10
STATIC STATIC Also known as Rayleigh or Classical Light Scattering Measures avg. intensity of scattered light for Absolute Molecular Weight STATIC STATIC Also known as Quasi-elastic Light Scattering (QUELS) or Photon Correlation Spectroscopy (PCS) Measures microsecond fluctuations of single photons Hydrodynamic Radius (Size) Light Scattering Light Scattering (DLS) M.L.Hackert (with figures from Precesion Detectors and Wyatt Tech.) What Do We Mean By ABSOLUTE? There are 4 Absolute Methods of Measuring MW 1) Membrane Osmometry (Number Average MW) 2) Light Scattering (Weight Average MW) 3) Sedimentation Equilibrium (Ultracentrifugation) (z-average MW) 4) Mass spectroscopy NO Reference to standards of mass NO assumptions of molecular model/conformation ALL parameters measured directly from 1st principles Refractive indices geometries of cell and detector wavelength concentrations detector response temperature dn/dc Wyatt Technologies Theory 101 – just the basics – Know why sky is blue, sunset burnt orange Rayleigh Scattering (Lord Rayleigh ~1871) / Rayleigh Ratio How does LS yield an “absolute” molecular weight? LS vs. RI What is polydispersity? How is it defined? Types of “Molecular Weight Averages” LS Instrument / Practical Considerations Static vs. Dynamic vs. X-ray Scattering (info from each) Goals for this unit: Calculation of Mw Averages C i M i y __________ C i M i y-1 M x = Y Mx Name______________ 0 (Mn) Avg. Molecular Number 1 (Mw) Avg. Molecular Weight 2 (Mz) Avg. Z (Velocity) 3 (Mz+1) Avg. Z+1 (Mw/Mn) Polydispersity (Mp) Peak Molecular Weight High Mw Low Mw C i Retention Volume (or Time) Mz+1 Mz Mw Mp Mn Pertinent Mw Averages for Biopolymers Mz & Mz+1 for industrial polymers

What Do We Mean By ABSOLUTE?

  • Upload
    others

  • View
    7

  • Download
    0

Embed Size (px)

Citation preview

Page 1: What Do We Mean By ABSOLUTE?

STATICSTATICAlso known as Rayleigh or Classical Light Scattering

Measures avg. intensity of scattered light for

Absolute Molecular Weight

Measures avg. intensity of scattered light for

Absolute Molecular Weight

STATICSTATIC

Also known as Quasi-elastic Light Scattering (QUELS)or Photon Correlation Spectroscopy (PCS)

Measures microsecond fluctuations of single photons

Hydrodynamic Radius (Size)

Measures microsecond fluctuations of single photons

Hydrodynamic Radius (Size)

Light Scattering

Light Scattering (DLS)

M.L.Hackert (with figures from Precesion Detectors and Wyatt Tech.)

What Do We Mean By ABSOLUTE?• There are 4 Absolute Methods of Measuring MW• 1) Membrane Osmometry (Number Average MW)• 2) Light Scattering (Weight Average MW)• 3) Sedimentation Equilibrium (Ultracentrifugation) (z-average MW)• 4) Mass spectroscopy• NO Reference to standards of mass• NO assumptions of molecular model/conformation• ALL parameters measured directly from 1st principles• Refractive indices• geometries of cell and detector• wavelength• concentrations• detector response• temperature• dn/dc

Wyatt Technologies

• Theory 101 – just the basics – Know why sky is blue, sunset burnt orange

• Rayleigh Scattering (Lord Rayleigh ~1871) / Rayleigh Ratio• How does LS yield an “absolute” molecular weight? LS vs.

RI• What is polydispersity? How is it defined?• Types of “Molecular Weight Averages”• LS Instrument / Practical Considerations• Static vs. Dynamic vs. X-ray Scattering (info from each)

Goals for this unit:

Calculation of Mw Averages

∑CiMiy

__________

∑CiMiy-1

Mx =

Y Mx Name______________0 (Mn) Avg. Molecular Number

1 (Mw) Avg. Molecular Weight

2 (Mz) Avg. Z (Velocity)

3 (Mz+1) Avg. Z+1

(Mw/Mn) Polydispersity

(Mp) Peak Molecular Weight

High Mw Low Mw

Ci

Retention Volume (or Time)

Mz+1

Mz

Mw

Mp

Mn

Pertinent Mw Averages for Biopolymers

Pertinent Mw Averages for Biopolymers

Mz & Mz+1 for industrialpolymers

Mz & Mz+1 for industrialpolymers

Page 2: What Do We Mean By ABSOLUTE?

