning direct or indirect impacts on health,
and which does not achieve its principal intended action by pharmacological, immunological or metabolic means, in or on the human body, but which may be assisted in its intended function by such means.
9/15/2014
Impact Tests to predict the likelihood patients will develop cancer or heart disease will be included, Life style tests for example tests to suggest dietary changes for health reasons will be included
20 Copyright © 2014 BSI. All rights reserved.
Scope
'in vitro diagnostic medical device’ means any medical device which is a reagent, reagent product, calibrator, control material, kit, instrument, apparatus, equipment, software or system, whether used alone or in combination, intended by the manufacturer to be used in vitro for the examination of specimens, including blood and tissue donations, derived from the human body, solely or principally for the purpose of providing information:
• concerning a physiological or pathological state;
• concerning a physical or mental impairments;
• concerning the predisposition to a medical condition or a disease;
• to determine the safety and compatibility with potential recipients;
• to predict treatment response or reactions;
• to define or monitor therapeutic measures.
In vitro diagnostic medical devices used for DNA-testing shall be subject to this Regulation
9/15/2014
21 Copyright © 2014 BSI. All rights reserved.
IVD Classification
15/09/2014
1.1. Application of the classification rules shall be governed by the intended purpose, novelty, complexity and inherent risk of the devices.
22 Copyright © 2014 BSI. All rights reserved.
Classification Class D (Blood screening)
• Devices intended to be used to detect the presence of, or exposure to,
• a transmissible agent in blood, blood components, cells, tissues or organs, or in any of their derivatives, in order to assess their suitability for transfusion or transplantation.
• a transmissible agent that causes a life-threatening disease with a high or currently undefined risk of propagation
• Blood grouping ABO, Rhesus, Kell, Kidd and Duffy systems
Class C
Devices intended for
• detecting the presence of, or exposure to, a sexually transmitted agent;
• detecting the presence in cerebrospinal fluid or blood of an infectious agent with a risk of limited propagation;
• detecting the presence of an infectious agent, if there is a significant risk that an erroneous result would cause death or severe disability to the individual, foetus or embryo being tested, or to the individual's offspring;
• pre-natal screening of women in order to determine their immune status towards transmissible agents;
• determining infective disease status or immune status, if there is a risk that an erroneous result would lead to a patient management decision resulting in an imminent life-threatening situation for the patient or for the patient's offspring;
15/09/2014
23 Copyright © 2014 BSI. All rights reserved.
Classification Class C (Continued)
• selection of patients, i.e.
• Devices intended to be used as companion diagnostics*; or
• Devices intended to be used for disease staging or prognosis;
• Devices intended to be used in screening for or in the diagnosis of cancer.
• human genetic testing;
• monitoring of levels of medicinal products, substances or biological components, when there is a risk that an erroneous result will lead to a patient management decision resulting in an imminent life-threatening situation for the patient or for the patient's offspring;
• management of patients suffering from a life-threatening infectious disease;
• screening for congenital disorders in the foetus or embryo
• Devices intended for self-testing are classified as class C, except those devices from which the result is not determining a medically critical status, or is preliminary and requires follow-up with the appropriate laboratory test in which case they are Class B.
• devices intended for blood gases and blood glucose determinations for near patient testing are class C. Other devices that are intended for near-patient testing shall be classified in their own right.
15/09/2014
*Companion diagnostic means a device specifically intended for and essential to the selection of patients with a previously diagnosed condition or predisposition as suitable or unsuitable for a specific therapy with a medicinal product or a range of medicinal products;
24 Copyright © 2014 BSI. All rights reserved.
Genetic Tests
‘ device for genetic testing’ means an in vitro diagnostic medical device the purpose of which is to identify a genetic characteristic of a person which is inherited or acquired during prenatal development;
New requirement
The following devices may only be supplied on a medical prescription:
1) Class D devices;
2) Class C devices in the following categories:
(a) devices for genetic testing;
(b) companion diagnostics.
