Wearable Chemical Sensors: Opportunities and Challenges

Somayeh Imani, Amay Bandodkar, Jayoung Kim, Joseph Wang, and Patrick P. Mercier

UC San Diego

Page 2: Wearable Chemical Sensors: Opportunities and Challengesefficiency.ucsd.edu/wp-content/uploads/2017/10/UCSD... · 2017-10-26 · Electrochemical Biosensors 8 Potentiometric Amperometric

Wearables: an exciting high-growth market

Source: Transparency Market Research

Industrial

Infotainment

Fitness

Medical

3 billion wearables

shipped by 2025*

*IDTechEx 2015 Report

Page 3: Wearable Chemical Sensors: Opportunities and Challengesefficiency.ucsd.edu/wp-content/uploads/2017/10/UCSD... · 2017-10-26 · Electrochemical Biosensors 8 Potentiometric Amperometric

Why aren’t we there now?

3

Battery Life:

Need ultra-low-power

and/or energy harvesting to

minimize re-charging

Utility:

Need to develop sensors

that are actually useful

Size & Usability:

Need to develop sensors

that are small & seamlessly

integrated into daily life

Page 4: Wearable Chemical Sensors: Opportunities and Challengesefficiency.ucsd.edu/wp-content/uploads/2017/10/UCSD... · 2017-10-26 · Electrochemical Biosensors 8 Potentiometric Amperometric

Wearable sensing opportunities

4

Physical attributes Electrical attributes

Biochemical attributes

• Motion (e.g., steps)

• Temperature

• Respiration

• ECG (heart)

• EEG (brain)

• EMG (muscles)

• Blood pressure

• Glucose

• Electrolytes

• Alcohol

• Lactate

• Many more!

Most of the

wearables market

today

Opportunity!

Page 5: Wearable Chemical Sensors: Opportunities and Challengesefficiency.ucsd.edu/wp-content/uploads/2017/10/UCSD... · 2017-10-26 · Electrochemical Biosensors 8 Potentiometric Amperometric

Chemical analysis: current practice

5

Painful / inconvenient blood draws

Page 6: Wearable Chemical Sensors: Opportunities and Challengesefficiency.ucsd.edu/wp-content/uploads/2017/10/UCSD... · 2017-10-26 · Electrochemical Biosensors 8 Potentiometric Amperometric

Example: lactate monitoring for athletes

6

Staying below the “lactate threshold”

important for endurance training

Current state-of-the-art testing

method:

Non-invasive and/or

continuous sensing is required

Page 7: Wearable Chemical Sensors: Opportunities and Challengesefficiency.ucsd.edu/wp-content/uploads/2017/10/UCSD... · 2017-10-26 · Electrochemical Biosensors 8 Potentiometric Amperometric

Non-Invasive Chemical

Biosensing Techniques

7

Electrochemical

Jean et. al. SAS IEEE 2008

Raman Spectroscopy

Bandodkar et. al.,

Anal. Chem. 2015Shih et. al. John, Wiley&Sons2009

Adamson et. al. Analyst 2012

Park et. al. J Diabetes Sci

Technol, 2009

Bioimpedance

Spectroscopy

Ultrasoundmm-wave Detection

Page 8: Wearable Chemical Sensors: Opportunities and Challengesefficiency.ucsd.edu/wp-content/uploads/2017/10/UCSD... · 2017-10-26 · Electrochemical Biosensors 8 Potentiometric Amperometric

Electrochemical Biosensors

8

Potentiometric Amperometric Electrochemical Impedance

Spectroscopy (EIS)

Bandodkar et. al.,

Analyst. 2013

Wang, Chem. Rev. 2008 Lien et. al., Biosensors &

Bioelectronics, 2011

Measures: Voltage Current Complex Impedance

Chemical communicate via ions, electronics via electrons.