Recent advances in genomics and proteomics have produced a proliferation of new proteins requiring characterization. Mass spectrometryis ideally suited for identification and primary structural purposes but is not well suited for determining conformational structures in solution. As these molecules are expressed in cell culture, purified and then formulated, rigorous production processes must be carefully evaluated to minimize impact on the protein structure and its long -term shelf life. Obtaining a conformational stability profile of protein or antibodies can help weed out bad drug candidates from good ones as environmental factors can change their tertiary and quaternary structure. Environmental factors include pH, ionic strength, temperature, and excipient composition.

AbstractHPLC is used in flow injection mode with a detector array composed of laser light scattering (static and dynamic modes) and a concentration source detector (RI or UV). This configuration determines the average molecular weight and average hydrodynamic radius with run times as short as 1 minute. Alternatively, a SEC guard column can be used isolate analytefrom excipients (eliminating blank runs) with run times under 3 minutes.

Chromatography with LS Set-up

Wyatt Technologies

Page 3: What Do We Mean By ABSOLUTE?

small

I ∝ 1/ r2 ; I ∝ 1/ λ4 ; I depends on scattering angle

Define “Raleigh Ratio” Rθ

Rθ = r2

Rθ = KCM or K*c

R(θ)= 1

M

Page 4: What Do We Mean By ABSOLUTE?

Basic Light Scattering Principles

• The amount of light scatteredis directly proportional to the product of the molar massand the molecular concentration

• The variation of scattered light with scattering angle is proportional to the average size of the scattering molecules.

Wyatt Technologies

ILS = C Mw (dn/dc)2 Pθ Kθ

K*c

R(θ)= 1 + <rg

2> + + 2 A2c 1 16π 2

M 3λ2[ ] sin2(θ/2) . . .

Basic Light Scattering Equation

Where:

K* = 4π2

(dn/dc)2

n02/ (NAλ0

4)

n0 is the refractive index of the solvent.

c is the concentration of the solute molecules (g/ml).

R(θ) is the fraction of light scattered per unit solid angle, in ex cess of the light scattered by the solvent, divided by the incident intensity.

NA is Avogodro ’s number.

λ0 is the vacuum wavelength of the incident light.

dn/dc is the refractive index increment, which tells how much the refractive index of the solution varies with solute concentration.

Mw is the weight-average molar mass.

K*cR(θ)

1MwP(θ) + 2A2c=

Wyatt Technologies

Excess Light Scattering Signal

Concentration of Solute

Specific Refractive Index Increment

Form Factor

ILS = C Mw (dn/dc)2 Pθ Kθ

Molecular Weight

Detector Constants

Static Light Scattering Detection

Rayleigh Equation

Ø Absolute Molecular Weight

Independent of Column Calibration

Ø Radius of Gyration (Rg)

> 10 nm to 150 nm

Determines

Page 5: What Do We Mean By ABSOLUTE?

RI signal = KRI (dn/dc) C

Rθ = C M (dn/dc)2 P(θ) K(θ) Rayleigh Equation

RI Equation

Rθ K(θ) Mw (dn/dc) P(θ)

RI signal KRI=

RI signal

=

~

• The light scattering signal is directly proportional to the Mw .• The concentration source signal (e.g. RI) is indirectly proportional to the Mw.

RI & Light Scattering Combined for Mw

For a truly effective measurement of molecularweight the static light scattering detector must becombined with a well matched refractometer

Mw

Mw =ILSRI Ktotal

1. Mw is directly proportional to the LS signal

2. Mw is indirectly proportional to the conc. source (RI)

3. The dn/dc is constant for the protein and it’s aggregates (0.186 mL/g)

4. Degree of aggregation can be approximated from visual inspection

Static LS Equation for Proteins Visualizing Aggregation State

Mw =ILSRI

Ktotal

Monomer

dimer

trimer

LS (90deg)

RI

LS and RI channels heightnormalized to BSA monomer

Using visual inspection

• Monomer (LS is equal to RI)

• Dimer (LS is 2X as large as RI)

• Trimer (LS is 3X as large as RI)

Using visual inspection

• Monomer (LS is equal to RI)

• Dimer (LS is 2X as large as RI)

• Trimer (LS is 3X as large as RI)

Page 6: What Do We Mean By ABSOLUTE?

Ø Low angles sensitive to large particles

Ø 90 degree angle more sensitive to smaller particles

Ø High angles less sensitive to larger particles

Ø Back angles better suited for opaque matrices

Why Multi-Angle Detection?