9/15/2014
25 Copyright © 2014 BSI. All rights reserved.
Classification
Class B
• Any IVD not listed under Classes D, C or A.
• Controls without an assigned value.
Class A
• Reagents, other articles with specific characteristics.
• Instruments intended specifically for use in IVD procedures.
• Specimen receptacles.
15/09/2014
26 Copyright © 2014 BSI. All rights reserved. Copyright © 2014 BSI. All rights reserved.
Exercise:
Classification
9/15/2014
27 Copyright © 2014 BSI. All rights reserved.
Classification Exercise
• In your groups, classify the IVD products on your assigned cards according to the proposed IVD Regulation
• Use the classification flow chart and the classification rules in Annex VII (see the handout)
• Record the classification and the applicable rule on the worksheet
30 minutes
Click here to start
28 Copyright © 2014 BSI. All rights reserved. Copyright © 2014 BSI. All rights reserved.
Routes to conformity
9/15/2014
29 Copyright © 2014 BSI. All rights reserved.
Conformity Assessment Routes
15/09/2014
30 Copyright © 2014 BSI. All rights reserved.
Designation of Notified Bodies and Special NBs
15/09/2014
Notified Body
Class B
& Class C
31 Copyright © 2014 BSI. All rights reserved.
Additional Requirements for Class D Devices
15/09/2014
Summary of safety and performance
High risk devices (Class C and D) devices will require a summary of safety and performance which will be available to the public and should be is clear to the intended user.
Greatest impact on devices which have been up classified or novel devices not included in a CTS
Maximum 175 days
32 Copyright © 2014 BSI. All rights reserved.
Quantum Leap
Do not require a Notified Body
80-90%
Require a Notified Body 80-90%
Require a Notified Body
Do not require a Notified Body
IVD Directive IVD Regulation
33 Copyright © 2014 BSI. All rights reserved. 33 Copyright © 2014 BSI. All rights reserved.
Clinical Expectations
34 Copyright © 2014 BSI. All rights reserved.
Clinical Requirements
• Increased expectation for clinical requirements
• Clinical evidence is to be kept up to date during the life time of the device
The GHTF documents now in the IMDRF archive best guidance
• Clinical Performance Studies for In Vitro Diagnostic Medical Devices
• Clinical Evidence for IVD Medical Devices – Key Definitions and Concepts
• Clinical Evidence for IVD Medical Devices – Scientific Validity Determination and Performance Evaluation
CLINICAL
EVIDENCE Analytical Performance
CLINICAL
UTILITY
SCIENTIFIC
VALIDITY
Clinical Performance
35 Copyright © 2014 BSI. All rights reserved.
Elements of Clinical Evidence
CLINICAL EVIDENCE
Scientific Validity
refers to the association of an analyte to a clinical condition
or physiological state
Clinical Performance
refers to its ability to yield results that relate to a
particular clinical condition/physiological state for the
intended use and in accordance with target population and, where
applicable, the intended user
Analytical Performance
refers to the ability of an IVD medical device to correctly
detect or measure a particular analyte
For established analytes, this may be from the literature; but for
companion diagnostics or novel analytes this needs to be
established
Sensitivity/ specificity similar to current essential requirements
Data to support reference ranges etc.
36 Copyright © 2014 BSI. All rights reserved.
Clinical Utility
The usefulness of the results obtained from testing with the IVD medical device and the value of the information to the individual being tested and/or the broader population
• Clinical Utility has never been a regulatory requirement in Europe
• Clinical Utility is still under discussion but likely to be required for companion diagnostics, according to the proposed amendment
37 Copyright © 2014 BSI. All rights reserved.
Responsibilities of the manufacturer, importer and distributers plus in-house manufacture
38 Copyright © 2014 BSI. All rights reserved.
Economic Operators Manufacturer
means the natural or legal person with responsibility for the design, manufacture, packaging and labelling of a device before it is placed on the market under that person's own name, regardless of whether those operations are carried out by that person or on that person's behalf by a third party. The obligations of this Regulation to be met by manufactures also apply to natural or legal persons who assemble, package, process, fully refurbish or label one or more ready-made products and/or assign to them their intended purpose as devices with a view to their being placed on the market under that person's own name or trademark.