Electrochemical biosensors operate by measuring electrons

as the output of redox reactions. Three main types of sensors:

Proportional to underlying concentration of chemical analyte

Page 9: Wearable Chemical Sensors: Opportunities and Challengesefficiency.ucsd.edu/wp-content/uploads/2017/10/UCSD... · 2017-10-26 · Electrochemical Biosensors 8 Potentiometric Amperometric

Biosensor Fabrication

9

Textiles Temporary Tattoos

Most electrochemical biosensors rely on two- or three-

electrode systems: a working electrode (WE), a reference

electrode (RE), and sometimes a counter electrode (CE).

Electrodes are functionalized for analyte specificity, and can

be fabricated on many different materials:

2000 25000

5

10

15

20

1000 1200 1400 1600 18000

10

20

30

A

B

C D

t, s

[Na+ ],

mM

Perspiration

begins

Perspiration

begins

Plastics

Page 10: Wearable Chemical Sensors: Opportunities and Challengesefficiency.ucsd.edu/wp-content/uploads/2017/10/UCSD... · 2017-10-26 · Electrochemical Biosensors 8 Potentiometric Amperometric

Instrumentation Circuits:Potentiometric Biosensors

10

Example: measure the

voltage between a

reference electrode and an

electrode with an ion-

selective membrane coating

Typical Instrumentation

Amplifier Requirements:

Voltage range: 0-400mV

Noise: several µV

Bandwidth: < 100Hz

+ ++

++

++

+

Ionic SolutionIon-selective membrane

Insulator

Substrate

Ag/AgCl electrode

++

+

Applications: pH, K+, Na+,

Cl-, etc…

Page 11: Wearable Chemical Sensors: Opportunities and Challengesefficiency.ucsd.edu/wp-content/uploads/2017/10/UCSD... · 2017-10-26 · Electrochemical Biosensors 8 Potentiometric Amperometric

Example: a wearable sodium sensor tattoo

Enables real-time measurement

of sodium in sweat

Bandodkar et al., Biosensors & Bioelectronics, 2014.

Page 12: Wearable Chemical Sensors: Opportunities and Challengesefficiency.ucsd.edu/wp-content/uploads/2017/10/UCSD... · 2017-10-26 · Electrochemical Biosensors 8 Potentiometric Amperometric

Amperometric biosensors:

sensing metabolites

12

Selectivity: working electrode functionalized with

an enzyme that catalyzes a reaction with the

metabolite to be sensed, which then transfers e- to

an ISE underneath

Operating principal: enzymatic reaction occurs at

specific voltage apply voltage, measure resulting

current proportional to analyte concentration

Circuit: potentiostat used to condition the cell; current

sensor (e.g., TIA) to sense current, often through the

counter electrode to not disturb cell potential

Typical Requirements:

Voltage range: ±1V

Current range: fA - µA

Noise: lower current range

Amp. voltage gain: > 104

Page 13: Wearable Chemical Sensors: Opportunities and Challengesefficiency.ucsd.edu/wp-content/uploads/2017/10/UCSD... · 2017-10-26 · Electrochemical Biosensors 8 Potentiometric Amperometric

13

J. Kim et al., Biosensors & Bioelectronics, 2015

Fitness applications Health applicationsMeasures Uric Acid for

Hyperuricemia

Example System:

A Saliva-Sensing Mouthguard (1)

CE WE

Ref Vol. &

Regulators

Microcontroller

I2CBluetooth

Radio

ADC

+

-A

+ -

TIAI2C

Potentiostat

VREFVBIAS

RE

Measure Lactate for Stress / Exertion

Page 14: Wearable Chemical Sensors: Opportunities and Challengesefficiency.ucsd.edu/wp-content/uploads/2017/10/UCSD... · 2017-10-26 · Electrochemical Biosensors 8 Potentiometric Amperometric

Example System:

A Saliva-Sensing Mouthguard (2)