Light Scattering Intensity of Particles Shows an Angular Dependence on Size

Flow Cell & Detector Geometry

incident beam

photodiode detectors

transmitted beam

Wyatt Technologies

ALS4000 Optical Platform Features

Diode laser: 230 mW @ 830 nm (adjustable)

Flow Cell (30 uL)

Peltier cooling/heating ( 0 to 80 degrees C )

Laser Beam Stop

DLS Detectors (Up to 4)

Static LS Detectors (Up to 8)

��

��

� � �

Detector Shutter

2nd Aperture Stop

1st Aperture Stop

Y-Z Translation Lens

DLS Detector

Mw & <rg2> determined by MALS

Intercept gives Mw

Slope gives <rg2>

sin2(θ/2)

K*c

R(θ)

K*c

R(θ)= 1 + <rg

2> + + 2 A2c sin2(θ/2)1 16π 2

M 3λ2[ ] . . .

Page 7: What Do We Mean By ABSOLUTE?

• <rg2> is the mean square radius, relating to

the distribution of mass within the molecule, given by

<rg2> = Σri

2mi

M

Wyatt Technologies

2.9 kD to 780 kDfor Globular Proteins

Rh range nm (log scale)

1 10 1000100

Rg range nm (log scale)

BSA

monom

er 66.

5 kD

BSA d

imer

133 kD

Thyro

globu

lin

669 kD

β-lacto

globu

lin

dimer

36.8

kD

1 102 3 4 5 6 7 8 9Rh range nm

Why use Rh instead of Rg for Biomolecules?

Diffusion Coefficient

Constants

Hydrodynamic Radius

Solvent Viscosity

Temperature

Boltzmann Constant

D0 = κ T(6π η0 Rh )-1

Dynamic Light Scattering Detection

Stokes-Einstein Equation

Ø Molecular or Particle Size

As Hydrodynamic Radius (Rh)

Ø Size Range

1 to 1000 nm

Determines

From g1(τ) the diffusion coefficient (D) for the scattering particles can be determined. From the diffusion coefficient, the hydrodynamic radius can be calculated.

Hydrodynamic Radius Determination

Stokes-Einstein Equation

D0 = κ T(6π η0 Rh )-1Diffus

ion Coe

fficien

t

Temper

ature (

Kelvin

)

Boltzm

ann C

onsta

nt

Solve

nt Visc

osity (

Poise

)

Hydrod

ynamic R

adius

(nm)

1.5 to 1000 nm Radius

Applicable DLS Size Range

Photon Fluctuations

Smaller Particles

Bigger ParticlesPhoton Fluctuations

24 usec

Microsecond Scale

100

0

% C

orr

elat

ion

Correlation Function

Highly correlated

Less correlated

No correlation

Page 8: What Do We Mean By ABSOLUTE?

• Measures Hydrodynamic Radius, Radius of Gyration, Molecular Weight. Particle Size Distribution.

• Instrument Measures

–Rh size from 1.7 nm to 1000 nm.

–Rg size from 9 nm to 150 nm.

–Molecular Weight 200 to 20,000,000 daltons

• Detects branching, aggregates and calculates Mw.

• DLS, does not require conc or dn/dc measurement for size.

• Batch Mode for non-flow system accessories.

Conclusions on Static and Dynamic LS

Page 9: What Do We Mean By ABSOLUTE?

Small angle X-ray scattering (SAXS)

Page 10: What Do We Mean By ABSOLUTE?

• Measures Hydrodynamic Radius, Radius of Gyration, Molecular Weight. Particle Size Distribution.

• Instrument Measures –Rh size from 1.7 nm to 1000 nm.

–Rg size from 9 nm to 150 nm.

–Molecular Weight 200 to 20,000,000 daltons

• Detects branching, aggregates and calculates Mw.

• DLS, does not require conc or dn/dc measurement for size.

• Batch Mode for non-flow system accessories.

Conclusions Part 1: Static and Dynamic LS

•Static Light Scattering •Provides absolute Mw in Solution..•Sensitive Aggregation Detection

•Dynamic Light Scattering •Provides Rh Sizing down to 1 nm for Biomolecules.•“On-the-fly Determinations”•Gives more insight on structure of material•Can be added to any existing static LS system from any manufacturer

•Flow Injection Analysis is Ideal for Rapid Screening for Avg. Mw and Size

Conclusions Part 2: Conclusions Part 3: Why Use Light Scattering Detection?

Ø Absolute Molecular Weight Calculations

Ø Non Destructive Technique

Ø Independent of Elution Time (no relative standards)

Ø Radius of Gyration – Rg

Ø Hydrodynamic Radius - Rh (Dynamic Light Scattering)

Ø Molecular Geometry