Importer
means any natural or legal person established within the Union who places a device from a third country on the Union market;
Distributor
means any natural or legal person in the supply chain, other than the manufacturer or the importer, who makes a device available on the market;
Economic operators
means the manufacturer, the authorised representative, the importer and the distributor;
15/09/2014
39 Copyright © 2014 BSI. All rights reserved.
Control of the Supply Chain
9/15/2014
Crucial Suppliers
OEM’s
Sub contractors
Distributers
Importers
Authorised Representatives
Manufacturer
• Increase expectation for virtual manufacturers and OBL to hold or have quick access to technical documentation during audits
• Notified bodies can now audit crucial suppliers as well as significant subcontractors including unannounced visits
• Changes to contracts will be required
40 Copyright © 2014 BSI. All rights reserved.
Increased Control of the Supply Chain
• Regulatory roles and requirements of
• Importers
• Distributers
• Authorised Representatives
These include
• Registration and keeping data up to date
• Mandate with the authorised representative
• Roles in vigilance and recall
• Required to have a Person responsible for regulatory compliance
• Manufacturers now have to have liability insurance but importers required to check this is adequate or take out their own
9/15/2014
41 Copyright © 2014 BSI. All rights reserved.
• Manufacturers shall have available within their organisation at least one person responsible for regulatory compliance who possesses the requisite expertise in the field of in vitro diagnostic medical devices.
• This will include:
• a degree or equivalent in natural sciences, medicine, pharmacy, engineering or law
• or 3 years of professional experience in regulatory affairs or in QMS relating to IVDs
• The person responsible for regulatory compliance is responsible for ensuring:
• that the conformity of the devices is appropriately assessed before a batch is released;
• that the technical documentation and the declaration of conformity are drawn up and kept up-to-date;
• that vigilance requirements have been fulfilled.
• for performance evaluation for interventional studies
• If compliance is share between more than one person responsibilities will be defined in writing
• This person should suffer no disadvantage by performing their role
• Authorised representatives will also be required to have a person responsible for regulatory compliance within their organisation.
15/09/2014
Person Responsible for Regulatory Compliance
42 Copyright © 2014 BSI. All rights reserved.
Health Institution Definition
• means an organisation whose primary purpose is the care or treatment of patients and which has the legal capacity to carry out such activities;
• commercial laboratories which provide diagnostic services shall not be considered to be health institutions;
43 Copyright © 2014 BSI. All rights reserved.
In-house Exemption for Class D IVDs
• If Class D devices are manufactured and used within a single health institution, they are exempt from the requirements of this Regulation, with the exception of vigilance requirements and general safety performance requirements where the following conditions are met:
(a) the recipient patient or patient group’s specific needs cannot be met by an available CE-marked device as such, and therefore, either a CE-marked device needs to be modified or a new device needs to be manufactured;
(b) the health institution is accredited to ISO standard 15189 quality management system, or any other equivalent recognised standard;
• The Commission shall verify that the devices on that list are eligible for exemption in accordance with the requirements under this paragraph.
• The information on exempt devices shall be made public.
• Member States shall retain the right to restrict the in-house manufacture and use of any specific type of in-vitro diagnostic device in relation to aspects that are not covered by this Regulation, and may also make the manufacture and use of the devices concerned subject to further safety requirements. In such cases, Member States shall inform the Commission and the other Member States accordingly.