14

Characterization of integrated

wireless mouthguard biosensor

Monitoring treatment of

hyperuricemia with Allopurinol®

Page 15: Wearable Chemical Sensors: Opportunities and Challengesefficiency.ucsd.edu/wp-content/uploads/2017/10/UCSD... · 2017-10-26 · Electrochemical Biosensors 8 Potentiometric Amperometric

Chemical biosensing

system considerations

15

Sensor

Calibration

Real-Time

Analysis

Data

Processing

Analog Front

End

Data

Transmission

Data

Acquisition

and Storage

Page 16: Wearable Chemical Sensors: Opportunities and Challengesefficiency.ucsd.edu/wp-content/uploads/2017/10/UCSD... · 2017-10-26 · Electrochemical Biosensors 8 Potentiometric Amperometric

Challenge: Efficient Calibration for

Operation in Complex Fluids

𝑉 = 𝑉0 +𝑅𝑇

𝑧𝑖𝐹ln 𝑎𝑖 +

𝑗

𝑘𝑖𝑗𝑎𝑗

𝑧𝑖𝑧𝑗

Selectivity: ISEs are not perfectly selective

other ions in solution will affected

measured potential

Finite selectivity coefficients

Challenge: may require significant

calibration with many analytes to

represent real (complex) fluids

Opportunity: leverage machine-learning

to quickly infer selectivity coefficients

Location: calibrate and/or process on-

device, in the fog, or in the cloud

Page 17: Wearable Chemical Sensors: Opportunities and Challengesefficiency.ucsd.edu/wp-content/uploads/2017/10/UCSD... · 2017-10-26 · Electrochemical Biosensors 8 Potentiometric Amperometric

Example: dopamine monitoring in

complex fluids

17

Bozorgzadeh et. al., VLSI, 2015

Leverage multi-variate analysis to improve specificity in

complex environments

Challenge: dopamine

sensors vary with pH and

background interferers

Solution: in-sensor

principal component

regression (PCR)-based

chemometrics to infer

dopamine concentration

Page 18: Wearable Chemical Sensors: Opportunities and Challengesefficiency.ucsd.edu/wp-content/uploads/2017/10/UCSD... · 2017-10-26 · Electrochemical Biosensors 8 Potentiometric Amperometric

Data analytics

• Real-time chemical data is generally not available

today. We don’t know what kind of correlations to

expect – analytics will help

18

Example: classifying user’s head motions to recognize activities

according to wearable sensor data from the smart glasses

Ho and Wang, User-Centric and Real-Time Activity Recognition Using Smart Glasses, 2015

Page 19: Wearable Chemical Sensors: Opportunities and Challengesefficiency.ucsd.edu/wp-content/uploads/2017/10/UCSD... · 2017-10-26 · Electrochemical Biosensors 8 Potentiometric Amperometric

0 500 1000 1500 2000

80

120

160

200

0

1

2

3

H.

R.

(B/m

in)

Time (s)

Cu

rre

nt

(A

)Simultaneous & continuous, real-time

lactate and ECG sensing

Opportunities for data analytics

S. Imani et al., Nature Communications, May 2016

Lactate

First demonstration of simultaneous

chemical+electrophysiological

sensing in a wearable patch

Page 20: Wearable Chemical Sensors: Opportunities and Challengesefficiency.ucsd.edu/wp-content/uploads/2017/10/UCSD... · 2017-10-26 · Electrochemical Biosensors 8 Potentiometric Amperometric

Conclusions

• Current wearable sensors have limited utility

• Introduction of chemical sensing can augment

utility across a number of applications spaces:• High-end athletes

• Diabetic patients

• General-purpose “tricorder-like” monitoring

• There are many challenges:• Sensor fabrication on soft wearable materials

• Miniaturized, low-power electronics

• Low-cost calibration

• Interference & longevity

• Analytics & inference

20

Acknowledgements: NIH NIBIB, the Air Force Research Laboratory, and SONIC

(one of the six SRC STARnet Centers sponsored by MARCO and DARPA).

Can be solved with big-data

analytics & machine learning