9/15/2014
44 Copyright © 2014 BSI. All rights reserved.
Conformity Assessment Routes
15/09/2014
Note Class D devices regardless of whether they are used in a single healthcare institution must meet the regulation with the exception of the requirements for economic operators unless there is no CE marked device Class A, B + C devices used within a single healthcare institution which have a single quality management system compliant with ISO 15189 (Medical laboratories - Particular requirements for quality and competence) may be exempt from the majority of the regulation; however, they must report adverse incidents.
45 Copyright © 2014 BSI. All rights reserved.
Emerging Pathogens
• In the case of urgent or unmet medical needs for patients, such as emerging pathogens and rare diseases, single health institutions should have the possibility of manufacturing, modifying and using devices in-house and thereby addressing, within a non-commercial and flexible framework, specific needs which cannot be met by an available CE-marked device.
• However, devices which are manufactured within non-health-institution laboratories and put into service without being placed onto the market should be subject to this Regulation.
46 Copyright © 2014 BSI. All rights reserved.
EUDAMED
15/09/2014
EUDAMED European Databank on Medical Devices
(as proposed by the European Commission)
Electronic
system
on
Registration
Medical devices / IVDs
economic operators,
incl.
Summary of Safety
and Clinical
Performance
(high risk devices)
Electronic
system
on
UDI
Device Identifier
data elements
Electronic
system
on
Certificates
Certificates issued
by notified bodies
&
Information on
certificates
refused
suspended
reinstated
restricted
withdrawn
Electronic
system
on
Vigilance
Serious incidents
&
Field safety
corrective actions
&
Field safety notices
Electronic
system
on
Market
surveillance
Measures taken
by Member States re.
devices presenting a
risk to health & safety
preventive health
protection measures
Electronic
system
on
Clinical
investigations
Sponsors
(& manufacturers)
description of:
investigational
device,
comparator,
purpose of CI,
status of CI
47 Copyright © 2014 BSI. All rights reserved.
UDI UDID Database
For DI part Only Bar-coding for every Medical Device
DI Device Information - Company - Product ID
PI Production Information -Life -Serial or Lot Information
DI -Company Name Address -Product Name . . . etc
-GMDN -code -term
48 Copyright © 2014 BSI. All rights reserved.
Unique Device Identifier
• The Commission shall be empowered to adopt delegated acts in accordance with Article 85:
• (a) determining the devices, categories or groups of devices, whose identification shall be based on the UDI system, … and the timelines for implementing this. Following a risk-based approach, implementation of the UDI system shall be gradual, starting with devices falling in the highest risk class;
• (b) specifying the data to be included in the production identifier which, following a risk-based approach, may vary depending on the risk class of the device;
• (c) defining the obligations of economic operators, of health institutions and of professional users, in particular regarding allocation of the numeric or alphanumeric characters, placement of the UDI on the label, storage of information in the electronic system on UDI, and use of the UDI in documentation and reporting related to the device provided for in this Regulation;
• (d) amending or supplementing the list of information … in the light of technical progress.
• UDI will ultimately be required
• Development of a UDI system is lagging behind US efforts
• The European Commission intends to make the system as similar as possible to the US system.
• For the time being, following US rules should prepare manufacturers adequately for European roll-out
9/15/2014
49 Copyright © 2014 BSI. All rights reserved.
Final Summary
• This is happening
• There is no grandfathering
• Requirements and expectations are increasing
• Keep up to speed and understand the impact to your organisation
• Talk to your notified body about their plans for designation and resource
• Classify your devices
• Look at the clinical data you have,
• is it enough? How can you get what you need?
• Discuss at management reviews
15/09/2014
50 Copyright © 2014 BSI. All rights reserved.
51 Copyright © 2014 BSI. All rights reserved. Copyright © 2014 BSI. All rights reserved.
Exercise 2:
Preparation for the IVDR
9/15/2014
52 Copyright © 2014 BSI. All rights reserved.
Instructions
• In your teams make a list of items to be considered when planning a strategy for implementing the IVDR
• Are there any dependencies?
• What order would you put them in?
9/15/2014
30 minutes
Click here to start